Neural implants and human safety: single-fault detection for DC-coupled recording front ends

Dimitris Antoniadis, Timothy Constandinou

eess.SP cs.HC ·submitted 2026-08-11 01:41 UTC · v1

Abstract

DC-coupled analogue front ends (AFEs) for neural implants provide a low-area solution. However, removing the coupling capacitor eliminates the intrinsic barrier that protects cortical tissue: a single-fault event, such as gate-oxide breakdown of a low-noise amplifier (LNA) input transistor, can open a direct DC path from the supply rail into the brain. On the stimulation side this hazard is well understood, and single-fault tolerance is enforced by a series DC-blocking capacitor; on the recording side, DC-coupled front ends discard the equivalent safeguard, yet their protection has gone almost unexamined. This paper presents a single-fault detection mechanism that monitors the LNA for the DC imbalance produced by such a failure and disables the amplifier before the resulting fault current can irreversibly damage tissue. The imbalance is encoded in the duty cycle of a current-starved relaxation oscillator and read out as a time-to-digital measurement. Designed in 65 nm, the mechanism resolves a worst-case fault of 6.4 nA across all corners within 0.81 ms - compliant with the ISO~14708-3 limit for an 8533 um2 electrode - opening a broader discussion of safety in DC-coupled recording.

Comment
Draft IEEE 4 + 1

Closest papers

Nearest neighbours by meaning across the whole index.

HDR-DTCP: High-Dynamic-Range Differential Tissue-Coupled Wireless Powering for Tunable Biphasic Stimulation in Injectable Implants

eess.SP eess.SY ·2026-11-14 ·Asif Iftekhar Omi, Anyu Jiang, Han Wu, Sultan Mahmud, Mali Halac +4 more
similarity 0.49
Abstract

Next-generation implantable neurostimulators require efficient wireless power delivery in smaller, less invasive form factors, yet variations in tissue coupling, implantation conditions, and stimulation loads can alter received power (or requirements) and degrade energy-harvesting performance. This work presents high-dynamic-range (HDR) differential tissue-coupled powering (DTCP), termed HDR-DTCP, a body-coupled architecture in which noninvasive differential electrodes establish an AC electric field in tissue and a thread-like implant extracts power from the longitudinal potential gradient for on-chip harvesting and stimulation. Distributed coupling along the implant length, rather than receiver cross-section, enables minimally invasive injectable geometries, with harvested power scaling with implant length. A 65-nm complementary metal-oxide-semiconductor (CMOS) prototype integrates a reconfigurable four-stage energy harvester with automatic stage selection and a tunable constant-current biphasic stimulator, achieving 63.5\% peak power-conversion efficiency (PCE) at $-16.3$ dBm and $>50\%$ PCE over a 9-dB input-power range. Preliminary in-vivo validation of DTCP was done with a second application-specific integrated circuit (ASIC) on an injectable flexible substrate, capable of being delivered through a 14-gauge needle adjacent to the sciatic nerve of Sprague-Dawley rats (n=4). External copper-pad DTCP powering evoked repeatable hind-limb twitches and electromyography (EMG) response; ultrasound confirmed sub-millimeter nerve proximity, video-based paw-tip tracking captured the response, and 8-day post-implant dissection verified placement along the sciatic-nerve trajectory. Overall, this work provides an integrated wireless powering, energy-harvesting, and stimulation platform toward fully battery-free injectable neuromodulation implants.

0.21

NORA: Neuro-headstage for Online Recording and Actuation in Freely Behaving Animal Models

eess.SP ·2026-11-11 ·Savanna Blade, Yuhan Hou, He Jia, Laura Kondrataviciute, Runtian Yang +2 more
published similarity 0.48
Abstract · 5 pages, 7 figures. Accepted for publication in IEEE BioCAS 2026. Author accepted manuscript

This paper presents NORA, a compact wireless neuro-headstage designed for simultaneous neural recording and stimulation in freely behaving animal models. Separate recording and stimulation interfaces, each managed with a dedicated MCU and wireless module, are integrated into a fully head-mounted device compatible with standard Omnetics connectors; this separation allows for simultaneous recording and stimulation without sample-dropping or stimulation waveform distortion. To minimize stimulation artifacts and cross-domain interference, the recording and stimulation subsystems are electrically isolated with independent power domains and dedicated wireless communication paths. NORA has been validated through in vivo animal experiments, demonstrating acquisition of meaningful neural features with results comparable to a commercial wireless headstage. Together, these features establish a practical platform for artifact-reduced neuromodulation research.

