Drape-Compatible Tool-Tip Localization for Hand-Held Laparoscopic Instruments via UWB Carrier-Phase Ranging and Trocar-Constrained Geometry

Jinseok Lee, Dongho Yee, Minsung Kim, Younghoon Noh, Seonho Shim, Juahn Oh, Yechan Seo, Jiyul Lee, Seong Jeong, Hyuk Choi, Hyoun-Joong Kong

cs.RO ·submitted 2026-09-30 07:01 UTC · v1

Abstract

A surgical robot policy needs to know where each instrument's working end sits relative to the camera and to the other instrument, yet a laparoscopic operation leaves only the endoscope video, and a draped instrument hides every optical path on itself. We estimate the tool tips from distances and an IMU alone. The distances are measured by ultra-wideband carrier phase between antenna nodes on the handle side of the instruments: one per hand-held instrument, one on the endoscope, all outside the sterile barrier. Take the endoscope antenna as the reference. The two instrument antennas carry six coordinates while the three pairs give three distances, so the problem is short by three, and averaging cannot close that gap because what is missing is information, not precision. Carrier phase adds one more unknown per pair, an integer that leaves distance fixed only modulo lambda/2 = 23.1 mm. The trocar settles both difficulties: each shaft passes through a port whose position is known and whose direction the on-board IMU measures, so its antenna keeps a single degree of freedom, the insertion depth. Two unknowns now face three distances -- two fix the depths and the third is left over, and that spare equation exposes a wrong integer or a wrong mounting offset instead of absorbing it. The same solve returns both tool tips in the endoscope frame, without the second carrier frequency that integer resolution usually needs, so the link never leaves a single PLL lock. In an 11-hole phantom among metal instruments the three distances hold sigma = 0.10-0.33 mm at 35-46 Hz, a sterile drape pressed onto the antennas leaves the link unchanged where an optical path would be blocked, and the logger was carried into a live rabbit appendectomy.

Comment
Submitted to IEEE ICRA 2027

Closest papers

Nearest neighbours by meaning across the whole index.

Recording Hand-Held Laparoscopic Instrument Motion in the Operating Room: Magnetometer-Free Fusion of Inertial, Range and Visual Sensing

cs.RO ·2026-09-22 ·Jiyul Lee, Dongho Yee, Juahn Oh, Jinseok Lee, Yechan Seo +7 more
hot direction similarity 0.60
Abstract · Submitted to IEEE ICRA 2027

Most minimally invasive procedures are still performed with hand-held laparoscopic instruments, yet only the endoscopic video is retained; the instrument motion that expresses surgical skill, and that could support skill assessment and robot learning, is lost. Pose from video alone remains millimeters to centimeters off, and an instrument-mounted inertial measurement unit (IMU) cannot rely on its magnetometer, whose field changed with tool pose and between sessions in our measurements. We present a surgical instrument-state logger that clips onto a conventional instrument without modifying the part that enters the patient and fuses a six-axis IMU and a time-of-flight (ToF) rangefinder with a markerless camera in an error-state Kalman filter under the remote center of motion (RCM) of the trocar. Heading comes from the shaft silhouette, segmented by a U-Net, in place of the magnetometer: the rotation-angle error is 0.200°, against 3.58° from the accelerometer and magnetometer alone. Against a Franka Research 3 manipulator, and without alignment to it, the displacement error over 300 translation trials was 1.21mm RMS and the relative-rotation error over 180 rotation trials 0.34° RMS. On continuous trajectories, tracked and displayed in real time, the absolute tip error was 1.22mm (programmed) and 3.04mm (teleoperated) after post-hoc tuning of three filter parameters, and the full fusion beat every sensor subset. Because the estimator uses no magnetic measurement, its accuracy does not rely on an undisturbed field. The same clip-on device could thus record metric tip trajectories during routine hand-held laparoscopy, while displaying the insertion depth and attitude that are hidden once the instrument is inside the patient.

0.38

ARTS: An Open-Source, Reproducible Three-Arm Platform for Autonomy Research in Minimally Invasive Surgery

cs.RO ·2026-11-11 ·Jiawei Ge, Chunwoo Kim, Antony Goldenberg, Douglas McDonough, Justin Liao +16 more
code/data similarity 0.49
Abstract · Submitted to the 2027 International Symposium on Medical Robotics (ISMR). † Equal contribution

