Effect of Photoplethysmogram Artifacts on Uncertainty in Atrial Fibrillation Detection

Andrius Sološenko, Enikö Vargova, Loïc Coquelin, Vaidotas Marozas, Andrius Petrėnas

eess.SP ·submitted 2026-08-10 07:19 UTC · v1

Abstract

Detection of atrial fibrillation (AF) from photoplethysmogram (PPG) is highly sensitive to artifacts, yet their effect on uncertainty of different AF detectors remains poorly understood. This work aims to quantify how different PPG artifact types affect the uncertainty of AF detectors. Two machine learning approaches to AF detection were explored: one using 25-s PPG signals as input ($\mathit{D}_r$) and another using AF-related rhythm irregularity features ($\mathit{D}_f$). The detectors were trained on wrist PPG signals acquired during cardiac rehabilitation and then systematically evaluated on 260,000 PPG signals containing controlled artifact types and durations. Uncertainty was quantified using a threshold-based error rate, conformal prediction, and Monte Carlo dropout. Using artifact-free PPG signals, $\mathit{D}_f$ outperforms $\mathit{D}_r$ with sensitivity/specificity of 0.94/0.94 versus 0.92/0.86. Relative to artifact-free performance, sensitivity/specificity drops by 0.52/0.03, 0.25/0.03, 0.16/0.02, and 0.07/0.02 using $\mathit{D}_r$ for 12-s artifacts of device displacement, forearm motion, hand motion, and poor contact respectively. For $\mathit{D}_f$, the respective drops are 0.31/0.07, 0.19/0.11, 0.14/0.15, and 0.13/0.21 for the same artifacts. $\mathit{D}_f$ is more robust to short artifacts but exhibits increasing uncertainty with longer artifact durations, whereas $\mathit{D}_r$ shows an abrupt performance drop when artifacts occur but is less sensitive to artifact duration. Applying conformal prediction with 90$\%$ coverage reduces the false-positive rate by up to 12$\%$ for $\mathit{D}_r$ and up to 64$\%$ for $\mathit{D}_f$. Artifact type and duration have detector-specific effects on AF detection uncertainty. Device displacement causes the largest increase in uncertainty.

Closest papers

Nearest neighbours by meaning across the whole index.

CardioFusion-AI: Robust ECG--PPG Fusion for Multimodal Physiological Monitoring Under Signal Degradation

eess.SP cs.LGq-bio.QM ·2026-08-26 ·Navaneetha Krishnan Kamalakannan, Janakiraman Kamalakannan
code/data similarity 0.60
Abstract · 7 pages, 4 figures. Under review at IEEE Journal of Biomedical and Health Informatics. Code: https://github.com/ka-cyber/CardioFusion-AI

Wearable electrocardiogram (ECG) and photoplethysmogram (PPG) sensors are complementary but individually fragile: motion artifact, poor contact, and sensor dropout can degrade one or both signals. Fusion strategies that assume both modalities are equally trustworthy can become less reliable than a single clean modality under degradation. We present CardioFusion-AI, a framework whose signal-processing front end, including R-peak and systolic-peak detection, an Orphanidou-type signal-quality index, and beat-by-beat pulse transit time estimation, is validated on 53 real intensive-care recordings (848 windows; heart-rate mean absolute error 1.61 bpm for ECG and 2.78 bpm for PPG) and a real annotated fetal ECG database (R-peak F1 0.89-0.98). We then conduct a controlled synthetic degradation study comparing eight ECG-PPG fusion strategies across six degradation regimes spanning graded corruption and complete modality loss, using five independent training seeds. Attention fusion achieved the lowest descriptive overall error (1.66+/-0.43 bpm). Both adaptive gates reallocated weight toward the healthy modality under complete modality loss, but showed near-zero correlation between gate weight and signal quality under graded degradation (r = 0.10-0.24). Signal-quality conditioning produced a specific improvement under missing-PPG conditions (1.56+/-0.59 bpm), approaching the 1.48 bpm unimodal ceiling. With only five training seeds, no pairwise comparison survives Holm-corrected significance testing; effect sizes and confidence intervals are therefore reported. These results indicate that modality availability and modality quality are functionally distinct problems for adaptive fusion.

