Selectively Actuated Fiberbots for High-precision Minimally Invasive Surgery

Jinshi Zhao, Libaihe Tian, Alex Ranne, Qindong Zheng, Yue Li, Xiaotong Guo, Ziqi Yang, Ali Anil Demircali, Mohamed E. M. K. Abdelaziz, Bruno Gil Rosa, Ferdinando Rodriguez y Baena, Burak Temelkuran

cs.RO ·submitted 2026-10-08 18:59 UTC · v1

Abstract

Long, slender instruments are essential for minimally invasive surgery (MIS), but achieving precise distal manipulation becomes increasingly difficult when actuation extends along the entire instrument. Here, we present an electrothermally actuated fiber-based robot designed for MIS procedures, in which Joule heating is concentrated in the distal active segment. The fiber is fabricated using a thermal drawing process and uses embedded metallic wires for thermal actuation. This robotic fiber, 1.9 mm in diameter and 1.2 m in length, can precisely manipulate surgical laser fibers with positional precision below 100 $μ$m. The compact size of the fiber robot allows it to pass through the working channel of a flexible endoscope, while keeping both devices mechanically uncoupled to allow independent motion. A phantom demonstration combining the robotic fiber with a da Vinci instrument conceptually demonstrated its compatibility with existing endoscopic systems for clinical applications. Our robotic fiber enables precise distal motion and holds the potential to improve clinical outcomes by enhancing precision in MIS procedures.

Comment
7 pages, 7 figures

Closest papers

Nearest neighbours by meaning across the whole index.

Helping Hand: A Continuum Robotic Grasper for Traction in Single-Channel Endoscopic Submucosal Dissection

cs.RO ·2026-11-11 ·Bendavid Maimon, Thompson Wong, Aedan Mangan, Sukjun Kim, Michael C. Yip +1 more
similarity 0.63
Abstract · 7 pages, 9 figures

Endoscopic submucosal dissection (ESD) is a minimally invasive approach for removing large polyps from the colon and offers many benefits compared to other surgical approaches. However, ESD is technically very challenging to perform, leading to limited adoption by physicians. We propose the Helping Hand, a two-segment continuum robotic grasper that, together with an electrosurgical knife, enables tissue traction and cutting within a single 3.7 mm working channel. We achieve this functionality by creating a slot at the base of the Helping Hand that lets the electrosurgical knife extend out, enabling countertraction and cutting through one instrument lumen. We evaluate the Helping Hand's workspace and its ability to provide traction forces from different starting configurations. Finally, we demonstrate the full workflow by isolating and removing tissue in an ex vivo chicken thigh, illustrating the potential for the Helping Hand to help improve ESD procedures.

0.29

Design and Characterization of a Unified Continuum Robot Actuation Platform

cs.RO ·2026-10-14 ·Yunti Xu, Aedan Mangan, Miheer Potdar, Alex Lu, Tania K. Morimoto
similarity 0.61
Abstract

Continuum robots enable dexterous access to confined spaces in applications such as minimally invasive surgery and industrial inspection. However, their actuation systems are often custom-designed for specific robot architectures and tasks, limiting reuse across publications, institutions, and applications. This paper presents a unified actuation platform that decouples actuation hardware and software from continuum robot design. The platform provides modular actuation across continuum robot classes, scales to a large number of degrees of freedom (DOFs), and integrates with conventional serial manipulators for hybrid architectures. We validate the platform on an open-sourced surgical robotic system and experimentally characterize the actuation modules, achieving sub-millimeter translational accuracy and less than $5^\circ$ rotational error over ranges of $30$~mm and $180^\circ$, respectively. We further demonstrate the platform in two surgical use cases: teleoperation of a continuum camera in a constrained environment and bimanual manipulation using $2.53$~mm concentric-tube-robot-based forceps alongside a conventional wristed instrument. All hardware designs and ROS-based software are open-sourced to facilitate reuse and further research in continuum and surgical robotics.