Research on surgical autonomy needs open platforms that any laboratory can build, yet the most widely shared platform depends on donated clinical hardware, and open alternatives on industrial arms each lack part of the surgical chain. We present the Autonomous Robotic Translational System (ARTS), an open-source three-arm platform for autonomy research in minimally invasive surgery, built entirely from purchasable parts for about USD 137,000. Two UR5e arms hold wristed da Vinci X/Xi instruments, a UR5 arm holds a stereo endoscope with two FLIR cameras, the surgeon console provides haptic handles, footswitches and a stereo viewer, and a hardware stop independent of the software lets a supervisor freeze all three arms in place. The ROS 2 software pins every dependency, installs from the same scripts on bare metal or in containers, and records stereo images with hardware-measured timing. On the real arms, ARTS commands each arm at its controller's full rate (e.g., 500 Hz on the UR5e), the two cameras expose within 12 microseconds of each other, hand motion reaches the arm in 58 ms, and the endoscope image reaches the monitor in 68 ms. The ISO 9283 pose repeatability is 0.14-0.25 mm, comparable to a da Vinci Research Kit (dVRK) Classic arm on the same protocol, and an optical tracker measured a mean remote-center-of-motion error of 0.34-0.37 mm. Three expert and three novice operators performed the Fundamentals of Laparoscopic Surgery (FLS) peg transfer through a laparoscopic trainer box, and one of the experts sutured on a suture pad. The hardware designs, firmware and software are available at https://github.com/MRE-Lab-UMD/arts.

0.56

DASH: A da Vinci Adapter for Serial-link and Humanoid Robots as an Accessible Platform for Surgical Robotics Research

cs.RO ·2026-10-05 ·Sara Wickenhiser, Junrong Zhou, Zekai Liang, Lizzie Peiros, Liu Shanglei +1 more
similarity 0.48
Abstract · 8 pages, 9 figures

Robotic minimally invasive surgery offers well-documented clinical benefits, but the cost and infrastructure requirements of purpose-built platforms limit access in rural and lower-resourced facilities. Recent work has teleoperated general-purpose robots for laparoscopic tasks and in vivo procedures, but relied on handheld instruments coupled through passive linkages rather than native robotic actuation. Instead, we adapt da Vinci Classic and Xi instruments onto general-purpose robots that can integrate in clinical workflows. We present DASH, a da Vinci Adapter for Serial-link and Humanoid platforms, consisting of two types of adapters that require no modification to the instruments themselves. Each adapter is compatible with many robotic platform load capacities, engages the instrument's native latch, and wirelessly identifies inserted tools to load instrument-specific kinematics and coupling matrices. Teleoperated experiments demonstrate the efficacy of DASH and its benefits for enabling research access to surgical robotic platforms.

0.45

Information-based Unified Design Framework for NMR Magnetic Tracking System

cs.RO eess.SY ·2026-10-13 ·Seçkin Eren Yetim, Mehmet Efe Tiryaki
similarity 0.48
Abstract · 6 pages, 6 figures, journal letter submission

Nuclear magnetic resonance (NMR) magnetic tracking offers a promising alternative to magnetic resonance imaging (MRI)-based localization for medical robotic guidance. However, scalar field measurements and field-aligned marker moments constrain position and orientation tracking, requiring careful configuration design. We present a unified information-based framework that uses local covariance analysis to evaluate receiver geometry, marker separation, and external source configuration in on/off-board NMR tracking. Numerical evaluation shows that planar-array precision depends jointly on receiver count and geometry, with triangular lattice arrays achieving the lowest best-case position uncertainty among the evaluated layouts for 9--20 receivers. Two spatially separated, field-aligned markers enable device-axis tracking, with predicted angular uncertainty below $2^\circ$ at the evaluated workspace center for suitable marker separations. We further establish point-dipole source--receiver duality, unifying the analysis of on/off-board NMR tracking. These

0.21

Physically Constrained Kinematic Calibration for High-Precision Robotic Needle Guidance

cs.RO eess.SYmath.OC ·2026-10-30 ·Prabhu Nithin Gollapudi, Andreas Maier, Franziska Mathis-Ullrich, Water Zhao, Jia Wang +1 more
similarity 0.48
Abstract · 6 pages, 4 figures, 4 tables

Percutaneous needle insertion is a common practice in minimally invasive procedures whose clinical success depends on accurate needle placement. However, achieving the necessary precision is a challenge for manual operators. Robotic assistance can improve targeting accuracy and reduce operator dependence, but residual millimeter-scale errors persist because nominal kinematic models do not fully represent the physical manipulator. This paper presents a robust bound-constrained kinematic calibration method designed to improve positional accuracy for a 6-Degree-of-Freedom (6-DoF) robotic needle-guidance manipulator. The approach uses physically plausible parameter ranges to regularize a compact 19-parameter geometric identification model. By treating these bounds as active optimization constraints rather than mere numerical safeguards, the framework effectively mitigates parameter coupling and overfitting, ensuring highly stable parameter identification. In simulation, the proposed method reduced end-effector Root Mean Square Error (RMSE) from 3.422 mm to 0.690 mm (a 79.8% improvement). In benchtop hardware experiments, the calibrated model achieved sub-millimeter validation accuracy, including 0.865 mm RMSE in experimental trajectory validation and 0.610 mm RMSE in a coordinate-lattice verification task. These results demonstrate that physically constrained kinematic calibration can enhance robotic needle-targeting accuracy, establishing a high-precision foundation for future clinical applications.