0.59

Pulse waveform morphology encodes haemodynamic signatures of impending stroke

cs.AI ·2026-11-07 ·Cheng Ding, Jiaming Liu, Jian Wu, Hongxia Xu, Daoqiang Zhang
similarity 0.57
Abstract · 23 pages, 10 figures

Stroke is a disorder of the circulation, yet the haemodynamic state that immediately precedes it has rarely been observed, because continuous physiological recordings before a stroke are seldom available. Photoplethysmography (PPG), recorded by every pulse oximeter and by hundreds of millions of wearable devices, captures the peripheral pulse beat by beat, but its morphology is largely discarded. Here we ask whether pulse waveform morphology encodes haemodynamic signatures of impending stroke. Using in-hospital stroke as a natural experiment that provides continuous PPG before physician-adjudicated stroke recognition, we studied intensive-care (MIMIC-III, $n=176$) and emergency-department (MC-MED, $n=154$) cohorts and described each beat with interpretable descriptors of pulse timing, amplitude, contour and variability. At a nominal 360-min horizon, morphology separated pre-recognition from stable windows with $F_1$ scores of 0.941 and 0.989 (sensitivity 0.998 and 0.980), with models transferred unchanged to the second cohort, and outperformed heart rate alone on identical windows. The signature was dominated by systolic crest time and its change from each patient's own baseline, feature attribution was concordant across cohorts (Spearman $ρ=0.865$), and morphology recorded in the first two hours of care stratified the subsequent clinical course ($F_1$ 0.714 and 0.691). In high-risk patients without stroke, 29% and 14% of windows were positive, but five-window persistence rules limited alerts to 17% and 2% of patients. The results demonstrate that the routinely recorded pulse thus carries haemodynamic imprint of the period preceding stroke, positioning pulse morphology as a candidate continuous cerebrovascular biomarker.

0.24

Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical Translation

cs.ET cs.LG ·2026-11-12 ·Karthika Shalini Potail, Harshitha Manjunatha, Hetvi Kotadia, Tooraj Nikoubin
surveybenchmark similarity 0.56
Abstract

Continuous blood pressure (BP) monitoring could improve hypertension management by capturing changes that intermittent cuff measurements miss. Photoplethysmography (PPG) is widely explored for cuffless BP estimation because optical sensors are inexpensive, wearable, and already present in many consumer and clinical devices. PPG, however, does not measure arterial pressure: BP is inferred from waveform characteristics or pulse timing, which also depend on vascular tone, arterial stiffness, sensor contact, motion, and individual physiology. This survey reviews PPG-based cuffless BP estimation along the full measurement chain, covering physiological mechanisms, waveform features, pulse arrival time (PAT), pulse transit time (PTT), machine-learning methods, datasets, calibration, validation, deployment, bias, and clinical translation. Studies are assessed by the strength of evidence they provide rather than by reported error alone, with particular attention to subject and temporal leakage, dataset imbalance, calibration dependence and drift, motion artifacts, and the gap between benchmark accuracy and reliable real-world measurement. The review finds that progress depends on more than better prediction algorithms. Future work should prioritize physiologically informed models, diverse longitudinal datasets, transparent reporting of calibration requirements and failure rates, stronger evaluation protocols, and studies that demonstrate practical value in clinical care.

