Data Archive for 2D and 3D Electrical Impedance Tomography with moderate conductivity contrast phantoms

S. J. Hamilton, A. Hauptmann, A. Hänninen, V. Kolehmainen, J. Niskanen, T. Savolainen, J. Toivanen

eess.IV cs.LG ·submitted 2026-11-12 18:19 UTC · v1
code/data

Abstract

This document describes open Electrical Impedance Tomography (EIT) measurement data that were collected from biomedical imaging inspired phantoms with moderate conductivity contrasts. The data were collected with the KIT4 and KIT5 EIT measurement devices at the University of Eastern Finland, and are intended for EIT image reconstruction algorithm development and testing. The data from thoracic imaging inspired cylindrical and chest-shaped phantoms are suitable for 2D algorithms, and the data from an intracerebral hemorrhage imaging inspired head-shaped phantom with resistive skull are suitable for 3D algorithms. In addition to the measurement data, computational meshes and example reconstruction codes are provided and sample reconstructions included.

Comment
11 pages, 6 figures. Data available at https://doi.org/10.5281/zenodo.24120439

Closest papers

Nearest neighbours by meaning across the whole index.

An Arduino Giga-Based Multi-Frequency Electrical Impedance Tomography System for Electrolyte Media: Hardware Design, Statistical Voltage Analysis, and System Performance Evaluation

eess.SP ·2026-10-11 ·Zahra Karimi, Natalie Verdugo, Joseph Schwab, Hamid Ghaednia
unusual similarity 0.61
Abstract

Multi-frequency electrical impedance tomography (EIT) often requires specialized hardware. This challenge is especially pronounced in biomedical applications. We developed a 16-electrode EIT system using an Arduino GIGA R1 WiFi for generation and acquisition, with a discrete current source, three analog multiplexers, and basic signal conditioning. Dual cores separate waveform timing from switching and acquisition. A 3D-printed tank with silver-wire electrodes was tested at 11 frequencies (1-180 kHz). Across the 10-180 kHz analysis band, median repeat-frame signal-to-noise ratio was 48 dB, acquisition rate was ~0.18 frames/s per frequency (~60 s per sweep), and cross-session baseline drift was ~6%. KCl phantoms contained an insulating ABS cylinder or KCl-filled dialysis bags (0.001-1.0 M) spanning positive and negative conductivity contrasts. Boundary voltages were analyzed without reconstruction to estimate inclusion position, contrast polarity, and relative magnitude. Leave-one-session-out matching located ABS inclusions in 29/30 runs at 160 kHz (87-97% across the band; chance 33%). For dialysis bags, polarity was correctly identified in all 36 nonzero-contrast runs. Signed voltage projections correlated with the cylindrical contrast factor (R^2=0.68 across runs; R^2=0.89 for the aggregate fit). Concentration was ordered (Spearman rho=0.85), despite saturation above ~10-fold contrast. Bag position accuracy ranged from 6/6 at the highest contrasts to 2/6 at matched conductivity. The matched-conductivity bag produced a ~1.7% RMS perturbation against 0.4-0.6% repeat-frame variability. These results demonstrate that microcontroller-based multi-frequency EIT can support selected detection and identification tasks through direct boundary-voltage analysis, supporting future studies of muscle electrochemistry.

0.25

Efficient 3D Whole-Body PET Image Denoising via Conditional Rectified Flow With Optimized Sampling Strategy

cs.CV ·2026-09-15 ·Jiale Shen, Guolin Wang, Chenhao Wang, Xinhui Su, Wei Luo +1 more
similarity 0.49
Abstract

Reducing radiation exposure in Positron Emission Tomography (PET) is important for patient safety; however, ultra-low-dose imaging suffers from severe noise, which may affect diagnostic interpretation without appropriate image enhancement. While current 3D deep generative models, particularly diffusion models, have shown strong reconstruction fidelity, their practical use can be limited by long inference times. In contrast, faster 2D-based alternatives may have difficulty maintaining volumetric consistency, an important consideration for whole-body PET imaging analysis. To bridge this gap, we propose a one-pass conditional 3D rectified flow (3D Flow) framework for whole-body PET image denoising that incorporates a novel optimized non-uniform sampling strategy. The model is trained with a one-pass linear-interpolant velocity-matching objective. This approach reconstructs a full 3D volume in approximately 30 seconds in our implementation, compared with multi-hour inference for the evaluated 3D DDPM baseline. Evaluations including zero-shot transfer to an independent clinical dataset show that our model achieves favorable global image quality and lesion conspicuity compared with the evaluated 3D DDPM and DDIM baselines, including on challenging short-acquisition data. Furthermore, the proposed method shows promising zero-shot transfer performance across the evaluated datasets and unseen dose levels (down to 1/100 of the standard dose), with artifact-focused visual comparisons supporting the need for further lesion-level validation. By balancing reconstruction fidelity and computational efficiency, this work presents a candidate approach for ultra-low-dose whole-body PET image denoising.

