Short-Range Microwave Interprobe Coupling in a Brain-Mimicking Phantom and Ex-Vivo Brain

Pouya Faeghi, Xiaoting Jia, Jeffrey S. Walling

eess.SP ·submitted 2026-10-12 00:52 UTC · v2

Abstract

Objectives: Short-range wireless links (microwave range) between closely spaced devices in brain tissue are relevant to neural interfaces and intracranial sensing, yet the change in coupling between two such devices caused by a finite volume of brain-like tissue has rarely been measured directly. Technology or Method: This work characterizes a printed, microstrip-fed, parasitic-strip-loaded tapered-slot coupling probe in which only a partially inserted 12-mm wire enters the medium, while the printed structure remains outside. Results: Two fabricated probes showed a measured free-space -10-dB reflection band of 5.78 to 6.86 GHz and a fractional bandwidth of approximately 17.1%, consistent with full-wave simulation. In a controlled two-probe experiment, a homogeneous agar-gelatin-saline brain-mimicking phantom was introduced and removed while the probe arrangement and measurement conditions were retained. Introducing the phantom increased the measured transmission coefficient near 6.3 GHz by approximately 16 dB, and this differential was reproduced across repeated trials. More than five measurements in ex vivo goat brain followed the phantom transmission response within approximately 1.5 dB across the 5.78-6.86-GHz band. Time-domain analysis of the phantom-present link gave a system root-mean-square delay spread of approximately 210 ps. Full-wave dosimetry predicted a peak 1-g-averaged specific absorption rate of 1.05 W/kg for approximately 0.49 W accepted power. Conclusions and Clinical or Biological Impact: These results provide an experimentally validated reference for short-range probe coupling in brain-like media and support phantom-based bench evaluation of minimally invasive intracranial probes before in vivo studies.

Comment
8 pages, 10 figures, 3 tables

Closest papers

Nearest neighbours by meaning across the whole index.

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

eess.SY ·2026-09-05 ·Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy
benchmark similarity 0.62
Abstract · 33 pages, 4 figures, 2 Tables

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

0.34

Sensitivity and Identifiability of Tumor-Related Changes in Longitudinal Microwave Breast Imaging

eess.SP physics.med-ph ·2026-10-19 ·Somayeh B. Shafiei, Ian Jeffrey, Vahab Khoshdel, Joe LoVetri
similarity 0.57
Abstract

Microwave imaging could monitor breast tumor response to neoadjuvant chemotherapy, but between-visit changes in breast positioning and normal tissue may confound tumor-related changes. We use eight patient-specific dielectric models from longitudinal MRI of two patients at four time points. For each model, 24-port S-parameters were simulated from 1 to 3 GHz, and Jacobians computed for tumor permittivity and loss, fibroglandular tissue (FGT) permittivity, and breast rotation and translation. 1 mm positioning changes produced changes 20 to 50 times larger than a 10% tumor permittivity change above 2 GHz. Nevertheless, tumor sensitivity patterns were weakly aligned with positioning patterns, tumor-FGT similarity exceeded 0.8 at only some visits, and the Jacobians remained well conditioned. Tumor responses are thus small but separable; monitoring will require positioning correction, frequency selection, and noise-aware detectability analysis.

0.23

Dielectric Models of Normal and Diseased Human Trabecular Bones at Microwave Frequencies

eess.SP physics.med-phphysics.bio-ph ·2026-11-03 ·Bilal Amin, Amra Mehboob, Eoghan Dunne, Ana González-Suárez, Finn Krewer +3 more
publishedhighly cited similarity 0.54
Abstract · 12 pages, 9 figures, 8 tables. Published in IEEE Open Journal of Antennas and Propagation (open access, CC BY 4.0)

