CS Paper Radar

Feasibility of Multifrequency Neural Network-Based Wave Inversion in Ultrasound Time-Harmonic Elastography

Héloïse Bustin, Tom Meyer, Paul Spiesecke, Jakob Jordan, Anja Hennemuth, Ingolf Sack

eess.IV cs.CV ·submitted 2026-11-20 16:25 UTC · v1

Abstract

Ultrasound time-harmonic elastography (THE) is a non-invasive imaging technique that measures soft-tissue stiffness. Relying on the same principle of time-harmonic shear wave excitation as established magnetic resonance elastography (MRE), THE offers full field-of-view stiffness mapping at short acquisition times and sub-second temporal resolution across organs. However, reconstruction of shear wave speed maps from acquired tissue wavefields remains challenging due to speckle noise and sparse spatial coverage of the lateral component. Deep learning methods have shown promising results for image reconstruction in transient-based ultrasound elastography modalities but remain unexplored in THE. In MRE, MF-ElastoNet, a 2D multifrequency neural network-based wave reconstruction method trained on synthetic physics-based wave data, has shown high generalizability across dynamic ranges and resolutions. We investigated the potential of a multifrequency neural network-based wave inversion for THE, by performing a first feasibility analysis. To this end, we adapted the processing pipeline of MF-ElastoNet from MRE to THE. We tested our approach on THE acquired phantom data from the algorithms for mapping elastic modulus Challenge, as well as on in vivo liver THE in 4 healthy volunteers and 14 patients with focal liver lesions, including benign focal nodular hyperplasia, and malignant hepatocellular carcinoma and adenocarcinoma. MF-ElastoNet was successfully adapted to THE with in vivo liver testing. Compared with k-MDEV, the state-of-the-art wave inversion method for THE, Bland-Altman analysis showed a mean bias of -0.02 m/s (95% limits of agreement: -0.36 to 0.33 m/s).

Closest papers

Nearest neighbours by meaning across the whole index.

Physics-Informed Deep Learning for Shear Wave Speed Estimation in MR Elastography

eess.IV cs.LGphysics.med-ph ·2026-11-19 ·Stefan Martin, Jakob Schattenfroh, Patrick Schuenke, Felix Frederik Zimmermann, Ingolf Sack +2 more
top venuecode/data similarity 0.74
Abstract · 13 pages (incl. cover page), 6 figures, 2 tables. Accepted version of the article published in IEEE Transactions on Biomedical Engineering. …

Objective: Magnetic Resonance Elastography (MRE) is a non-invasive imaging technique for mapping biomechanical properties of in vivo tissue, including shear wave speed (SWS), but involves intrinsically slow data acquisition and an ill-posed wave inversion. Instead of relying on handcrafted image priors, we propose a data-driven approach jointly combining image reconstruction and MRE inversion for robust SWS estimation from undersampled k-space data. Methods: Our physics-informed reconstruction framework comprises two blocks: a model-based neural network (NN)-regularized reconstruction module and a phase-gradient inversion (k-MDEV) calculating SWS from the reconstructed images. Concatenating both blocks yields an end-to-end trainable method to estimate SWS directly from measured k-space data. We evaluated the method on retrospectively highly undersampled brain MRE data and compared it to a total variation (TV) minimization-based approach. We assessed the impact of end-to-end training (qualitative images and SWS maps as targets) versus pre-training (qualitative images as targets) and applied the method also to in vivo data. Results: Our approach significantly reduces NRMSE by 30% compared to TV. End-to-end training improves SWS estimation over separate image reconstruction and SWS calculation. Conclusion: Accurate SWS quantification is possible at acceleration factors up to 19. Our method significantly outperforms TV, highlighting the need for data-driven regularization in this challenging MR problem. Further, our approach successfully generalizes to in vivo data. Significance: We present the first end-to-end trainable MRE reconstruction method for estimating SWS maps directly from k-space. NN-based reconstruction can enable rapid stiffness mapping for dynamic studies, functional imaging, and real-time clinical feedback.

