Multi-Sensor Fusion for False-Alarm Reduction by Timed Electromechanical Corroboration

Zixuan Fu, Anthony Raffaele, Unnikrishna Pillai

eess.SP physics.med-ph ·submitted 2026-11-21 20:20 UTC · v1

Abstract

A physician checks the clinical electrocardiogram against other observations. A wearable monitor has no physician. This paper shows that a mechanical sensor worn beside the electrocardiogram can take over that check, and that the right added sensors lower false alarms further. A beat is accepted only when a mechanical contraction follows at the delay learned for the wearer. Beat-shaped artifacts injected into resting and exercise recordings passed the electrocardiogram alone almost every time. A seismocardiogram or phonocardiogram as the second channel rejected most while keeping nearly all true beats, strongly at rest and partly during exercise. Shifting the mechanical signal in time removed the benefit. A detection criterion scores an added sensor by the part of its response that the sensors in place cannot predict. With four sensors, this score ranked the sets as their measured false-alarm rates did. With the two sensors it selected, 6.2 percent of the artifacts passed while 95 percent of true beats were kept, half the rate of the best single sensor. In integer arithmetic on a low-cost microcontroller, the detector made the same decisions as the software. These rates come from injected artifacts, not alarms in patients. Natural motion and clinical events remain untested.

Comment
8 pages, 8 figures, 3 tables. Submitted to IEEE Journal of Biomedical and Health Informatics

Closest papers

Nearest neighbours by meaning across the whole index.

Sensor-Free Shockable Rhythm Classification During CPR Using the Existing ECG Channel of Public-Access Automated External Defibrillators

eess.SP ·2026-11-19 ·Jaechan Lim, David Hicks, Matt Valentine, Ki H. Chon
similarity 0.65
Abstract

Automated external defibrillators (AEDs) are public-access devices intended for use by lay rescuers, yet they must pause chest compressions to analyze the rhythm because compressions corrupt the electrocardiogram (ECG). Removing this pause requires a method that uses the hardware already present in the device, fits within its processor budget, and meets the rhythm-recognition performance required for defibrillator approval. Existing approaches rely either on an auxiliary sensor, which entails a hardware redesign, or on an additional network that reconstructs the ECG before classification. This work classifies the compression-contaminated ECG directly, using only the existing ECG channel of the AED. A cascade of three compact convolutional networks trained with different class ratios verifies each decision with a second network biased toward the proposed outcome; disagreement yields an indeterminate outcome. Evaluated on an independent manufacturer-provided cohort of 464 subjects over 15 train-validation partitions, the cascade satisfies both adjudicable American Heart Association criteria under all six noise conditions at an eight-second window and a signal-to-artifact ratio of 0 dB, with sinus-rhythm specificity of 99.20-99.43% and other non-shockable specificity of 95.48-96.20%, leaving only 1.07-1.48% of segments indeterminate. The criteria are not met at -3 dB. Single networks matched in parameter count and inference cost do not reproduce this performance. Each decision evaluates two of the three networks and requires 6.68 ms on a single thread of a general-purpose processor, and eight-bit quantization reduces the model to 1.45 MiB without changing any compliance verdict. Thus, rhythm analysis during ongoing CPR can be implemented with the ECG hardware already present in public-access AEDs, without auxiliary sensing hardware, subject to validation on the target device.

0.28

CardioFusion-AI: Robust ECG--PPG Fusion for Multimodal Physiological Monitoring Under Signal Degradation

eess.SP cs.LGq-bio.QM ·2026-08-26 ·Navaneetha Krishnan Kamalakannan, Janakiraman Kamalakannan
code/data similarity 0.63
Abstract · 7 pages, 4 figures. Under review at IEEE Journal of Biomedical and Health Informatics. Code: https://github.com/ka-cyber/CardioFusion-AI

Wearable electrocardiogram (ECG) and photoplethysmogram (PPG) sensors are complementary but individually fragile: motion artifact, poor contact, and sensor dropout can degrade one or both signals. Fusion strategies that assume both modalities are equally trustworthy can become less reliable than a single clean modality under degradation. We present CardioFusion-AI, a framework whose signal-processing front end, including R-peak and systolic-peak detection, an Orphanidou-type signal-quality index, and beat-by-beat pulse transit time estimation, is validated on 53 real intensive-care recordings (848 windows; heart-rate mean absolute error 1.61 bpm for ECG and 2.78 bpm for PPG) and a real annotated fetal ECG database (R-peak F1 0.89-0.98). We then conduct a controlled synthetic degradation study comparing eight ECG-PPG fusion strategies across six degradation regimes spanning graded corruption and complete modality loss, using five independent training seeds. Attention fusion achieved the lowest descriptive overall error (1.66+/-0.43 bpm). Both adaptive gates reallocated weight toward the healthy modality under complete modality loss, but showed near-zero correlation between gate weight and signal quality under graded degradation (r = 0.10-0.24). Signal-quality conditioning produced a specific improvement under missing-PPG conditions (1.56+/-0.59 bpm), approaching the 1.48 bpm unimodal ceiling. With only five training seeds, no pairwise comparison survives Holm-corrected significance testing; effect sizes and confidence intervals are therefore reported. These results indicate that modality availability and modality quality are functionally distinct problems for adaptive fusion.

