A Miniaturized Wi-Fi HaLow-Based Wireless Imaging Transfer System for Capsule Endoscopy

Michele Bertozzi, Lorenzo Amelio, Sebastiano Strangio, Nikita J. Greenidge, Benjamin Calmé, Pietro Valdastri, Giuseppe Iannaccone Fellow, Alessandro Catania

eess.IV ·submitted 2026-10-09 16:27 UTC · v1

Abstract

Wireless capsule endoscopy is constrained by severe limitations in size, available power, and reliable data transmission through biological tissues. Although recent advances in miniature imaging sensors have improved image quality, the wireless transmission of color images at practical frame rates remains a challenging task under the tight capsule constraints. We present the design and the experimental validation of a miniaturized Wireless Imaging Transfer (WIT) system for capsule endoscopy based on a sub-GHz IEEE 802.11ah (Wi-Fi HaLow) wireless link. In addition, a preliminary in-vivo validation of Wi-Fi HaLow communication is reported, showing that a stable through-body wireless link can be established across biological tissues in a clinically relevant porcine model. These in-vivo results provide direct experimental support for the adoption of sub-GHz Wi-Fi HaLow for capsule endoscopy applications. Experimental results demonstrate stable wireless image transmission at a resolution of 320x320 pixels, with measured end-to-end data rates above 12 Mbps and a transmission energy efficiency of 20.7 nJ/bit. We also report a detailed power consumption analysis, with a breakdown of the contributions from the imaging sensor, illumination, digital processing, and RF transmission. When all active subsystem contributions are considered, including the camera, digital processing, RF transmission, and illumination at 60% LED brightness, the total power consumption is approximately 477 mW, corresponding to a complete-chain energy consumption of approximately 39.6 nJ/bit. Overall, the measured performance confirms that sub-GHz Wi-Fi HaLow provides a promising trade-off among throughput, robustness, and energy efficiency for wireless imaging in capsule endoscopy.

Closest papers

Nearest neighbours by meaning across the whole index.

Evaluating Wi-Fi HaLow for Next-Generation Capsule Endoscopy

eess.SY ·2026-10-25 ·Ziyao Zhou, Chen Shen, Tiancheng Cao, Hen-Wei Huang
similarity 0.83
Abstract

Capsule endoscopy is limited by its wireless link: insufficient throughput restricts image resolution and frame rate, whereas high-rate radios typically use higher carrier frequencies that experience greater attenuation through tissue. This work evaluates Wi-Fi HaLow (IEEE 802.11ah), a sub-GHz Wi-Fi standard, as an image link for capsule endoscopy. Across all tested combinations of modulation and coding scheme and bandwidth, the saturation TCP throughput reached 15.55 Mbps, and the fastest configuration also achieved the lowest energy per bit. Through power optimization strategies, the combined power consumption of the HaLow module and microcontroller at 1 frame per second (FPS) was reduced from 101.70 to 14.00 mW, which is comparable to that of Bluetooth Low Energy systems. In a simulated in vivo setting, the default rate control selected the highest tested MCS and bandwidth, indicating sufficient link quality; this configuration also had the lowest measured power in both the 1 FPS mode (13.59 mW) and the 30 FPS mode (174.21 mW). The capsule-to-receiver latency was 20.68 ms for a 24 KB frame, while the receiver-to-capsule latency was 4.77 ms for a command; this low-latency bidirectional communication supports real-time detection and manipulation for next-generation capsule endoscopy.

