The paper presents a numerical approach for the end-effector trajectory smoothing of a parallel robot designed for minimally invasive pancreatic surgery. The approach is tailored for real-time master-slave control architecture and uses a 3D space mouse for command input for velocity control. The trajectory smoothing is achieved by generating S-curves in the end-effector velocity fields, thus controlling the accelerations, which in turn reduces tissue trauma in the minimally invasive procedures. Real-time control is enabled by segmenting the S-curves based on the command inputs from the 3D space mouse. A special case is considered where the acceleration time is constant for all command inputs. Numeric results demonstrate stable transitions (without abrupt changes) in both the end-effector parameter space and in the active joints parameters, thereby validating the proposed approach. Further work aims to test the approach on an experimental model and integrate it into AI-based training modules.

Journal ref
8th International Workshop on Computational Kinematics, Nov 2025, Madras (Chennai), India

Closest papers

Nearest neighbours by meaning across the whole index.

A Master-to-Simulation Interface with Demonstration-Trajectory Guidance for Vascular Intervention

cs.RO ·2026-10-14 ·Mingyu Zou, Shuxiang Guo, Sheng Cao, Pengcheng Li, Jian Guo +2 more
similarity 0.61
Abstract · 6 pages, 3 figures, submitted to IEEE journals

Robotic vascular intervention can reduce radiation exposure to physicians and improve manipulation stability, but effective guidance remains challenging because conventional fluoroscopic images provide limited spatial information. This paper developed a vascular intervention interactive system that connects a real robotic master controller with the SOFA(Simulation Open Framework Architecture) physical simulation platform and reused successful operation trajectories as visual guidance in physical phantom experiments. Experiments were conducted to validate the master-to-SOFA control loop and to compare three guidance conditions: no guidance, centerline guidance, and successful-trajectory guidance. The control validation results showed reliable command delivery under velocity-mode operation. In the physical phantom experiment, centerline guidance showed no obvious improvement over no guidance. In contrast, successful trajectory guidance provided a clear performance benefit. Compared with the no-guidance condition, it reduced task completion time, collision number, rollback number, and rotation number by approximately 66.8%, 34.3%, 55.6%, and 66.7%, respectively. These results indicate that successful trajectories obtained under instrument-vessel interaction can provide practical guidance cues, offering a simple and effective visual assistance strategy for robotic vascular intervention experiments.

0.32

An Open-Source ROS 2 Framework for Shared Control in Surgical Teleoperation: Integrating the Haply Inverse3 with the da Vinci Research Kit

cs.RO eess.SY ·2026-11-19 ·Balint Varga, Sebastian Fritz, Gergely Kiss, Kristof Takacs, Peter Galambos
benchmarkcode/data similarity 0.59
Abstract

Surgical teleoperation deprives surgeons of touch, forcing visual inference of tissue forces; the commercial platforms now restoring force feedback remain closed, preventing open research on haptic shared control. This paper provides a system contribution: it presents an open-source, ROS 2-native framework that integrates the compact, three-degree-of-freedom (3-DOF) Haply Inverse3 haptic device with the da~Vinci Research Kit (dVRK). The framework includes a novel mapping strategy that combines 3-DOF translational device input with separately tracked VerseGrip orientation into 6-DOF patient-side manipulator commands. Two virtual-fixture-inspired guidance modes serve as reference implementations to evaluate the framework in a within-subjects user study on a teleoperated hot-wire benchmark task in simulation, and the architecture is further demonstrated on physical dVRK hardware. Relative to unassisted manual control, linear guidance significantly reduced both the number of wire contacts and the task completion time, whereas the progressive-stiffness guidance produced consistent but statistically non-significant trends in the same direction. Subjective ratings of workload, agency, and satisfaction showed favorable descriptive trends for the guidance modes. The results demonstrate the feasibility of haptic shared control with a compact device coupled to the dVRK and provide an openly available, reproducible platform for future research. They further suggest that simpler linear feedback may be more robust for inexperienced operators than more complex profiles.

