Fast Slice-Wise Deep Learning Motion Correction for Spinal Cord Functional MRI

Thomas Dagonneau, Sandrine Bédard, Nontharat Tucksinapinunchai, Uten Yarach, Loïc Le Folgoc, Pietro Gori, Julien Cohen-Adad

cs.CV ·submitted 2026-11-10 19:32 UTC · v1
code/data

Abstract

Respiratory and bulk motion displace the spinal cord between the volumes of a functional MRI time series. Current methods such as the Spinal Cord Toolbox (SCT) take several minutes per time series, and general-purpose brain tools register whole volumes, whereas the cord moves mainly slice by slice. We propose a lightweight convolutional neural network that regresses the in-plane translation aligning each axial slice to a reference. Two models, trained without motion labels, correct large and then residual motion. On a 3T (120 scans) and a 7T (59 scans) dataset, the proposed method improves on SCT on every metric (tSNR +10 to +25%, DVARS -20 to -38%), leaves almost no slice displaced by more than one voxel (<0.1%), performs on par with a slice-wise AFNI variant and better than slice-wise FSL FLIRT, and runs in about 10 s per scan, about 30 times faster than SCT. The code is available at https://github.com/ivadomed/moco-dl and the method is available in SCT v7.4.

Comment
4 pages, 3 figures

Closest papers

Nearest neighbours by meaning across the whole index.

Motion Artifact-Aware Self-Supervised Representation Learning for 3D Brain MRI Motion Artifact Reduction

cs.CV physics.med-ph ·2026-08-10 ·Mojtaba Safari, Shansong Wang, Zach Eidex, Matthew Goette, Tonghe Wang +2 more
top venue similarity 0.64
Abstract

Patient motion remains a source of image degradation in brain MRI, leading to signal loss, blurring, and geometric distortion that compromise quantitative analysis. Existing deep learning methods for motion correction typically rely on paired clean-corrupted data or k-space acquisitions, which are rarely available in clinical settings. We propose SSRL-MAR, a motion artifact-aware unpaired representation learning framework for motion artifact reduction that requires neither paired training data nor explicit motion labels. SSRL-MAR employed a three-stage training strategy: (1) contrastive learning on 3D patches to extract motion representations by contrasting clean and synthetically corrupted images, (2) a motion artifact-aware synthesis network to generate motion artifacts from clean scans, and (3) a motion artifact-aware generator to restore clean volumes using the learned degrader for self-supervised supervision. On in-silico dataset, SSRL-MAR achieved PSNR 23.81dB, SSIM 91.55%, and NMSE 0.79%. On in-vivo MR-ART dataset, the pretrained model reduced motion distortion, and unsupervised domain adaptation further improved anatomical fidelity. Against a source-only supervised model trained on the same simulated pairs, SSRL-MAR improved PSNR by up to 2.0 dB on MR-ART after unsupervised domain adaptation, and remained within 0.25-0.47 dB of an oracle supervised model that requires real paired data unavailable in practice. At the milder motion level, volumetric error in structures such as the corpus callosum and ventricular system decreased by more than 50%, confirming improved neuroanatomical consistency. These results indicate that SSRL-MAR provides a robust and scalable image-domain solution for 3D brain MRI motion correction, enabling reliable structural quantification in large-scale neuroimaging studies without requiring prospectively acquired pairs or acquisition-specific calibration.

0.29

ContiLNN: Mitigating Slice Sampling Discontinuity with Liquid Neural Networks for Medical Image Restoration

cs.CV cs.AI ·2026-10-08 ·Jialei He, Enhe Liu, Sifan Song, Pengfei Jin, Jionglong Su +8 more
similarity 0.62
Abstract

Anatomical continuity provides complementary information for medical image restoration, but its use requires accounting for local anatomy and variations in slice sampling. We introduce ContiLNN, which augments two-dimensional restoration backbones with bidirectional closed-form continuous-time (Bi-CfC) modules for cross-slice modeling while retaining in-plane feature extraction. Slice-index intervals modulate gates determined by local features and hidden states, enabling propagation to respond to sampling variations without numerical ODE integration. Reference-guided consistency aligns first- and second-order cross-slice intensity differences to preserve anatomical variation, while distillation from a frozen backbone helps retain in-plane fidelity. Across five training seeds, ContiLNN improves mean PSNR over Restore-RWKV by 0.1907, 1.0176, and 1.2482 dB for CT denoising, MRI super-resolution, and reduced-count PET restoration, respectively, with lower RMSE in all three tasks. CT results are descriptive for one held-out patient. PET ablations support ordered propagation beyond additional pointwise capacity. Under contiguous training, Bi-CfC achieves higher fidelity than a Bi-GRU with similar parameter counts and arithmetic costs across all tested sampling conditions. Matched seven-slice profiling shows 52.8% lower latency and 57.0% lower peak GPU memory use than Bi-GRU. Mixed-gap training improves sparse and irregular-context performance for both operators, without a uniform ranking across metrics and contexts. Experiments with fewer training patients and a second backbone further support data efficiency and backbone compatibility.

