CS Paper Radar

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy

eess.SY ·submitted 2026-09-05 00:43 UTC · v1
benchmark

Abstract

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

Comment
33 pages, 4 figures, 2 Tables

Closest papers

Nearest neighbours by meaning across the whole index.

Short-Range Microwave Interprobe Coupling in a Brain-Mimicking Phantom and Ex-Vivo Brain

eess.SP ·2026-10-12 ·Pouya Faeghi, Xiaoting Jia, Jeffrey S. Walling
similarity 0.62
Abstract · 8 pages, 10 figures, 3 tables

Objectives: Short-range wireless links (microwave range) between closely spaced devices in brain tissue are relevant to neural interfaces and intracranial sensing, yet the change in coupling between two such devices caused by a finite volume of brain-like tissue has rarely been measured directly. Technology or Method: This work characterizes a printed, microstrip-fed, parasitic-strip-loaded tapered-slot coupling probe in which only a partially inserted 12-mm wire enters the medium, while the printed structure remains outside. Results: Two fabricated probes showed a measured free-space -10-dB reflection band of 5.78 to 6.86 GHz and a fractional bandwidth of approximately 17.1%, consistent with full-wave simulation. In a controlled two-probe experiment, a homogeneous agar-gelatin-saline brain-mimicking phantom was introduced and removed while the probe arrangement and measurement conditions were retained. Introducing the phantom increased the measured transmission coefficient near 6.3 GHz by approximately 16 dB, and this differential was reproduced across repeated trials. More than five measurements in ex vivo goat brain followed the phantom transmission response within approximately 1.5 dB across the 5.78-6.86-GHz band. Time-domain analysis of the phantom-present link gave a system root-mean-square delay spread of approximately 210 ps. Full-wave dosimetry predicted a peak 1-g-averaged specific absorption rate of 1.05 W/kg for approximately 0.49 W accepted power. Conclusions and Clinical or Biological Impact: These results provide an experimentally validated reference for short-range probe coupling in brain-like media and support phantom-based bench evaluation of minimally invasive intracranial probes before in vivo studies.

0.23

HDR-DTCP: High-Dynamic-Range Differential Tissue-Coupled Wireless Powering for Tunable Biphasic Stimulation in Injectable Implants

eess.SP eess.SY ·2026-11-14 ·Asif Iftekhar Omi, Anyu Jiang, Han Wu, Sultan Mahmud, Mali Halac +4 more
similarity 0.62
Abstract

Next-generation implantable neurostimulators require efficient wireless power delivery in smaller, less invasive form factors, yet variations in tissue coupling, implantation conditions, and stimulation loads can alter received power (or requirements) and degrade energy-harvesting performance. This work presents high-dynamic-range (HDR) differential tissue-coupled powering (DTCP), termed HDR-DTCP, a body-coupled architecture in which noninvasive differential electrodes establish an AC electric field in tissue and a thread-like implant extracts power from the longitudinal potential gradient for on-chip harvesting and stimulation. Distributed coupling along the implant length, rather than receiver cross-section, enables minimally invasive injectable geometries, with harvested power scaling with implant length. A 65-nm complementary metal-oxide-semiconductor (CMOS) prototype integrates a reconfigurable four-stage energy harvester with automatic stage selection and a tunable constant-current biphasic stimulator, achieving 63.5\% peak power-conversion efficiency (PCE) at $-16.3$ dBm and $>50\%$ PCE over a 9-dB input-power range. Preliminary in-vivo validation of DTCP was done with a second application-specific integrated circuit (ASIC) on an injectable flexible substrate, capable of being delivered through a 14-gauge needle adjacent to the sciatic nerve of Sprague-Dawley rats (n=4). External copper-pad DTCP powering evoked repeatable hind-limb twitches and electromyography (EMG) response; ultrasound confirmed sub-millimeter nerve proximity, video-based paw-tip tracking captured the response, and 8-day post-implant dissection verified placement along the sciatic-nerve trajectory. Overall, this work provides an integrated wireless powering, energy-harvesting, and stimulation platform toward fully battery-free injectable neuromodulation implants.

