CS Paper Radar

Velocity Optimization via Thermal Imaging Enhances Autonomous Tumor Resection

Naveed D. Riaziat, Jiawei Ge, Ethan Kilmer, Michael W. Mikula, Paul Maria Scheikl, Michael Kam, Leila J. Mady, Axel Krieger, Jeremy D. Brown

cs.RO ·submitted 2026-10-20 03:22 UTC · v1

Abstract

Objective: To demonstrate and evaluate THERMO, a thermal-velocity optimization algorithm for autonomous electrosurgery, in ex vivo porcine tissue using the UR10e-based ASTR system. Methods: Thermography for Electrosurgical Rate Modulation via Optimization (ThERMO), a closed-loop algorithm that uses real-time thermal imaging data to minimize thermal injury while ensuring accurate cut margins, was tested on ex vivo porcine tissue. Histological analysis was performed to detect thermal injury extent at a cellular level, and 3D photogrammetric measurement was used to verify margin accuracy. ThERMO's performance was compared against two previously used fixed velocity methods. Results: ThERMO outperformed both fixed velocity comparison methods without requiring data-driven training or iterative tuning, reducing thermal injury by up to 52%, improving cut accuracy by up to 59%, and maintaining task completion speed. Conclusions: Closed-loop thermal feedback via ThERMO improves autonomous robotic electrosurgery by simultaneously reducing thermal injury and improving margin accuracy, reinforcing its value over fixed velocity approaches for tumor resection applications. Significance: This work advances autonomous robotic tumor resection by demonstrating that real-time closed-loop thermal feedback control can substantially reduce collateral tissue damage and improve surgical margin accuracy.

Comment
submitted to IEEE Access

Closest papers

Nearest neighbours by meaning across the whole index.

Occupancy Network-Guided Autonomous Robotic Partial Nephrectomy

cs.RO cs.CV ·2026-09-14 ·Ethan Kilmer, Pit Henrich, Jiawei Ge, Paul M. Scheikl, Laura Connolly +11 more
top venue similarity 0.60
Abstract

Autonomous soft-tissue cancer surgery has been limited to interventions on organ surfaces, because current systems cannot perceive and adapt to anatomy once it deforms or is cut. We introduce the first vision-guided autonomous system capable of performing complete tumor resections for partial nephrectomy. Our system integrates conditional occupancy networks, trained entirely in a physics-based simulation, that infer full 3-D anatomy (tumor, margin tissue, and kidney) from single-view partial point clouds. These occupancy networks maintain intraoperative tracking even as tissue is cut and deformed, enabling adaptive planning and execution. The surgical platform combines a depth camera for capturing surface point clouds, dual robotic arms for electrosurgical cutting and vacuum-based tissue manipulation, and an autonomous control strategy for tumor resection. In patient-derived hydrogel phantoms under an open partial nephrectomy setting, the robot performed eight consecutive autonomous tumor resections comprising 77 electrosurgical cuts, with all cuts achieving negative surgical margins and 1.61 $\pm$ 0.48 mm mean absolute margin error. This work demonstrates, for the first time, a foundation for supervised autonomous closed-loop, imaging-driven, margin-negative tumor removal in phantoms.

0.30

Incision trajectory tracing for electrosurgical navigation by CNN-based knife contacting frames extraction method

eess.IV cs.CV ·2026-08-14 ·Yu Chun Wang, Kaixu Chen, Naoto Ienaga, Yoshihiro Kuroda
published similarity 0.58
Abstract

Background and Objective: Image-guided surgical navigation has been actively studied because of its advantage of identifying subsurface targets and critical structures, whereas it requires incision trajectories to update the preoperative three-dimensional model dynamically during the surgery. The novelty of this study is the thermal feature distinguishment of whether the electric tools contacting the tissue by Convolutional Neural Network (CNN), and the extraction of the knife contacting frames, to form incision trajectories which can meet with the requirement during the surgery. Methods: This study firstly verified that CNN can classify the thermal images of electric knife and ultrasonic cutter operations separately, and can raise the accuracy of the incision trajectories derived from the connection of the thermal intensity centroid of the frames predicted by CNN as contacting. Results: Our results obtained by employing the electric knife not only reveal a remarkably high accuracy 97.2 % in CNNs identification, but also can achieve an error reduction as high as more than 2.5 times of the incision trajectory prediction as compared to those proceeded in the conventional method. Besides electric knife, the results obtained by employing another electric tool, ultrasonic cutter, reveal a high accuracy up to 93.7 %. Conclusion: In this study, we ensured the possibility of CNN in distinguishing electric tools contacting with the tissue, and confirmed that the proposed method has not only overcome the problem of missing trajectories which usually occurs in the convolutional long-short term memory method but also achieved a remarkable improvement of the accuracy with less limitation.

