Unified Motion Correction and Reconstruction for Accelerated Zero Echo Time MRI

Yan Chen, Curtis A. Corum, James H. Holmes, Mathews Jacob

eess.IV ·submitted 2026-11-03 16:38 UTC · v1

Abstract

Zero echo time (ZTE) MRI is near-silent, but dead time gaps, head motion, and undersampling jointly degrade reconstruction. To develop an accelerated ZTE MRI reconstruction framework that jointly estimates the 3D image and rigid-motion parameters, this work introduces Energy-based Motion Correction (EnMoCo), a regularized maximum a posteriori (MAP) formulation that also compensates for missing k-space center data and angular undersampling artifacts. Solved by alternating between k-space rigid-motion updates and conjugate-gradient data-consistency updates, the framework incorporates regularization via a pre-trained Multi-Scale Energy-based model (MuSE) implemented in a data-efficient 2.5D form. Evaluations on simulated and in vivo 3D ZTE data, spanning dead time gaps of 1.5 to 5 samples, rotations up to 17 degrees and translations up to 10 mm, demonstrate that EnMoCo significantly improves PSNR and SSIM relative to CG-SENSE and Autofocusing baselines (achieving roughly 10 to 15 dB higher PSNR for gaps >= 3.5) while reducing motion blur and tissue contrast inconsistencies.

Closest papers

Nearest neighbours by meaning across the whole index.

Consistency Models for Fast MRI Reconstruction Using Regularization by Denoising

eess.IV cs.AIcs.CVcs.LG+1 ·2026-08-20 ·Merve Gülle, Junno Yun, Yaşar Utku Alçalar, Mehmet Akçakaya
code/data similarity 0.67
Abstract

Diffusion models (DMs) have emerged as powerful generative priors for MRI reconstruction with promising results. Yet DM-based methods require extensive iterative refinement, limiting their practical deployment. Consistency models (CMs) provide a compelling alternative, aiming to map out the diffusion trajectory in a single pass, enabling faster generation. In this work, we propose CM-RED, a novel MRI reconstruction method that integrates a pretrained CM into the regularization by denoising (RED) scheme. Our method builds on accelerated proximal gradient RED (RED-APG), and further incorporates controlled noise injection during the update steps to enhance generative diversity and accelerate convergence. Extensive experiments on the fastMRI knee and brain datasets demonstrate that CM-RED achieves high-quality reconstructions across multiple anatomies, contrast weights, acceleration factors, and undersampling patterns, using only 4 network function evaluations (NFEs). The proposed method consistently outperforms existing DM- and CM-based approaches in both quantitative metrics and visual fidelity, and exhibits strong robustness to hyperparameter variations, highlighting CM-RED as an efficient and effective generative framework for accelerated MRI reconstruction. The source code and pretrained models are publicly available at https://github.com/MerveGulle/CM-RED.

0.31

Multi-Stage Phase-Aware Self-Supervised Reconstruction for Multi-Echo Gradient-Echo MRI

cs.AI ·2026-11-09 ·Juan Liu, Nian Wang
similarity 0.66
Abstract · Submit to a journal; under review

Purpose: To develop a phase-aware self-supervised framework for accelerated high-resolution multi-echo gradient-echo (mGRE) MRI that preserves anatomical detail and complex echo information for quantitative mapping. Methods: We developed multi-stage phase-aware self-supervised reconstruction (MSPR-GRE), integrating progressive k-space expansion with phase estimation and refinement. Sequential unrolled networks reconstruct complex multi-echo images and update echo-dependent phase priors. A relaxation-informed temporal subspace is used for acquisitions with at least four echoes, while shorter echo trains are reconstructed in the image domain. Training uses held-out acquired k-space samples without fully sampled targets. Evaluation included 9.4 T ex vivo mouse-brain data 30um eight-echo and 18um dual-echo) and retrospectively undersampled 7 T AHEAD human-brain data, with supervised ablations. Results:MSPR-GRE reduced artifacts in regions of rapid phase variation and preserved fine mouse-brain anatomy. Across five AHEAD test datasets at R = 4, 8, and 12, it generally improved magnitude-image, T2* map, and QSM metrics, particularly at higher acceleration. At R = 12, fourth-echo magnitude NRMSE was 12.09$\pm$0.60% versus 21.55$\pm$1.12% for SSDU. Ablations indicated benefits of phase refinement, while supervised experiments supported multi-stage phase-aware reconstruction. Conclusion: MSPR-GRE improves accelerated mGRE reconstruction and quantitative-map agreement without fully sampled training targets. Further validation of generalizability and phase-estimation accuracy is needed.