0.38

Design and Characterization of a Unified Continuum Robot Actuation Platform

cs.RO ·2026-10-14 ·Yunti Xu, Aedan Mangan, Miheer Potdar, Alex Lu, Tania K. Morimoto
similarity 0.46
Abstract

Continuum robots enable dexterous access to confined spaces in applications such as minimally invasive surgery and industrial inspection. However, their actuation systems are often custom-designed for specific robot architectures and tasks, limiting reuse across publications, institutions, and applications. This paper presents a unified actuation platform that decouples actuation hardware and software from continuum robot design. The platform provides modular actuation across continuum robot classes, scales to a large number of degrees of freedom (DOFs), and integrates with conventional serial manipulators for hybrid architectures. We validate the platform on an open-sourced surgical robotic system and experimentally characterize the actuation modules, achieving sub-millimeter translational accuracy and less than $5^\circ$ rotational error over ranges of $30$~mm and $180^\circ$, respectively. We further demonstrate the platform in two surgical use cases: teleoperation of a continuum camera in a constrained environment and bimanual manipulation using $2.53$~mm concentric-tube-robot-based forceps alongside a conventional wristed instrument. All hardware designs and ROS-based software are open-sourced to facilitate reuse and further research in continuum and surgical robotics.

0.31

Contact-Stable Deformable Tissue Simulation Using Implicit Integration and Live-Pose Grasp Constraints for Laparoscopic Surgery Robot Policy Evaluation

cs.RO ·2026-09-22 ·Juahn Oh, Dongho Yee, Jinseok Lee, Jiyul Lee, Yechan Seo +7 more
hot direction similarity 0.46
Abstract · Submitted to IEEE ICRA 2027

Closed-loop evaluation of surgical robots requires tissue that deforms, can be grasped and lifted, and reproduces the anatomy in which the robot will operate. We present a simulator in which this tissue is reconstructed from a fixed-view RGB-D recording of the surgical field, composited to remove the instruments, closed into watertight volumes and tetrahedralised; the pipeline was applied unchanged to three specimens of two species (thirteen organs, 146,061 tetrahedra, no inverted elements). For one specimen, the organs are placed in a bimanual cell in which two Franka FR3 arms operate motorised instruments through 6 mm trocars. The core contribution is the numerical and contact design that keeps this cell stable: implicit integration, simulation meshes separate from collision meshes, numerical guards, and a grasp constraint captured at the live tissue pose. In 45 repeated grasp-lifts, a friction grasp held the tissue in 0 of 15 trials and each constraint grasp in 13 of 15; on displaced tissue, a rest-pose constraint produced one-step snaps of up to 17.8 mm, which live-pose capture eliminates. Against the recording, front-surface depth error is 1.33 to 1.41 mm, organ silhouette IoU is 0.80, and in five grasp-lifts reproduced from video the landmark displacement RMSE is 11.8 mm against 14.2 mm for a static prediction. Biofidelity is not claimed; the environment is intended for closed-loop feasibility, safety, contact and policy screening.

0.32

Descriptor: UWB Anchor Self-Calibration Dataset (UWB-ASC)

eess.SP ·2026-11-08 ·Paul Schwarzbach
survey similarity 0.45
Abstract

Anchor self-calibration estimates the positions of fixed nodes of an indoor radio positioning or localization system from observations among the nodes and to mobile devices. It replaces the survey of every anchor, which is costly and goes stale when anchors move. The UWB Anchor Self-Calibration Dataset (UWB-ASC) is a testing ground for such methods, recorded in one industrial hall with ultra-wideband (UWB) two-way ranging. It has three parts. The first is two measured deployments. In the line-of-sight (LOS) deployment, 10 anchors range with a walking tag tracked by motion capture. In the non-line-of-sight (NLOS) deployment, 12 nodes range among obstructions, and one of them roves over surveyed markers in 20 quasi-static sessions. Both deployments also hold static sessions, and a total station surveyed every fixed node. The second part is a seeded range simulator that draws synthetic ranges on the surveyed geometry with adjustable error components, from noise-free ranges to NLOS errors, for benchmarks at controlled error levels. Its 81 frozen draws in 5 conditions serve as the reference. The third part is a set of readers that convert the ranges among the static nodes of two public UWB datasets, TWR-CLOUD-Bern and selfloc, into the same format without redistribution. All parts share open Parquet and YAML files, in which topology views select the ranges for each class of calibration method. Raw recordings with rebuild scripts, quality reports, and field labels of the line of sight complete the dataset.

0.28