Cross-Device Transfer of a Hospital-Trained Fetal Head Circumference Model to a Low-Cost Handheld Ultrasound Probe

Patrick Gao, Lara Hart, Ken Sullivan

eess.IV ·submitted 2026-10-28 23:56 UTC · v1
benchmark

Abstract

Automated fetal head circumference (HC) measurement reaches near-expert accuracy on the HC18 benchmark, but published models are trained and evaluated on hospital-grade equipment. We quantify what happens when such a model is carried to a low-cost handheld probe. A lightweight segmentation network (CSM, 0.13M parameters) was trained on HC18 across five independent runs and applied without modification to images from a $1,350 handheld probe (SonoQue C3), acquired at the same visit as the reference hospital scan in 18 patients. In-domain accuracy was 2.34 +/- 0.43 mm; on handheld images the same model gave 16.12 mm against the hospital record, a 6.9-fold degradation. Input preprocessing, a choice rarely reported, accounted for about a third of that gap: direct resizing gave 49.45 mm against 16.12 mm for an aspect-preserving pipeline. A leave-one-out data-efficiency curve showed that fine-tuning on two images significantly worsened accuracy, and that from four images upward error fell by 0.75 mm per doubling of the local dataset (p = 0.0045), although no training-set size up to 27 images significantly improved on the unmodified model. Much of the residual gap belongs to the reference rather than the model: scored against a manual tracing of the same image the model was already within 10.48 mm, and its disagreement with sonographer calipers on the same frame was a near-constant offset, negative in 10 of 10 cases. A single calibration constant estimated leave-one-out reduced that disagreement from 8.12 mm to 3.49 mm, reaching the intraobserver reproducibility reported for experienced sonographers, but did not improve agreement with the hospital record. Cross-device evaluations should therefore report the input preprocessing used and characterise the precision of their reference standard.

Comment
24 pages, 5 figures, 4 tables

Closest papers

Nearest neighbours by meaning across the whole index.

FETALFusion: Resolution-Aware State-Space Modeling for Multi-Domain Fetal Head Segmentation

cs.CV cs.AI ·2026-10-13 ·Ammar Bhilwarawala, Mainak Bandyopadhyay
published similarity 0.72
Abstract · Accepted at the 15th International Conference on Image Processing Theory, Tools and Applications, 2026

Head circumference (HC) measured from 2D ultrasound is a routine marker of fetal growth, and automating it needs segmentations whose boundaries hold up on individual scans, not only on average. We present FETALFusion, a dual-path CNN-Mamba encoder-decoder for fetal head segmentation across acquisition types. It applies four-directional state-space scans only at the two highest-resolution stages, routes bottleneck features through domain-specific channel gates that are averaged at test time, and adds a weak landmark shape prior. We train jointly on transabdominal (HC-18) and transperineal (PSFHS) images and compare ten methods on a held-out test set of 471 images, each trained under one schedule and scored by surface distances at native resolution. FETALFusion reaches a Dice score of 0.963 and an HD95 of 2.10 mm without test-time augmentation or a domain label. It improves on U-Mamba on all five reported metrics (HD95 reduced by 0.31 mm; Holm-corrected ($p < 10^{-4}$), a lead that holds under test-time augmentation and across two independent training runs, and in the mean on U-Net, Attention U-Net and TransUNet. Residual encoders close the gap: a ResNet-34 UNet++ exceeds FETALFusion, and an ImageNet initialised ResNet-34 U-Net is best overall (HD95 1.75 mm). The ablation agrees. Calibrated against run-to-run variation, it attributes FETALFusion's measurable gains to the landmark shape prior and, on transperineal images, to the domain gates, while removing the state-space branch does not measurably change accuracy. HC from the predicted masks agrees with the clinical reference at an intraclass correlation of 0.9986.

