SIFPBPNet: A Dual-Path Network for Wearable and Cuffless Blood Pressure Estimation via Individualized Steady-state Representation

Shuailong Tang, Xiaoyu Li, Donglin Xie, Wei Chen, Guangpu Zhu, Yelei Li, Yali Zheng

cs.LG ·submitted 2026-09-11 10:36 UTC · v1
published

Abstract

Continuous and cuffless blood pressure (BP) monitoring using photoplethysmography (PPG) is of great interest for low-cost and personalized cardiovascular health management. However, significant population heterogeneity and the "one-to-many mapping" problem, where similar waveforms across individuals correspond to different BP levels, limit the accuracy of conventional population-based models. To address this challenge, we propose a dual-path architecture termed SIFPBPNet, which separately represents steady-state and instantaneous features, through a Steady-state Feature Path (SFP) and an Instantaneous Feature Path (IFP). The SFP employs a Graph Attention Network (GAT) to extract individual-specific and long-term characteristics from multi-day historical PPG trajectories. In parallel, the IFP captures short-term dynamics from current PPG segments and incorporates the steady-state prior via a cross-attention mechanism. Experiments on a large-scale wearable dataset demonstrate that SIFPBPNet achieves a Mean Absolute Error (MAE) of 8.57 and 5.97 mmHg for systolic and diastolic BP, respectively, outperforming state-of-the-art models. Furthermore, the SFP module consistently improves performance when integrated into various backbone architectures, yielding 2.8-13.1% relative MAE reductions for systolic BP. These results highlight the strong generalizability and plug-and-play transferability of the SFP module, underscoring its great potential for accurate cuffless BP monitoring.

Comment
Accepted for publication at the 48th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2026), Toronto, Canada

Closest papers

Nearest neighbours by meaning across the whole index.

Evaluating Generalization and Long-Term Stability of Smartwatch-Based Cuffless Blood Pressure Monitoring

eess.SP ·2026-10-29 ·Hongda Huang, Xiao Peng, Xiaoyu Li, Shuailong Tang, Guangpu Zhu +4 more
published similarity 0.73
Abstract · Accepted by the 2026 IEEE EMBS International Conference on Biomedical and Health Informatics (BHI 2026)

Smartwatch-based cuffless blood pressure (BP) monitoring is a promising low-burden solution for long-term cardiovascular management, but practical deployment remains constrained by model complexity, uncertain cross-dataset generalization, and insufficient evidence on long-term tracking stability. This study presents an efficient Signal-Context Gated Fusion Network (SCGBPNet) for cuffless BP estimation from photoplethysmography (PPG) signals. SCGBPNet integrates signal-level representations through self-attention and further incorporates subject-specific contextual information via a compact gated fusion module. On Aurora-BP, SCGBPNet achieved SBP/DBP mean absolute errors (MAEs) of 7.08/4.90 mmHg, compared with 6.44/4.89 mmHg under within-dataset validation on the private smartwatch dataset, indicating a modest within-dataset discrepancy. Cross-dataset transfer from Aurora-BP to the smartwatch dataset increased the MAEs to 8.98/6.47 mmHg, demonstrating substantial cross-dataset degradation. For long-term stability, the proposed method showed low week-to-week performance variation and a low daily decay rate, indicating good stability over a follow-up period of more than one month. Nevertheless, estimation accuracy deteriorated when BP deviated substantially from an individual's baseline. These findings indicate that favorable aggregate MAE alone should not be interpreted as evidence of deployment-ready performance. More systematic and stratified evaluation across dataset settings, BP ranges, and baseline-deviation levels is needed to assess the practical reliability of smartwatch-based cuffless BP monitoring.

0.25

Personalized and Explainable Blood Pressure Estimation from PPG via Hybrid CNN--Morphological Features

cs.CV ·2026-08-11 ·Myung-Kyu Yi, Jongshill Lee, Jeyeon Lee, In Young Kim
published similarity 0.73
Abstract

Continuous cuffless blood pressure (BP) monitoring using photoplethysmography (PPG) offers a promising solution for personalized healthcare. However, existing methods have two major limitations. Handcrafted feature-based approaches rely on precise fiducial point detection and are limited to short-term analysis, while deep learning models, despite their accuracy, often operate as black boxes with limited physiological interpretability. To address these challenges, we propose a physiology-guided hybrid framework for personalized BP estimation that couples a convolutional neural network (CNN) branch capturing global and local waveform dynamics with a morphology-prior branch that explicitly encodes person-specific vascular characteristics. By embedding a morphology-based feature set that explicitly encodes individual vascular characteristics, the proposed framework enhances personalization and reduces dependence on large-scale training datasets. Evaluated on a subset of the MIMIC-III database under a subject-specific (personalized) testing protocol, the proposed personalized physiology-guided hybrid approach achieved mean absolute errors (MAEs) of 3.77 mmHg for systolic BP and 2.36 mmHg for diastolic BP, corresponding to relative improvements of 43.7% and 32.4% over a subject-specific (personalized) CNN-only baseline. SHAP-based analysis confirmed that the introduced morphology-prior features align with individual vascular characteristics, reinforcing per-subject interpretability. These findings highlight the potential of personalized, physiology-guided hybrid learning with novel morphological descriptors for accurate and explainable BP monitoring in real-world settings.

