Hall Effect-Based Tactile Force Detection Sensor for Robot-Assisted Minimally Invasive Surgery

Charles DeLorey, Michael Brockdorff, Adam Metcalf, Benjamin Calmé, Pietro Valdastri, Dominic Jones

cs.RO ·submitted 2026-10-07 16:24 UTC · v1

Abstract

Robotic minimally invasive surgery has revolutionized surgical practice. However, as the surgeons are mechanically separated from the surgical tools, loss of tactile feedback occurs. We present a compact force sensor with three, millimeter scale sensing elements (taxels) that can be affixed to the grasping face of a robotic surgical tool. The sensor features a miniaturized design using Hall effect sensors and magnets embedded in a deformable elastomer contact layer. Sensor calibration is achieved with a 6 degree of freedom (DoF) parallel robot. Each taxel can detect applied normal and shear forces with average RMSE of 2.133 kPa. The calibrated sensor was validated on phantom and ex vivo porcine tissue, demonstrating tactile sensing during retraction as well as slip. This work introduces a small-scale Hall effect based tactile sensor, enabling sensing for force-sensitive robotic surgical applications.

Comment
9 pages, 12 figures

Closest papers

Nearest neighbours by meaning across the whole index.

A robust single-sensing-element tactile sensor for concurrent pressure and tackiness detection with real-time signal decoupling capability

cs.RO cond-mat.mtrl-sci ·2026-09-29 ·Ying Yang, Mingwei Gu, Jia-Sen Xie, Xingyu Ma, Yan-Na Lu +6 more
similarity 0.69
Abstract

Integrating tackiness sensation into the artificial skin of humanoid robots significantly enhances their cognitive and operational capabilities. However existing tactile sensors face challenges in decoupling of the multimodal signal and stability. Here we present a surface-soft tactile sensor that incorporates a Hall effect sensor and a soft magnetic composite within a robust elastic framework. The sensor surface indents under pressure and bulges prominently when retracted from sticky surfaces dynamically altering the Hall sensor-magnet distance. This generates whole-process-traceable and baseline-separated signals enabling real-time differentiation between pressure and pull-off force. This single-sensing-element design facilitates bimodal sensing at the same contact spot while eliminate stress cross-talk enhancing both accuracy and sensitivity. The fusion of a robust framework and magneto-mechanical sensing mechanism equips the sensor with exceptional reliability and excellent signal baseline stability. This tactile sensor holds substantial potential for advancing robotic capabilities in evaluating adhesive properties monitoring rubber aging precisely handling lightweight objects and cognizing natural objects surface characteristics.

0.20

VecTile: A Low-Cost Flexible Tactile Array for 3-DoF Force Sensing

cs.RO ·2026-10-22 ·Jian Hou, Xin Zhou, Adam J. Spiers
similarity 0.69
Abstract · 8 pages, 8 figures. Submitted to IEEE International Conference on Robotics and Automation (ICRA) 2027

Tactile sensing provides robots with contact information difficult to obtain from vision alone. Shear force is particularly important for understanding tangential contact, guiding corrective motions, and maintaining grasp stability, yet flexible and low-cost tactile arrays often measure only normal force. We introduce VecTile, a low-cost, 6 mm thin, flexible tactile array capable of three-axis force sensing. VecTile combines a piezoresistive layer with passive silicone nubs that convert normal and shear forces into distinguishable pressure patterns. The sensor can be fabricated in-house without specialized equipment. The complete system costs less than 35 USD for a 4x4 array. We validate the sensing mechanism through finite element analysis and hardware experiments. In controlled single-taxel tests over 1-5 N normal force and up to 1.5 N shear force, VecTile achieves root-mean-square errors of 0.39 N and 0.16 N for normal and shear forces. A VecTile calibrated while flat retains accuracy when curved, while a five-minute manual calibration is sufficient for practical use. Two applications demonstrate its utility. First, shear readings are sensitive enough to stop a sharp scalpel from puncturing a balloon after contact. Second, the sensor substantially improves Diffusion Policy-guided power-plug insertion. All designs and code will be released upon paper acceptance. Project website: https://jiannnh.github.io/VecTile/.