0.13

High-Resolution Quantitative MRI Using Contrast-Domain Multiscale Energy (MuSE) Regularization

eess.IV ·2026-11-03 ·Rushdi Rusho, Jyothi Rikhab Chand, Steve R. Kecskemeti, James H. Holmes, Mathews Jacob
similarity 0.65
Abstract · 5 pages, 3 figures

Quantitative MRI methods can offer tissue parameter maps (e.g. T1) that are comparable across scanners, sessions and subjects, together with multiple contrast-weighted images from a single scan. However, most data are acquired at 1 mm, which limits their use in neurodegenerative diseases that affect fine brain structures. We introduce, to our knowledge, the first diffusion-model-based reconstruction for quantitative MRI at 0.5 mm isotropic resolution from scans of less than 5 minutes, which has eightfold smaller voxel volume than clinical protocols. The key idea is to apply the prior to the synthesized contrast images rather than to the subspace coefficients. The prior is then independent of the temporal basis and the sequence, which improves generalization and allows a 2-D multiscale energy (MuSE) model trained on grayscale CelebA face images to be used without any MRI training data. The explicit energy makes reconstruction MAP estimation by a step-size-free majorize-minimize algorithm that regularizes a rotating subset of contrasts along different axes, which makes 3-D multi-contrast reconstruction tractable. For 3-D radial MPnRAGE at 7 T in 4.3 min, MuSE reduced the phantom NRMSE from 0.133 (CG-SENSE) and 0.091 (L1-wavelet) to 0.074 and the WM T1 interquartile range from 117 and 47 ms to 38 ms, and it was demonstrated in vivo at 7 T.

0.24

Training-Free Motion-Corrected 3D Brain MRI Reconstruction via Event-Guided References

cs.CV ·2026-11-01 ·Wei Jiang, Hongli Chen, Tianyi Ding, Feng Liu, Nan Ye +1 more
code/data similarity 0.65
Abstract

Head motion causes artefacts in 3D brain MRI and can bias structural measurements. Model-based correction alternates between image and motion estimation, but artefacts in the reconstructed reference can reinforce motion errors. Learned priors can reduce these artefacts but require pretraining and may add test-time computation. We propose EventMC (event-guided motion correction), a training-free method for interleaved Cartesian acquisitions. It iteratively detects motion events and estimates motion using reference images reconstructed from stable intervals. It then refines the image and motion from low to high spatial frequencies. In a fair comparison in simulation (same reconstruction), EventMC's motion estimates improve mean PSNR by up to 4.0 dB over those of the learning-based MotionTTT. A single run of EventMC is up to 7.5 times faster than MotionTTT with less than half the GPU memory. On in vivo motion-corrupted scans, EventMC achieves similar mean image quality to MotionTTT without any training. Code is available at https://github.com/bayesbreeze/EventMC.

0.36

MOSAIC: A Self-supervised Dynamic Multi-encoding Reconstruction Framework for 3D Late Gadolinium Enhancement MRI

eess.IV ·2026-08-19 ·Muhammad A. Sultan, Yingmin Liu, Katherine Binzel, Katarzyna E. Gil, Karolina M. Zareba +1 more
similarity 0.65
Abstract

Purpose: To develop and evaluate a self-supervised dynamic reconstruction framework for highly undersampled dual-echo three-dimensional late gadolinium enhancement (3D LGE) MRI. Methods: MOSAIC jointly models multi-echo image content, coil sensitivity maps, and beat-specific nonrigid motion directly from acquired undersampled data, without requiring fully sampled training datasets or accurate precomputed sensitivity maps. Unlike existing methods that bin the acquired data into different motion states, with or without motion compensation, MOSAIC reconstructs a motion-resolved 3D LGE image from each heartbeat. The method was evaluated using digital phantoms with simulated myocardial scars and in vivo animal and human studies. Results: In phantom experiments, MOSAIC achieved higher peak signal-to-noise ratio and structural similarity index measure than low-rank deep image prior reconstruction and ablation variants of MOSAIC. In animal and human studies, MOSAIC achieved higher blinded expert image-quality scores than inline image-navigated compressed-sensing and low-rank deep image prior reconstructions. Conclusion: MOSAIC demonstrated the feasibility of motion-resolved free-breathing dual-echo 3D LGE MRI at acceleration factors exceeding 1,000, with improved detail preservation and artifact suppression relative to the state-of-the-art comparison methods.

