CS Paper Radar

NORA: Neuro-headstage for Online Recording and Actuation in Freely Behaving Animal Models

Savanna Blade, Yuhan Hou, He Jia, Laura Kondrataviciute, Runtian Yang, Suneil K. Kalia, Xilin Liu

eess.SP ·submitted 2026-11-11 19:32 UTC · v1
published

Abstract

This paper presents NORA, a compact wireless neuro-headstage designed for simultaneous neural recording and stimulation in freely behaving animal models. Separate recording and stimulation interfaces, each managed with a dedicated MCU and wireless module, are integrated into a fully head-mounted device compatible with standard Omnetics connectors; this separation allows for simultaneous recording and stimulation without sample-dropping or stimulation waveform distortion. To minimize stimulation artifacts and cross-domain interference, the recording and stimulation subsystems are electrically isolated with independent power domains and dedicated wireless communication paths. NORA has been validated through in vivo animal experiments, demonstrating acquisition of meaningful neural features with results comparable to a commercial wireless headstage. Together, these features establish a practical platform for artifact-reduced neuromodulation research.

Comment
5 pages, 7 figures. Accepted for publication in IEEE BioCAS 2026. Author accepted manuscript

Closest papers

Nearest neighbours by meaning across the whole index.

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

eess.SY ·2026-09-05 ·Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy
benchmark similarity 0.62
Abstract · 33 pages, 4 figures, 2 Tables

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

0.34

An Implant-to-Wearable IR-UWB Transmitter-Receiver Architecture and Layered Protocol for High-Density Brain-Computer Interfaces

eess.SP cs.AI ·2026-09-01 ·G. D. Su, Y. Mo
similarity 0.53
Abstract · 18 pages,14 figures

High-rate neural telemetry requires an implant-towearable link whose circuit partition, waveform mapping, packet processing, and failure semantics are mutually consistent. This paper specifies a chip-oriented, one-way impulse-radio ultrawideband (IR-UWB) uplink comprising an implanted packet engine and free-running transmitter, a tissue-proxy channel, and a wearable mixed-signal receiver and digital baseband.

0.27

Short-Range Microwave Interprobe Coupling in a Brain-Mimicking Phantom and Ex-Vivo Brain

eess.SP ·2026-10-12 ·Pouya Faeghi, Xiaoting Jia, Jeffrey S. Walling
similarity 0.50
Abstract · 8 pages, 10 figures, 3 tables

Objectives: Short-range wireless links (microwave range) between closely spaced devices in brain tissue are relevant to neural interfaces and intracranial sensing, yet the change in coupling between two such devices caused by a finite volume of brain-like tissue has rarely been measured directly. Technology or Method: This work characterizes a printed, microstrip-fed, parasitic-strip-loaded tapered-slot coupling probe in which only a partially inserted 12-mm wire enters the medium, while the printed structure remains outside. Results: Two fabricated probes showed a measured free-space -10-dB reflection band of 5.78 to 6.86 GHz and a fractional bandwidth of approximately 17.1%, consistent with full-wave simulation. In a controlled two-probe experiment, a homogeneous agar-gelatin-saline brain-mimicking phantom was introduced and removed while the probe arrangement and measurement conditions were retained. Introducing the phantom increased the measured transmission coefficient near 6.3 GHz by approximately 16 dB, and this differential was reproduced across repeated trials. More than five measurements in ex vivo goat brain followed the phantom transmission response within approximately 1.5 dB across the 5.78-6.86-GHz band. Time-domain analysis of the phantom-present link gave a system root-mean-square delay spread of approximately 210 ps. Full-wave dosimetry predicted a peak 1-g-averaged specific absorption rate of 1.05 W/kg for approximately 0.49 W accepted power. Conclusions and Clinical or Biological Impact: These results provide an experimentally validated reference for short-range probe coupling in brain-like media and support phantom-based bench evaluation of minimally invasive intracranial probes before in vivo studies.

0.23

iEEG Can Help EEG: Improving Clinical Scalp EEG Analysis through Intracranial Priors

cs.LG eess.SPq-bio.NC ·2026-11-16 ·Xianglong Bao, Yi Ding, Zengxiang Li, Nan Liu, Kewen Wang +2 more
top venuecode/data similarity 0.48
Abstract · Accepted at KDD 2027

Scalp electroencephalography (EEG) is a cornerstone of clinical brain monitoring, making the quality of its signal fundamental to downstream tasks and models. However, that signal is heavily contaminated by non-neural interference and blurred by volume conduction. Existing methods operate almost exclusively within the scalp modality, so they inherit the contamination they would need to overcome. In practice, intracranial EEG (iEEG) observes the neural activity with far higher fidelity, yet it is invasive and therefore scarce, and the two modalities are rarely recorded as pairs at scale, share no common channel space, and offer no clean ground truth to directly supervise a mapping. Motivated by the neuronal synchronization mechanism shared by the two modalities, we propose a reprogramming approach that fits a signal-to-signal model on iEEG alone and freezes it as a cross-modal prior. We introduce ReBrain, a framework to transfer an iEEG-learned prior to scalp EEG, which re-expresses any scalp signal through that prior and can be inserted upstream of any downstream model with no pairing or fine-tuning. ReBrain consistently improves host models across recordings from multiple hospitals and across architecture families, and our analysis traces this gain to genuine neural structure carried by iEEG rather than to the architecture itself. This work provides evidence that a scarce clean modality can serve as a transferable prior for an abundant noisy one, offering a general recipe for cross-modal transfer among biosignals. Our code is available at https://github.com/Xianglong-Bao/ReBrain.

