CS Paper Radar

Physically Constrained Kinematic Calibration for High-Precision Robotic Needle Guidance

Prabhu Nithin Gollapudi, Andreas Maier, Franziska Mathis-Ullrich, Water Zhao, Jia Wang, Siming Bayer

cs.RO eess.SYmath.OC ·submitted 2026-10-30 08:35 UTC · v1

Abstract

Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.

Comment
6 pages, 4 figures, 4 tables

Closest papers

Nearest neighbours by meaning across the whole index.

Compressed Orientation Reachability Maps for Sub-Second Robotic Needle Entry Point Planning

cs.RO cs.CGeess.SY ·2026-10-29 ·Prabhu Nithin Gollapudi, Andreas Maier, Franziska Mathis-Ullrich, Water Zhao, Jia Wang +1 more
similarity 0.61
Abstract · 6 pages, 4 figures, 2 tables

In the context of multi-stage trajectory planning for robotic percutaneous liver interventions, identifying a kinematically feasible entry point remains a critical computational bottleneck. Existing methods rely on exhaustive inverse kinematics or randomized searches, incurring delays of 13 to 27 seconds per target and limiting real-time intraoperative adaptation. This paper introduces an Orientation-Based Reachability Map (ORM) that precomputes the kinematic feasibility of a 6 Degrees of Freedom (DoF) serial manipulator across 188,873 workspace positions and encodes orientation feasibility into a compact, indexed 83 MB map. A Just-In-Time (JIT) compiled runtime pipeline replaces live optimization with near-instantaneous array lookups. Experimental evaluation demonstrates a reduction in planning latency from 13.5-26.9 s to 0.289-0.349 s per target while maintaining high geometric fidelity, achieving a mean convex hull volume overlap ratio of 0.933 and a mean alpha-shape volume ratio of 0.976 relative to the baseline reachable point cloud. The proposed ORM framework enables sub-second entry point planning and provides an efficient computational framework for robotic needle insertion in percutaneous liver ablation.

0.37

Active Constraints for Simultaneous Guidance and Safety in Handheld Micromanipulation

cs.RO ·2026-10-27 ·Roger Ho, Robert A. MacLachlan, Cameron N. Riviere
similarity 0.60
Abstract · 8 pages, 3 figures, submitted to ICRA 2027

In micromanipulation, erroneous hand motion limits unaided manipulation accuracy, which can affect task performance and safety in critical applications like microsurgery. Handheld robots (micromanipulators) enable motion accuracy beyond human limits by providing robotic assistance in a low-cost, compact, intuitive form factor. Particularly, active constraints are an effective form of robotic assistance, which aids the operator by restricting the robot's permissible motion based on task-specific requirements or objectives, thereby providing task guidance and enforcing safety. However, existing formulations of active constraints for handheld systems are limited, most notably because of the inability to provide guidance and enforce safety simultaneously. Implementing active constraints with optimization-based methods can address these limitations, but existing implementations are designed for free-standing robots, and neglect that handheld robots actuate relative to the operator's hand with a limited workspace. In this work, we propose an optimization-based active constraint framework for handheld micromanipulation that enables simultaneous guidance and safety enforcement. The formulation was implemented in a handheld robot and demonstrated on a medically relevant artificial task. Results show that the proposed formulation reduced safety violation rate by more than two orders of magnitude relative to baselines, while recovering the threefold accuracy loss from enforcing safety alone.

0.38

Force-Based Offset Estimation for Keyed Peg-in-Hole Assembly Using Local Gaussian Process Regression

cs.RO ·2026-08-18 ·Chandra Yuvesh Aubeeluck, Abilash Philip Madavath, Augustin Raju, Nicolas Pyschny, Felix Hackelöer +1 more
publishedhot direction similarity 0.59
Abstract · 6 pages, 11 figures. Accepted and presented at the 2026 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM 2026), …