0.57

Calibration-Free Cuffless Blood Pressure Estimation Using Multimodal ECG-PPG Fusion on a Google Pixel Watch

cs.HC ·2026-08-26 ·Jathushan Kaetheeswaran, Boyi Ma, Ali Abedi, Shehroz S. Khan, Milad Lankarany
hot direction similarity 0.70
Abstract

Inadequate blood pressure (BP) monitoring and management outside of clinical settings can worsen major cardiovascular risk factors such as hypertension. While cuff-based devices are commonly used for at-home monitoring, these devices can be inconvenient for daily use due to their sensitivity to body positions, upper-arm constrictions, and limited portability. A promising alternative is emerging in the form of consumer-grade smartwatches, where physiological signals related to cardiac activity can be used to estimate BP non-invasively and continuously across daily living conditions. In this work, we use data collected from a Google Pixel Watch in 40 participants to develop and compare several algorithm approaches for BP estimation. We found that our proposed deep learning model achieved the strongest overall performance, and that fusing smartwatch signals with demographic information improved model generalizability to unseen individuals. However, we also identified that model accuracy was not consistent across participant subgroups, with obese individuals yielding higher estimation errors than others. This study highlights the feasibility of consumer-grade smartwatches as accessible platforms for deploying robust BP estimation algorithms, though clinical reliability will require larger, more diverse populations and additional sensing modalities.

0.22

Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical Translation

cs.ET cs.LG ·2026-11-12 ·Karthika Shalini Potail, Harshitha Manjunatha, Hetvi Kotadia, Tooraj Nikoubin
surveybenchmark similarity 0.69
Abstract

Continuous blood pressure (BP) monitoring could improve hypertension management by capturing changes that intermittent cuff measurements miss. Photoplethysmography (PPG) is widely explored for cuffless BP estimation because optical sensors are inexpensive, wearable, and already present in many consumer and clinical devices. PPG, however, does not measure arterial pressure: BP is inferred from waveform characteristics or pulse timing, which also depend on vascular tone, arterial stiffness, sensor contact, motion, and individual physiology. This survey reviews PPG-based cuffless BP estimation along the full measurement chain, covering physiological mechanisms, waveform features, pulse arrival time (PAT), pulse transit time (PTT), machine-learning methods, datasets, calibration, validation, deployment, bias, and clinical translation. Studies are assessed by the strength of evidence they provide rather than by reported error alone, with particular attention to subject and temporal leakage, dataset imbalance, calibration dependence and drift, motion artifacts, and the gap between benchmark accuracy and reliable real-world measurement. The review finds that progress depends on more than better prediction algorithms. Future work should prioritize physiologically informed models, diverse longitudinal datasets, transparent reporting of calibration requirements and failure rates, stronger evaluation protocols, and studies that demonstrate practical value in clinical care.

0.10

A Multidimensional Data-Driven Hybrid Transformer Framework for Non-invasive Continuous Blood Pressure Prediction

cs.LG ·2026-08-24 ·Yuexin Ma, Jingqi Hou, Yuxuan Kang, Zhaoying Liu
similarity 0.68
Abstract · 21 pages, 5 figures, 6 tables. Corrected typographical errors in two author email addresses; no changes to the scientific content

Objective. To develop and evaluate a cuffless continuous blood pressure (BP) estimator using temporal physiological and demographic features. We propose a hybrid Transformer framework to estimate diastolic and systolic BP from ECG/PPG-derived feature sequences. Approach. Rather than raw waveforms, the framework models 10-step sequences of six physiological descriptors and two demographic covariates. A Multi-Source Temporal Encoder Module combines Transformer, Kolmogorov-Arnold Network, and XGBoost branches to capture complementary temporal, nonlinear, and tabular information. A Dynamic Conditional Fusion-Decoder applies differential multi-head attention, token-weighted aggregation, and gated residual correction. A robust composite objective jointly optimizes DBP and SBP. Main results. Using the MIMIC-III Waveform and Clinical Databases, the source pool comprised 28,486 waveform segments from 203 subjects, and feature generation retained 53,621 observations from 166 subjects. On 2,431 segment-level held-out test windows, mean error +/- standard deviation was 0.41 +/- 3.74 mmHg for diastolic BP and -1.60 +/- 5.95 mmHg for systolic BP, with 95% limits of agreement of [-6.93, 7.74] and [-13.25, 10.06] mmHg, respectively. The proportions within 10 mmHg were 98.48% and 94.36%. The framework achieved the lowest standard deviations and narrowest limits of agreement among the locally retrained baselines. Significance. The feature-sequence fusion framework improved agreement with reference BP and fell within numerical AAMI and BHS Grade A thresholds on this split. This retrospective analysis is not formal device validation; subject-disjoint and external evaluation remain necessary before clinical use.