The dielectric properties of biological tissues are used to investigate the interaction of electromagnetic (EM) fields with the human body. These properties are integrated into EM numerical simulations, such as the finite difference time domain (FDTD) method, using dielectric models such as Cole-Cole and Debye models. These computational tools are employed to study the propagation of EM waves within the human body. This paper has developed the one-pole and two-pole Cole-Cole and Debye models for the dielectric properties of normal and diseased human trabecular bone tissue at microwave frequencies (0.5-8.5 GHz). The dielectric properties of normal trabecular and cortical bone were derived based on the literature, while those of diseased human trabecular bones (osteoporotic and osteoarthritis) were acquired from a previous study conducted by the authors, which characterised the properties within the 0.5-8.5 GHz frequency range. This study has employed a weighted least squares method to optimise the one-pole and two-pole dielectric models' parameters. The accuracy of the developed dielectric models is validated by comparing the simulated results with the experimental data already published by the authors. The comparative analysis of the models demonstrated that the one-pole Cole-Cole model achieved a better fit compared to the one-pole Debye model across all bone types. Additionally, the two-pole Cole-Cole model exhibited a better fit than the two-pole Debye model. The proposed models can be used in numerical simulations to investigate the interaction of EM fields with bone tissue and develop and optimise microwave bone health monitoring and treatment techniques.

0.65

An Arduino Giga-Based Multi-Frequency Electrical Impedance Tomography System for Electrolyte Media: Hardware Design, Statistical Voltage Analysis, and System Performance Evaluation

eess.SP ·2026-10-11 ·Zahra Karimi, Natalie Verdugo, Joseph Schwab, Hamid Ghaednia
unusual similarity 0.53
Abstract

Multi-frequency electrical impedance tomography (EIT) often requires specialized hardware. This challenge is especially pronounced in biomedical applications. We developed a 16-electrode EIT system using an Arduino GIGA R1 WiFi for generation and acquisition, with a discrete current source, three analog multiplexers, and basic signal conditioning. Dual cores separate waveform timing from switching and acquisition. A 3D-printed tank with silver-wire electrodes was tested at 11 frequencies (1-180 kHz). Across the 10-180 kHz analysis band, median repeat-frame signal-to-noise ratio was 48 dB, acquisition rate was ~0.18 frames/s per frequency (~60 s per sweep), and cross-session baseline drift was ~6%. KCl phantoms contained an insulating ABS cylinder or KCl-filled dialysis bags (0.001-1.0 M) spanning positive and negative conductivity contrasts. Boundary voltages were analyzed without reconstruction to estimate inclusion position, contrast polarity, and relative magnitude. Leave-one-session-out matching located ABS inclusions in 29/30 runs at 160 kHz (87-97% across the band; chance 33%). For dialysis bags, polarity was correctly identified in all 36 nonzero-contrast runs. Signed voltage projections correlated with the cylindrical contrast factor (R^2=0.68 across runs; R^2=0.89 for the aggregate fit). Concentration was ordered (Spearman rho=0.85), despite saturation above ~10-fold contrast. Bag position accuracy ranged from 6/6 at the highest contrasts to 2/6 at matched conductivity. The matched-conductivity bag produced a ~1.7% RMS perturbation against 0.4-0.6% repeat-frame variability. These results demonstrate that microcontroller-based multi-frequency EIT can support selected detection and identification tasks through direct boundary-voltage analysis, supporting future studies of muscle electrochemistry.

0.25

Split-Ring Couplers for On-Skin Wi-Fi Communication via an SSPP Waveguide Textile

eess.SP physics.app-ph ·2026-11-09 ·Akihito Noda, Ryo Takahashi
similarity 0.51
Abstract · 4 pages, 6 figures