0.31

DASH: A da Vinci Adapter for Serial-link and Humanoid Robots as an Accessible Platform for Surgical Robotics Research

cs.RO ·2026-10-05 ·Sara Wickenhiser, Junrong Zhou, Zekai Liang, Lizzie Peiros, Liu Shanglei +1 more
similarity 0.56
Abstract · 8 pages, 9 figures

Robotic minimally invasive surgery offers well-documented clinical benefits, but the cost and infrastructure requirements of purpose-built platforms limit access in rural and lower-resourced facilities. Recent work has teleoperated general-purpose robots for laparoscopic tasks and in vivo procedures, but relied on handheld instruments coupled through passive linkages rather than native robotic actuation. Instead, we adapt da Vinci Classic and Xi instruments onto general-purpose robots that can integrate in clinical workflows. We present DASH, a da Vinci Adapter for Serial-link and Humanoid platforms, consisting of two types of adapters that require no modification to the instruments themselves. Each adapter is compatible with many robotic platform load capacities, engages the instrument's native latch, and wirelessly identifies inserted tools to load instrument-specific kinematics and coupling matrices. Teleoperated experiments demonstrate the efficacy of DASH and its benefits for enabling research access to surgical robotic platforms.

0.20

Electrospun Fields: 3D Nano-Fiber Material Computation as Design Method

cs.GR cond-mat.softcs.RO ·2026-08-24 ·Wai Lok Wan, Ayah Mahmoud, Sergio Mutis, Avantika Velho, Annie Xing +1 more
publishedcode/dataunusual similarity 0.55
Abstract · 15 pages, 10 figures. Published in Proceedings of the ACM on Computer Graphics and Interactive Techniques (SIGGRAPH 2026). Open-source hardw…

We present a robotic electrospinning platform and design method for depositing nanofiber membranes onto non-planar, three-dimensional conductive geometries. Conventional electrospinning relies on fixed emitters and planar grounded collectors, which restricts deposition to flat substrates: on concave geometries, field shielding prevents fibers from reaching recessed regions, and material bridges across elevated features instead. We address this with a custom end-effector integrated with a six-axis UR20 arm. The tool carries a localized stepper-driven syringe pump that maintains consistent polymer flow independent of orientation, and routes high-voltage DC (up to 25 kV) directly to a robot-mounted needle, turning the arm into a mobile emitter with full kinematic control over position, orientation, working distance, and traversal velocity. Toolpaths that continuously reorient the emitter along surface normals give access to concave topologies unreachable by fixed-axis systems. We characterize the resulting deposition behavior in two parts. A catalog of four bio-compatible polymer systems (PEO, PVA, keratin-PEO, silk-PEO) establishes the operating envelope, reporting deposition speed, jet stability, fiber size, alignment, and durability for each. A taxonomy of 3D-printed conductive scaffolds spanning geometric primitives, hybrid compositions, and square, triangular, and hexagonal lattices links collector geometry to fiber alignment, density, and cross-void bridging. We also demonstrate programmable grounding, in which selectively energized pins in an array steer deposition without changing physical geometry. Assembly instructions and toolpath-generation code are released as an open-source repository.

0.49

Design and Modeling of a Single-Port Three-Arm Robotic Tool for Minimally Invasive Neurosurgery

cs.RO ·2026-09-22 ·Nazia H. Dana, Harith S. Gallage, Ismail A. Auta, Dhanvi Yuvaraj, Ronghuai Qi
similarity 0.54
Abstract · 6 pages, 11 figures, submitted to IEEE ROBIO 2026

Surgical robots require highly dexterous and compact robotic systems capable of operating effectively within confined anatomical spaces. However, due to limited access provided by a single incision, the miniaturization and maneuverability of these robots still need to be improved. In this paper, we propose the design and modeling of a single-port three-arm robotic tool containing one major cannula (7.14 mm outer diameter (OD)) and three steerable minor cannulas (1.93 mm OD). By integrating the proposed 12 degrees-of-freedom (DoFs) steerable robotic tool with a 7-DoF robotic arm, this robotic system can potentially achieve multi-arm manipulation capability. We present the design of the steerable robotic tool consisting of tendon-driven joints controlled by a compact actuation system, derive the kinematic model, and validate both the static and kinematic models through experiments. The performance is evaluated with the root mean square error (RMSE) and mean absolute error (MAE) computed between the experimental data and the kinematic model.

