CS Paper Radar

RIQE: a NIQE-style reference model for Computed Tomography

Fabio Mattiussi

eess.IV cs.CV ·submitted 2026-09-30 09:35 UTC · v1
code/data

Abstract

The Natural Image Quality Evaluator (NIQE) scores an image by its statistical distance from a model fitted on pristine images, and its distributed model is fitted on photographs. We release the Radiology Image Quality Evaluator (RIQE), a NIQE-style model fitted on 3,792 full-dose slices from 158 patients of the public LDCT-and-Projection-data collection, with a declared intensity mapping, a manifest of every slice and a script that reproduces the fit. On 40 held-out patients, RIQE ranks reduced-dose reconstructions, simulated by projection-domain noise insertion, worse than the full-dose reconstruction of the same slice in 240 of 240 chest and 230 of 240 abdominal pairs, and ranks images with 20% more noise worse than their source in 97.5-100% of cases. Its preferences among filtered images, however, do not follow lesion signal. With a 4 mm, +10 HU lesion inserted in noisy abdominal slices, RIQE prefers bilateral filtering to the unfiltered image in every image up to a 32 HU residual, at which 29% of the lesion's matched-filter signal remains and its detectability index falls from 0.51 to 0.33; it never prefers Gaussian smoothing, which at the same 32 HU residual leaves 70% of the signal and a detectability index of 0.48. Fitted on photographs with the parameters published for NIQE, the same code ranks every simulated reduced-dose abdominal image better than its full-dose counterpart. RIQE is suited to ranking a degraded image against its source; under the conditions tested it should not be the sole criterion for selecting, comparing or tuning denoisers.

Comment
17 pages, 6 figures, 6 tables. Code and model: https://github.com/Metallogik/RIQE, archived at doi:10.5281/zenodo.23055559

Closest papers

Nearest neighbours by meaning across the whole index.

DoseBridge: Denoising Diffusion Bridge Model for Dose Prediction in Lung Intensity-Modulated Proton Therapy

cs.CV ·2026-08-10 ·Zerun Zhang, Xiaoda Cong, Xiangkun Xu, Peter Y. Chen, Xuanfeng Ding
similarity 0.66
Abstract

Most radiotherapy dose-prediction models use only CT images and anatomical structures, although intensity-modulated proton therapy (IMPT) dose also depends strongly on beam geometry and available clinical datasets are often small. We present DoseBridge, a denoising diffusion bridge model that uses the patient CT as a structured bridge endpoint and encodes plan-specific beam geometry in a spatially aligned beam mask. Multiscale fusion combines CT, target, organ-at-risk, and beam-mask representations with 1.95% additional parameters. DoseBridge was retrospectively evaluated on single-institution CT images and treatment plans from 52 patients with advanced-stage lung cancer treated with 60 Gy in 30 fractions; 42 cases were used for training and 10 for testing. Performance was assessed using image-similarity, dose-volume, and Lyman-Kutcher-Burman normal-tissue complication probability (NTCP) metrics and compared with two deep-learning models. On the test cohort, DoseBridge achieved a mean absolute error of 4.170 Gy, peak signal-to-noise ratio of 23.06 dB, and structural similarity index of 0.798, outperforming both comparison models on these metrics. Clinical target volume D95 differed from the reference dose by 0.62 +/- 1.6 Gy; signed organ-at-risk mean-dose differences ranged from -0.32 to 0.24 Gy, and NTCP differences were -0.40 +/- 2.2 and 0.52 +/- 3.4 percentage points for acute esophagitis and radiation pneumonitis, respectively. Changing only the beam mask redirected predicted low-dose entrance regions while preserving the high-dose target region. To our knowledge, DoseBridge is the first denoising diffusion bridge model for radiotherapy dose prediction. These results support its feasibility as a beam-aware planning prior for lung IMPT, pending evaluation in larger external cohorts.

0.04

AgentCT: A Vision-Language Agent for Universal CT Imaging

cs.CV ·2026-11-08 ·Zhihao Chen, Qi Gao, Chuang Niu, Qiegen Liu, Yi Zhang +1 more
code/datahot direction similarity 0.66
Abstract · 14 pages, 5 figures

