VNA-Only Calibration of Waveguide Thermistor Mounts with Self-Consistent Determination of the Thermistor-Port Impedance

Patricio Greco

eess.SP physics.ins-det ·submitted 2026-11-09 13:15 UTC · v1
unusual

Abstract

A method is presented for determining the calibration factor and the effective efficiency of waveguide thermistor mounts from vector-network-analyzer reflection data alone, without a microcalorimeter, extending the three reflection-coefficient method of the coaxial case. In waveguide the thermistor is coupled to the line through a matching structure and the reference impedance of the thermistor port, $Z_0$, is not known a priori; it is inferred through a power-consistency relation between pairs of states, under the convention that it is real. The method reproduces the factory certificates (U = 2 %) of four K486A mounts within 1.0 % rms and the PTB-traceable values (U = 0.85 to 0.98 %) of a fifth within 0.34 % rms; its expanded measurement uncertainty is 0.5 to 1.16 % (k = 2) for the best specimen and below 2 % for the five. A mount with its waveguide-to-coaxial adapter is calibrated directly, without characterizing the adapter, from the model factor of the mount and the geometric factor measured through the adapter. The scattering matrix implied by the solution is not passive; a minimum-norm correction of the thermistor impedances along the two directions invisible to the calibration factor - the ratios of the efficiencies of the three states - brings it to the floor of passivity, where the efficiency depends on the load only through its mismatch factor, leaves the result unchanged, and gives the sensitivity of the efficiency to the bias resistance in closed form: below $4\times10^{-5}$ $Ω^{-1}$ in band mean, proportional to the mismatch of the operating state and to the loss of the mount.

Comment
60 pages, 21 figures. Submitted to Metrologia (IOP Publishing / BIPM), November 2026
DOI
10.5281/zenodo.24038232

Closest papers

Nearest neighbours by meaning across the whole index.

Passivity Correction of the Thermistor Model in Bolometric Power Sensors and Its Dependence on Bead Temperature

eess.SP physics.ins-det ·2026-11-20 ·Patricio Greco
similarity 0.62
Abstract · 18 pages, 19 figures, 4 tables. This work has been submitted to the IEEE for possible publication. Copyright may be transferred without noti…

This paper closes a series of works on thermistor bolometric power sensors, coaxial and waveguide, by treating what the two have in common. The per-state transformation m introduced in the waveguide work, which corrects the thermistor impedances of the model in each bridge state so that the two-port of the mount is passive, is analyzed on the eleven sensors of the preceding papers: the corrections are a few tenths of a percent in the coaxial mounts and one to two percent in the waveguide mounts, follow one pattern within each family, and the hottest state is the one whose efficiency departs most from the others. The transformation is tested by an independent measurement: the substituted powers of the three states, read with the manufacturer's procedure against a leveled source whose reflection is obtained from the closure of three Apollonius circles, give efficiency ratios of 0.999 +/- 0.016 and 0.997 +/- 0.008 (k = 2), which contain the values at the passivity floor (0.996, 0.998) and exclude those of the first pass (0.941, 1.023). Three sensors were swept in the bridge resistance, 23 and 31 states, and a fourth is treated apart: the correction reduces to one constant per state, the parallel resistance of the bead, represented by a law with one term scaling with the bridge resistance and one exponential in the bead temperature, with one thermal coefficient per family. The same property explains why the cross-ratio of four measured reflection coefficients, invariant under any state-independent two-port, departs from that of the four bridge resistances.

0.30

Calibration electronics for the 30 and 40 GHz instrument (TFGI) of the QUIJOTE experiment

eess.SY astro-ph.IMphysics.ins-det ·2026-09-17 ·Jorge Luis Díaz-Acosta, Roger John Hoyland, Francisco Javier Casas, Enrique Martinez-González
publishedunusual similarity 0.38
Abstract · 4 pages; 6 figures; Accepted conference paper submitted to URSI La Laguna 2026; Original conference paper submitted in Spanish; English and …

The 30 and 40 GHz instruments of the QUIJOTE radio astronomy experiment use very sensitive receivers which need to be characterized both to ensure their proper functioning and for their calibration. Given the age of the PXI-1031DC device used up until now for this purpose, the design of a new device, the Calibrator TFGI QUIJOTE, has been proposed. This new device has been designed with the idea of being simple and modular, easy to use and upgrade. It is built on a Raspberry Pi 5 system, using an MCC 118 as analog reading module. This system, widely known among both makers and professionals, not only meets the requirements of modularity and ease of use, but also ensures that the device is much more economically competitive than the PXI. A first version of the device has been built and tested, obtaining good results in accuracy and ease of use.