0.40

Contact-Stable Deformable Tissue Simulation Using Implicit Integration and Live-Pose Grasp Constraints for Laparoscopic Surgery Robot Policy Evaluation

cs.RO ·2026-09-22 ·Juahn Oh, Dongho Yee, Jinseok Lee, Jiyul Lee, Yechan Seo +7 more
hot direction similarity 0.54
Abstract · Submitted to IEEE ICRA 2027

Closed-loop evaluation of surgical robots requires tissue that deforms, can be grasped and lifted, and reproduces the anatomy in which the robot will operate. We present a simulator in which this tissue is reconstructed from a fixed-view RGB-D recording of the surgical field, composited to remove the instruments, closed into watertight volumes and tetrahedralised; the pipeline was applied unchanged to three specimens of two species (thirteen organs, 146,061 tetrahedra, no inverted elements). For one specimen, the organs are placed in a bimanual cell in which two Franka FR3 arms operate motorised instruments through 6 mm trocars. The core contribution is the numerical and contact design that keeps this cell stable: implicit integration, simulation meshes separate from collision meshes, numerical guards, and a grasp constraint captured at the live tissue pose. In 45 repeated grasp-lifts, a friction grasp held the tissue in 0 of 15 trials and each constraint grasp in 13 of 15; on displaced tissue, a rest-pose constraint produced one-step snaps of up to 17.8 mm, which live-pose capture eliminates. Against the recording, front-surface depth error is 1.33 to 1.41 mm, organ silhouette IoU is 0.80, and in five grasp-lifts reproduced from video the landmark displacement RMSE is 11.8 mm against 14.2 mm for a static prediction. Biofidelity is not claimed; the environment is intended for closed-loop feasibility, safety, contact and policy screening.

0.32

Surgical Master Console Using General-Purpose Robot Arms and a Separable Articulated Distal Interface: Porcine In-Vivo Evaluation

cs.RO ·2026-09-27 ·Minsung Kim, Seonho Shim, Dongho Yee, Younghoon Noh, Juahn Oh +7 more
similarity 0.54
Abstract · Submitted to IEEE ICRA 2027

High-performance surgical master consoles offer intuitive articulated manipulation but are expensive and difficult to reproduce, while accessible commercial haptic devices lack built-in interfaces for surgical wrist articulation and continuous grasp. We present a laparoscopic master in which general-purpose robot manipulators provide the programmable base and surgical-specific interaction is concentrated in a separable distal adapter. Each 6-DoF arm carries a custom 2-DoF direct-drive wrist, continuous grasp sensing, and clutch-based workspace management; the complete bimanual hardware costs approximately US$5,500, below common bimanual commercial haptic references. To support sustained hand-guided use, we characterize current scaling, gravity, and break-away friction and use them for posture-dependent gravity compensation and motion-gated assistance, reducing continuous actuation and thermal loading. A common master state drives either Isaac Sim or an RCM-constrained dual-FR3 platform. Master-state publication to FR3 target update adds at most 8 ms, and tip tracking shows 0.31-0.66 mm mean delay-aligned residuals. In a porcine cholecystectomy experiment, five common operator-side metrics and hand/tip-speed distributions were compared descriptively with human in-vivo telemetry from 120 clinical Versius recordings of 99 cholecystectomy procedures; no interruption originated from the master console, and no abnormal clinical signs were recorded during the 7-day postoperative observation period. These results demonstrate a surgical master built from general-purpose robot arms with surgical-specific interaction concentrated in a separable distal interface.

0.32

An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research

cs.RO ·2026-11-16 ·Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang +4 more
code/data similarity 0.54
Abstract · Submitted to the 2027 International Symposium on Medical Robotics (ISMR). † Equal contribution. ‡ Equal supervision

Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each activation request. In bench testing, the interposer recorded all 12 activation and deactivation events from six pedal presses. Ex vivo demonstrations included monopolar and bipolar coagulation of splenic tissue and monopolar cutting during gallbladder dissection. Optical tracking of the Fenestrated Bipolar Forceps showed that calibration brought pitch and yaw gains from 0.904 and 1.14 to 1.00 and 1.02, respectively. The module enables researchers to record instrument motion, video, and the timing and source of electrosurgical commands during laparoscopic tasks.

0.32