CS Paper Radar

An MRI-Guided Robotic System to Improve Hippocampal Access for Epilepsy Interventions

John E. Peters, Abby M. Grillo, Mahshid Mansouri, Daniel S. Esser, Sarah Garrow, Nithin S. Kumar, Dario J. Englot, Joseph S. Neimat, William A. Grissom, Robert J. Webster, Eric J. Barth

cs.RO ·submitted 2026-09-11 22:26 UTC · v1
unusual

Abstract

This paper presents an MRI-guided robotic system that improves hippocampal access by delivering a curved needle-like laser ablator through the foramen ovale, a natural opening in the base of the skull. Both the delivery path and the curved nature of the needle improve upon current clinical straight-line laser interstitial thermal therapy (LITT) as measured by hippocampal cannulation percentage. We describe the design of the robotic system which includes positioning, aiming, and curved needle deployment stages, followed by experimental results assessing cannulation percentages using MRI images in phantoms. In three curvilinear, single-insertion experiments, we achieved cannulation percentages of 93.5%, 96.5%, and 60.4%, exceeding reported clinical averages of 50-60%. Seizure control is believed by physicians to be a function of hippocampal volume treated, and our system provides a means of treating a greater volume of the hippocampus with LITT-based interventions.

Closest papers

Nearest neighbours by meaning across the whole index.

Velocity Optimization via Thermal Imaging Enhances Autonomous Tumor Resection

cs.RO ·2026-10-20 ·Naveed D. Riaziat, Jiawei Ge, Ethan Kilmer, Michael W. Mikula, Paul Maria Scheikl +4 more
similarity 0.53
Abstract · submitted to IEEE Access

Objective: To demonstrate and evaluate THERMO, a thermal-velocity optimization algorithm for autonomous electrosurgery, in ex vivo porcine tissue using the UR10e-based ASTR system. Methods: Thermography for Electrosurgical Rate Modulation via Optimization (ThERMO), a closed-loop algorithm that uses real-time thermal imaging data to minimize thermal injury while ensuring accurate cut margins, was tested on ex vivo porcine tissue. Histological analysis was performed to detect thermal injury extent at a cellular level, and 3D photogrammetric measurement was used to verify margin accuracy. ThERMO's performance was compared against two previously used fixed velocity methods. Results: ThERMO outperformed both fixed velocity comparison methods without requiring data-driven training or iterative tuning, reducing thermal injury by up to 52%, improving cut accuracy by up to 59%, and maintaining task completion speed. Conclusions: Closed-loop thermal feedback via ThERMO improves autonomous robotic electrosurgery by simultaneously reducing thermal injury and improving margin accuracy, reinforcing its value over fixed velocity approaches for tumor resection applications. Significance: This work advances autonomous robotic tumor resection by demonstrating that real-time closed-loop thermal feedback control can substantially reduce collateral tissue damage and improve surgical margin accuracy.

0.23

A Master-to-Simulation Interface with Demonstration-Trajectory Guidance for Vascular Intervention

cs.RO ·2026-10-14 ·Mingyu Zou, Shuxiang Guo, Sheng Cao, Pengcheng Li, Jian Guo +2 more
similarity 0.51
Abstract · 6 pages, 3 figures, submitted to IEEE journals

Robotic vascular intervention can reduce radiation exposure to physicians and improve manipulation stability, but effective guidance remains challenging because conventional fluoroscopic images provide limited spatial information. This paper developed a vascular intervention interactive system that connects a real robotic master controller with the SOFA(Simulation Open Framework Architecture) physical simulation platform and reused successful operation trajectories as visual guidance in physical phantom experiments. Experiments were conducted to validate the master-to-SOFA control loop and to compare three guidance conditions: no guidance, centerline guidance, and successful-trajectory guidance. The control validation results showed reliable command delivery under velocity-mode operation. In the physical phantom experiment, centerline guidance showed no obvious improvement over no guidance. In contrast, successful trajectory guidance provided a clear performance benefit. Compared with the no-guidance condition, it reduced task completion time, collision number, rollback number, and rotation number by approximately 66.8%, 34.3%, 55.6%, and 66.7%, respectively. These results indicate that successful trajectories obtained under instrument-vessel interaction can provide practical guidance cues, offering a simple and effective visual assistance strategy for robotic vascular intervention experiments.