0.23

STORK: Spatio-Temporal Observation of uterine contRactions via neural networKs

cs.CV ·2026-10-07 ·Melissa Schween, Tristan Gottwald, Jordina Aviles Verdera, Lisa Story, Mary Rutherford +1 more
published similarity 0.62
Abstract · Accepted at the PIPPI Workshop at MICCAI 2026 and will appear in the workshop proceedings (Springer)

Uterine contractions in fetal MRI are typically identified manually and discarded, limiting insights into contraction dynamics. We formalize Uterine Contractile Activity Detection (UCAD) as a weakly-supervised learning problem and introduce STORK, a multi-instance learning model trained on dynamic MRI series using only coarse, series-level labels. STORK factorizes 3D spatio-temporal convolutions into parallel branches across temporal hyperplanes to capture coherent tissue motion without the cost of full 4D convolutions. Per-frame embeddings, combining intensity and Demons-estimated displacement fields, are aggregated by a linear mean-pooling head. This ensures that frame-level contraction scores can be recovered post-hoc without frame-level training supervision. Evaluated on around 700 multi-vendor dynamic fetal MRI series, STORK achieves a series-level AUROC of 95.0% and AUPRC of 94.6%, substantially outperforming 3D ResNet and ConvNeXt baselines. Grad-CAM analysis suggests that the model draws on predictive features extending beyond the placenta into the uterine tissue, offering an automated tool for richer phenotyping of uterine behavior.

0.25

MRI-to-CT Translation of Spinal Images via Anatomy-Guided Pseudo-Pairing and Adversarial Diffusion

cs.CV cs.AI ·2026-10-27 ·Eduardo Nieto, Ibai Genua, Davide Scorza, Karen López-Linares
similarity 0.62
Abstract

Magnetic Resonance Imaging (MRI) is routinely used to diagnose spinal disorders, but surgical planning often requires Computed Tomography (CT) for better visualization of vertebral bone anatomy. Synthesizing CT from MRI could avoid additional radiation exposure and streamline preoperative assessment. We propose a three-stage framework for spine MRI-to-CT synthesis. First, an anatomy-guided vertebral matching system identifies similar sagittal MRI and CT slices using spine segmentations and aligns them through rigid registration, creating pseudo-paired data from unpaired scans. Second, these pairs train an adversarial diffusion model for MRI-to-CT translation. Third, synthesized slices are assembled into 3D volumes and refined by a network that improves inter-slice continuity and structural coherence. The framework was trained on over 10,000 unpaired sagittal MRI and CT slices and evaluated on 1,911 slices from 83 MRI volumes spanning all spinal regions. Assessment combined distribution-based measures with vertebral segmentation metrics. Compared with diffusion alone, the proposed approach increased binary 3D vertebral Dice from 0.666 to 0.739. Volumetric post-processing reduced visible discontinuities between stacked slices and improved vertebral coherence in axial and coronal views. Combining anatomy-guided pseudo-pairing, adversarial diffusion and 3D refinement improves spine MRI-to-CT synthesis. Vertebral matching primarily enhanced anatomical alignment and fidelity, while volumetric refinement improved 3D consistency. These findings support further development of MRI-derived CT-like visualization for preoperative planning, particularly to preserve fine intensity details while maintaining volumetric coherence.

0.16

FlowMoDL: Model-Based Deep Learning with Conjugate-Gradient Data Consistency for Highly Accelerated 4D Flow MRI Reconstruction

cs.CV cs.LG ·2026-08-26 ·Tristan Gottwald, Michelle Bruch, Mubashir-Ul Hassan, Fatma Alickovic, Milan Kloiber +4 more
similarity 0.61
Abstract

We present FlowMoDL, an unrolled neural network for highly accelerated 4D flow MRI reconstruction that directly optimizes for both anatomical magnitude and phase-derived velocity accuracy. Building on the MoDL framework, FlowMoDL alternates a learned (3+1)D spatiotemporal denoiser with conjugate-gradient data-consistency updates based on the SENSE forward model. A novel dual-pathway conditioning scheme adapts the denoiser features and data-consistency weighting, enabling a single model to handle varying acceleration factors ($10\times$ to $50\times$). To ensure physiological accuracy, the network is trained using a deep-supervision composite loss that explicitly penalizes velocity magnitude and angular errors, stabilized by a curriculum schedule. We evaluate FlowMoDL on the multi-center CMRx4DFlow dataset against classical and deep-learning baselines (CG-SENSE, MoDL, FlowVN, and FlowMRI-Net). A key advantage of FlowMoDL is its superior gradient step efficiency. When evaluated under an equivalent, limited budget of gradient steps, competing flow-specific networks degrade significantly. In contrast, FlowMoDL robustly converges and strictly outperforms all competitors across all acceleration factors in magnitude SSIM, nRMSE, relative velocity error, and angular error, successfully recovering sharp structural details and temporally coherent velocity fields.