0.08

A Two-Element Capsule Antenna for Near-Field Polarization MIMO in Wireless Capsule Endoscopy

eess.SP ·2026-10-02 ·Pyry Kiviharju, Lauri Vähä-Savo, Clemens Icheln, Kenichi Takizawa, Katsuyuki Haneda
unusual similarity 0.62
Abstract · 12 pages, 21 figures. Submitted to IEEE Open Journal of Antennas and Propagation

Multiple-Input-Multiple-Output (MIMO) technology is considered to be a viable method for spectral efficiency and data rate enhancement in wireless communications. This article presents a two-element capsule antenna with orthogonal current distributions for polarization multiplexing in a wireless capsule endoscopy (WCE) application at 915 MHz. The proposed antenna was evaluated in a MIMO link where it served as the transmitter and an existing on-body antenna was the receiver. The performance was assessed in two distinct environments: an anatomically accurate human body model using full-wave (FW) simulations, and a homogeneous liquid tissue phantom, using both FW simulations and measurements. The former was used to evaluate the performance in a realistic environment and to demonstrate the superiority of the proposed capsule antenna against existing two-element antennas in the literature, while the latter aimed at verifying the antenna design by comparison of simulated and measured MIMO links. We studied channel capacity and eigenvalues for the comparisons. The MIMO link with the proposed antenna provided nearly doubled median channel capacity relative to a baseline single-input-single-output link, consistently in the simulations and measurements. Our results prove the suitability of the proposed antenna for polarization multiplexing and, in terms of channel capacity, establish it as the best conformal two-element capsule antenna design at 915 MHz.

0.31

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

eess.SY ·2026-09-05 ·Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy
benchmark similarity 0.51
Abstract · 33 pages, 4 figures, 2 Tables

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

0.34

Short-Range Microwave Interprobe Coupling in a Brain-Mimicking Phantom and Ex-Vivo Brain

eess.SP ·2026-10-12 ·Pouya Faeghi, Xiaoting Jia, Jeffrey S. Walling
similarity 0.50
Abstract · 8 pages, 10 figures, 3 tables

Objectives: Short-range wireless links (microwave range) between closely spaced devices in brain tissue are relevant to neural interfaces and intracranial sensing, yet the change in coupling between two such devices caused by a finite volume of brain-like tissue has rarely been measured directly. Technology or Method: This work characterizes a printed, microstrip-fed, parasitic-strip-loaded tapered-slot coupling probe in which only a partially inserted 12-mm wire enters the medium, while the printed structure remains outside. Results: Two fabricated probes showed a measured free-space -10-dB reflection band of 5.78 to 6.86 GHz and a fractional bandwidth of approximately 17.1%, consistent with full-wave simulation. In a controlled two-probe experiment, a homogeneous agar-gelatin-saline brain-mimicking phantom was introduced and removed while the probe arrangement and measurement conditions were retained. Introducing the phantom increased the measured transmission coefficient near 6.3 GHz by approximately 16 dB, and this differential was reproduced across repeated trials. More than five measurements in ex vivo goat brain followed the phantom transmission response within approximately 1.5 dB across the 5.78-6.86-GHz band. Time-domain analysis of the phantom-present link gave a system root-mean-square delay spread of approximately 210 ps. Full-wave dosimetry predicted a peak 1-g-averaged specific absorption rate of 1.05 W/kg for approximately 0.49 W accepted power. Conclusions and Clinical or Biological Impact: These results provide an experimentally validated reference for short-range probe coupling in brain-like media and support phantom-based bench evaluation of minimally invasive intracranial probes before in vivo studies.

0.23

CapsuleMotion: A Lightweight Real-Time Visual Motion Predictor for Capsule Endoscopy

cs.CV ·2026-09-14 ·Oliver Bause, Julia Werner, Oliver Bringmann
published similarity 0.49
Abstract · 4 pages, 3 figures, 2 tables, accepted at the 22nd Annual International Conference on Body Sensor Networks (BSN2026)