0.59

ARTA: Graph Combinatorial Optimization for Robust Interbeat Interval Estimation from Wearable Electrocardiogram Signals

eess.SP cs.CLcs.MS ·2026-09-15 ·Shovito Barua Soumma, Asiful Arefeen, Hassan Ghasemzadeh
similarity 0.59
Abstract · Under Review

Accurate heart rate (HR) and interbeat interval (IBI) estimation from wearable electrocardiogram (ECG) is essential for continuous physiological monitoring, but motion artifacts can create false QRS-like peaks and obscure true R-peaks. Conventional detectors and local interval-repair rules process detections independently and may become unreliable when several consecutive beats are corrupted. We present ARTA (Artifact-Robust Temporal Alignment), a graph-based combinatorial framework that estimates the heartbeat sequence through constrained temporal optimization. ARTA combines candidates from multiple ECG detectors, represents them as nodes in a directed acyclic graph and connects candidates through physiologically plausible RR intervals. Candidate confidence uses detector agreement and local ECG quality, while history-dependent transition costs capture recent IBI consistency and a soft heart-rate prior. Beam-search dynamic programming selects a coherent low-cost path across the recording, suppressing false detections and recovering missed beats when plausible auxiliary candidates are available. On the IEEE Signal Processing Cup exercise dataset with electrode-motion artifacts from -2 to 12 dB, ARTA achieved an HR mean absolute error of 1.79 beats per minute, 54% lower than the strongest single-detector baseline, with 92.38% of windows within +/-5 beats per minute, an R-peak F1 score of 0.95, and an IBI mean absolute error of 6.52 ms. Cross-dataset evaluation on the European ST-T Database further assessed transfer to long clinical ambulatory recordings and a label-free reliability analysis examined conditional performance without using reference annotations for segment selection. ARTA is single-channel, training-free, interpretable and computationally lightweight, operating approximately 27 times faster than real time with peak memory below 2.1 MB in a single-thread CPU simulation.

0.17

IoMT-SecAlarmBench: A Counterfactual Benchmark for Integrity Attacks in IoMT

cs.CR ·2026-08-26 ·Emmanuel C. Ugwuabonyi, Dmitri Perkins
benchmarkcode/dataunusual similarity 0.58
Abstract · 10 pages, 2 figures, 5 tables. Submitted to IEEE International Conference on Big Data (BigData) 2026. Benchmark, generator, evaluation harne…

The Internet of Medical Things (IoMT) combines clinical physiological data with cyber-system information, creating challenges in determining whether an abnormal reading reflects a genuine physiological event, a device fault, or a cyber-attack within the expected physiological range. Answering this requires counterfactual ground truth, which no existing dataset provides. We present IoMT-SecAlarmBench, a semi-synthetic benchmark that injects controlled integrity attacks into genuine coupled ECG+PPG recordings using a structured experimental design combining four attack morphologies, four severity levels, two physiological plausibility conditions, and replay attacks. Each injected window retains its cause, attack subtype, and the clean signal that would have been observed without the attack. We evaluate six detectors from five method families using threshold-independent measures and a matched false-alarm budget. Results show no method consistently detects the most difficult cases: replay attacks and low-amplitude transient spikes remain close to chance-level performance across detectors. Results also reveal a trade-off between detecting attacks and distinguishing them from sensor faults: the best-performing detector on hard cases flags fault/artifact windows at 5.3 times its false-alarm rate on normal data. Three-way classification performs poorly for genuine physiological events, and a leakage audit of a dual-modality network dataset indicates previously reported IoMT intrusion-detection performance is partly driven by identifying information. Benchmark, generation code, preprocessing, evaluation tools, and datasheet are released.

0.43

Certified AI Triage of ICU Alarms

cs.LG ·2026-09-11 ·Mohammed Sameer Syed, Rozhin Yasaei
benchmark similarity 0.57
Abstract

In the VTaC benchmark 71% of ventricular-tachycardia alarms are false, but silencing a real one can delay recognition of a dangerous arrhythmia. We reframe alarm reduction as three-way triage (retain, suppress, or defer) and bound the decision this analysis treats as harmful: among suppressed alarms, the fraction that were genuine stays below a user-set budget with 95% confidence, under i.i.d. event sampling. Alarms sharing a waveform record are dependent, so the clustered analysis is a sensitivity check. On the official split a 5% budget certifies in all three seeds, suppressing 74.8% of false alarms while silencing 1.5% of genuine ones, at AUROC 0.953 and Challenge Score 83.33, numerically comparable to the strongest of the eleven published systems. Our central finding measures what multiplicity costs: the correction charges for every candidate, so a finer grid can certify strictly less. Under held-out calibration the 885-cell grid we declared certifies 1 of 15 fold-runs, while choosing the grid on a separate selection partition certifies 8. We project the calibration volume each budget needs, making an uncertifiable budget a design parameter. Finally, adding a learned reliability dimension to the policy grid did not sharpen the certified frontier.