Key-keyway assembly tasks impose strict geometric constraints and are highly sensitive to grasp pose deviations in uncertain environments. This work presents a force-based offset estimation method for keyed peg-in-hole assembly, embedded within a perception-validation-insertion pipeline. Residual misalignment is estimated directly from wrist force/torque measurements using a local KNN-Gaussian Process hybrid regressor. The framework distinguishes between two contact regimes, hard collision and guided chamfer insertion, and routes inference to a dedicated model for each. Regime classification is achieved via a contact-window duration threshold. KNN combined with a deterministic search using the results of a post-grasp monocular visual validation contributes to an increased accuracy of the regressor model. This approach achieves accurate radial offset estimation in chamfered peg insertion, during a keypoint detection-based pick and place application. Experiments using the integrated force/torque sensor of a collaborative robot arm showed an increase in insertion success rate from 67% to 87% after the pipeline was applied.

0.46

Before the Tipping Point: Force-Guided Active Perception for Shape-Agnostic Estimation of 3D Centers of Mass

cs.RO ·2026-09-11 ·Steven M. Hyland, Jing Xiao, Cagdas D. Onal
published similarity 0.56
Abstract · For associated video, visit https://stevenmhyland.com/assets/videos/IROS2026.mp4

Estimating the 3D center of mass of unknown objects is challenging when grasping is infeasible, geometry is irregular, or mass distribution is uneven. We present a force-based method that estimates CoM height and mass from a single sub-critical tipping experiment by a robot manipulator. The robot applies a quasistatic elevated push and retract motion, using force-angle measurements recorded during tipping to identify parameters from the object trajectory. Our proposed push-retract cycle mitigates frictional bias, enabling generalized fitting. We experimentally validate our method using a robot manipulator with a six-axis force torque sensor on varying types of objects without prior shape information and without specific models. We also propose a method to prevent toppling, keeping the object in a sub-critical tipping regime by leveraging a safety margin. In experimental studies, our method recovers mass, CoM height, and toppling angle with relative errors below 5.0 percent across all unknown objects. This work demonstrates reliable 3D inertial parameter estimation under proper safety thresholds in tipping. Our proposed method informs and enables reliable non-prehensile manipulation and robotic grasping of challenging objects that were previously infeasible.

0.59

Shaft-Configuration-Adaptive Catheter Tip Position Estimation via Motor-History Conditioned Residual Learning

cs.RO eess.SY ·2026-09-22 ·Peihan Zhang, Michael C. Yip, Ankur Kapoor, Young-Ho Kim
similarity 0.54
Abstract

Tendon-driven continuum manipulators are widely used in medical applications, where accurate tip-position estimation is essential for precise navigation and instrument positioning. However, patient anatomy and procedural setup impose task-dependent unknown shaft configurations, while friction, slack, and compliance introduce hysteresis, making tip estimation challenging. This paper presents a motor-history-conditioned gated recurrent unit (GRU) residual estimator for three-dimensional catheter tip estimation without direct shaft-configuration sensing. First, an initial multidirectional sweep strategy is applied to calibrate a geometric catheter model backbone, and encode the motor-angle and drive-torque response into a shaft-configuration context vector. During subsequent motion, the context conditions a GRU that predicts a task-space residual correcting this backbone, relying on motor measurements alone. The context remains fixed for the current shaft configuration, while the recurrent state captures the evolving actuation history. Across four disposable intra-cardiac echocardiography catheters and 16 bent shaft configurations, the method achieves 3.3mm open-loop tip RMSE, a 59% reduction relative to the constant-curvature baseline.