0.25

Surgical Master Console Using General-Purpose Robot Arms and a Separable Articulated Distal Interface: Porcine In-Vivo Evaluation

cs.RO ·2026-09-27 ·Minsung Kim, Seonho Shim, Dongho Yee, Younghoon Noh, Juahn Oh +7 more
similarity 0.59
Abstract · Submitted to IEEE ICRA 2027

High-performance surgical master consoles offer intuitive articulated manipulation but are expensive and difficult to reproduce, while accessible commercial haptic devices lack built-in interfaces for surgical wrist articulation and continuous grasp. We present a laparoscopic master in which general-purpose robot manipulators provide the programmable base and surgical-specific interaction is concentrated in a separable distal adapter. Each 6-DoF arm carries a custom 2-DoF direct-drive wrist, continuous grasp sensing, and clutch-based workspace management; the complete bimanual hardware costs approximately US$5,500, below common bimanual commercial haptic references. To support sustained hand-guided use, we characterize current scaling, gravity, and break-away friction and use them for posture-dependent gravity compensation and motion-gated assistance, reducing continuous actuation and thermal loading. A common master state drives either Isaac Sim or an RCM-constrained dual-FR3 platform. Master-state publication to FR3 target update adds at most 8 ms, and tip tracking shows 0.31-0.66 mm mean delay-aligned residuals. In a porcine cholecystectomy experiment, five common operator-side metrics and hand/tip-speed distributions were compared descriptively with human in-vivo telemetry from 120 clinical Versius recordings of 99 cholecystectomy procedures; no interruption originated from the master console, and no abnormal clinical signs were recorded during the 7-day postoperative observation period. These results demonstrate a surgical master built from general-purpose robot arms with surgical-specific interaction concentrated in a separable distal interface.

0.32

Contact-Stable Deformable Tissue Simulation Using Implicit Integration and Live-Pose Grasp Constraints for Laparoscopic Surgery Robot Policy Evaluation

cs.RO ·2026-09-22 ·Juahn Oh, Dongho Yee, Jinseok Lee, Jiyul Lee, Yechan Seo +7 more
hot direction similarity 0.58
Abstract · Submitted to IEEE ICRA 2027

Closed-loop evaluation of surgical robots requires tissue that deforms, can be grasped and lifted, and reproduces the anatomy in which the robot will operate. We present a simulator in which this tissue is reconstructed from a fixed-view RGB-D recording of the surgical field, composited to remove the instruments, closed into watertight volumes and tetrahedralised; the pipeline was applied unchanged to three specimens of two species (thirteen organs, 146,061 tetrahedra, no inverted elements). For one specimen, the organs are placed in a bimanual cell in which two Franka FR3 arms operate motorised instruments through 6 mm trocars. The core contribution is the numerical and contact design that keeps this cell stable: implicit integration, simulation meshes separate from collision meshes, numerical guards, and a grasp constraint captured at the live tissue pose. In 45 repeated grasp-lifts, a friction grasp held the tissue in 0 of 15 trials and each constraint grasp in 13 of 15; on displaced tissue, a rest-pose constraint produced one-step snaps of up to 17.8 mm, which live-pose capture eliminates. Against the recording, front-surface depth error is 1.33 to 1.41 mm, organ silhouette IoU is 0.80, and in five grasp-lifts reproduced from video the landmark displacement RMSE is 11.8 mm against 14.2 mm for a static prediction. Biofidelity is not claimed; the environment is intended for closed-loop feasibility, safety, contact and policy screening.

0.32

A Unified Dynamic Force Guidance Framework for Performance-Optimized Kinesthetic Teaching

cs.RO ·2026-09-18 ·Chunxin Li, Jianhua Wu, Zhenhua Xiong, Xiangyang Zhu
similarity 0.58
Abstract

Collaborative robots are increasingly deployed in industrial scenarios characterized by frequent product changeovers. As an intuitive programming method, kinesthetic teaching facilitates rapid robot deployment. However, users may overlook the configuration of the robot during kinesthetic teaching, leading to degradation in operational performance. Operational performance refers to the capability of the robot to generate motion and can be quantified by the Minimum Singular Value of the Jacobian matrix. To address this issue, this paper proposes an online dynamic force guidance method that integrates performance constraint and optimization mechanisms. Specifically, variable admittance control maintains the operational performance of the robot above a predefined threshold, while a virtual force actively guides the user to drag the robot towards configurations with improved performance. Experiments are conducted on a 6-DOF collaborative robot, comparing three typical paths in the task space. To evaluate the quality of the taught trajectories, trajectory playback experiments are conducted to analyze the relationship between the operational performance of the robot and the work efficiency. The results demonstrate that the proposed method effectively enhances the operational performance of the robot and consequently improves the work efficiency, holding significant value for reducing production takt time in industrial deployment.