0.31

Interplay between Excitability and Noise in Analog Spiking Neurons

eess.SY q-bio.NCcond-mat.stat-mech ·2026-10-05 ·Léopold Van Brandt, Alon Ascoli, Michele Bonnin, Grégoire Brandsteert, Denis Flandre +1 more
publishedtheory similarity 0.48
Abstract · Accepted for presentation at IEEE CDC 2026

A spiking neuron is a dynamical system exhibiting a limit cycle (a spike) when subject to sufficiently excitatory input. Famous neuroscience experiments by Bryant and Segundo as well as Mainen and Sejnowski revealed that the spike times of some biological neurons are more reliable when subject to time-varying stimuli compared to constant ones. Provided with an industrial physics-based transient noise SPICE simulation framework compatible with foundry transistor compact models, we have demonstrated similar behaviours for a CMOS analog spiking neuron designed for neuromorphic computing hardware platforms. Under constant excitation current, the neuron operates in oscillatory regime and the empirical periods (interspike intervals) suffer from accumulated phase noise (jitter) during successive cycles, mainly due to the thermal noise of transistors. A time-varying input is reported to be capable of triggering output spikes according to a controlled statetransition mechanism, thereby significantly reducing the spike time jitter in the excitability regime. The physical limitation of the oscillatory or rate-coding regime in terms of reliabilitydissipation tradeoff is further evidenced by a stochastic thermodynamic uncertainty relation, a theorem that applies to general overdamped dissipative stochastic dynamical systems.

0.49

Single-visit screening for intracochlear tissue growth validated against multi-visit electrical measurement changes

eess.SP ·2026-11-02 ·Minh Q. Vu, Erin L. Bratu, Jared A. Rybarczyk, Jack H. Noble
similarity 0.47
Abstract

Purpose: Tissue growth (TG) around cochlear implant electrodes raises impedance but is not reliably visible on CT scans. Routinely recorded stimulus current-induced non-stimulating electrode voltages (SCINSEVs) carry this information, but a single recording has no patient-specific reference. We present an algorithm for detecting signs of significant TG from a single SCINSEV measurement. Methods: TG is assumed to concentrate basally near the electrode insertion site, leaving the apical region mostly unaffected. A patient-specific forward model built from CT is fitted to the apical measurements and predicts the basal ones. The misfit score, T, measures each basal contact's signed relative misfit against the apical misfit, accounting for apical fit quality and measurement noise. An ear is flagged when two adjacent basal contacts cross a predetermined threshold. We compare T with T-delta, a model-free reference, tracking basal relative to apical SCINSEV change since the intra-operative recording, persisting over the last two visits. Results: T had a false-positive rate of 3% on growth-free ears. It agreed with T-delta in 81% of 68 ears (sensitivity 71%; specificity 86%; p = 6e-6). Over 200 patient-level two-fold splits, median held-out agreement was 78% at a false-positive rate of 5.1%. Conclusion: An apically fitted patient-specific model predicts basal SCINSEVs well enough to flag signs of significant TG at a single visit, without an intra-operative baseline. Detections largely agree with persistent multi-visit basal-versus-apical change, a surrogate ground truth for significant TG.

0.43

Analog Pin Directionality as an Exfiltration Attack Surface in Mixed-Signal ICs

cs.CR cs.AR ·2026-09-16 ·Ramana Ranganatham, Chirag Adiga, Michael Zuzak, Tejasvi Das
similarity 0.46
Abstract · Submitted to IEEE Transactions on Information Forensics and Security (TIFS)