0.32

OCTN: Neural OCT Representations for Robot-Guided Precision Intervention

cs.RO ·2026-09-06 ·Ravi Prakash, Ryan P. McNabb, Patrick J. Codd, Shan Lin
code/data similarity 0.49
Abstract

Optical coherence tomography (OCT) offers compact, contactless, micron-scale imaging suitable for intraoperative guidance, but native OCT volumes are discretely sampled, anisotropic, and currently inefficient for downstream geometric reasoning and robot integration. We present OCTN (pronounced "octane"), an implicit neural representation framework that converts volumetric OCT scans into a continuous, differentiable, and spatially faithful tissue-intensity field. OCTN uses a two-stage hybrid training strategy that combines supervision from acquired voxels with inter-slice interpolations, preserving B-scan fidelity while improving continuity in sparsely sampled regions. For versatility, we first show that OCTN enables fast volumetric reasoning by storing the learned tissue representation natively on the GPU, supporting intensity-based spatial queries with up to 43x speedup over conventional CPU processing. We then demonstrate OCTN-enabled OCT-guided robotic laser surgery where the continuous tissue representation supports implicit surface discovery and surface-constrained path planning via multiple optimization strategies, including Newton- and SGD-based optimization. Next, OCTN enables reconstruction of dense volumetric structure from sparsely acquired B-scans, while reducing acquisition time by 4x and preserving clinically relevant structures. Across the newly generated Duke TissueOCT dataset and public OCT datasets, OCTN achieves robust, high-fidelity reconstruction with PSNR > 30 dB and training time < 10 s, while preserving surface consistency within 10 $μ$m Chamfer distance relative to baseline reconstruction. The TissueOCT dataset and code are publicly available at raprakashvi.github.io/octn

0.20

Selectively Actuated Fiberbots for High-precision Minimally Invasive Surgery

cs.RO ·2026-10-08 ·Jinshi Zhao, Libaihe Tian, Alex Ranne, Qindong Zheng, Yue Li +7 more
similarity 0.49
Abstract · 7 pages, 7 figures

Long, slender instruments are essential for minimally invasive surgery (MIS), but achieving precise distal manipulation becomes increasingly difficult when actuation extends along the entire instrument. Here, we present an electrothermally actuated fiber-based robot designed for MIS procedures, in which Joule heating is concentrated in the distal active segment. The fiber is fabricated using a thermal drawing process and uses embedded metallic wires for thermal actuation. This robotic fiber, 1.9 mm in diameter and 1.2 m in length, can precisely manipulate surgical laser fibers with positional precision below 100 $μ$m. The compact size of the fiber robot allows it to pass through the working channel of a flexible endoscope, while keeping both devices mechanically uncoupled to allow independent motion. A phantom demonstration combining the robotic fiber with a da Vinci instrument conceptually demonstrated its compatibility with existing endoscopic systems for clinical applications. Our robotic fiber enables precise distal motion and holds the potential to improve clinical outcomes by enhancing precision in MIS procedures.

0.33

Occupancy Network-Guided Autonomous Robotic Partial Nephrectomy

cs.RO cs.CV ·2026-09-14 ·Ethan Kilmer, Pit Henrich, Jiawei Ge, Paul M. Scheikl, Laura Connolly +11 more
top venue similarity 0.47
Abstract

Autonomous soft-tissue cancer surgery has been limited to interventions on organ surfaces, because current systems cannot perceive and adapt to anatomy once it deforms or is cut. We introduce the first vision-guided autonomous system capable of performing complete tumor resections for partial nephrectomy. Our system integrates conditional occupancy networks, trained entirely in a physics-based simulation, that infer full 3-D anatomy (tumor, margin tissue, and kidney) from single-view partial point clouds. These occupancy networks maintain intraoperative tracking even as tissue is cut and deformed, enabling adaptive planning and execution. The surgical platform combines a depth camera for capturing surface point clouds, dual robotic arms for electrosurgical cutting and vacuum-based tissue manipulation, and an autonomous control strategy for tumor resection. In patient-derived hydrogel phantoms under an open partial nephrectomy setting, the robot performed eight consecutive autonomous tumor resections comprising 77 electrosurgical cuts, with all cuts achieving negative surgical margins and 1.61 $\pm$ 0.48 mm mean absolute margin error. This work demonstrates, for the first time, a foundation for supervised autonomous closed-loop, imaging-driven, margin-negative tumor removal in phantoms.