0.33

From Instrument-Mounted Demonstrations to In-Vivo Execution: Learning Bimanual Laparoscopic Appendectomy Without Robot-Collected Demonstrations

cs.RO ·2026-09-22 ·Dongho Yee, Juahn Oh, Jinseok Lee, Jiyul Lee, Yechan Seo +8 more
code/data similarity 0.58
Abstract · Submitted to IEEE ICRA 2027. Project page & Dataset: https://rosota-research.github.io/instrument-mounted-demonstrations-in-vivo/

Most minimally invasive surgery is still performed with hand-held laparoscopic instruments, and the surgeon's instrument kinematics are lost when the operation ends; only the endoscope video is kept. This paper presents an end-to-end pipeline that captures this motion in the operating room and uses it to train a surgical robot policy, validated on live animals. We introduce a surgical instrument-state logger that mounts on the shaft of a standard laparoscopic instrument and recovers its pose and jaw state from an inertial sensor, a time-of-flight sensor and a Hall sensor, with no external camera or tracker. A data pipeline measures the latency of every sensor channel against a robot ground truth and aligns the channels before forming observation-action pairs. On these demonstrations we train a diffusion policy with a fine-tuned DINOv3 backbone, selecting its design by closed-loop rollouts in a physics simulator reconstructed from depth maps of an ex-vivo rabbit appendix. The policy is then retrained on 849 in-vivo demonstrations from four live rabbits and deployed on four additional live rabbits with electrosurgery armed. With the surgeon selecting the surgical phase, the policy completed the appendectomy in three of the four animals. The results show that demonstrations recorded from a surgeon's own instruments are sufficient to train, select and deploy a bimanual surgical policy in vivo. The robot serves only as the timing reference for sensor calibration and as the executor, and collects no demonstrations. Both demonstration corpora are released to support future surgical robot learning research.

0.34

Supervised Autonomous Bimanual Suction and Cauterization for Bleeding Control in Low Grade Splenic Trauma

cs.RO ·2026-10-30 ·Francis Fan, Mackenzie A Eagleson, Yvonne Zhang, Hongting Pan, Lujia Yang +8 more
similarity 0.58
Abstract

Uncontrolled hemorrhage remains a leading cause of preventable trauma mortality. Timely intervention is not always available, making autonomous systems that can aid in bleeding control a promising alternative. Given the safety-critical nature of the task, supervised autonomy, in which the system executes the technical steps while a human retains oversight, represents a necessary intermediate step toward this goal. In this work, we present a supervised autonomous bimanual robotic framework for injury cauterization (eletrosurgical coagulation) in low-grade splenic trauma. The system integrates real-time bleeding segmentation, markerless tool keypoint tracking, and deep tool-tissue contact detection to coordinate two robotic arms for synchronized suction and cauterization. Evaluated on a novel, cannulated ex vivo splenic hemorrhage model, the system achieved an average bleeding surface cauterization coverage of 82.9% across seven distinct injuries. This work demonstrates the feasibility of using a supervised autonomous bimanual system to deliver targeted coagulation across an injury site.

0.51

A VR Training System for Robot-Assisted Vascular Intervention with Dual-Mode Control and Model-Based Interaction Response

cs.RO ·2026-10-14 ·Yining Shang, Shuxiang Guo, Sheng Cao, Pengcheng Li, Bin Wang +1 more
similarity 0.58
Abstract · 8 pages, 7 figures, submitted to a IEEE journal

Virtual reality training for robot-assisted vascular intervention requires accurate robotic control and effective feedback on guidewire--vessel interactions. This paper presents a VR training system combining asynchronous dual-mode command mapping with sensorless interaction estimation. The system supports velocity-controlled navigation and position-driven incremental manipulation. Guidewire deformation and a Hunt--Crossley-type contact response are combined and temporally filtered to estimate interaction intensity, while trajectory-based visualization supports post-procedure review. Tests confirmed distinct command-dispatch patterns across three velocity gears and the position mode. During hybrid-scene navigation, the estimated response remained low during straight insertion and increased as the guidewire bent within the vascular model. Trajectory analysis yielded a model-derived cumulative trauma work (CTW) value of 15.166 in normalized units and localized regions of elevated interaction. In a preliminary study with 15 novices, physical-phantom task success increased from 53.3% to 93.3% after VR training, while mean completion time decreased from 132.4 s to 96.7 s. These results demonstrate the system's dual-mode operation, deformation-sensitive feedback, and potential to improve subsequent physical-phantom task performance.

0.28