0.21

An Implant-to-Wearable IR-UWB Transmitter-Receiver Architecture and Layered Protocol for High-Density Brain-Computer Interfaces

eess.SP cs.AI ·2026-09-01 ·G. D. Su, Y. Mo
similarity 0.62
Abstract · 18 pages,14 figures

High-rate neural telemetry requires an implant-towearable link whose circuit partition, waveform mapping, packet processing, and failure semantics are mutually consistent. This paper specifies a chip-oriented, one-way impulse-radio ultrawideband (IR-UWB) uplink comprising an implanted packet engine and free-running transmitter, a tissue-proxy channel, and a wearable mixed-signal receiver and digital baseband.

0.27

NORA: Neuro-headstage for Online Recording and Actuation in Freely Behaving Animal Models

eess.SP ·2026-11-11 ·Savanna Blade, Yuhan Hou, He Jia, Laura Kondrataviciute, Runtian Yang +2 more
published similarity 0.62
Abstract · 5 pages, 7 figures. Accepted for publication in IEEE BioCAS 2026. Author accepted manuscript

This paper presents NORA, a compact wireless neuro-headstage designed for simultaneous neural recording and stimulation in freely behaving animal models. Separate recording and stimulation interfaces, each managed with a dedicated MCU and wireless module, are integrated into a fully head-mounted device compatible with standard Omnetics connectors; this separation allows for simultaneous recording and stimulation without sample-dropping or stimulation waveform distortion. To minimize stimulation artifacts and cross-domain interference, the recording and stimulation subsystems are electrically isolated with independent power domains and dedicated wireless communication paths. NORA has been validated through in vivo animal experiments, demonstrating acquisition of meaningful neural features with results comparable to a commercial wireless headstage. Together, these features establish a practical platform for artifact-reduced neuromodulation research.

0.31

Data-Centric Neuromotor Interfaces for Portable Human-Machine Interaction

cs.RO ·2026-08-31 ·Jiaxuan Li, Di Wu, Jianhua Liu, Yuxin Zhao, Jinnuo Li +5 more
publishedbenchmark similarity 0.57
Abstract

Dexterous human-machine interaction requires intuitive and expressive interfaces that can be efficiently deployed on constrained edge devices. Flexible material-based neuromotor interfaces hold considerable promise, as they decode human movement intention into natural control. Although emerging flexible electronic skins enable wearable high-fidelity data acquisition, practical deployment inevitably involves trade-offs between computational resources and portability. We present a data-centric paradigm where physiological features yield fundamental separability, providing sufficient discriminative cues for recognition. A wireless, high-bandwidth system developed for collecting various electrophysiological signals, when integrated with muscle-specific electrodes, forms a surface electromyography-based interface. Exploiting highly separable data, a 2,210-parameter model achieves 94.36% accuracy across 34 gestures and can be rapidly deployed on edge devices, establishing a new thousand-parameter benchmark for dexterous decoding. The underlying data-algorithm interactions in the data-centric paradigm are further clarified, demonstrating its feasibility in real-world scenarios. This study provides a principled and validated pathway for practical deployment of reliable neuromotor interfaces.

0.22

Towards an Earpiece for Wearable and Dry Ultrasonic Vagus Nerve Stimulation

eess.SY ·2026-10-05 ·Sebastian Frey, Farouk Ghezraoui, Marc Nölke, Marco Giordano, Luca Benini +1 more
unusual similarity 0.52
Abstract · 4 pages, 3 figures

Ultrasound (US)-based auricular vagus nerve stimulation (aVNS) is emerging as a versatile, noninvasive modality for neuromodulation - spanning clinical therapies (e.g., pain management, depression) and general well-being applications. However, most existing US-based aVNS systems remain bulky and require wet coupling media (e.g., ultrasound gel) to achieve both spatially localized stimulation and efficient acoustic transmission. We present a wearable US stimulation system combining a personalized, dry-coupled earpiece with a battery-powered, wireless US probe. The earpiece integrates a 2 MHz piezoelectric transducer fully encapsulated in Humimic material designed for anatomical fit and stable coupling without externally applied gel. Hydrophone measurements of the fully encapsulated earpiece in water give a peak rarefactional pressure of 0.24 MPa at 15 mm for a 15 V unipolar excitation. We further demonstrate in vivo feasibility through fully dry, in-ear 2 MHz pulse-echo acquisitions under realistic wearing conditions, confirming effective US transmission without the need for coupling gel. These results demonstrate the integration of a personalized dry acoustic interface with battery-powered stimulation electronics, providing a wearable platform for investigating US-based aVNS. Future studies will focus on functional outcomes, including heart rate variability and pupillary responses during stimulation.