0.30

An Anatomy-guided 3D Segmentation Pipeline for Prostate Surgical Phantoms from CT Scans

cs.CV ·2026-11-05 ·Iven Chen, Kuan-I Chung, Ayberk Acar, Susheela Sharma Stern, Alejandro O. Chara +2 more
similarity 0.47
Abstract · 11 pages, 2 figures, 3 tables, 10 algorithms. Extended abstract accepted at SPIE Medical Imaging 2027: Image Processing

Full autonomy has been a long-term goal in robotic surgery and a near-to-medium-term goal for prostatectomy. Due to the relatively closed structure of the procedure, its standard-of-care robotic/laparoscopic solutions, and procedure economics, prostate procedures are an optimal early target for surgical automation. Prototyping and testing of possible solutions is done in constructed laboratory phantoms, either synthetic tissues or animal phantoms. To enable measurement of perceptual and robotic motion error, phantoms undergo volumetric imaging; these images are critical to both a potential deployed automatic system, used in planning motion and treatment (cutting, ablation, suturing, etc.), as well as post-hoc assessment of system performance. Rapid iteration on these phantom models yields small, inconsistent datasets, while the current workflow for human annotation is often anatomically unreliable and prohibitively time-consuming, rendering deep learning unsuitable at reasonable scales. To address these limitations, we present an automated 3D segmentation pipeline that uses classical medical image-processing techniques to encode the known anatomy of the phantoms as explicit geometric and intensity priors, which requires no training corpus and is transferred to each new phantom revision without re-annotation. Across 57 scans spanning multiple phantom revisions, our pipeline recovers the capsule interior at 0.83 median Dice on 11 contrast-suppressed scans and detects all 32 annotated focal lesions without false positives, completing a full scan in 2-8 minutes on commodity CPU hardware versus 1-2 hours by hand.

0.35

Is Deformable Image Registration Ready for Brain Metastasis Reirradiation Dose Accumulation? A Longitudinal MRI Benchmark of Registration Accuracy

eess.IV cs.CV ·2026-08-27 ·Hengjie Liu, Manju Sharma, Xinyi Fu, Di Xu, Ke Sheng
benchmark similarity 0.47
Abstract

Dose accumulation is increasingly important in adaptive radiation therapy and reirradiation, but its clinical validity depends on the performance of deformable image registration (DIR). Reirradiation of brain metastases (BMs) with stereotactic radiosurgery (SRS) provides a controlled but clinically meaningful DIR test case: intra-subject brain deformation is usually limited after rigid alignment, yet recurrent lesions can undergo substantial local shape and volume changes that rigid registration cannot capture and can affect dose accumulation. We benchmarked a wide range of learning-based and optimization-based DIR methods on 87 manually screened longitudinal contrast-enhanced T1-weighted MRI lesion pairs from an institutional BM SRS retreatment cohort. Learning-based methods pretrained on healthy-brain MRI were evaluated zero-shot and after instance-specific optimization (ISO) or tumor-proximity target-specific optimization (TSO). Registration was assessed using lesion overlap (Dice), surface distance metrics (HD95 and sASD), target-volume recovery, and runtime and memory. Pretrained learning-based methods showed variable zero-shot performance, while ISO/TSO improved all tested learning-based families. However, optimization-based methods remained the best-performing approach while maintaining reasonable runtime. These findings suggest that even state-of-the-art DIR methods do not yet provide sufficiently accurate and consistent registration for unmonitored use in brain metastasis reirradiation dose accumulation. Because accurate registration is a prerequisite for deformable dose accumulation, clinical application will require case-level quality control and direct assessment of how registration uncertainty affects downstream dose metrics.

0.21

Autonomous Robotic Navigation for Endovascular Brain-Computer Interface Access

cs.RO cs.LG ·2026-10-02 ·Harry Robertshaw, Weijie Qi, Nikola Fischer, Alejandro Granados, Thomas C. Booth +1 more
similarity 0.46
Abstract

Endovascular brain-computer interfaces (BCIs) avoid craniotomy but require precise device delivery through anatomically variable cerebral veins. This work presents the first demonstration of in vitro autonomous robotic navigation for endovascular BCI access in the cerebral venous system. Soft Actor-Critic controllers were trained in silico for two sequential tasks spanning the right internal jugular vein to the superior sagittal sinus, using geometric augmentation of one training anatomy. Navigation was evaluated in a training anatomy and an anatomically unseen hold-out model over 250 in silico episodes and five fluoroscopy-guided in vitro robotic runs per task-anatomy condition, comprising 1,000 simulated episodes and 20 physical runs overall. Task recurrent predictors were also evaluated for online identification of impending navigation failure. In silico success rates for Tasks A and B were 85.6% and 98.4% in the training anatomy and 42.0% and 91.6% in the hold-out anatomy, respectively. Fourteen of 20 physical runs were successful (70% overall), including 80% success for Task B in the hold-out phantom. In silico the predictors detected 99.3-100.0% of failures with false-alarm rates of 0.8-6.7%. During in vitro evaluation, predicted risk increased before failed episodes, but elevated probabilities during some successful runs showed reduced calibration after transfer. These results demonstrate the feasibility of autonomous cerebral venous access and show how online failure prediction could support human oversight, while also identifying anatomical generalization and sim-to-real calibration as priorities before preclinical translation.

0.24