An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research

Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang, Mackenzie A Eagleson, Ravi Prakash, Axel Krieger, Chunwoo Kim

cs.RO ·submitted 2026-11-16 19:03 UTC · v1
code/data

Abstract

Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each activation request. In bench testing, the interposer recorded all 12 activation and deactivation events from six pedal presses. Ex vivo demonstrations included monopolar and bipolar coagulation of splenic tissue and monopolar cutting during gallbladder dissection. Optical tracking of the Fenestrated Bipolar Forceps showed that calibration brought pitch and yaw gains from 0.904 and 1.14 to 1.00 and 1.02, respectively. The module enables researchers to record instrument motion, video, and the timing and source of electrosurgical commands during laparoscopic tasks.

Comment
Submitted to the 2027 International Symposium on Medical Robotics (ISMR). † Equal contribution. ‡ Equal supervision

Closest papers

Nearest neighbours by meaning across the whole index.

DASH: A da Vinci Adapter for Serial-link and Humanoid Robots as an Accessible Platform for Surgical Robotics Research

cs.RO ·2026-10-05 ·Sara Wickenhiser, Junrong Zhou, Zekai Liang, Lizzie Peiros, Liu Shanglei +1 more
similarity 0.68
Abstract · 8 pages, 9 figures

Robotic minimally invasive surgery offers well-documented clinical benefits, but the cost and infrastructure requirements of purpose-built platforms limit access in rural and lower-resourced facilities. Recent work has teleoperated general-purpose robots for laparoscopic tasks and in vivo procedures, but relied on handheld instruments coupled through passive linkages rather than native robotic actuation. Instead, we adapt da Vinci Classic and Xi instruments onto general-purpose robots that can integrate in clinical workflows. We present DASH, a da Vinci Adapter for Serial-link and Humanoid platforms, consisting of two types of adapters that require no modification to the instruments themselves. Each adapter is compatible with many robotic platform load capacities, engages the instrument's native latch, and wirelessly identifies inserted tools to load instrument-specific kinematics and coupling matrices. Teleoperated experiments demonstrate the efficacy of DASH and its benefits for enabling research access to surgical robotic platforms.

0.20

ARTS: An Open-Source, Reproducible Three-Arm Platform for Autonomy Research in Minimally Invasive Surgery

cs.RO ·2026-11-11 ·Jiawei Ge, Chunwoo Kim, Antony Goldenberg, Douglas McDonough, Justin Liao +16 more
code/data similarity 0.60
Abstract · Submitted to the 2027 International Symposium on Medical Robotics (ISMR). † Equal contribution

Research on surgical autonomy needs open platforms that any laboratory can build, yet the most widely shared platform depends on donated clinical hardware, and open alternatives on industrial arms each lack part of the surgical chain. We present the Autonomous Robotic Translational System (ARTS), an open-source three-arm platform for autonomy research in minimally invasive surgery, built entirely from purchasable parts for about USD 137,000. Two UR5e arms hold wristed da Vinci X/Xi instruments, a UR5 arm holds a stereo endoscope with two FLIR cameras, the surgeon console provides haptic handles, footswitches and a stereo viewer, and a hardware stop independent of the software lets a supervisor freeze all three arms in place. The ROS 2 software pins every dependency, installs from the same scripts on bare metal or in containers, and records stereo images with hardware-measured timing. On the real arms, ARTS commands each arm at its controller's full rate (e.g., 500 Hz on the UR5e), the two cameras expose within 12 microseconds of each other, hand motion reaches the arm in 58 ms, and the endoscope image reaches the monitor in 68 ms. The ISO 9283 pose repeatability is 0.14-0.25 mm, comparable to a da Vinci Research Kit (dVRK) Classic arm on the same protocol, and an optical tracker measured a mean remote-center-of-motion error of 0.34-0.37 mm. Three expert and three novice operators performed the Fundamentals of Laparoscopic Surgery (FLS) peg transfer through a laparoscopic trainer box, and one of the experts sutured on a suture pad. The hardware designs, firmware and software are available at https://github.com/MRE-Lab-UMD/arts.

0.44

Precision Through Play: Enhancing Fundamental Laparoscopic Skill Acquisition Via Gamified XR and Pseudo-Haptics

cs.HC cs.GR ·2026-10-31 ·Bita koohestani, Ashkan Rezaei, Azam Bastanfard
similarity 0.59
Abstract

Extended Reality (XR) technologies have shown potential to improve laparoscopic skills acquisition. Among fundamental skills in laparoscopic surgery is accurate instrument handling. Appropriate instrument alignment and a perpendicular path ensure movement efficiency and improved performance. This work introduces an interactive XR platform designed to support an immersive, engaging, and accessible environment that facilitates continuous practice. This platform incorporates a gamification layer enriched by pseudo-haptics that generate a motivational feedback loop, allowing trainees to continuously practice instrument manipulation and angle control. The platform monitors instrument orientation, computes angular deviation from an ideal reference vector, and assigns scores based on user performance. A point-based scoring component reinforces correct technique and penalizes deviations in tool alignment beyond predefined thresholds. Furthermore, a pseudo-haptic feedback mechanism is integrated to provide real-time tactile feedback when deviation exceeds tolerance. Combined with low-cost hardware and accessible deployment, the framework aims to increase training scalability, engagement, and improve skill acquisition. Preliminary evaluations indicate that integrating gamified feedback significantly improves precision, reduces angular variance, and encourages sustained practice.