Computed tomography (CT) restoration methods often use a single network for low-dose CT (LDCT), sparse-view CT (SVCT), and metal artifact reduction (MAR), despite the distinct physics of these degradations. We propose AgentCT, a vision-language agent that identifies the degradation and routes images to task-specific restoration models. A LoRA-adapted Qwen3-VL-4B controller predicts degradation type and severity, selects a specialist, and reassesses the restored image. The LDCT specialist uses an image-domain diffusion model, whereas the SVCT and MAR specialists combine projection-domain correction, data consistency, differentiable filtered back-projection, and image refinement. We evaluated the controller and specialists separately on patient-disjoint Mayo 2020 abdominal CT data across three severity levels for each degradation. The controller achieved 95.7% accuracy for degradation type and 93.9% for severity, with a 100% JSON parse rate. Using ground-truth degradation labels for routing, AgentCT achieved an overall PSNR/SSIM of 46.18 dB/0.984, compared with 44.93 dB/0.976 for AMIR and 44.29 dB/0.979 for TAMP, and attained the highest PSNR in all nine settings. For SVCT, its dual-domain specialist achieved 46.04 dB with 4.2 million parameters, compared with 44.43 dB for single-task AMIR and 43.57 dB for all-in-one AMIR with 23.5 million parameters. These results support degradation-specific CT restoration with controller-based model selection; evaluation under automatic routing and on independent clinical data remains necessary. Code is available at https://github.com/hao1635/AgentCT.

0.42

Coarse-to-Fine Diffusion with Residual Prior for Low-Dose CT Denoising

eess.IV ·2026-09-30 ·Ping Wu, Pengcheng Zheng, Jiahao Zhang, Bohao Chen, Yazhou Ren +1 more
similarity 0.63
Abstract · 10 pages

Low-dose CT (LDCT) reduces radiation exposure but often introduces severe noise and streak artifacts that obscure subtle anatomical structures. Existing diffusion-based denoising methods commonly rely on implicit assumptions about the noise distribution, which may not hold in practical LDCT scenarios and can degrade denoising performance. To address this limitation, we propose a Residual-Guided Diffusion framework that effectively incorporates an explicit difference prior to inform the diffusion process of noise distribution. The difference prior is computed as the residual between the LDCT image and a coarse clean estimate, constraining diffusion sampling to a plausible residual space. We further adopt a coarse-to-fine strategy in which a pre-trained RED-CNN provides the coarse estimate to anchor global anatomy, and the diffusion model refines the remaining details under the prior guidance. For efficiency, we employ accelerated sampling with 20 steps at inference. Experiments on three representative datasets demonstrate that our approach outperforms existing methods in both quantitative and qualitative performance measures.

0.14

Conditioning noise is a free regularizer for LoRA fine-tuning: no pathology encoder required for diffusion-based artifact detection in histopathology

eess.IV cs.CV ·2026-09-11 ·Konstantinos Moutselos, Ilias Maglogiannis
code/data similarity 0.63
Abstract · 23 pages, 3 figures. Code and laboratory record: https://github.com/kmouts/condnoise-histoqc (doi:10.5281/zenodo.22702198). Data: doi:10.528…

Diffusion-based artifact detectors score whole-slide image patches by reconstruction error under a model fine-tuned on clean tissue. We show that conditioning this fine-tuning on random Gaussian embeddings -- resampled at every step from approx. 200 KB of precomputed embedding statistics, with no encoder, no cache, and no change to inference -- consistently widens the clean/artifact separation. A four-step ablation chain shows the benefit requires neither content (shuffled real embeddings), provenance (synthetic Gaussians), a tuned intensity (flat across an 8x variance range), nor per-patch identity (fresh per-step noise); a LoRA-dropout control shows the conditioning pathway specifically, not generic weight perturbation, carries the effect. Patch-level gains of +0.25-0.48 Cohen's d replicate across nine trainings; honest leave-one-slide-out evaluation clears a pre-registered bar in 2/2 seeds; and two pre-registered external endpoints on a 281-case set confirm pooled Delta F1 = +0.0073 (95% CI) and +0.0129 (97.5% CI, two-look corrected). We release the full evaluation protocol, including measured seed noise and selection-optimism pricing.

0.50

Anatomy-preserving unpaired cone-beam CT refinement for image-guided radiotherapy using pseudo-label guided diffusion

cs.CV ·2026-10-05 ·Qi Lai, Yutong He
code/data similarity 0.62
Abstract

Cone-beam computed tomography (CBCT) is widely used in image-guided radiotherapy, but scatter, beam hardening, noise, truncation, and other artifacts limit image quality and CT number accuracy. Paired CBCT and CT data are difficult to obtain clinically because of motion, anatomical changes, and acquisition mismatch. We present RefineCBCT, an unpaired CBCT refinement framework that uses pseudo-label guidance and short-step diffusion to reduce artifacts while preserving patient-specific anatomy. RefineCBCT was trained and evaluated on unpaired CBCT and planning CT data from public LUNG TCIA and PELVIC TCIA datasets and compared with representative GAN and diffusion based methods. On LUNG TCIA, it achieved the best results across all metrics, with MAE 19.411, RMSE 62.758, PSNR 30.845 dB, and SSIM 0.931. On PELVIC TCIA, it achieved the best MAE, PSNR, and SSIM, with values of 14.905, 36.671 dB, and 0.876. The refined images showed fewer streaking and shading artifacts, clearer anatomical boundaries, and improved soft tissue uniformity, with line profile and ROI analyses showing closer agreement with planning CT. These results suggest that RefineCBCT provides efficient and effective CBCT refinement under clinically realistic unpaired training conditions and may support more reliable CBCT use in image-guided radiotherapy workflows. Code is publicly available on GitHub, and the evaluated datasets are available from The Cancer Imaging Archive.