0.27

PPG-Bench: A Unified Benchmark for PPG Foundation Models

cs.LG ·2026-10-09 ·Arvasu Kulkarni, Mahir Jain, Aditya Ray Mishra, Sandeep Singh, Siddharth Panwar
benchmarkcode/data similarity 0.67
Abstract

Photoplethysmography (PPG) is increasingly recorded by wearable and clinical devices at enormous scale, and a growing number of PPG foundation models (FMs) have been developed to compute reusable representations for various downstream tasks. These models are difficult to compare: different evaluations use different datasets, task definitions, preprocessing, downstream heads, hyperparameter tuning, and baselines. We present PPG-Bench, a unifying open-source benchmark evaluating five public PPG FMs across nine open-access datasets and 26 tasks spanning cardiovascular state estimation, activity, sleep, screening, and identification. We evaluate models under four downstream regimes of increasing adaptability (linear probing, random forests, frozen MLP probes, and end-to-end finetuning), with model-specific hyperparameter optimization and subject-disjoint cross-validation. We also evaluate against three controls: random/constant baseline, a deterministic pulse-rate-variability (PRV) feature set, and a generic time-series FM (MOMENT). Across all 68 task-regime combinations, 35.3% show no meaningful FM advantage over the random/constant baseline, and in 83.8% no PPG-specific FM outperforms both PRV and MOMENT. We also find that added downstream head capacity generally improves performance, but can cause significant degradation if the model is performing at chance/constant level. Current PPG FMs therefore cannot yet claim to provide generalised and transferrable representations of PPG signals, and future evaluation in this field should include checks against simple, non-learned controls. Code for the benchmark and instructions to evaluate more models can be found here - https://github.com/mannas-ai/ppg-bench.

0.31

ExpertoRhythm: Morphology-Aware Learning for Waveform Reconstruction and Cuffless Blood Pressure Estimation from Single-Channel PPG

cs.LG ·2026-09-28 ·Amir Arjomand, Kenneth B. Kent, Georgiy Krylov
published similarity 0.65
Abstract

Continuous cuffless blood pressure (BP) monitoring from photoplethysmography (PPG) has strong potential for wearable health and telemonitoring, but accurate estimation remains difficult because PPG-to-BP mapping must preserve subtle waveform morphology and pressure-range-dependent dynamics. We introduce ExpertoRhythm, an attention-enhanced 1D U-Net that reconstructs the arterial blood pressure (ABP) waveform from a single-channel PPG signal and derives systolic and diastolic BP directly from the reconstructed waveform. The central contribution is a composite morphology-aware learning objective that integrates range-weighted SmoothL1 reconstruction with a window-range regularizer to emphasize high-dynamic BP segments and reduce amplitude under/over-shoot. On the UCI cuff-less BP dataset with 942 subjects, ExpertoRhythm achieves 2.46/1.46 mmHg MAE for systolic/diastolic BP (SBP/DBP), while obtaining a 30.4% average relative error reduction over pure MSE across waveform reconstruction and BP estimation metrics. Clinical-style evaluation further demonstrates low bias and strong agreement across the BP range, including high-pressure windows up to 200 mmHg, satisfying AAMI criteria and achieving BHS Grade A. These results suggest that morphology-aware waveform reconstruction from a single PPG channel can provide an accurate and practical pathway toward continuous cuffless BP monitoring in wearable and remote-care settings.

0.21

Change Point--Aware Evaluation and Re-Calibration of PPG-Based Blood Pressure Estimation

eess.SP cs.AIcs.LG ·2026-08-19 ·Yunwon Tae, Minje Park, Gyunho Rho, Dongjoon Yoo, Sunghoon Joo
similarity 0.65
Abstract · MLHC 2026. The first two authors contribute equally

Non-invasive continuous blood pressure (BP) monitoring using photoplethysmography (PPG) is a promising alternative to cuff-based measurements. However, existing PPG-based BP estimation studies predominantly rely on aggregated performance metrics (e.g., mean absolute error) computed over entire evaluation intervals, which can obscure model failures during rapid BP fluctuations and limit clinical relevance. In this work, we propose a fluctuation-aware evaluation framework for PPG-based BP estimation based on time-series change point detection. Instead of heuristic BP thresholding (e.g., $Δ\mathrm{BP} > 10\mathrm{mmHg}$), we identify BP change points by capturing abrupt distributional shifts in BP trajectories and evaluate estimation performance specifically during these fluctuation periods. Our analysis shows that several state-of-the-art models exhibit substantial performance degradation around BP change points, and that periodic test-time calibration is insufficient to handle such dynamic BP variations. To address this limitation, we introduce a targeted re-calibration framework triggered by detected BP change points, improving robustness without modifying model architectures. To the best of our knowledge, this is the first systematic evaluation of PPG-based BP estimation from a BP change point perspective, highlighting the importance of fluctuation-aware evaluation and calibration for real-world continuous BP monitoring.

0.25