Per-Bundle, Pre-Reconstruction Inversion for Temporal CT: Flux-Compensation Design and Training-Free Empirical-Bayes Estimation

Guy M Besson

eess.IV ·submitted 2026-10-13 21:46 UTC · v2

Abstract

In Temporal computed tomography (TCT) each detector reading sums the photons of up to three rays. Earlier work set the resulting per-bundle limits and tested a learned prior on one patient at one incident count $N_0$; this paper lifts both. N$_0$FC ($N_0$ flux compensation), a bow-tie and tube-current-modulation co-design, equalizes each bundle's detected flux. SNN2 is a per-bundle Poisson maximum-likelihood estimator. V8c is a neural-network corrector trained on 19 patients stratified by water-equivalent diameter (WED). T2bp, a training-free empirical-Bayes post-processor, combines per-bundle Poisson likelihoods with a prior built from the patient's own bundles. Results use that work's Dose Inflation Factor (DIF), charging bias as error. On 23 non-test volumes, N$_0$FC lowers the per-volume median middle-path Cramer-Rao floor, as a DIF, from 23-48 under a flat-fluence (no bow-tie) baseline to 3.3-6.3 at 35-62% of its incident photons, near the equal-attenuation value 13/3 and 1.4-1.7 times below conventional bow-ties; the bow-tie alone achieves this median, and tube-current modulation trims the tail. For bundles with $X_2<6$ (83%), SNN2 attains this floor (median mean-squared-error ratio 1.02); training-free T2bp$\circ$SNN2 lowers the ratio to 0.89 (0.74-1.26). It is best or tied on three of four N$_0$FC test cells (WED 207-342 mm) and trails T2bp$\circ$V8c by 3% on the fourth. V8c beats SNN2 in distribution under baseline acquisition, gains at most 9% under N$_0$FC (losing 14% on the largest in-distribution test patient), and fails on a patient acquired below its training count level. Acquisition co-design removes most of the attenuation-contrast penalty, and training-free estimation attains the resulting limit. The remaining three- to six-fold gap to single-source CT at equal fluence, and the high-attenuation tail, are targets for inter-bundle priors.

Comment
v2: title changed from v1 (was "...: Flux-Compensation Design and Multi-Patient Generalization"); corpus results recomputed with the truth-free T2bp configuration; conventional bow-tie comparison and acquisition and bias-structure tests added. 72 pages (30 main text, 42 supplementary material), 8 figures, 19 tables. Paper 2 of a series; companion to arXiv:2604.10934

Closest papers

Nearest neighbours by meaning across the whole index.

Generative Priors for Total-Body PET Reconstruction

eess.IV cs.CV ·2026-09-18 ·Movindu Dassanayake, George Webber, Denis Prokopenko, Julia A. Schnabel, Andrew J. Reader
published similarity 0.59
Abstract · MICCAI RIME 2026 (Accepted Poster)

Total-body positron emission tomography (TB-PET) enables fast, low-dose, and dynamic imaging over a long axial field of view, but reconstruction remains challenging due to reduced counts, large image volumes, and increased data dimensionality. Generative models offer powerful learned priors for low-count PET, yet explicit PET measurement conditioning at TB scale has not been demonstrated. Inspired by PET decomposed diffusion sampling (PET-DDS), we propose TB-PET-DDS, a reconstruction framework that combines a CT-conditioned generative prior with patient-specific list-mode PET data. During reverse diffusion, PET consistency is enforced using a closed-form 3D TB-PET maximum-a-posteriori expectation maximisation fusion update, rather than relying solely on image-domain conditioning. The synthetic measurements are generated from the low-dose vendor reconstruction, providing a controlled inverse problem and retaining a full-dose reference. Proof-of-concept evaluation was performed on a single clinical TB-PET acquisition, with reconstruction hyperparameters selected against the full-dose reconstruction from the same case for all models. TB-PET-DDS improved NRMSE and SSIM relative to the DDIM and Image-DDS baselines, improving through-plane consistency in coronal views while enhancing agreement with the measured PET data. Although smoothed low-dose vendor reconstructions achieved the best global voxel-wise metrics, TB-PET-DDS produced the lowest liver coefficient of variation. These results demonstrate the feasibility of measurement conditioned diffusion reconstruction at TB scale and provide an initial route towards learned, projector-consistent generative reconstruction for low-count TB-PET imaging.