0.08

An Evidence-Guided Framework for Tissue-Barrier Impedance Review: Deletion and Resistance Sensitivity

eess.SP ·2026-10-23 ·Leo Yang, Yingtong Zhou
similarity 0.49
Abstract · 5 pages, 3 figures, 3 tables

Reviewing tissue-barrier impedance requires connecting spectral anomalies with resistance estimates and experimental context. We present an evidence-guided framework that coordinates numerical checks and traceable actions for human review. In-house simulations assess the effect of removing flagged frequencies, public datasets test model adaptation, and local measurements demonstrate the review workflow. Removing flagged frequencies reduced resistance error under quadratic objectives, whereas robust fitting showed little average benefit from deletion and a mean error increase for endpoint-pair perturbations. High errors clustered around a few fitted circuit parameter sets, highlighting the value of examining parameter identifiability. Under the tested noise and perturbation conditions, these findings support robust full-spectrum estimation, flag-guided inspection, and matched sensitivity checks.

0.26

Short-Range Microwave Interprobe Coupling in a Brain-Mimicking Phantom and Ex-Vivo Brain

eess.SP ·2026-10-12 ·Pouya Faeghi, Xiaoting Jia, Jeffrey S. Walling
similarity 0.49
Abstract · 8 pages, 10 figures, 3 tables

Objectives: Short-range wireless links (microwave range) between closely spaced devices in brain tissue are relevant to neural interfaces and intracranial sensing, yet the change in coupling between two such devices caused by a finite volume of brain-like tissue has rarely been measured directly. Technology or Method: This work characterizes a printed, microstrip-fed, parasitic-strip-loaded tapered-slot coupling probe in which only a partially inserted 12-mm wire enters the medium, while the printed structure remains outside. Results: Two fabricated probes showed a measured free-space -10-dB reflection band of 5.78 to 6.86 GHz and a fractional bandwidth of approximately 17.1%, consistent with full-wave simulation. In a controlled two-probe experiment, a homogeneous agar-gelatin-saline brain-mimicking phantom was introduced and removed while the probe arrangement and measurement conditions were retained. Introducing the phantom increased the measured transmission coefficient near 6.3 GHz by approximately 16 dB, and this differential was reproduced across repeated trials. More than five measurements in ex vivo goat brain followed the phantom transmission response within approximately 1.5 dB across the 5.78-6.86-GHz band. Time-domain analysis of the phantom-present link gave a system root-mean-square delay spread of approximately 210 ps. Full-wave dosimetry predicted a peak 1-g-averaged specific absorption rate of 1.05 W/kg for approximately 0.49 W accepted power. Conclusions and Clinical or Biological Impact: These results provide an experimentally validated reference for short-range probe coupling in brain-like media and support phantom-based bench evaluation of minimally invasive intracranial probes before in vivo studies.

0.23

Matching two-dimensional statistics does not certify three-dimensional transport in battery electrodes

cs.CE cond-mat.mtrl-sci ·2026-11-15 ·Maziar Raissi, Devesh Upadhyay
similarity 0.49
Abstract · 13 pages, 4 figures, 4 tables; includes Supplementary Information

Three-dimensional battery electrode structures reconstructed from two-dimensional micrographs are routinely used to predict transport, assuming that matching the measured 2D statistics reproduces the relevant 3D properties. We test this premise on X-ray tomograms of three all-solid-state cathodes. With a differentiable conductivity solver, we optimise voxel structures that match the real electrode in phase fractions, boundary density, two- and three-point correlations and lineal-path functions, on all section planes, slab by slab and layer by layer, within measurement noise. Every sample has statistical twins conducting 10-88-fold less than the real electrode, robust to 5^3 majority filtering, whereas the most conductive are at most 3.3-fold above it. The twins rearrange rather than remove material, severing or funnelling through-plane paths. A classifier on 2D sections detects them, as it does current 2D-to-3D generators: the information is in the images but not in the statistics. Matching standard 2D statistics therefore does not certify 3D transport.

0.41

EPC-RawDC: Equivariant Physics Constrained 3D Diffusion with Raw Data Consistency for CBCT to CT Image Synthesis

cs.CV cs.AIphysics.med-ph ·2026-10-29 ·Alzahra Altalib, Chunhui Li, Alessandro Perelli
code/datahot direction similarity 0.46
Abstract · 10 pages, 4 figures, 8 tables