0.09

Impact of Multiple Non-Invasive Biosignals on Cardiovascular Biomarker Estimation via Simulation-Based Inference

cs.LG physics.med-phstat.AP ·2026-08-26 ·Shusaku Maeda, Masahiro Nakano, Tomoharu Iwata, Kenji Komiya, Ryo Nishikimi +1 more
similarity 0.56
Abstract · 8 pages, 5 figures, 2 tables

As the population ages, the number of patients with cardiovascular diseases continues to increase, highlighting the need for early detection before progression to severe and irreversible functional decline. Consequently, estimating cardiovascular biomarkers from non-invasive biosignals, such as photoplethysmography (PPG) and arterial pressure wave (APW) signals, has attracted increasing attention. These signals can be measured using wearable and cuff-type devices. Previous studies have used PPG and APW signals to estimate cardiovascular biomarkers. However, these signals exhibit strong similarities in both the temporal and frequency domains and primarily reflect peripheral and arterial pulse waveforms. Therefore, they may provide limited information about cardiac mechanical function. In contrast, the quantitative impact of additional biosignals, such as ballistocardiography (BCG), which reflect the body's minute mechanical responses to cardiac ejection, remains unclear. In this study, we generated synthetic PPG, APW, and BCG signals from a unified whole-body cardiovascular circulation model and evaluated the complementary contribution of BCG to probabilistic cardiovascular biomarker estimation. We estimated posterior distributions of cardiovascular biomarkers using neural posterior estimation and simulation-based inference. The results showed that adding BCG signals significantly improved estimation performance. Furthermore, even in ill-posed cases where PPG and APW alone produced multimodal posterior distributions, adding BCG yielded unimodal posterior distributions. These findings provide fundamental insights into signal selection for estimating cardiovascular dynamics.

0.28

PPG-Bench: A Unified Benchmark for PPG Foundation Models

cs.LG ·2026-10-09 ·Arvasu Kulkarni, Mahir Jain, Aditya Ray Mishra, Sandeep Singh, Siddharth Panwar
benchmarkcode/data similarity 0.56
Abstract

Photoplethysmography (PPG) is increasingly recorded by wearable and clinical devices at enormous scale, and a growing number of PPG foundation models (FMs) have been developed to compute reusable representations for various downstream tasks. These models are difficult to compare: different evaluations use different datasets, task definitions, preprocessing, downstream heads, hyperparameter tuning, and baselines. We present PPG-Bench, a unifying open-source benchmark evaluating five public PPG FMs across nine open-access datasets and 26 tasks spanning cardiovascular state estimation, activity, sleep, screening, and identification. We evaluate models under four downstream regimes of increasing adaptability (linear probing, random forests, frozen MLP probes, and end-to-end finetuning), with model-specific hyperparameter optimization and subject-disjoint cross-validation. We also evaluate against three controls: random/constant baseline, a deterministic pulse-rate-variability (PRV) feature set, and a generic time-series FM (MOMENT). Across all 68 task-regime combinations, 35.3% show no meaningful FM advantage over the random/constant baseline, and in 83.8% no PPG-specific FM outperforms both PRV and MOMENT. We also find that added downstream head capacity generally improves performance, but can cause significant degradation if the model is performing at chance/constant level. Current PPG FMs therefore cannot yet claim to provide generalised and transferrable representations of PPG signals, and future evaluation in this field should include checks against simple, non-learned controls. Code for the benchmark and instructions to evaluate more models can be found here - https://github.com/mannas-ai/ppg-bench.

0.31

Unlocking the Specialization Capability of PPG Foundation Models for Wearable Signals through Structured Pruning and Adaptive Fine-Tuning

cs.AI cs.LG ·2026-10-24 ·Huaming Du, Zhixian Xie, Gaoyang Zhen, Gang Kou, Guisong Liu +4 more
benchmark similarity 0.55
Abstract