0.13

Taming Neural Fields with a Frequency-Lipschitz Curriculum for Sparse-View CT Reconstruction

cs.CV cs.LG ·2026-11-11 ·Songtao Hu, Jingrui Wu, Hong Ding, Fanhao Zhu, Yunhao Liu +1 more
similarity 0.59
Abstract

Sparse-view computed tomography (SVCT) reduces radiation dose by acquiring fewer projection views, but the resulting under-determined inverse problem requires careful regularization to prevent reconstruction algorithms from fitting streak artifacts and hallucinated structures. Existing implicit neural representation (INR) methods rely on implicit spectral bias or iterative external denoising, yet they provide limited control over when high-frequency capacity is exposed during optimization. To address this limitation, we propose FLIC (Frequency-Lipschitz INR Curriculum), a capacity-controlled self-prior INR framework that makes the network's representational capacity explicit and schedulable. The core of our method is a joint frequency-Lipschitz curriculum that couples random Fourier feature encoding with spectrally normalized MLP decoding under non-uniform Lipschitz budgets, guiding reconstruction along a coarse-to-fine trajectory from stable global anatomy to fine details. The terminal capacity is automatically calibrated from FBP gradient statistics, enabling scan-adaptive regularization without per-scan manual tuning of the terminal Lipschitz budget. We further introduce a prior-guided adaptive ray sampling strategy that concentrates integration samples near high-gradient anatomical boundaries, improving forward-model fidelity. The entire framework operates in a single training stage without iterative denoising, enabling efficient per-image reconstruction. Extensive experiments on four public CT datasets under noise-free and low-dose settings show that our method achieves state-of-the-art unsupervised reconstruction quality with consistent performance across diverse anatomical structures.

0.13

Improving Calibration of Black-Box Radiology AI Using Test-Time Augmentation

cs.LG cs.HC ·2026-09-24 ·Nathan Le, Magdalini Paschali, Arogya Koirala, Andrew Johnston, Zhongnan Fang +3 more
publishedcode/data similarity 0.58
Abstract · 11 pages, 3 figures, 1 table. Accepted at the MICCAI 2026 Workshop on Uncertainty for Safe Utilization of Machine Learning in Medical Imagin…

Radiology AI systems increasingly inform clinical decisions such as triage, follow-up imaging, and treatment planning. For these decisions to be made safely, model outputs must be well calibrated, meaning predicted probabilities accurately reflect true risk. Many standard techniques for improving calibration, such as MC Dropout and Deep Ensembles, require access to model parameters or retraining. However, proprietary clinical AI systems operate as black boxes, preventing access to the model's internals. To that end, we propose a model-agnostic framework for improving calibration of black-box models using clinically grounded test-time augmentation (TTA). Our framework applies geometric and physics-inspired 3D CT perturbations and learns probability-level aggregation strategies without access to model internals or the original training data. Across pulmonary embolism and intracranial hemorrhage detection tasks, DualTTA achieved the strongest overall calibration among TTA methods, reducing the Expected Calibration Error by 54% (0.239 -> 0.109) and 43% (0.051 -> 0.029), respectively, while requiring only input-output access. Additionally, DualTTA outperformed uncertainty estimation techniques that require access to model internals, such as Temperature Scaling, MC Dropout, and Deep Ensembles, in most calibration metrics. These results demonstrate that learned TTA aggregation can improve the calibration of clinical AI systems, providing a practical approach for improving the reliability of black-box medical AI.

0.45

Extending TotalSegmentator: Predicting Patient and Acquisition Characteristics from CT and MR Images

cs.AI ·2026-08-29 ·Jakob Wasserthal, Joshy Cyriac, Michael Bach, Kimia Mozahheb Yousefi, Minh-Son To +5 more
code/data similarity 0.58
Abstract