0.51

A hybrid CNN-adjoint optimization framework for reconstruction of viscoelastic tissue properties in magnetic resonance elastography

cs.LG math.OCmath.NA ·2026-10-02 ·Anwesa Dey, Johann Rudi, Elena Cherkaev
similarity 0.71
Abstract

Magnetic resonance elastography (MRE) is a noninvasive imaging modality for quantifying the viscoelastic properties of soft tissues from shear wave propagation. Recovering the complex-valued shear modulus from measured displacement fields leads to a severely ill-posed inverse problem, particularly in the presence of noise and limited boundary excitations. We investigate adjoint-based optimization, convolutional neural network (CNN) reconstruction, and a hybrid framework combining both approaches. The forward model is based on a scalar form of the modified stationary Stokes system with a complex shear modulus. We establish well-posedness of the forward problem, existence of minimizers, and first-order optimality conditions for the adjoint-based formulation, and implement a nonlinear conjugate-gradient method with Armijo line search. While PDE-constrained optimization can accurately refine coefficient reconstructions, its performance depends strongly on initialization. We therefore construct a two-dimensional CNN that maps complex-valued displacement measurements to spatially varying complex shear modulus fields and provides rapid, informative initial reconstructions. The proposed hybrid method uses the CNN reconstruction to initialize the adjoint-based optimization, yielding faster convergence and improved accuracy. The CNN is trained on coefficient fields containing individual perturbations and tested on both individual and previously unseen combined configurations. Numerical experiments demonstrate that the CNN generalizes to these more challenging configurations, while subsequent PDE-constrained optimization further refines the reconstructed coefficient. These results demonstrate the potential of combining data-driven initialization with physics-based optimization for efficient and accurate MRE reconstruction.

0.12

Initial condition recovery in nonlinear damped viscous photoacoustic tomography using a convolutional neural network-guided gradient-free optimization framework

cs.LG math.OC ·2026-10-01 ·Madhu Gupta, Anwesa Dey, Prapti Tala, Souvik Roy
similarity 0.65
Abstract

Photoacoustic tomography (PAT) is a hybrid imaging modality that combines high optical contrast with high ultrasonic resolution for biomedical imaging applications. In this work, we investigate the inverse problem of recovering the initial pressure distribution from boundary measurements in the presence of nonlinear acoustic propagation and viscous attenuation effects. To model these phenomena more accurately, we consider a nonlinear damped viscoelastic wave equation incorporating spatially varying sound speed, temporal attenuation, and nonlinear propagation mechanisms. We first establish the well-posedness of the corresponding forward problem using a Galerkin approximation combined with energy estimates and a fixed-point argument. For the inverse problem, we derive existence, uniqueness, and local uniqueness results under suitable assumptions through a harmonic extension reduction, spectral Laplace transform techniques, and observability estimates. To numerically reconstruct the initial pressure field, we develop a hybrid reconstruction framework that combines a convolutional neural network (CNN) with a gradient-free optimization strategy based on the sequential quadratic Hamiltonian (SQH) method derived from Pontryagin's maximum principle. The CNN is used to generate an informative initial guess, while the SQH framework enforces the governing PDE dynamics during the reconstruction process. Numerical experiments demonstrate that the proposed hybrid strategy significantly improves reconstruction quality, contrast, and robustness compared to standalone time-reversal and CNN-based approaches.

0.29

EL-DIP: A Deep Image Prior Framework for Accurate,Efficient, and Noise-Robust Elastography

eess.IV cs.CE ·2026-10-22 ·Hao Liu, Fuyi Li, Dong Liu, Yue Mei, Sheng Mao
similarity 0.65
Abstract

Elastography is a key technique for non-invasive imaging, yet it remains severely ill-posed and sensitive to measurement noise, causing blurring and artifacts. We propose EL-DIP, a training-data-free framework based on deep image prior (DIP) that reconstructs heterogeneous shear modulus distributions. Inverse distance weighting interpolation enables arbitrary-shaped domains, and adjoint sensitivity analysis provides efficient gradient computation. Numerical and experimental studies show that EL-DIP accurately recovers inclusion geometry and modulus across different noise levels. Under extreme noise, image fidelity improves by approximately 97.1% over conventional methods. We believe that this method has great potential as a robust tool for biomedical diagnosis,biomechanical characterization, and non-destructive evaluation.

0.20

Ultrasound Elastography: Excitation Methods and Low-Cost Alternative

cs.CV eess.IV ·2026-11-11 ·Corin F. Otesteanu
published similarity 0.64
Abstract

Ultrasound elastography (or elasticity imaging) is a method that characterizes the tissue mechanical properties with the purpose of aiding the diagnosis of certain pathologies. Many novel clinical applications use the assumption that these tissue properties are changed in the presence of pathological processes. Characterization of tissue mechanical properties is achieved by imaging the response of the tissue to an induced stress or excitation. A plethora of approaches have been used in the scientific community to induce excitation in tissue, which usually require additional tools or devices than are costly. Furthermore these are usually integrated in the imaging pipeline using additional software or hardware for control. An alternative that is simpler, faster to integrate (and low-cost) is proposed in this work, by using readily available smartphones as a source of mechanical excitation. The induced displacements are used to solve the elasticity inverse problem (IP) using a finite element method (FEM) approach. Preliminary elasticity reconstructions of a phantom that mimics tissue are similar to when using a dedicated actuator as means of inducing harmonic displacements, illustrating the feasibility of our low-cost ultrasound elasography approach.