0.22

Dynamic infrared thermography and computed tomography angiography fusion for DIEP flap perforator selection: a novel measurement system

eess.IV physics.med-ph ·2026-11-16 ·Warre Clarys, Roanne Van Steen, Simon Verspeek, Filip Thiessen, Gunther Steenackers
publishedunusual similarity 0.56
Abstract · 6 pages, 6 figures

Accurate perforator selection is critical for successful deep inferior epigastric artery perforator (DIEP) flap breast reconstruction. While computed tomography angiography (CTA) provides high-resolution anatomical detail, it lacks functional perfusion assessment. Dynamic infrared thermography (DIRT), in contrast, captures superficial rewarming dynamics but lacks objective anatomical validation. This study presents a standardized multimodal measurement system enabling automated fusion of DIRT with CTA-derived perforator coordinates. The framework integrates controlled thermographic acquisition, real-time hotspot detection, metric spatial calibration, and global one-to-one assignment of thermal hotspots to CTA perforators within a predefined periumbilical region of interest. The workflow was evaluated in 12 patients. Of 46 CTA perforators within the region of interest, 84.8% were successfully linked to thermographic hotspots, with a median spatial offset of 14.74 mm. Automated assignment excluded 76.1% of detected hotspots due to lack of anatomical correspondence, thereby reducing subjective interpretation. The proposed system enables quantitative multimodal validation of functional and anatomical imaging and provides objective support for perforator selection in clinical practice.

0.42

ProCut: Probabilistic Cutting Topology for Autonomous Electrosurgical Tissue Dissection

cs.RO ·2026-10-05 ·Xiao Liang, Fei Liu, Florian Richter, Genie Rubaiyat, Michael Yip
similarity 0.54
Abstract

Accurately modeling and tracking the deformation of soft tissue is critical for a wide range of interventional and surgical procedures. However, current methods struggle in scenarios involving topological changes, such as cutting and dissection, due to the inherent non-linearity and discontinuity introduced by explicit changes in connectivity. In this work, we present a novel, fully differentiable framework that enables robust estimation and modeling of topological changes during deformable tracking. Our method introduces a continuous, sigmoid-based formulation to smooth the otherwise discrete event of tissue cutting, making it amenable to gradient-based optimization within a differentiable Position-Based Dynamics (PBD) simulation. To account for uncertainty and improve robustness in the presence of noisy visual data, we incorporate Stein Variational Gradient Descent (SVGD) for particle-based probabilistic inference, generating multiple hypotheses for topological state estimation. Building on this foundation, we develop an autonomous dissection algorithm for thin-shell tissues that leverages topological updates to guide closed-loop cutting trajectory control. We evaluate our approach in both simulated and real-world electrosurgical environments, demonstrating significant improvements in topological estimation accuracy and dissection precision over existing methods. Our results highlight the potential of this framework to advance automation in soft-tissue surgical procedures by enabling reliable perception and control in the presence of complex structural changes.

0.16

An MRI-Guided Robotic System to Improve Hippocampal Access for Epilepsy Interventions

cs.RO ·2026-09-11 ·John E. Peters, Abby M. Grillo, Mahshid Mansouri, Daniel S. Esser, Sarah Garrow +6 more
unusual similarity 0.53
Abstract

This paper presents an MRI-guided robotic system that improves hippocampal access by delivering a curved needle-like laser ablator through the foramen ovale, a natural opening in the base of the skull. Both the delivery path and the curved nature of the needle improve upon current clinical straight-line laser interstitial thermal therapy (LITT) as measured by hippocampal cannulation percentage. We describe the design of the robotic system which includes positioning, aiming, and curved needle deployment stages, followed by experimental results assessing cannulation percentages using MRI images in phantoms. In three curvilinear, single-insertion experiments, we achieved cannulation percentages of 93.5%, 96.5%, and 60.4%, exceeding reported clinical averages of 50-60%. Seizure control is believed by physicians to be a function of hippocampal volume treated, and our system provides a means of treating a greater volume of the hippocampus with LITT-based interventions.