0.20

MagCilia: A Compact Magnetociliary Tactile Sensor with 3D Force Sensing for Robotic Contact Perception and Grasping Feedback

cs.RO ·2026-10-07 ·Yu Feng, Hao Wu, Haotian Guo, Haoming Liu, William Su +5 more
highly cited similarity 0.66
Abstract

Robotic grasping and surface exploration benefit from simultaneous measurement of normal and tangential forces and from surface information obtained through contact. Here, we present a compact magnetociliary tactile sensor (MagCilia) that combines a flexible magnetic-cilia structure with a Hall sensor for 3D force sensing. Quasi-static finite element analysis is used to investigate structural deformation and magnetic responses under multidirectional loading. To reconstruct forces from the coupled magnetic channels, we propose causal history fusion regression (CHFR), which combines current magnetic-field measurements with their recent changes. Five-fold cross-validation grouped by calibration record yields root-mean-square errors of 0.40, 0.57, and 0.69 N for Fx, Fy, and Fz, respectively, with corresponding coefficients of determination of 0.93, 0.90, and 0.92. Robotic experiments demonstrate tangential-force-guided gripper adjustment and multi-axis load monitoring under external perturbations. Frequency-domain features of the reconstructed forces distinguish six surface categories with 99.39% accuracy in three-fold cross-validation grouped by acquisition session. An online robotic demonstration additionally identifies all six tested surfaces. These results demonstrate 3D force reconstruction, grasping feedback, and surface recognition using a single compact tactile unit.

0.58

MISTac: A Vision-Based Tactile Sensor for Minimally Invasive Surgery

cs.RO ·2026-08-14 ·Robin Koch, Annabella Mascot, Rayan Younis, Martin Wagner, Stefanie Speidel +3 more
code/data similarity 0.64
Abstract · This work has been submitted to the IEEE for possible publication

Minimally invasive and robot-assisted surgery offer many advantages over traditional open surgery, but deprive surgeons of tactile feedback and the ability to palpate tissue with their fingers. To address this lack of tactile feedback, we introduce the MISTac, a high resolution vision-based tactile sensor specifically designed for palpation in MIS. The sensor has a replaceable sensor tip with a diameter of 8 mm which allows it to fit through the trocars used in minimally invasive surgery. Its modular 3D-printed case design allows the use of bulky off-the-shelf illumination and imaging hardware that can easily be exchanged and upgraded. The sensor has an optical resolution of 176.68 $μm$, a tactile resolution of 250 $μm$, and can resolve forces as little as 24.3 mN. An in vivo study with the sensor shows its usability in minimally invasive surgery. We trained a machine learning model with the tactile data collected in the trial on a tissue classification task achieving an aggregate accuracy of ~84% in a leave-one-out cross validation. Tactile sensors have the potential to one day aid surgeons during minimally invasive surgery with tasks such as tissue classification or intra-operative tumor localization; MISTac is a small step towards this vision. We open-source MISTac at https://github.com/lasr-lab/mistac

0.26

MagPillar: A Soft Magnetopillar Array Tactile Sensor for Robotic Surface Perception

cs.RO ·2026-11-01 ·Haotian Guo, Yu Feng, Hao Wu, Jiazhi Tang, Ningbin Zhang +1 more
similarity 0.64
Abstract

Robotic surface exploration benefits from tactile interfaces that accommodate contact and capture variations in surface characteristics. Herein, we report MagPillar, a tactile sensor inspired by spider tactile hairs that combines a magnetic pillar array with three-axis Hall sensing. A full-array mechanical and magnetic-field model characterizes how pillar deformation affects the superposed fields measured across the array, providing a physical basis for interpreting its contact responses. The compliant pillar architecture translates contact with surface textures and geometric features into distinct magnetic response patterns. Then, in robotic surface perception experiments, spectral features and logistic regression distinguish six material surfaces with an offline accuracy of 99.34% under within-session temporally blocked cross-validation. Continuous scans over periodic protrusions produce amplitude and peak-spacing variations consistent with changes in surface geometry. These findings connect compliant contact mechanics with distributed magnetic sensing, paving a new path for material recognition and qualitative surface-profile perception.