0.29

BeatFlow-ECG: Rectified Flow for ECG Reconstruction from Indirect Wearable Signals

cs.LG eess.SP ·2026-10-06 ·Mohamed Kamel, Sahar Selim, Walaa Medhat, Tamer Nadeem
similarity 0.56
Abstract · 27 pages. Preprint. Under review

Continuous cardiac monitoring outside clinical settings requires signals that are both informative and practical to collect during daily life. Electrocardiography (ECG) provides rich information about cardiac rhythm and waveform morphology, while wearable photoplethysmography (PPG) is easier to acquire continuously but is only an indirect cardiovascular measurement and is highly sensitive to motion. We present BeatFlow-ECG, a conditional rectified-flow model for reconstructing single-channel ECG from synchronized PPG and inertial measurements. BeatFlow-ECG models reconstruction as conditional transport from noise to ECG using a convolutional encoder-decoder with a transformer bottleneck and explicit flow-time conditioning. Motion information is incorporated through IMU-derived conditioning features, motion-dependent loss weighting, and an easy-to-hard training curriculum. We evaluate the model under leave-one-subject-out protocols on PPG-DaLiA and WESAD. BeatFlow-ECG achieves the best results among the evaluated deterministic, adversarial, and diffusion-based baselines across all reported waveform and beat-timing metrics, with Pearson correlations of 0.983 and 0.986 and R-peak F1 scores of 0.946 and 0.955, respectively. Compared with Conditional DDPM-1D, L1 error decreases from 0.085 to 0.062 on PPG-DaLiA and from 0.074 to 0.055 on WESAD. Additional analyses on PPG-DaLiA show higher correlation in fixed R-peak-relative waveform regions and lower reconstruction error across low-, medium-, and high-motion subsets.

0.25

Learning Cardiac Features: ECG Biometrics Across Time and~Exercise

cs.AI q-bio.TO ·2026-09-18 ·Luca Thiebaud, Paul Chauchat, Mustapha Ouladsine, Stéphane Delliaux
published similarity 0.56
Abstract

Electrocardiograms (ECGs) carry subject-specific patterns enabling reliable individual discrimination, forming the basis of ECG biometrics. Beyond authentication, this paradigm holds significant potential to secure sensitive cardiac data and to serve as a pretext task in self-supervised learning. Yet, most studies remain confined to singlesession, resting data, leaving robustness to temporal and physiological variations largely untested. We address this gap by evaluating ECG biometrics under realistic conditions involving exercise-induced stress and cross-session variability. A Siamese ResNet with late multi-lead fusion strategy is trained on a large ECG dataset extracted from cardiopulmonary exercise tests and evaluated with a exercise-and time-aware protocol, as well as on public benchmarks. This first extensive assessment of ECG biometrics under combined physiological and temporal variability achieves an intra-session rest-to-peak EER of 1.7% and stateof-the-art 3.9% on the CYBHi dataset. Findings support the presence of an intrinsic cardiac signature resilient to physiological and temporal drift.

0.32

A Synchronized Tri-Modality Wireless System for Multi-Point Non-Contact Cardiac Monitoring

cs.AR eess.SPeess.SY ·2026-11-05 ·Wei Ren, Kaili Zhao, Huansheng Yi, Miaoyang Hu, Jianqi Wang +2 more
similarity 0.55
Abstract · 11 pages, 9 figures

Cardiac physiological monitoring is fundamental to evaluating cardiovascular health and diagnosing related diseases. Non-contact radar sensing can capture chest-wall vibrations with spatial specificity, enabling localized physiological monitoring and cardiac assessment. Progress, however, has been limited by the absence of a reliable, research-grade hardware platform for rigorous validation of radar-based methods. Prior systems typically pair radar, namely Doppler cardiography (DCG), with only electrocardiography (ECG) or other coarse references, providing no direct mechanical ground truth and thus constraining fine-grained evaluation of radar outputs. We present a tri-modality wireless system that integrates ECG, seismocardiography (SCG), and DCG under a unified acquisition and control framework for time-aligned recording and analysis. The tri-modality system combines miniaturized wearable ECG/SCG modules with a high-resolution mmWave radar module, supporting spatially resolved sensing and subsequent feature extraction. To our knowledge, this is the first integrated ECG--SCG--DCG hardware system for comprehensive cardiac monitoring; it enables multi-point thoracic acquisition, achieves precise cross-modal synchronization, and maintains specifications suitable for wearable operation. Experiments on a linear-motion platform show close agreement between DCG- and SCG-derived displacements, confirming accurate displacement measurement. Human-subject studies further demonstrate cross-modal consistency and robust extraction of feature points, establishing ECG as the electrical anchor and SCG as the mechanical ground truth for DCG. Overall, the system provides a rigorous, spatially resolved validation platform that advances non-contact cardiac sensing and supports next-generation radar-based monitoring.

0.24