0.36

FlowMoDL: Model-Based Deep Learning with Conjugate-Gradient Data Consistency for Highly Accelerated 4D Flow MRI Reconstruction

cs.CV cs.LG ·2026-08-26 ·Tristan Gottwald, Michelle Bruch, Mubashir-Ul Hassan, Fatma Alickovic, Milan Kloiber +4 more
similarity 0.65
Abstract

We present FlowMoDL, an unrolled neural network for highly accelerated 4D flow MRI reconstruction that directly optimizes for both anatomical magnitude and phase-derived velocity accuracy. Building on the MoDL framework, FlowMoDL alternates a learned (3+1)D spatiotemporal denoiser with conjugate-gradient data-consistency updates based on the SENSE forward model. A novel dual-pathway conditioning scheme adapts the denoiser features and data-consistency weighting, enabling a single model to handle varying acceleration factors ($10\times$ to $50\times$). To ensure physiological accuracy, the network is trained using a deep-supervision composite loss that explicitly penalizes velocity magnitude and angular errors, stabilized by a curriculum schedule. We evaluate FlowMoDL on the multi-center CMRx4DFlow dataset against classical and deep-learning baselines (CG-SENSE, MoDL, FlowVN, and FlowMRI-Net). A key advantage of FlowMoDL is its superior gradient step efficiency. When evaluated under an equivalent, limited budget of gradient steps, competing flow-specific networks degrade significantly. In contrast, FlowMoDL robustly converges and strictly outperforms all competitors across all acceleration factors in magnitude SSIM, nRMSE, relative velocity error, and angular error, successfully recovering sharp structural details and temporally coherent velocity fields.

0.09

MoSAIK: Motion Search with Artificial Intelligence in K-space

cs.AI physics.med-ph ·2026-10-19 ·Oscar Dabrowski, Abijith Jagannath, Frédéric Grouiller, Sébastien Courvoisier
similarity 0.65
Abstract

Object: Motion artefacts remain a major limitation of routine two-dimensional T2-weighted turbo spin-echo (TSE) brain MRI, reducing diagnostic confidence and often necessitating repeated acquisitions. MoSAIK (Motion Search with AI in K-space) retrospectively estimates in-plane rigid-body motion directly from standard multi-coil TSE k-space. Without a motion-free reference, external navigators, or additional acquisitions, MoSAIK infers motion from k-space self-consistency violations and incorporates predicted parameters into model-based, motion-aware reconstruction rather than synthesizing corrected images. Materials and Methods: Selected complex coil data were represented as log-magnitude and phase channels. A decoupled convolutional embedding and Transformer encoder estimated phase-encoding-line-indexed rotation and two-dimensional translations. Training used motion-free clinical acquisitions with simulated motion; four held-out subjects were evaluated. Component-wise errors assessed estimation; SSIM and PSNR in silico and entropy focus criterion (EFC) in vivo assessed reconstruction. Results: Motion estimation errors were low for both rotation (MSE $<$ 0.2 deg$^2$) and translation (MSE $<$ 0.2 pixels$^2$), with Pearson correlations of approximately 0.96 and 0.94, respectively. In every in-vivo acquisition, more than 75% of slices showed information gain relative to uncorrected reconstructions. MoSAIK reduced visible artefacts while preserving anatomical detail. Discussion: Transformer-based k-space motion-trajectory estimation enables fast retrospective correction of standard clinical TSE brain MRI.

0.21

Beyond Training from Scratch: Foundation Models for Data-Efficient and Generalizable Cardiac MRI Reconstruction

cs.CV ·2026-10-06 ·Anam Hashmi, Mayug Maniparambil, Julia Dietlmeier, Kathleen M. Curran, Noel E. O'Connor
publishedcode/data similarity 0.64
Abstract · Accepted at ECCVW 2026

Cardiac magnetic resonance imaging reconstruction aims to recover high-quality images from undersampled acquisitions, enabling faster scans while preserving diagnostic fidelity. Recent reconstruction methods are typically trained from scratch and often require large amounts of task-specific data, limiting their robustness under data scarcity and distribution shifts. In this work, we investigate whether pretrained vision foundation models can serve as effective priors for accelerated cardiac MRI reconstruction. We propose a reconstruction framework that integrates frozen and parameter-efficiently adapted visual encoders, including CLIP, BiomedCLIP, and DINOv2, within a transformer-based reconstruction architecture. Extensive experiments on the CMRxRecon2023 and CMRxRecon2024 benchmarks demonstrate that pretrained representations consistently outperform a transformer trained from scratch across multiple acceleration factors. We further evaluate performance under limited supervision and cross-dataset transfer, showing that foundation models provide superior data efficiency and generalization. While frozen representations are particularly effective in extreme low-data regimes, Low-Rank Adaptation (LoRA) yields additional gains when moderate amounts of training data are available. Among the evaluated backbones, DINOv2 achieves the strongest overall performance. These findings highlight the potential of vision foundation models as robust and transferable priors for cardiac MRI reconstruction.

0.31