Mixed-signal SoCs rely on nominally input-only analog pins to acquire off-chip signals, but the directionality of these interfaces is generally treated as a functional property rather than explicitly verified as a security property. This work identifies and experimentally demonstrates a directionality-based class of analog and mixed-signal (AMS) exfiltration attacks in which data-dependent circuit-offset modulation converts a nominally input-only pin into an outbound information channel. We analytically model the attack mechanism and identify three enabling host conditions: a closed-loop amplifier, an exposed amplifier input, and sufficiently high impedance at that pin. This attack class is validated through a representative silicon case study using a photoplethysmography (PPG) analog front-end (AFE) fabricated in a commercial 55-nm CMOS process. The payload incurs $<$0.001\% area overhead relative to typical biosensing AFEs. Under the evaluated conditions, payload activation reduces the filtered PPG-output SNR by only 0.03~dB, while the maximum HT-induced perturbation of 5.9\% of the PPG amplitude remains within the 34.3\% benign variation at the exposed sensor-input pin across process and temperature. The raw exfiltration SINR remains below -20~dB, while targeted filtering increases it above 14~dB and enables signal recovery. Silicon measurements demonstrate data exfiltration through the input pin at bit rates up to 10~kbps and error-free recovery of a PRBS message. These results expose a conventional test-observability gap and establish analog pin directionality as an AMS security property requiring explicit verification, test coverage, and defense rather than being inferred from nominal signal flow.

0.30

A Voltage-controlled MTJ-CMOS Neuron Emulating Tunable Izhikevich-Inspired Dynamics

cs.ET cs.ARcs.NE ·2026-09-25 ·Kayode Oluwaseyi Adebunmi, Jordan Athas, Allison Fleming, Hamed Poursiami, Giorgio A. Ascoli +3 more
similarity 0.46
Abstract

Biological neurons exhibit diverse firing dynamics that enable adaptive and stimulus-dependent signalling, yet reproducing these dynamics in hardware has remained an enduring challenge. In this work, we present an Izhikevich- inspired reconfigurable neuron that co-designs voltage-controlled magnetic tunnel junction (V-MTJ) dynamics with CMOS circuitry. The proposed architecture combines V-MTJ excitability dynamics, enabled by a tunable energy landscape, with CMOS recovery dynamics to generate five distinct neuronal firing pat- terns with different spiking, bursting and response characteristics. Our results, based on measured V-MTJ characteristics and circuit simulations using com- mercial GlobalFoundries 22-nm FD-SOI CMOS technology, show an average energy consumption of 145.44 fJ per spike. Algorithmic simulations further show that these firing dynamics reduce inference spike activity by up to 88.6% while maintaining baseline classification accuracy. These results highlight the potential of V-MTJ/CMOS reconfigurable neurons to reduce computational activity and enable compact, energy-efficient brain-inspired computing systems.

0.09

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

eess.SY ·2026-09-05 ·Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy
benchmark similarity 0.44
Abstract · 33 pages, 4 figures, 2 Tables

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

0.34

EarLog: Opportunistic Ear Physiology Sensing toward Personalized Hearing Risk Awareness

eess.AS cs.LGcs.SDeess.SP ·2026-09-30 ·Yongzhi Huang, Jiayi Zhao, Kaishun Wu
top venuebenchmark similarity 0.43
Abstract · 22 pages. Accepted at ACM MobiCom 2026. Includes appendix

Many headphone users risk irreversible hearing damage during high-load listening, yet current protection still relies on sound-pressure limits. Clinical tests detect damage after it becomes measurable, leaving the risk-accumulating process unobserved. The acoustic reflex (AR), an involuntary response to high acoustic load, marks when the ear enters a protective state. AR traces are weak, and existing measurements require dedicated probes and controlled conditions that are difficult to maintain during everyday audio. We present EarLog, which turns playback, the dominant interference, into an opportunistic probe. EarLog introduces content-conditioned residual sensing to suppress variations caused by nonstationary content, individual ear acoustics, and device response. It identifies weak AR through loudness-dependent changes and localized spectral irregularities in the residual. A resonance-constrained sparse finite impulse response (FIR) model captures the short-window response. Dual-microphone observations, mild-distortion recovery, and cross-session reference updates keep detection usable under low signal-to-noise ratio (SNR), fit changes, and environmental drift. EarLog runs in real time on an embedded controller using a low-cost dual-microphone add-on. We evaluate it with 81 participants, 16 headphone models, and more than 400 hours of recordings. EarLog achieves 99.0% sensitivity and 98.7% specificity on a clinically referenced benchmark. EarLog records listener-specific AR events during everyday listening, providing data for future longitudinal studies of hearing risk.

0.45