0.27

Streaming P300 Acquisition and Statistical Signal Validation Across Five EEG Platforms: A Hardware-Agnostic BrainFlow/LSL Pipeline

cs.HC ·2026-09-09 ·Isabella Guan, Rui Liu, Fusheng Wang
similarity 0.51
Abstract

P300 spellers offer people with severe motor impairment, such as ALS, an effective communication channel and remain one of the most established surgery-free alternatives to intracortical interfaces. Advanced language models have made spellers faster and more robust, yet the hardware beneath them is under-studied. We present a hardware-agnostic, real-time P300 acquisition pipeline built on BrainFlow and Lab Streaming Layer (LSL) that runs unchanged across consumer- and research-grade EEG headsets, with permutation tests of signal separability. Using a standard 6 x 6 row/column paradigm, we piloted five configurations: a custom dry system, a custom wet/gel system, Emotiv Flex, Emotiv EPOC X, and Muse 2. The custom systems and EPOC X showed weak or inconsistent signal separability, Muse 2 had the highest acquisition reliability despite limited centro-parietal coverage, and Flex showed the most promising signal. In 20 further Flex sessions varying subject, timing, and phrase length (131 target characters), a peak-amplitude permutation test and a cross-validated xDAWN decoder both detected a significant target response under two channel-exclusion policies, with decoder AUC reaching about 0.72 after 15 repetitions. Character accuracy depended heavily on evaluation methodology: in-sample majority voting reached 94.7%, whereas character-held-out accuracy was 31.3% with evidence accumulated across repetitions, about three times that of held-out majority voting. These analyses indicate that Flex captured a detectable, if still weak, P300 under the studied conditions, while broader participant-level validation and improved decoding remain necessary.

0.32

A Miniaturized Wi-Fi HaLow-Based Wireless Imaging Transfer System for Capsule Endoscopy

eess.IV ·2026-10-09 ·Michele Bertozzi, Lorenzo Amelio, Sebastiano Strangio, Nikita J. Greenidge, Benjamin Calmé +3 more
similarity 0.51
Abstract

Wireless capsule endoscopy is constrained by severe limitations in size, available power, and reliable data transmission through biological tissues. Although recent advances in miniature imaging sensors have improved image quality, the wireless transmission of color images at practical frame rates remains a challenging task under the tight capsule constraints. We present the design and the experimental validation of a miniaturized Wireless Imaging Transfer (WIT) system for capsule endoscopy based on a sub-GHz IEEE 802.11ah (Wi-Fi HaLow) wireless link. In addition, a preliminary in-vivo validation of Wi-Fi HaLow communication is reported, showing that a stable through-body wireless link can be established across biological tissues in a clinically relevant porcine model. These in-vivo results provide direct experimental support for the adoption of sub-GHz Wi-Fi HaLow for capsule endoscopy applications. Experimental results demonstrate stable wireless image transmission at a resolution of 320x320 pixels, with measured end-to-end data rates above 12 Mbps and a transmission energy efficiency of 20.7 nJ/bit. We also report a detailed power consumption analysis, with a breakdown of the contributions from the imaging sensor, illumination, digital processing, and RF transmission. When all active subsystem contributions are considered, including the camera, digital processing, RF transmission, and illumination at 60% LED brightness, the total power consumption is approximately 477 mW, corresponding to a complete-chain energy consumption of approximately 39.6 nJ/bit. Overall, the measured performance confirms that sub-GHz Wi-Fi HaLow provides a promising trade-off among throughput, robustness, and energy efficiency for wireless imaging in capsule endoscopy.

0.22