Photoplethysmography (PPG) plays a central role in wearable health monitoring and clinical decision support. However, existing approaches to universal PPG representation learning are predominantly trained on high-quality large-scale datasets obtained from oximeters in hospital settings and often struggle to adapt to sparse and low-quality real-world datasets, thereby limiting their generalization to clinically relevant tasks on prevalent PPG data such as collected from wearable devices. Surprisingly, even after full fine-tuning, PPG foundation models cannot consistently outperform general-purpose time-series models. Therefore, a key question is how to effectively adapt PPG foundation models to target prediction tasks. Through empirical studies on various PPG foundation models, we find that pretrained models often exhibit inherent sparsity and computational redundancy, suggesting that PPG foundation models have learned to activate task-relevant network substructures for different prediction tasks. To preserve this valuable prior knowledge, we propose a structured pruning-and-fine-tuning framework, SPAFT, which regularizes the adaptive fine-tuning process by constraining it to a more relevant and compact parameter space based on parameter importance and structural redundancy. Extensive experiments across four PPG foundation models, five benchmark datasets, and ten downstream tasks demonstrate that SPAFT achieves win rates of 82.8% and 87.5% on regression and classification tasks, respectively. This prune-then-finetune paradigm often enables PPG foundation models to achieve state-of-the-art performance, laying a promising foundation for the real-world clinical deployment of PPG foundation models and the development of personalized PPG foundation models.

0.04

Enhancing Atrial Fibrillation Detection Through Hyperparameter Optimization

cs.LG ·2026-10-29 ·Sara Gjorgjieva, Ilinka Ivanoska, Marjan Gusev
publishedhot direction similarity 0.55
Abstract · 11 pages, 3 figures, 6 tables. Presented at the ICT Innovation 2025 Conference, Ohrid, North Macedonia. Accepted for publish by Springer in …

This study explores the role of hyperparameter optimization (HPO) in enhancing the detection of Atrial Fibrillation (AF) using both machine learning and deep learning models. We evaluated a range of standard classifiers - Decision Tree, Random Forest, and XGBoost classifiers, alongside deep learning architectures including Recurrent Neural Networks, Deep Neural Networks, and Siamese Neural Networks. To fine-tune these models, we applied four HPO methods: Genetic Algorithm, Hyperband, Optuna, and SMAC3. The experiments were conducted using inter-patient and intra- patient data splits to assess model generalization and robustness in the presence of class imbalance, which is reflective of real-world AF occurrence. Our results, based on PhysioNet electrocardiogram datasets, showed that the Hyperband-optimized XGBoost model outperformed all others, achieving a peak F1 score of 96.95%. These findings highlight the critical role of suitable HPO strategies in enhancing the accuracy and robustness of AF classification outcomes in clinical settings.

0.39

Property-Specific Recoverability from Contact PPG to Camera rPPG under Heterogeneous Observation Conditions

eess.SP cs.AI ·2026-08-27 ·Timothy Oladunni, Farouk Ganiyu-Adewumi
similarity 0.55
Abstract

Camera-derived remote photoplethysmography (rPPG) is commonly validated through endpoint accuracy, but endpoint performance does not establish whether other physiological properties of source contact photoplethysmography (PPG) remain preserved recording by recording. We evaluated property-specific PPG-to-rPPG recoverability on 655 recordings from the Multi-Domain Mobile Video Physiology Dataset using CHROM as a fixed camera-rPPG observation pathway. The pathway reproduced the published CHROM correlation regime, with heart-rate MAE of 15.26 bpm and Pearson correlation of 0.0801. Matched-versus-shuffled validation revealed modest recording-specific autocorrelation correspondence, while spectral and recurrence-rate measures showed little matched discrimination. Maximal Lyapunov exponents showed essentially no recording-specific PPG-to-rPPG correspondence, with correlation of 0.0231 and permutation p-value of 0.5584, despite population-level overlap. Endpoint discrepancy exhibited Fitzpatrick-associated heterogeneity after adjustment for lighting and motion, including a Fitzpatrick VI versus III contrast of 9.32 bpm, while dynamical discrepancy showed no corresponding gradient. Aggregate RGB signal-to-noise ratio did not materially account for the endpoint contrast. In subject-held-out analysis, adding motion and lighting consistently reduced MAE across linear, ridge, and random-forest learners relative to rPPG-HR-only calibration, with reductions up to 13.32 percent. These findings show that recoverability is property-specific: physiological properties differ in recording-specific preservation and dependence on observation conditions, and population-level plausibility does not establish preservation of individual recordings.

0.27