0.35

A Reusable Four-Port S-Parameter to Link-Level Signal-Integrity Analysis Framework for High-Speed Detector Interconnects

eess.SY hep-ex ·2026-08-10 ·D M S Sultan, D. Bortoletto, R. Plackett, A. E. McDougall, A. S. Rotelli +2 more
similarity 0.37
Abstract

A reusable MATLAB signal-integrity (SI) framework is presented that converts compatible four-port S-parameter data, measured by VNA or obtained from electromagnetic simulation, into traceable link-level evidence rather than a single loss metric. Its contribution is the checked, traceable automation of established SI operations while keeping source data, model assumptions, analytical projections, and finite-record observations distinct. The framework is demonstrated on the four 1.25 Gbps differential routes (DP1-DP4) of the PPCB-1347-MuPix11 probe card using PTSL CST Microwave 3D-Solver-derived four-port S-parameters and a virtual time-domain solver. The automated pipeline preflights file structures, performs a power-normalized mixed-mode transformation, applies route-length-aware loss decomposition, constructs a causally loaded channel model, and propagates full PRBS-31 sequences into eye-diagram, conditional-BER, and 8b10b-coded-link analyses. At the 1.25 Gbps data rate (Nyquist 0.625 GHz), DP1-DP4 exhibit differential insertion loss (SDD21) from -0.350 to -0.300 dB, differential-to-common conversion from -31.038 to -28.236 dBc, and modeled FEB-input eye openings from 0.586 to 0.588 V. The comparison shows that path length alone is not an adequate SI ranking variable: DP3 has the lowest Nyquist insertion loss, DP1 the strongest differential-to-common isolation, and DP4 the largest modeled eye. All analytical BER values remain below the reporting floor and therefore do not support a BER ranking. By preserving the distinction between route-dependent waveform behavior, model projections, and finite-record observations, the framework provides an extensible basis for comparative high-speed-interconnect SI analysis from design review through calibrated VNA measurement interpretation.

0.04

CMS ECAL Phase-2 upgrade: characterisation and stability analysis of a three-stage low-voltage power-conversion system

eess.SY physics.ins-dethep-ex ·2026-11-04 ·Liv Anke Richter, Günther Dissertori, Werner Lustermann, Krzysztof Stachon
unusual similarity 0.34
Abstract · 7 pages, 6 figures. Proceedings of the Topical Workshop on Electronics for Particle Physics (TWEPP 2026), 28 September - 2 October 2026, Cas…

As part of the High-Luminosity upgrade of the Large Hadron Collider, the front-end electronics of the Compact Muon Solenoid electromagnetic calorimeter barrel are being replaced. A three-stage low-voltage power-conversion system supplies the upgraded electronics. As the cascade dynamics require characterisation to ensure stable operation, interactions between the three converter stages are assessed at their two interfaces using the Nyquist criterion. A validated shunt-injection setup using a Bode 100 frequency-response analyser provides in-situ impedance measurements under load to assess the accuracy of the averaged converter models. Results show that both interfaces are stable over the evaluated configurations, with modelled minimum distances to the critical point of 0.76 and 0.79, respectively. A turn-on study identifies capacitor charging at Stage 3 as the main cause of inrush currents triggering Stage 2 overcurrent protection. Two Stage 3 board variants stagger converter enables within and between boards, keeping the measured two-tower current below 5 A, within the 8.0 A protection threshold.

0.37

Pattern Multiplication Underestimates the Sidelobe Level of a Planar Slotted-Waveguide Array by 1.1 to 2.8 dB Across a 2% Band

eess.SP ·2026-08-21 ·William Khalili
similarity 0.34
Abstract

The standard shortcut in slotted-waveguide array design -- solve one radiating stick, multiply its embedded pattern by the array factor of the intended aperture distribution, and accept the product as the array pattern -- is tested against a full-wave solution of the same metal. For a 16x16-slot planar array in WR-90 synthesised to a -30 dB Taylor illumination, the shortcut underestimates the transverse-plane sidelobe level by 1.06 dB at the design frequency and by 2.22 and 2.84 dB at the edges of a 2% band: a spread of 1.78 dB, larger than the mid-band error itself. Every sidelobe figure carries +/-0.11 dB of mesh spread from a convergence study on the production geometry. The error is signed, is worst at the band edges where the design has least margin, and falls only from 1.58 dB to 1.06 dB when the aperture is doubled from eight sticks to sixteen, which identifies the edge elements as its source. The array was made analysable by a thirteen-point single-slot full-wave calibration: over the design's offset range the handbook closed form for resonant length moves 3.9 micrometres where the measurement moves 143.9 micrometres, and Stevenson's conductance is shown to overestimate by a nearly constant 5.16%.