Background: Patient details and acquisition metadata are important for clinical decisions, image quality control, and automated research pipelines, but may be missing or unreliable in imaging archives. Purpose: To develop and evaluate a fast open-source model that predicts patient and acquisition characteristics directly from CT and MR images. Materials and Methods: Separate 3D ResNet-10 ensembles for CT and MR were trained on 57,291 and 43,200 clinical examinations acquired from 2011 to 2025. Both predicted weight, height, age, sex, contrast presence, vertebral coverage, and image noise. The CT model additionally predicted scanner manufacturer, tube voltage, tube current, convolution kernel, and post-injection time; the MR model predicted sequence class. Performance was evaluated on internal CT (n=501) and MR (n=636) test sets and an external CT dataset (n=54). Results: Internal CT MAEs were 3.90 kg, 3.68 cm, and 4.42 years for weight, height, and age, with sex F1=0.990; corresponding MR results were 4.34 kg, 4.62 cm, 7.13 years, and F1=0.970. The CNN outperformed a segmentation-derived XGBoost baseline for all four core targets in both modalities (adjusted P<=.042). F1 scores were 0.963 for CT contrast, 0.953 for MR sequence, and 0.823 for MR contrast. External CT MAEs were 4.45 kg, 4.05 cm, and 5.17 years, with sex F1=0.971. CPU inference required 20 seconds for CT and 12 seconds for MR. Conclusion: One 3D multitask model per modality can rapidly recover patient and acquisition characteristics from heterogeneous CT and MR examinations. Models are available in TotalSegmentator: https://github.com/wasserth/TotalSegmentator

0.43

Image-Condition-Controlled Diffusion Network with Large-Scale CT Pretraining for CT-Free Attenuation Correction in Dynamic Cardiac PET

cs.CV ·2026-11-18 ·Zhejia Zhang, Scott M. Leonard, Jinkui Hao, Ryan J. Avery, Richard L. Weinberg +1 more
similarity 0.57
Abstract · 11 pages, 6 figures, 3 tables. Submitted to IEEE Transactions on Radiation and Plasma Medical Sciences

Positron emission tomography (PET) provides quantitative information on myocardial perfusion. In dynamic PET, tracer kinetics are tracked through a series of reconstructed time frames. Reliable attenuation correction is essential for accurate measurements. Clinical attenuation maps ($μ$-maps) are typically derived from computed tomography (CT), which adds radiation exposure and may introduce PET-CT misregistration due to respiratory and cardiac motion. Moreover, supervised CT-free methods rely on paired non-attenuation-corrected (NAC) PET and CT data, which are scarce, whereas large-scale unpaired CT data are more readily available. To address these challenges, we propose Large-scale Unpaired CT-Informed Diffusion (LUCID), a framework that transfers anatomical priors learned from 10,000 chest CT volumes in the public CT-RATE dataset to $μ$-map synthesis for dynamic cardiac $^{82}$Rb PET. First, a convolutional neural network (CNN) predicts an initial $μ$-map from each NAC PET frame. Second, the CT-pretrained diffusion backbone refines a low-noise perturbation of the CNN prediction through a single denoising step, conditioned on the corresponding NAC PET frame via ControlNet. Finally, frame-wise predictions are averaged voxel-wise to obtain the final $μ$-map. We further introduce the voxel-wise temporal standard deviation (TSD) to quantify inter-frame variability without requiring a reference $μ$-map. We evaluated LUCID on an in-house dataset of 102 patients and found that it achieved the best quantitative accuracy while maintaining competitive inter-frame consistency. Ablation studies showed that refinement with a diffusion backbone pretrained on in-house CT data from the training patients was less accurate than the CNN prediction alone, highlighting the importance of large-scale CT pretraining.

0.22

EPC-RawDC: Equivariant Physics Constrained 3D Diffusion with Raw Data Consistency for CBCT to CT Image Synthesis

cs.CV cs.AIphysics.med-ph ·2026-10-29 ·Alzahra Altalib, Chunhui Li, Alessandro Perelli
code/datahot direction similarity 0.57
Abstract · 10 pages, 4 figures, 8 tables