0.34

Risk Control Certifies Rates, Not Magnitudes: An Audit of Selective Fetal Biometry

cs.AI cs.CV ·2026-11-04 ·Zhengshuai Yang, Linna Jiang, Changyun Li, Guihu Zhao
theory similarity 0.63
Abstract · 10 pages, 2 figures, 3 tables

Prenatal ultrasound biometry must return a head circumference (HC) and refuse when the image cannot support one. Selective prediction trades coverage against error, but its finite-sample guarantees bound the rate of gross errors, not their magnitude. We audit a tool-grounded biometry pipeline on two separately pre-registered four-way splits of the 999 labelled HC18 images, retraining the segmenter three times on each. Calibrating the contour-completeness threshold by exact Clopper-Pearson risk control returns a certificate indistinguishable from a hand-set floor, yet the pipeline still emits a 271.3 mm error: a small, geometrically complete ring fitted to an image whose true HC is 324.6 mm. The accepted candidate scores 0.94 completeness, so the score cannot see the failure, and one outlier among $\approx 150$ answers leaves the certified risk bound ($α= 0.10$) intact. A second, untouched split reproduces the mechanism on a different image (233.3 mm). Constraining the admissibility region---the anatomical range in which the measurand is defined---removes the family instead: raising the lower bound from 50 to 55 mm costs 0.4 coverage points and cuts the worst-case error on both splits to 11.4 and 11.1 mm, whereas suppressing it through the completeness threshold costs 19.0 and 12.1 points. At an anatomically defined 70 mm the frozen tests give 1.91 and 1.81 mm MAE.

0.30

USB Gym: A Gymnasium Testbed for Closed-Loop Ultrasound View Matching

cs.CV cs.LGcs.RO ·2026-11-05 ·Thanapong Intharah, Zhichao Gao, Yiliang Liu, Hao Dong
benchmarkcode/data similarity 0.63
Abstract · 75 pages; supplementary material (62 pages) as an ancillary file. Submitted to Computer Methods and Programs in Biomedicine. Code: https://g…

Background and Objective: Our main goal is ultrasound guidance: moving a probe until its image matches a target view. Published methods for each part of a guide are evaluated in separate simulators. We asked whether these methods, combined into one guide, are good enough. Methods: We built USB Gym, a testbed whose standard engine simulates ultrasound from public patient CT, with a network drawing organ textures from CT labels. Its guide has three replaceable parts (slots), an organ segmenter, a controller and a reward, which received 20 published methods, each run on the same 60 episodes of six held-out patients. In each slot, a published method replaced ours only if detectably better on these patients. The better of this combination and our original guide on 20 new CTs is the benchmark baseline. A pose oracle, reading the exact probe pose available only in simulation, set the reference. The results below are exploratory. Results: On these CTs, the probe started on average 16.0 mm from the target view; the pose oracle brought it to 1.5 mm. Of the results known when the combination was fixed, only an image-similarity reward (SSIM) was slightly but detectably better, so the combination was our guide with that reward. This combination was the benchmark baseline, ending at 13.7 mm (16% of the oracle's progress). Conclusions: Combining untuned published methods with ours did not guide the probe near the target view; the task remains under-explored. The testbed accepts new methods; no guide was tested on hardware.

0.44

Open ultrasound foundation model for robust segmentation and clinical measurement across heterogeneous settings

cs.CV ·2026-09-16 ·Chao Qin, Fahad Shahbaz Khan, Salman Khan, Sarim Ather, Siddiq Anwar +2 more
code/data similarity 0.61
Abstract · The PDF includes the Supplementary Information. Code is available at https://github.com/AlfredQin/sonobase

Ultrasound is the most widely deployed imaging modality worldwide, yet clinical AI remains fragmented into narrow single-task models that fail when device, operator, or anatomy changes. Here we present SonoCorpus, an open resource unifying 456,963 images and 1,626,085 expert masks from 53 public datasets spanning 24 clinical applications and 17 countries, and SonoBase, an interactive segmentation foundation model pretrained on it. Across fifteen evaluation datasets introducing new organs, devices, operators, and geographies, SonoBase outperforms SAM2, MedSAM2, and the concept-promptable MedSAM3 on every dataset and matches per-dataset specialist models trained on the same data; on fully external data it exceeds the accuracy these baselines achieve on their own in-distribution benchmarks. Ejection fraction derived from its segmentations falls within inter-observer variability (6.63\% error), with fewer misclassifications at the defibrillator-candidacy threshold than either promptable baseline (13\% versus 18--42\%); fetal head-circumference (1.81~mm) and gestational-age (1.2 days) errors fall below inter-observer variability. Where a baseline fails outright, one in four test cases, SonoBase recovers a usable segmentation in 81\% of them, including on handheld probes operated by minimally trained users in two low- and middle-income countries (Sierra Leone and Tanzania). Five labeled examples can help the model adapt to a new setting, and the identical training protocol transfers well to newer models such as SAM3, locating the advantage in ultrasound-specific pretraining rather than any single architecture. To ensure reproducibility and enable the community to build on SonoBase as a platform, we release all checkpoints, optimizer states, data-split indices, deduplication hashes, and starter code at https://github.com/AlfredQin/sonobase.