0.21

Measurement-Aware Multimodal LLM for No-Reference Low-Dose CT Quality Assessment

eess.IV cs.CV ·2026-11-05 ·Kerem Basim, Tolga Demiroglu, Mehmet Ozan Unal, Metin Ertas, Isa Yildirim
hot direction similarity 0.62
Abstract · 5 pages, 2 figures, 3 tables. Submitted to IEEE ISBI 2027

No-reference image quality assessment (IQA) of low-dose CT predicts the score that radiologists would assign to a slice without a full-dose reference, as needed to compare dose protocols and denoisers. Trained CT IQA networks agree well with radiologists but output only a score, whereas multimodal large language models (MLLMs) can explain a score but perform poorly on CT when only prompted. We propose a measurement-aware MLLM, built on Qwen3.8-27B, that rates a slice from its image together with six image-domain measurements of noise, sharpness, high-frequency content and streaks, given to the language model as percentiles in text. On the LDCT-IQA 2023 test set, it reaches an Overall score of 2.735, on par with the retrained challenge winner (2.735) and MedSigLIP-FiLM (2.713), and substantially above its base MLLM used as a zero-shot agent. Shuffling the measurements at test time reduces the score by 0.048, mostly through noise, but the same measurements also lead the model to assign higher scores to over-smoothed denoised images.

0.34

Can saliency maps flag chest radiograph classifier errors at a new hospital? An external validation study

cs.CV cs.LGeess.IV ·2026-11-03 ·Aashish Chapagai, Angad Singh Gurtatta, Erik Cambria
code/data similarity 0.62
Abstract · 21 pages, 5 figures, 1 table, 6 Extended Data figures and 4 Extended Data tables. Code: https://github.com/aashish-chapagai/CXR-Saliency-Ext…

Saliency maps are offered as a safeguard for medical image classifiers, but whether they reveal individual errors at a new hospital remains largely untested. We trained a convolutional network and a vision transformer on NIH ChestX-ray14 and evaluated 303,768 decisions per model at the training hospital and at two external sites, in the USA and Vietnam, with radiologist-consensus labels. We scored each explanation by its anatomical enrichment, the area-normalized concentration of attribution in a segmented thoracic region, and tested whether it detects errors beyond the model's confidence. Grad-CAM localized findings less often than radiologists (hit rate at most 0.575 versus 0.888). Adding enrichment to a confidence-based error detector calibrated at the training hospital never raised its area under the curve by more than 0.011, and for the convolutional network the relation between enrichment and correctness reversed between hospitals, degrading the transferred detector. Model confidence, not saliency, should anchor deployment monitoring.

0.31

Noise2Match: Self-Supervised Low-Dose CT Denoising via Adaptive Inter-Slice Matching

cs.CV ·2026-10-13 ·Langrui Zhou, Guang Li, Xinyu Huang, Ziteng Zhou, Zuyang Huang +3 more
top venue similarity 0.61
Abstract

With the rapid development of deep learning, supervised convolutional neural network methods have greatly improved the performance of low-dose computed tomography (CT) denoising. However, they require paired noisy-clean training data, which is difficult to obtain in practice. Some classic self-supervised denoising methods rely on the assumption of pixel-wise independent noise, making them ineffective for correlated noise in low-dose CT (LDCT) images. Considering the high similarity and weaker noise correlation between adjacent CT slices, several recently proposed self-supervised methods use adjacent LDCT slices for training. Nevertheless, these approaches fail to effectively account for the non-rigid misalignment between slices caused by anatomical variation, hindering them from satisfying the equal-mean condition required by the classic Noise2Noise framework. In this paper, we propose Noise2Match, a self-supervised LDCT denoising method based on adaptive inter-slice matching. Noise2Match introduces an inter-slice matching network to model the label transition process between adjacent slices, and then applies correction to the denoiser's outputs to better satisfy the equal-mean condition. We theoretically prove that, under idealized conditions, training Noise2Match using adjacent LDCT slices is equivalent to supervised training with paired noisy-clean images. We further find that Noise2Match can also effectively perform low-dose conebeam CT (CBCT) denoising, which could be attributed to its correction operation that reduces inter-slice noise correlation. To the best of our knowledge, Noise2Match is the first effective image-domain self-supervised method for low-dose CBCT. Extensive experiments demonstrate that Noise2Match outperforms existing image-domain selfsupervised methods, offering broader applicability and higher practical value.

0.33