Cone beam computed tomography (CBCT) is widely used for image guided and adaptive radiotherapy, but scatter, noise, reconstruction artifacts, and inaccurate Hounsfield units limit its quantitative use. Recent diffusion approaches for CBCT to CT synthesis remain largely image driven, with limited use of the underlying acquisition physics. We propose EPC-RawDC, a conditional 3D latent diffusion framework that combines volumetric generative modeling with two complementary constraints derived from the same scanner specific CBCT forward model. First, geometry equivariant projection constraint exploits the rotational equivariance of the CBCT forward operator by supervising the synthesized CT through correspondingly transformed reference projections. Second, raw data consistency uses noisy projection measurements of the individual acquisition as scan specific supervision during training. The framework was evaluated on two independent clinical cohorts with distinct scanner geometries and projection sources. For JUST, the physics constraints used measured clinical raw CBCT projections acquired with an Elekta system, whereas for SynthRAD2025 the projection data were simulated from the reference CT using a Varian geometry with Poisson photon statistics. EPC-RawDC achieved an SSIM of 0.97 in both cohorts and improved PSNR over DDPM by 2.66 dB on SynthRAD2025 and 2.10 dB on JUST, while reducing mean MAE by 20.3% and 19.0%, respectively. On SynthRAD2025, projection consistency and equivariant errors were reduced by 55.0% and 41.4% relative to DDPM. These results demonstrate that combining a volumetric diffusion prior with scanner geometry and scan specific projection measurements improves quantitative CBCT to CT synthesis across both simulated and measured projection domains. Code available after publication at https://github.com/ALZAHRAALTALIB/EPC-RawDC.

0.37

Simulating Vein Compressibility for Ultrasound-Guided Femoral Access: Speed, Accuracy, and Phantom Validation

cs.RO ·2026-11-09 ·Amit Manicka, Laurent Coulon, Darwin Ang, Michael Loushin, Timothy M. Kowalewski
similarity 0.46
Abstract · Submitted to the Journal of Medical Robotics Research (JMRR) for the International Symposium of Medical Robotics 2027 (ISMR 2027) 11 Pages (…

In ultrasound-guided femoral access, probe pressure deforms the tissue and changes the vascular anatomy a robot observes. A simulator must reproduce these mechanics fast enough for robotic interaction, yet the computational effort needed for a clinically meaningful signal like vein collapse before artery is unclear. We test the feasibility of a physics-based simulation in NVIDIA's Isaac Sim for Healthcare, with Stable Neo-Hookean tissue solved by Vertex Block Descent on one GPU. A simulation of a low-fidelity cylindrical benchtop phantom with known geometry and mechanical properties, pressed by a robot, predicted its artery within 5 percentage points at every depth. It matched the vein within 3 points through 11 mm of a 25.4 mm press and followed the collapse behavior of the low-fidelity physical phantom. Across element size, time step, and solver iterations, wall clock time per frame scaled linearly with solver work, but accuracy did not improve with more work. The chosen setting reduced solver time per frame by a factor of 976, from 29,700 ms to 30.4 ms, while remaining within 1.03 points of the finest reference setting. The physics solver alone took 30.4 ms per frame, while the full stack, including ultrasound and graphics rendering, took 56.4 ms per frame. Finally, when seeded with realistic literature-based mechanical properties, the digital leg model compressed the vein far more than the artery at every setting, reproducing the expected clinical signal of vein collapse before artery, which is used to distinguish them. In total, these results suggest that the NVIDIA Isaac Sim platform may be suitable for augmenting reinforcement learning training or online simulation for autonomous femoral access with ultrasound.

0.22

Extending TotalSegmentator: Predicting Patient and Acquisition Characteristics from CT and MR Images

cs.AI ·2026-08-29 ·Jakob Wasserthal, Joshy Cyriac, Michael Bach, Kimia Mozahheb Yousefi, Minh-Son To +5 more
code/data similarity 0.45
Abstract

Background: Patient details and acquisition metadata are important for clinical decisions, image quality control, and automated research pipelines, but may be missing or unreliable in imaging archives. Purpose: To develop and evaluate a fast open-source model that predicts patient and acquisition characteristics directly from CT and MR images. Materials and Methods: Separate 3D ResNet-10 ensembles for CT and MR were trained on 57,291 and 43,200 clinical examinations acquired from 2011 to 2025. Both predicted weight, height, age, sex, contrast presence, vertebral coverage, and image noise. The CT model additionally predicted scanner manufacturer, tube voltage, tube current, convolution kernel, and post-injection time; the MR model predicted sequence class. Performance was evaluated on internal CT (n=501) and MR (n=636) test sets and an external CT dataset (n=54). Results: Internal CT MAEs were 3.90 kg, 3.68 cm, and 4.42 years for weight, height, and age, with sex F1=0.990; corresponding MR results were 4.34 kg, 4.62 cm, 7.13 years, and F1=0.970. The CNN outperformed a segmentation-derived XGBoost baseline for all four core targets in both modalities (adjusted P<=.042). F1 scores were 0.963 for CT contrast, 0.953 for MR sequence, and 0.823 for MR contrast. External CT MAEs were 4.45 kg, 4.05 cm, and 5.17 years, with sex F1=0.971. CPU inference required 20 seconds for CT and 12 seconds for MR. Conclusion: One 3D multitask model per modality can rapidly recover patient and acquisition characteristics from heterogeneous CT and MR examinations. Models are available in TotalSegmentator: https://github.com/wasserth/TotalSegmentator

0.43