0.09

Native-Space 3D CarveMix for Multi-Site T1w Stroke Segmentation

eess.IV cs.CV ·2026-08-24 ·Dexter Wen Jie Teo, Kumaradevan Punithakumar
published similarity 0.61
Abstract · Accepted at the SWITCH+ Workshop (ISLES 2026 Challenge), MICCAI 2026. Springer LNCS

Segmenting ischemic stroke lesions on T1-weighted (T1w) MRI acquired across different scanners and protocols without intensity standardization is difficult because lesions are subtle and share intensity characteristics with cerebrospinal fluid. Standard deep learning architectures trained across multiple centers plateau around Dice 0.66, with acute lesions ($\le 7$ days post-stroke) performing substantially worse due to severe sample scarcity. We combine a MedNeXt-L ($k=5$) backbone with on-the-fly 3D CarveMix augmentation that pastes real lesion patches into healthy brain regions during training. By generating synthetic lesion placements dynamically within each fold with subject-level split isolation, the model sees more diverse lesion patterns without requiring pre-generated copies on disk. We evaluate on 1,453 native T1w scans from 55 clinical centers in the ISLES 2026 challenge. Our method achieves a mean 5-fold cross-validation Dice of 0.648 at 500 epochs, a +0.018 improvement over the MedNeXt-L backbone at a matched training budget (0.630)

0.31

A foundation for systematic analysis of transformers and RNNs for tractography

cs.LG eess.IVq-bio.NC ·2026-10-01 ·Emmanuelle Renauld, Philippe Poulin, Hugo Larochelle, Antoine Théberge, Maxime Descoteaux
published similarity 0.60
Abstract

Machine learning (ML) has emerged as a promising approach for improving diffusion MRI (dMRI) tractography, a task that remains limited by the intrinsic tension between local diffusion information and global anatomical plausibility. In this work, we systematically evaluate recurrent neural networks (RNNs) and Transformer models for iterative tractography, with particular attention to training strategies, input representations (including convolutional neural network (CNN)-based embeddings and end-of-sequence (EOS) tokens), and hyperparameter selection. We introduce a generation-validation phase enabling supervision at the streamline level during training, allowing supervision despite the mismatch between local loss functions and global streamline quality. Using the ISMRM2015 tractography challenge dataset, our models achieve the highest reported performance to date. Through controlled experiments, we quantify the impact of missing bundles, noisy or imperfect training streamlines, and invalid fibers in the training set. Finally, we demonstrate the applicability of our best-performing models for in vivo data from the Tractoinferno database. Overall, our results highlight both the potential and the limits of sequence-based deep learning models such as Transformers and RNNs for tractography, and emphasize the need for improved phantoms and evaluation methods for in vivo validation. We provide takeaways and recommendations for future researchers training and validating sequence-based supervised methods for tractography.

0.30

Unified Motion Correction and Reconstruction for Accelerated Zero Echo Time MRI

eess.IV ·2026-11-03 ·Yan Chen, Curtis A. Corum, James H. Holmes, Mathews Jacob
similarity 0.60
Abstract

Zero echo time (ZTE) MRI is near-silent, but dead time gaps, head motion, and undersampling jointly degrade reconstruction. To develop an accelerated ZTE MRI reconstruction framework that jointly estimates the 3D image and rigid-motion parameters, this work introduces Energy-based Motion Correction (EnMoCo), a regularized maximum a posteriori (MAP) formulation that also compensates for missing k-space center data and angular undersampling artifacts. Solved by alternating between k-space rigid-motion updates and conjugate-gradient data-consistency updates, the framework incorporates regularization via a pre-trained Multi-Scale Energy-based model (MuSE) implemented in a data-efficient 2.5D form. Evaluations on simulated and in vivo 3D ZTE data, spanning dead time gaps of 1.5 to 5 samples, rotations up to 17 degrees and translations up to 10 mm, demonstrate that EnMoCo significantly improves PSNR and SSIM relative to CG-SENSE and Autofocusing baselines (achieving roughly 10 to 15 dB higher PSNR for gaps >= 3.5) while reducing motion blur and tissue contrast inconsistencies.

0.24