0.35

Tactile Reconstruction of Contact Task Frames and Forces for Hybrid Force/Motion Control

cs.RO ·2026-10-07 ·Antonio Rapuano, Simone Orelli, Barbara Bazzana, Antonio Franchi, Alessandro De Luca
similarity 0.62
Abstract · 14 pages, 9 figures, 7 tables. Video: https://youtube.com/playlist?list=PLejKMZmW8AvI. Dataset: https://doi.org/10.5281/zenodo.23209032

Hybrid force/motion control requires knowledge of the interaction force and of a task frame defining the force- and motion-controlled directions. These quantities are usually obtained from force/torque sensing or model-based residuals, often assuming also a nominal environment model. This work addresses the online estimation of the contact force and a possibly time-varying task frame using only soft optical tactile sensing, under the assumption of locally planar contact with a negligible contact moment. The proposed method maps a single image of the deformed elastomer of a soft optical tactile sensor to observable contact variables: indentation depth, two surface-to-sensor tilt angles, and 3D contact force, each with a per-sample uncertainty estimate. The mapping is learned through a self-labeling acquisition procedure, in which a manipulator imposes controlled contacts while an auxiliary Force/Torque sensor is used offline to provide ground-truth labels. The tactile measurement is then fused with robot proprioceptive data in an Extended Kalman Filter, producing a continuously updated estimate of the contact task frame and of the interaction force. Control experiments with a DigiTac sensor mounted on a UR10 manipulator demonstrate closed-loop contact force regulation against a flat rigid board in linear and angular motion by a human operator, with touch as the only exteroceptive feedback.

0.23

GlowTact: Simple and Compact Vision-Based Tactile Sensing with High Sensitivity and Spatial Resolution

cs.RO ·2026-09-26 ·Yuxiang Ma, Megha Tippur, Pengfei Ye, Sandra Q. Liu, Haonan Chen +2 more
similarity 0.62
Abstract · Project webiste: https://glowtact.github.io/

Vision-based tactile sensors (VBTS) provide rich contact information for robotic manipulation, but existing designs can be hard to simplify and adapt to the size and constraints of humanoid fingertips. We introduce \textbf{GlowTact}, a pressure-responsive vision-based tactile sensing mechanism that directly visualizes contact pressure. \textbf{GlowTact} requires only single-color, non-directional illumination, and the raw tactile image directly represents the pressure distribution without explicit geometry reconstruction. This simple sensing principle enables compact, customizable tactile sensors while preserving high sensitivity and rich spatial detail. We demonstrate gram-scale contact detection, accurate normal-force estimation, and reconstruction of fine contact geometry, including M1 screw threads. These results establish \textbf{GlowTact} as a practical new sensing technology for compact humanoid fingertips, combining a durable nitrile membrane and simple optical design with sensitive and information-rich tactile perception.

0.29

Capstan-driven Continuum Surgical Robot: Design, Modeling, and Perception

cs.RO ·2026-08-13 ·Gang Zhang, Yufu Qiu, Junyan Yan, Wenhui Zeng, Wenlong Lu +1 more
similarity 0.61
Abstract

Shape and force sensing have long been critical bottlenecks in the development of compact capstan-driven continuum surgical robots, primarily due to the difficulty of obtaining cable tension information within the confined capstan assembly. To overcome these challenges, this paper presents an integrated design-modeling-sensing approach based on the concept of actuation-perception co-design. A compliant element is introduced into the motor mounting bracket of the drive system, enabling micro-deformation under the cable reaction force and thereby allowing real-time cable tension measurement without occupying the compact capstan space. To address the modeling complexity arising from unconventional joint configurations introduced by the spatial cable routing strategy, a parallel computation framework based on a multibody short-thick-beam model is proposed, which captures shear effects in short beam segments and synergistic multi-cable interactions while achieving real-time performance. Building on this framework, stable shape and force sensing is achieved by incorporating a proximal multi-axis force/torque sensor as an additional measurement anchor. Following this design-modeling-sensing framework, capstan-driven continuum surgical robots with single- and dual-segment configurations are developed. Experimental results validate the proposed framework in both single- and dual-segment continuum robots, demonstrating real-time tip pose estimation together with contact force and location perception. By enabling cable tension feedback without compromising the compact capstan architecture, the proposed framework makes integrated perception feasible for capstan-driven continuum surgical robots.

0.04