Textile spoof surface plasmon polariton (SSPP) waveguides offer guided on-body links, but their body-facing ground conductors obstruct coupling from on-skin wireless nodes. This letter presents a split-ring coupling interface using local ground recesses at the textile terminations, without patterned garment-side launchers. The capacitively loaded rings are supported by polydimethylsiloxane carriers with a thickness of 2~mm. An image-current model and full-wave simulations examine the effect of backing-ground spacing and determine the recess geometry. For a 450~mm textile waveguide on a water-filled body phantom, the measured $S_{21}$ at 2.45 GHz was $-$18.8~dB, versus $-$60.9~dB without the textile. A 625-position scan showed a maximum 5.1 dB reduction relative to the nominal position within the measured joint $\pm$5~mm placement region. A separate two-node IEEE~802.11n test demonstrated a 65~Mb/s physical-layer (PHY) rate with zero packet errors among 300 frames at $-$20~dBm transmit power through the coupler link. These results establish high-rate, placement-tolerant on-skin Wi-Fi communication through an SSPP waveguide textile.

0.34

An Implant-to-Wearable IR-UWB Transmitter-Receiver Architecture and Layered Protocol for High-Density Brain-Computer Interfaces

eess.SP cs.AI ·2026-09-01 ·G. D. Su, Y. Mo
similarity 0.51
Abstract · 18 pages,14 figures

High-rate neural telemetry requires an implant-towearable link whose circuit partition, waveform mapping, packet processing, and failure semantics are mutually consistent. This paper specifies a chip-oriented, one-way impulse-radio ultrawideband (IR-UWB) uplink comprising an implanted packet engine and free-running transmitter, a tissue-proxy channel, and a wearable mixed-signal receiver and digital baseband.

0.27

An Evidence-Guided Framework for Tissue-Barrier Impedance Review: Deletion and Resistance Sensitivity

eess.SP ·2026-10-23 ·Leo Yang, Yingtong Zhou
similarity 0.50
Abstract · 5 pages, 3 figures, 3 tables

Reviewing tissue-barrier impedance requires connecting spectral anomalies with resistance estimates and experimental context. We present an evidence-guided framework that coordinates numerical checks and traceable actions for human review. In-house simulations assess the effect of removing flagged frequencies, public datasets test model adaptation, and local measurements demonstrate the review workflow. Removing flagged frequencies reduced resistance error under quadratic objectives, whereas robust fitting showed little average benefit from deletion and a mean error increase for endpoint-pair perturbations. High errors clustered around a few fitted circuit parameter sets, highlighting the value of examining parameter identifiability. Under the tested noise and perturbation conditions, these findings support robust full-spectrum estimation, flag-guided inspection, and matched sensitivity checks.

0.26

A Two-Element Capsule Antenna for Near-Field Polarization MIMO in Wireless Capsule Endoscopy

eess.SP ·2026-10-02 ·Pyry Kiviharju, Lauri Vähä-Savo, Clemens Icheln, Kenichi Takizawa, Katsuyuki Haneda
unusual similarity 0.50
Abstract · 12 pages, 21 figures. Submitted to IEEE Open Journal of Antennas and Propagation

Multiple-Input-Multiple-Output (MIMO) technology is considered to be a viable method for spectral efficiency and data rate enhancement in wireless communications. This article presents a two-element capsule antenna with orthogonal current distributions for polarization multiplexing in a wireless capsule endoscopy (WCE) application at 915 MHz. The proposed antenna was evaluated in a MIMO link where it served as the transmitter and an existing on-body antenna was the receiver. The performance was assessed in two distinct environments: an anatomically accurate human body model using full-wave (FW) simulations, and a homogeneous liquid tissue phantom, using both FW simulations and measurements. The former was used to evaluate the performance in a realistic environment and to demonstrate the superiority of the proposed capsule antenna against existing two-element antennas in the literature, while the latter aimed at verifying the antenna design by comparison of simulated and measured MIMO links. We studied channel capacity and eigenvalues for the comparisons. The MIMO link with the proposed antenna provided nearly doubled median channel capacity relative to a baseline single-input-single-output link, consistently in the simulations and measurements. Our results prove the suitability of the proposed antenna for polarization multiplexing and, in terms of channel capacity, establish it as the best conformal two-element capsule antenna design at 915 MHz.

0.31