0.64

Neural implants and human safety: single-fault detection for DC-coupled recording front ends

eess.SP cs.HC ·2026-08-11 ·Dimitris Antoniadis, Timothy Constandinou
similarity 0.48
Abstract · Draft IEEE 4 + 1

DC-coupled analogue front ends (AFEs) for neural implants provide a low-area solution. However, removing the coupling capacitor eliminates the intrinsic barrier that protects cortical tissue: a single-fault event, such as gate-oxide breakdown of a low-noise amplifier (LNA) input transistor, can open a direct DC path from the supply rail into the brain. On the stimulation side this hazard is well understood, and single-fault tolerance is enforced by a series DC-blocking capacitor; on the recording side, DC-coupled front ends discard the equivalent safeguard, yet their protection has gone almost unexamined. This paper presents a single-fault detection mechanism that monitors the LNA for the DC imbalance produced by such a failure and disables the amplifier before the resulting fault current can irreversibly damage tissue. The imbalance is encoded in the duty cycle of a current-starved relaxation oscillator and read out as a time-to-digital measurement. Designed in 65 nm, the mechanism resolves a worst-case fault of 6.4 nA across all corners within 0.81 ms - compliant with the ISO~14708-3 limit for an 8533 um2 electrode - opening a broader discussion of safety in DC-coupled recording.

0.04

Data-Centric Neuromotor Interfaces for Portable Human-Machine Interaction

cs.RO ·2026-08-31 ·Jiaxuan Li, Di Wu, Jianhua Liu, Yuxin Zhao, Jinnuo Li +5 more
publishedbenchmark similarity 0.48
Abstract

Dexterous human-machine interaction requires intuitive and expressive interfaces that can be efficiently deployed on constrained edge devices. Flexible material-based neuromotor interfaces hold considerable promise, as they decode human movement intention into natural control. Although emerging flexible electronic skins enable wearable high-fidelity data acquisition, practical deployment inevitably involves trade-offs between computational resources and portability. We present a data-centric paradigm where physiological features yield fundamental separability, providing sufficient discriminative cues for recognition. A wireless, high-bandwidth system developed for collecting various electrophysiological signals, when integrated with muscle-specific electrodes, forms a surface electromyography-based interface. Exploiting highly separable data, a 2,210-parameter model achieves 94.36% accuracy across 34 gestures and can be rapidly deployed on edge devices, establishing a new thousand-parameter benchmark for dexterous decoding. The underlying data-algorithm interactions in the data-centric paradigm are further clarified, demonstrating its feasibility in real-world scenarios. This study provides a principled and validated pathway for practical deployment of reliable neuromotor interfaces.

0.22

Streaming P300 Acquisition and Statistical Signal Validation Across Five EEG Platforms: A Hardware-Agnostic BrainFlow/LSL Pipeline

cs.HC ·2026-09-09 ·Isabella Guan, Rui Liu, Fusheng Wang
similarity 0.47
Abstract

P300 spellers offer people with severe motor impairment, such as ALS, an effective communication channel and remain one of the most established surgery-free alternatives to intracortical interfaces. Advanced language models have made spellers faster and more robust, yet the hardware beneath them is under-studied. We present a hardware-agnostic, real-time P300 acquisition pipeline built on BrainFlow and Lab Streaming Layer (LSL) that runs unchanged across consumer- and research-grade EEG headsets, with permutation tests of signal separability. Using a standard 6 x 6 row/column paradigm, we piloted five configurations: a custom dry system, a custom wet/gel system, Emotiv Flex, Emotiv EPOC X, and Muse 2. The custom systems and EPOC X showed weak or inconsistent signal separability, Muse 2 had the highest acquisition reliability despite limited centro-parietal coverage, and Flex showed the most promising signal. In 20 further Flex sessions varying subject, timing, and phrase length (131 target characters), a peak-amplitude permutation test and a cross-validated xDAWN decoder both detected a significant target response under two channel-exclusion policies, with decoder AUC reaching about 0.72 after 15 repetitions. Character accuracy depended heavily on evaluation methodology: in-sample majority voting reached 94.7%, whereas character-held-out accuracy was 31.3% with evidence accumulated across repetitions, about three times that of held-out majority voting. These analyses indicate that Flex captured a detectable, if still weak, P300 under the studied conditions, while broader participant-level validation and improved decoding remain necessary.

0.32

iBrain: A Unified Foundation Model Reading the Brain from Surface to Spikes

cs.AI ·2026-09-07 ·Ying Chen, Tiou Wang, Zhifeng Yue
similarity 0.47
Abstract

Invasive neural recordings provide high-fidelity measurements of brain activity, with signals such as intracranial EEG (iEEG) and intracortical spiking activity capturing neural dynamics at different spatial and temporal scales. Yet existing neural foundation models have largely been developed independently for different invasive recording paradigms, leaving joint pretraining across heterogeneous invasive signals underexplored. In this work, we introduce iBrain, a unified foundation model that jointly learns from iEEG and spiking activity. iBrain employs signal-specific encoders to accommodate their distinct signal characteristics and a shared spatiotemporal Transformer backbone to model dependencies across recording channels and time. We pretrain iBrain on over 7,000 hours of heterogeneous neural recordings using masked signal reconstruction and channel-view alignment, promoting contextual modeling of neural dynamics and robustness across different channels. iBrain consistently outperforms single-signal pretraining baselines and achieves state-of-the-art performance on multiple benchmarks. Further experiments demonstrate that iBrain exhibits transferability and data efficiency across diverse recording settings. These results highlight the potential of joint pretraining on heterogeneous invasive neural recordings to support scalable neural modeling and transferable representations across recording settings and downstream tasks.

0.24