Cone beam computed tomography (CBCT) is widely used for image guided and adaptive radiotherapy, but scatter, noise, reconstruction artifacts, and inaccurate Hounsfield units limit its quantitative use. Recent diffusion approaches for CBCT to CT synthesis remain largely image driven, with limited use of the underlying acquisition physics. We propose EPC-RawDC, a conditional 3D latent diffusion framework that combines volumetric generative modeling with two complementary constraints derived from the same scanner specific CBCT forward model. First, geometry equivariant projection constraint exploits the rotational equivariance of the CBCT forward operator by supervising the synthesized CT through correspondingly transformed reference projections. Second, raw data consistency uses noisy projection measurements of the individual acquisition as scan specific supervision during training. The framework was evaluated on two independent clinical cohorts with distinct scanner geometries and projection sources. For JUST, the physics constraints used measured clinical raw CBCT projections acquired with an Elekta system, whereas for SynthRAD2025 the projection data were simulated from the reference CT using a Varian geometry with Poisson photon statistics. EPC-RawDC achieved an SSIM of 0.97 in both cohorts and improved PSNR over DDPM by 2.66 dB on SynthRAD2025 and 2.10 dB on JUST, while reducing mean MAE by 20.3% and 19.0%, respectively. On SynthRAD2025, projection consistency and equivariant errors were reduced by 55.0% and 41.4% relative to DDPM. These results demonstrate that combining a volumetric diffusion prior with scanner geometry and scan specific projection measurements improves quantitative CBCT to CT synthesis across both simulated and measured projection domains. Code available after publication at https://github.com/ALZAHRAALTALIB/EPC-RawDC.

0.37

PEEL: Physics-Enabled Evidential Learning for Identifiable Uncertainty in CT Imaging

cs.AI physics.med-ph ·2026-08-27 ·Ge Wang
similarity 0.57
Abstract · 10 pages, 3 figures, 1 table. Proof-of-concept study of physics-enabled identification of NIG uncertainty in CT imaging

Normal-inverse-gamma (NIG) regression is not uniquely identifiable from its marginal Student-t likelihood: the likelihood determines three combinations of four NIG parameters and is constant along a one-dimensional fiber. We identify that fiber using independent physical measurement. As an initial embodiment, a reconstruction network receives one noisy filtered-backprojection (FBP) image and is first trained only by Student-t negative log-likelihood to estimate the three identifiable coordinates (gamma, alpha, c). The network is then frozen; repeated physical-noise realizations propagated through its reconstruction output form a Monte Carlo (MC) teacher label for output-domain aleatoric variance. An aleatoric head attached to frozen features learns this label, after which (beta, nu) are recovered algebraically. On 30 held-out simulated objects at five photon levels, one-image predictions achieved pooled Spearman correlations of 0.832-0.951 against independent 400-repeat references, median within-image correlations were 0.834-0.947, and 98.81-99.55% of evaluated pixels satisfied the algebraic admissibility condition. The method needs no KL term, reference prior, evidence regularizer, or cross-loss weight.

0.48

Impact of Pretraining Length, Model Size, and Annotation Budget on Downstream Clinical Tasks of Foundation Models in Radiotherapy

cs.CV ·2026-11-17 ·Alejandro Cortina Uribe, Sidsel Højklint Poulsen, Ivan Richter Vogelius, Jens Petersen
code/data similarity 0.57
Abstract · 9 pages, 5 figures, 2 tables. Supplementary Material included

Cone-beam CT (CBCT) scans acquired during radiotherapy capture changes such as tumor shrinkage and radiation-induced tissue changes, offering great potential for foundation models that help monitor individual treatment responses. However, to our knowledge, no large-scale pretrained models for radiotherapy CBCT exist, and evaluating modeling choices is difficult since downstream tasks are not known in advance. We examine how pretraining length, model size, and finetuning strategy relate to downstream clinical performance, and how metrics tracked during pretraining correlate with it. We pretrain vision transformer encoders with masked image modeling on about 150,000 raw 3D radiotherapy CBCT scans, tracking representation quality via validation loss, effective rank, and linear probing. Downstream tasks are classification of feeding tube presence ($n=51,135$ CBCTs, $1,562$ patients) and clinical characteristics (T-, N-stage, sex, age, tumor site; $n=1,278$ patients) in head and neck cancer. We compare low-rank adaptation (LoRA) and full finetuning against no pretraining, varying annotated dataset size. Pretext validation loss was the most consistent predictor of downstream performance; effective rank and linear probing were also associated, but not consistently across model sizes and finetuning methods. LoRA performed comparably to full finetuning while training only 0.5-2.5% of the parameters. Fully finetuned models consistently outperformed models without pretraining across all sizes, with AUROC 0.91-0.97 vs. 0.58-0.85 for feeding tube classification and 0.99-1.00 vs. 0.71-0.97 for sex classification. Model size had only a small, inconsistent effect, suggesting smaller models suffice at lower computational cost. Code and pretrained models (non-commercial license): https://github.com/alexCortinaU/RT-CBCT_SSL

0.48