Video Capsule Endoscopy (VCE) is a non-invasive medical examination that allows for the observation of the small intestine, which is otherwise difficult to access. A fundamental challenge persists in the form of their limited size in order to still be swallowable. The resulting restricted battery capacity, however, contradicts with the power-intensive nature of image capture and transmission. Therefore, we propose CapsuleMotion, a patient-specific dynamic capsule behavior that utilizes the available energy in a goal-oriented manner to increase the likelihood of a complete screening of the gastrointestinal tract. By investigating and combining metrics from the on-device image compression, CapsuleMotion predicts the motion between two successive frames. The camera's frame rate will be modified in accordance with the predicted magnitude of motion. Furthermore, prior to entering the small intestine, the capsule operates in a low power mode with a significantly reduced frame rate. In this mode, the LocalizationNet is employed to determine the current organ, provided that motion was predicted. The proposed framework is evaluated on the Rhode Island VCE dataset and deployed on an ultra-low power single-core RISC-V demonstrator with an integrated hardware accelerator. CapsuleMotion demonstrated the capability to reduce electric energy consumption by up to 20.66% in comparison with conventional capsules that lack a dynamic frame rate. Additionally, the accuracy of detecting the entry point of the small intestine has been improved.

0.41

An Implant-to-Wearable IR-UWB Transmitter-Receiver Architecture and Layered Protocol for High-Density Brain-Computer Interfaces

eess.SP cs.AI ·2026-09-01 ·G. D. Su, Y. Mo
similarity 0.45
Abstract · 18 pages,14 figures

High-rate neural telemetry requires an implant-towearable link whose circuit partition, waveform mapping, packet processing, and failure semantics are mutually consistent. This paper specifies a chip-oriented, one-way impulse-radio ultrawideband (IR-UWB) uplink comprising an implanted packet engine and free-running transmitter, a tissue-proxy channel, and a wearable mixed-signal receiver and digital baseband.

0.27

Sensitivity and Identifiability of Tumor-Related Changes in Longitudinal Microwave Breast Imaging

eess.SP physics.med-ph ·2026-10-19 ·Somayeh B. Shafiei, Ian Jeffrey, Vahab Khoshdel, Joe LoVetri
similarity 0.45
Abstract

Microwave imaging could monitor breast tumor response to neoadjuvant chemotherapy, but between-visit changes in breast positioning and normal tissue may confound tumor-related changes. We use eight patient-specific dielectric models from longitudinal MRI of two patients at four time points. For each model, 24-port S-parameters were simulated from 1 to 3 GHz, and Jacobians computed for tumor permittivity and loss, fibroglandular tissue (FGT) permittivity, and breast rotation and translation. 1 mm positioning changes produced changes 20 to 50 times larger than a 10% tumor permittivity change above 2 GHz. Nevertheless, tumor sensitivity patterns were weakly aligned with positioning patterns, tumor-FGT similarity exceeded 0.8 at only some visits, and the Jacobians remained well conditioned. Tumor responses are thus small but separable; monitoring will require positioning correction, frequency selection, and noise-aware detectability analysis.

0.23

Mobility Enhancement of Patients Body Monitoring based on WBAN with Multipath Routing

cs.NI ·2026-10-02 ·Yasna Ghanbari Birgani, Nastooh Taheri Javan, Mohsen Tourani
publishedunusualhighly cited similarity 0.45
Abstract

One of the promising applications of wireless sensor networks (WSNs) is monitoring of the human body for health concerns. For this purpose, a large number of small sensors are implanted in the human body. These sensors altogether provide a network of wireless sensors (WBANs) and monitor the vital signs and signals of the human body; these sensors will then send this information to the doctor. The most important application of the WBAN is the implementation of the monitoring network for patient safety in the hospital environment. In this case, supporting patients' mobility is one of the basic needs, which has been underestimated in recent studies. The problem that involves providing the required energy for the units used in this type of network is challenging; for this reason, sent/received units with very low power consumption and with a very small radius are used in order to save energy. The resulting small sending range leads to the lack of support for patients' mobility. In this paper, the ad hoc mode is suggested for use to establish a network and a multipath routing algorithm for the purpose of supporting patients' mobility in a hospital setting. The results of the simulation show that, in addition to supporting patients' mobility, the use of the proposed idea instead of previously presented protocols reduces delays in data transmission and energy consumption; it also increases the delivery rate depending on the destination and the lifetime of the network, while increasing routing overhead.

0.62