0.55

From Instrument-Mounted Demonstrations to In-Vivo Execution: Learning Bimanual Laparoscopic Appendectomy Without Robot-Collected Demonstrations

cs.RO ·2026-09-22 ·Dongho Yee, Juahn Oh, Jinseok Lee, Jiyul Lee, Yechan Seo +8 more
code/data similarity 0.59
Abstract · Submitted to IEEE ICRA 2027. Project page & Dataset: https://rosota-research.github.io/instrument-mounted-demonstrations-in-vivo/

Most minimally invasive surgery is still performed with hand-held laparoscopic instruments, and the surgeon's instrument kinematics are lost when the operation ends; only the endoscope video is kept. This paper presents an end-to-end pipeline that captures this motion in the operating room and uses it to train a surgical robot policy, validated on live animals. We introduce a surgical instrument-state logger that mounts on the shaft of a standard laparoscopic instrument and recovers its pose and jaw state from an inertial sensor, a time-of-flight sensor and a Hall sensor, with no external camera or tracker. A data pipeline measures the latency of every sensor channel against a robot ground truth and aligns the channels before forming observation-action pairs. On these demonstrations we train a diffusion policy with a fine-tuned DINOv3 backbone, selecting its design by closed-loop rollouts in a physics simulator reconstructed from depth maps of an ex-vivo rabbit appendix. The policy is then retrained on 849 in-vivo demonstrations from four live rabbits and deployed on four additional live rabbits with electrosurgery armed. With the surgeon selecting the surgical phase, the policy completed the appendectomy in three of the four animals. The results show that demonstrations recorded from a surgeon's own instruments are sufficient to train, select and deploy a bimanual surgical policy in vivo. The robot serves only as the timing reference for sensor calibration and as the executor, and collects no demonstrations. Both demonstration corpora are released to support future surgical robot learning research.

0.34

HumynexSurg-1: A Curated Expert Liposuction Dataset

cs.RO cs.LG ·2026-09-20 ·Rhea Huang, David L. Matlock, Laurence Reich
similarity 0.57
Abstract · 15 pages, 7 figures, 13 tables. Version 0 dataset release

Robot foundation models learn manipulation from large demonstration corpora, but surgery is missing from those corpora: across the 780-hour Open-H surgical collection, one dataset carries synchronized force and none covers an aesthetic procedure. Liposuction is the hard case, because the instrument works under the skin and the surgeon operates by feel and by judgment. Humynex Robotics builds curated expert datasets for this kind of procedure. HumynexSurg-1 is the first release: a master liposuction surgeon performing on porcine abdominal tissue while narrating every decision, recorded with synchronized suction pressure, six-axis hand force/torque, top-down RGB-D video, side video and a lavalier microphone -- 14 episodes, 42,738 frames, 35.6 minutes, 356 utterances of which 95% compile into a liposuction-specific label schema. The capture follows a patent-pending sensing plan organized around the quantities a policy needs, so a channel captured today by a model can be upgraded to a sensor tomorrow without changing the data format. This release captures the instrument motion as a tool-hand track in the side video and provides the force channel as state; the funded capture adds a measured 6-DoF handle pose, a validated force channel, ultrasound imaging of the fat layer, and palpation sensing. As a proof of concept, NVIDIA Isaac GR00T N1.7 fine-tunes on the dataset with no custom code in under an hour per run and learns the recorded sessions; scaling probes on the same episodes show where further gains come from: every new session lowers the error on an unseen session. The dataset, its label schema, its quality-assurance reports and its evaluation protocol are the product; the next capture, many short sessions across fat regions with the sensors named here, is what the probes point to.