0.36

Pulse-Echo Ultrasound Methods for Speed-of-Sound Estimation: A Review

eess.IV physics.med-ph ·2026-08-23 ·Can Deniz Bezek, Orcun Goksel
survey similarity 0.64
Abstract

Speed of sound (SoS) is fundamental to ultrasound beamforming and a promising quantitative biomarker for tissue characterization. Although transmission- and reflector-based methods have demonstrated the diagnostic value of SoS, their need for dedicated hardware or constrained acquisition geometries limits clinical usability. Pulse-echo SoS imaging addresses these limitations by estimating SoS with conventional ultrasound probes and data acquisition paradigms. This review provides a structured overview of pulse-echo SoS estimation methods, categorized by the source of information used, the estimated SoS representation, and the reconstruction strategy. We review methods by their locality, e.g., global, layer-wise, and spatially-resolved predictions, as well as in terms of prediction strategies including gradient-free search, optimization-based solutions, analytical closed-form approximations, and deep learning-based methods. Finally, we summarize emerging diagnostic applications in the breast, liver, and muscle, while discussing challenges and possibilities with SoS imaging.

0.28

Deep-Learning-Assisted Full Waveform Inversion for Attenuation Reconstruction in Ultrasound Computed Tomography

eess.SP physics.med-ph ·2026-11-17 ·Mingrui Liu, Michael L. Oelze
similarity 0.63
Abstract

Acoustic attenuation is more challenging to reconstruct and has received less attention than sound speed in ultrasound computed tomography using full-waveform inversion (FWI). In this study, a deep-learning (DL)-assisted FWI framework was proposed for attenuation reconstruction based on our previous work. Some numerical experiments were conducted to investigate the strategy and the effectiveness of using DL to help accelerate and improve the reconstruction quality of attenuation imaging in FWI. The results indicated that DL saved computational cost by 28% and improved attenuation reconstruction accuracy by 70% when combined with sufficiently accurate sound speed estimates. The result supports the potential of DL initialization to improve the computational efficiency and reconstruction quality of joint SOS and attenuation FWI.

0.38

Identifying the Shear Piezoelectric Coefficient using Physics-Informed Neural Networks under Exact Boundary and Initial Constraints

cs.LG physics.comp-phcond-mat.mtrl-sciphysics.app-ph ·2026-11-02 ·Suhas Suresh Bharadwaj
similarity 0.61
Abstract · 19 pages, 9 figures

The shear piezoelectric coefficient $e_{15}$ governs the shear-mode electromechanical coupling in piezoelectric micromachined ultrasonic transducers (PMUTs) and surface acoustic wave (SAW) devices, but measuring it typically requires destructive resonance testing on dedicated specimens. On simulated data, $e_{15}$ can instead be recovered non-destructively from sparse, noisy subsurface wavefield observations using a constrained physics-informed neural network (PINN). By building exact spatial distance functions and temporal multipliers directly into the network output, the clamped, grounded boundaries and resting initial state hold automatically, leaving the loss with only the interior equations of coupled electro-elastodynamics and the data misfit. Non-dimensional scaling brings all governing coefficients within one decade of O(1), and the setup removes the boundary and initial-condition penalty weights that soft-constraint methods must tune. In a six-way comparison (plane-strain PZT-5H, manufactured wavefield, 30 dB noise), a soft-penalty baseline with adaptive weights converges to an erroneous $e_{15}$ in four of the five runs (51.7% mean error), and fixed penalties reach an error of 0.85% only after a second-order optimization stage. The constrained formulations reach mean $e_{15}$ errors between 0.15% and 0.51% with held-out displacement errors below 1%. Within these variants, the spatial multiplier gives rapid first-order convergence, and the quadratic temporal multiplier lowers the early-time electrostatic residual by a factor of 6.8 versus a linear one. On finite-element data along one subsurface line, the full formulation recovers $e_{15}$ with 1.19% mean error at 30 dB and keeps it between 0.65% and 2.71% across a 40 to 0 dB noise sweep. Finally, training on displacements alone still recovers $e_{15}$ with mean errors of 0.45% on finite-element and 1.64% on manufactured data.

0.24