0.39

Human-Like Redundancy Resolution for Real-Time Bimanual Humanoid Teleoperation

cs.RO ·2026-10-12 ·Chao Tan, Ming Shin Cindy Sung, Shuxian Liang, Chaoqun Wan, Zhen Huang +1 more
similarity 0.54
Abstract · 8 pages, 10 figures, 3 tables. Project website: https://human-like-redundancy.github.io/

Real-time bimanual humanoid teleoperation requires accurate wrist tracking while keeping arm configurations predictable. Because a 7-DoF arm admits multiple elbow configurations for the same wrist pose, wrist-only tracking can cause elbow drift or abrupt configuration changes. Fixed anthropomorphic constraints can stabilize elbow posture, but they can also block the redundant adjustments needed near singular configurations. We propose a human-like redundancy-resolution framework that combines adaptive elbow guidance with singularity-robust wrist tracking. Shoulder-wrist geometric retargeting generates robot-scale wrist targets. An adaptive elbow-offset criterion uses the offset direction to guide posture and scales this guidance by the offset magnitude, so the redundant degree of freedom is progressively released as the human arm straightens. A hierarchical velocity solver prioritizes wrist tracking and projects the adaptive elbow guidance into the wrist-task null space. In a fixed-wrist-target ablation, adaptive guidance reduces mean elbow-direction error by 54.4% compared with removing this guidance. In a workspace-boundary stress test, singularity-robust tracking reduces peak raw joint-velocity demand by 97.9% compared with the undamped pseudoinverse. Real-world bimanual tasks and cross-platform retargeting show applicability across robot morphologies.

0.23

ActiveReg: Information-Driven Active Regional Probing for Partial-to-Full Bone Registration

cs.RO ·2026-10-07 ·Tiancheng Li, Yingyu Wang, Peter Walker, Liang Zhao, Shoudong Huang
similarity 0.54
Abstract · Preprint

Accurate bone registration is essential for orthopedic surgical navigation and robotic assistance. In a common workflow, the surgeon needs to identify and probe a number of prescribed locations on the exposed bone surface to obtain data points for registration, which can be difficult and time-consuming under limited surgical exposure. Reducing the number of required points while providing clear probing guidance would ease the surgeon's acquisition task intra-operatively. We present ActiveReg, a closed-loop framework that recommends probing regions rather than individual points, allowing flexibility in the exact contact location. A D-optimal planner that takes into account the already acquired points and the uncertainty of current registration information helps achieve accurate registration with fewer points. An online assessment combines pre-update innovation consistency with uncertainty at predefined surgical task locations improves the reliability of completion decision, without requiring ground truth. Simulations across four anatomical scenarios demonstrated that ActiveReg can achieve competitive accuracy compared with baseline registration methods while using substantially fewer acquired points. Real phantom experiments with optical and electromagnetic tracking demonstrated comparable mean target registration error to the Gradient-SDF framework, using only 24-28 acquired points instead of 771-1263.

0.25

Force/Torque-Based Kinematic Adaptation for Robotic Manipulation Tasks

cs.RO ·2026-08-21 ·Carl Glen Henshaw
hot direction similarity 0.54
Abstract · 29 pages including appendices, 3 figures

Contact-rich robotic manipulation requires an accurate model of the kinematic relationship between a robot's joints and the task features it senses. This relationship is rarely known exactly: it changes with each tool the robot picks up and shifts, sometimes almost instantaneously, as contact modes change --- especially for multi-fingered hands that make and break contact at points that are not exactly prescribed, as in full-hand grasping. This paper develops an adaptive scheme that estimates that relationship online, using only joint-angle sensing and a wrist-mounted force/torque sensor, with no exteroceptive measurement of the tool tip. We derive a provably stable kinematic update law that identifies the kinematics of an unknown tool from force/torque feedback alone, and prove stability of both the rigid case and the case with a compliance controller as an inner loop. We show that identification is confined to the directions the motion excites --- so that, for example, a tool's length is unobservable under a rigid insertion push, while a compliant loop's passive yielding partially excites it; and that with a second-order admittance the compliant certificate holds unconditionally in continuous time. We also pose the combined control and estimation problem as a Quadratic Program (QP): the formulation yields the prediction term of the update law exactly but, instructively, cannot reproduce the tracking adaptation term. We validate the scheme in simulation on a peg-in-hole insertion. This work is the first step in a research program aimed at factoring manipulation learning into a task policy which can be learned in isolation of the robot, for instance by reinforcement learning, and an adaptive kinematic component that adapts online to the particular robot, hand, or tool in use.

0.29