0.26

Selectively Actuated Fiberbots for High-precision Minimally Invasive Surgery

cs.RO ·2026-10-08 ·Jinshi Zhao, Libaihe Tian, Alex Ranne, Qindong Zheng, Yue Li +7 more
similarity 0.52
Abstract · 7 pages, 7 figures

Long, slender instruments are essential for minimally invasive surgery (MIS), but achieving precise distal manipulation becomes increasingly difficult when actuation extends along the entire instrument. Here, we present an electrothermally actuated fiber-based robot designed for MIS procedures, in which Joule heating is concentrated in the distal active segment. The fiber is fabricated using a thermal drawing process and uses embedded metallic wires for thermal actuation. This robotic fiber, 1.9 mm in diameter and 1.2 m in length, can precisely manipulate surgical laser fibers with positional precision below 100 $μ$m. The compact size of the fiber robot allows it to pass through the working channel of a flexible endoscope, while keeping both devices mechanically uncoupled to allow independent motion. A phantom demonstration combining the robotic fiber with a da Vinci instrument conceptually demonstrated its compatibility with existing endoscopic systems for clinical applications. Our robotic fiber enables precise distal motion and holds the potential to improve clinical outcomes by enhancing precision in MIS procedures.

0.33

Contact-Stable Deformable Tissue Simulation Using Implicit Integration and Live-Pose Grasp Constraints for Laparoscopic Surgery Robot Policy Evaluation

cs.RO ·2026-09-22 ·Juahn Oh, Dongho Yee, Jinseok Lee, Jiyul Lee, Yechan Seo +7 more
hot direction similarity 0.52
Abstract · Submitted to IEEE ICRA 2027

Closed-loop evaluation of surgical robots requires tissue that deforms, can be grasped and lifted, and reproduces the anatomy in which the robot will operate. We present a simulator in which this tissue is reconstructed from a fixed-view RGB-D recording of the surgical field, composited to remove the instruments, closed into watertight volumes and tetrahedralised; the pipeline was applied unchanged to three specimens of two species (thirteen organs, 146,061 tetrahedra, no inverted elements). For one specimen, the organs are placed in a bimanual cell in which two Franka FR3 arms operate motorised instruments through 6 mm trocars. The core contribution is the numerical and contact design that keeps this cell stable: implicit integration, simulation meshes separate from collision meshes, numerical guards, and a grasp constraint captured at the live tissue pose. In 45 repeated grasp-lifts, a friction grasp held the tissue in 0 of 15 trials and each constraint grasp in 13 of 15; on displaced tissue, a rest-pose constraint produced one-step snaps of up to 17.8 mm, which live-pose capture eliminates. Against the recording, front-surface depth error is 1.33 to 1.41 mm, organ silhouette IoU is 0.80, and in five grasp-lifts reproduced from video the landmark displacement RMSE is 11.8 mm against 14.2 mm for a static prediction. Biofidelity is not claimed; the environment is intended for closed-loop feasibility, safety, contact and policy screening.

0.32

Anatomy-Aware Dexterity-Driven Design Optimization of Surgical Continuum Robots

cs.RO cs.AI ·2026-09-25 ·Tony Qin, Peter Connor, Khoa Dang, Carter Hatch, Caleb Rucker +2 more
similarity 0.52
Abstract

Performing complex medical procedures with continuum robots requires careful selection of their geometric design parameters. The robot should have high dexterity in the specific anatomical environment of its procedure. This work presents a design optimization method that considers both dexterity and anatomy. We introduce the Reachable Volumetric Dexterous Solid Angle (RVDSA) metric as our objective, which measures the ability of a robot's end effector to reach the points in a goal volume from different directions via collision-free paths from a start configuration. We present a computationally efficient motion planner to compute this objective function for a given robotic design, and we use an asymptotically optimal simulated annealing optimizer to compute an optimized design. We applied our new method to optimize the design of a bimanual dexterous sheaths robot for performing procedures on cancerous polyps in colon anatomies, achieving a 78% higher RVDSA on average than optimizing for 3D voxel coverage alone.

0.31