0.25

The Paired Thermal Channel Is Two Orthogonal Axes: A Controlled Attribution Study of Calibration Cost and Detection Gain in Infrared Person Detection

cs.CV cs.MM ·2026-10-23 ·Tian Zhang, Yang Chen, Bing Bai, Weikai He, Xiang Gao +1 more
theoryunusual similarity 0.34
Abstract

The calibrated thermal channel in infrared person detection comprises two structurally orthogonal axes. True calibrated $ΔT$ outperforms scale-matched grayscale pseudo-$ΔT$ by $+0.0221$ mAP (paired Wilcoxon $p=0.0098$, Cohen's $d=+1.08$), confirming that the calibrated channel carries information the grayscale channel lacks. The calibration axis is driven by the joint pairing-scale condition. Preserving both paired correspondence and absolute calibrated scale raises expected calibration error (ECE) from $\approx 0.05$ to $0.142$. The $2\times 2$ interaction ($p=0.0059$) localizes the rise to this condition. Breaking either factor alone returns ECE to the broken-factor level (true-vs-const $p=0.002$; true-vs-shuffled $p=0.002$). Under post-hoc temperature scaling, the pairing cost survives recalibration ($p=0.002$), while the scale cost is largely absorbed ($p=0.16$). As regards the detection axis, the gain stems from structural de-structuring. Phase randomization, which destroys spatial structure while preserving per-sample variance, yields $+0.0185$ mAP over true $ΔT$ on FLIR ADAS ($p=0.014$, $d=+0.96$). The FLAME~3 gain is $+0.0064$ mAP ($p=0.078$, n=8). On the calibration axis, the same intervention is insignificant on FLIR ADAS ($p=0.47$) but significant on FLAME~3 ($p=0.0078$). These findings rest on a five-arm controlled attribution with temperature-evidential learning (TEL) as the testbed. TEL obtains a single evidence-derived routing scalar together with a complementary diagnostic entropy, and establishes a finite-evidence degeneracy theorem proving that the acquisition product degenerates to this scalar with a closed-form bound. A physical $0.5^\circ$C threshold admits cross-dataset physical readings, whereas a gray threshold of $20$ inverts without physical interpretation.

0.44

Feasibility and Convex Design of Probe-Position Matching in a Scanning X-Band Radar Array

eess.SP physics.app-ph ·2026-08-12 ·William Khalili
similarity 0.34
Abstract

The probe position of a microstrip element is normally chosen from the isolated-element input resistance. This paper replaces that procedure with a decision available before any array optimisation. A single Floquet unit-cell solution at one arbitrary probe position is decomposed into a feed inductance, a transformer ratio carrying the probe position, and an array-loaded resonator. Three closed-form results follow. The set of input impedances reachable by probe position and resonant length is a disk in the impedance plane. An exact match to a real reference Z0 exists if and only if Rp >= Z0 + Xp^2/Z0. When this fails, the best attainable reflection follows from the image of that disk under the bilinear map. Admitting one series reactance as a second variable, the worst-case reflection over a scan sector is shown to be quasiconvex and globally solvable by bisection. The results are applied to a sixteen-element X-band array and the corresponding 16x24 lattice. For the array considered, the isolated interior element is matched to -16.94 dB while the broadside active reflection is -6.6 dB. For the two-dimensional lattice the criterion is violated at every scan angle, and the minimax design improves the worst-case sector reflection from -2.6 to -11.1 dB. Both matching designs are predictions of the extracted circuit model and have not been re-solved in the full-wave model.

0.04

Agentic TCAD Calibration Workflow for Oxide Semiconductor Transistors

cs.AI physics.ins-detcs.LGphysics.app-ph ·2026-09-10 ·Gyujun Jeong, Junmo Lee, Sungwon Cho, Woohyun Hwang, Kwangyou Seo +9 more
similarity 0.33
Abstract · Preprint. Submitted to an IEEE journal for review

Experimental TCAD calibration is essential for predictive technology modeling of emerging oxide semiconductor transistors. However, it remains time-consuming and expert dependent because of model ambiguity. Multiple physical models and parameter sets can reproduce the same measured transfer characteristics, while local fitting alone cannot uniquely identify the underlying device physics. We present the first demonstration of an agentic TCAD calibration workflow for a fabricated bottom-gate In--W--O (BG-IWO) transistor. Starting from the measured transfer curve and device information, the workflow uses measurement--TCAD residuals and local sensitivity tests to select bounded parameter corrections or evaluate additional physical models, and accept only updates that improve device metrics. The LLM agent orchestrates the workflow, while Sentaurus governs the device physics. For the 2\%-W reference device, five agent-suggested updates yield a fixed calibrated model, reducing the multi-metric device objective $J$ by 14.3$\times$. Maximum $V_{\mathrm{th}}$/$I_{\mathrm{on}}$ errors are 36.1~mV/0.022 decade for varying-drain-bias tests and 46.2~mV/0.062 decade for varying-channel-length tests, demonstrating model transferability across bias and geometry rather than a local parameter fit. W-composition tests provide process-sensitive insight. This agentic workflow provides a faster route to model development for emerging device technologies.

0.30