0.25

Recording Hand-Held Laparoscopic Instrument Motion in the Operating Room: Magnetometer-Free Fusion of Inertial, Range and Visual Sensing

cs.RO ·2026-09-22 ·Jiyul Lee, Dongho Yee, Juahn Oh, Jinseok Lee, Yechan Seo +7 more
hot direction similarity 0.56
Abstract · Submitted to IEEE ICRA 2027

Most minimally invasive procedures are still performed with hand-held laparoscopic instruments, yet only the endoscopic video is retained; the instrument motion that expresses surgical skill, and that could support skill assessment and robot learning, is lost. Pose from video alone remains millimeters to centimeters off, and an instrument-mounted inertial measurement unit (IMU) cannot rely on its magnetometer, whose field changed with tool pose and between sessions in our measurements. We present a surgical instrument-state logger that clips onto a conventional instrument without modifying the part that enters the patient and fuses a six-axis IMU and a time-of-flight (ToF) rangefinder with a markerless camera in an error-state Kalman filter under the remote center of motion (RCM) of the trocar. Heading comes from the shaft silhouette, segmented by a U-Net, in place of the magnetometer: the rotation-angle error is 0.200°, against 3.58° from the accelerometer and magnetometer alone. Against a Franka Research 3 manipulator, and without alignment to it, the displacement error over 300 translation trials was 1.21mm RMS and the relative-rotation error over 180 rotation trials 0.34° RMS. On continuous trajectories, tracked and displayed in real time, the absolute tip error was 1.22mm (programmed) and 3.04mm (teleoperated) after post-hoc tuning of three filter parameters, and the full fusion beat every sensor subset. Because the estimator uses no magnetic measurement, its accuracy does not rely on an undisturbed field. The same clip-on device could thus record metric tip trajectories during routine hand-held laparoscopy, while displaying the insertion depth and attitude that are hidden once the instrument is inside the patient.

0.38

Contact-Stable Deformable Tissue Simulation Using Implicit Integration and Live-Pose Grasp Constraints for Laparoscopic Surgery Robot Policy Evaluation

cs.RO ·2026-09-22 ·Juahn Oh, Dongho Yee, Jinseok Lee, Jiyul Lee, Yechan Seo +7 more
hot direction similarity 0.56
Abstract · Submitted to IEEE ICRA 2027

Closed-loop evaluation of surgical robots requires tissue that deforms, can be grasped and lifted, and reproduces the anatomy in which the robot will operate. We present a simulator in which this tissue is reconstructed from a fixed-view RGB-D recording of the surgical field, composited to remove the instruments, closed into watertight volumes and tetrahedralised; the pipeline was applied unchanged to three specimens of two species (thirteen organs, 146,061 tetrahedra, no inverted elements). For one specimen, the organs are placed in a bimanual cell in which two Franka FR3 arms operate motorised instruments through 6 mm trocars. The core contribution is the numerical and contact design that keeps this cell stable: implicit integration, simulation meshes separate from collision meshes, numerical guards, and a grasp constraint captured at the live tissue pose. In 45 repeated grasp-lifts, a friction grasp held the tissue in 0 of 15 trials and each constraint grasp in 13 of 15; on displaced tissue, a rest-pose constraint produced one-step snaps of up to 17.8 mm, which live-pose capture eliminates. Against the recording, front-surface depth error is 1.33 to 1.41 mm, organ silhouette IoU is 0.80, and in five grasp-lifts reproduced from video the landmark displacement RMSE is 11.8 mm against 14.2 mm for a static prediction. Biofidelity is not claimed; the environment is intended for closed-loop feasibility, safety, contact and policy screening.

0.32

Surgical Master Console Using General-Purpose Robot Arms and a Separable Articulated Distal Interface: Porcine In-Vivo Evaluation

cs.RO ·2026-09-27 ·Minsung Kim, Seonho Shim, Dongho Yee, Younghoon Noh, Juahn Oh +7 more
similarity 0.55
Abstract · Submitted to IEEE ICRA 2027

High-performance surgical master consoles offer intuitive articulated manipulation but are expensive and difficult to reproduce, while accessible commercial haptic devices lack built-in interfaces for surgical wrist articulation and continuous grasp. We present a laparoscopic master in which general-purpose robot manipulators provide the programmable base and surgical-specific interaction is concentrated in a separable distal adapter. Each 6-DoF arm carries a custom 2-DoF direct-drive wrist, continuous grasp sensing, and clutch-based workspace management; the complete bimanual hardware costs approximately US$5,500, below common bimanual commercial haptic references. To support sustained hand-guided use, we characterize current scaling, gravity, and break-away friction and use them for posture-dependent gravity compensation and motion-gated assistance, reducing continuous actuation and thermal loading. A common master state drives either Isaac Sim or an RCM-constrained dual-FR3 platform. Master-state publication to FR3 target update adds at most 8 ms, and tip tracking shows 0.31-0.66 mm mean delay-aligned residuals. In a porcine cholecystectomy experiment, five common operator-side metrics and hand/tip-speed distributions were compared descriptively with human in-vivo telemetry from 120 clinical Versius recordings of 99 cholecystectomy procedures; no interruption originated from the master console, and no abnormal clinical signs were recorded during the 7-day postoperative observation period. These results demonstrate a surgical master built from general-purpose robot arms with surgical-specific interaction concentrated in a separable distal interface.

0.32