0.36

Validation of Smartphone-Based Photogrammetric 3D Body Scanning for Automated Anthropometric Measurements Compared with a Commercial Depth-Sensor-Based Body Scanner

cs.CV ·2026-08-13 ·Ruting Cheng, Boyuan Feng, Chuhui Qiu, Joaquin A. Calderon, Qing Pan +2 more
published similarity 0.61
Abstract · 16 pages, 5 figures, 4 tables

3D body scanning has become an important tool in healthcare applications because of its rapid and non-invasive nature. While smartphone-based photogrammetric reconstruction provide a low-cost and accessible alternative to commercial 3D body scanners, their performance for whole-body scanning remains insufficiently validated. Thus, we designed this study to comprehensively validate the photogrammetric 3D scanning application by evaluating automatically extracted whole-body measurements and longitudinal body-shape monitoring. We evaluated a representative application, PolyCam, against the commercial depth-sensor-based Fit3D ProScanner using 144 pregnant participants scanned longitudinally throughout pregnancy. We designed an automatic circumference extraction pipeline to get measurements at four anatomical landmarks from paired 3D scans. A linear mixed-effects model was used to evaluate scanner effects and longitudinal body-shape changes. Measurement consistency was assessed using repeated PolyCam scans and tape measurements on a rigid mannequin. PolyCam demonstrated strong agreement with Fit3D, with average biases below 16 mm, intraclass correlation coefficients above 0.8, and Pearson correlation coefficients above 0.9 across all landmarks. Both systems captured comparable longitudinal body-shape changes. Mannequin experiments showed mean biases below 3.5 mm and no significant differences from tape measurements. These findings support smartphone photogrammetry as a potential accessible alternative to commercial body scanners and applicable for longitudinal 3D body-shape assessment.

0.28

Evaluation Choices Shape Biomedical ML Claims: A Pediatric Pneumonia Benchmark Case Study

cs.CV cs.LG ·2026-09-29 ·Bhanu Prakash Vangala, Sowmya Guda, Latha Peddi, Navya Vangala
benchmark similarity 0.57
Abstract

Biomedical machine learning papers often compress model performance into one headline number. That number can look like a property of the model even when it depends strongly on how the benchmark was evaluated. We study this problem on the widely used Kermany pediatric chest radiograph dataset using nine image classifiers and a controlled evaluation protocol. Under the same protocol, the eight pretrained backbones differ by only 0.026 AUROC. In contrast, changing whether the backbone is frozen or fine-tuned changes AUROC by 0.044 on average, and changing the decision threshold changes balanced accuracy by 0.090 on average. The official test split is also measurably different from the training pool: a partition classifier distinguishes them at AUC 0.697, rising to 0.898 for normal radiographs. Most strikingly, a classifier using only file properties, with no image anatomy, reaches 0.992 balanced accuracy within the training pool but falls to 0.496 on the official test split. Validation-fitted thresholds and calibration also transfer imperfectly. These results show that a high benchmark score can support different conclusions when the split, training policy, threshold, metric, calibration, and uncertainty are not communicated with it. We end with a seven-item reporting recommendation in which each item is tied to an effect measured in the study

0.11

Anatomy-Guided Foundation Model Adaptation with Within-Case Prototype Supervision for Standard Plane Detection in Fetal Ultrasound Blind Sweeps

cs.CV eess.IV ·2026-08-27 ·Yuzhe Zhao
benchmark similarity 0.57
Abstract

Detecting the fetal abdominal circumference standard plane in low-cost obstetric blind sweeps is a highly imbalanced frame-classification problem: positive frames account for under 3% of a sequence, form short contiguous segments, and are poorly handled by off-the-shelf ultrasound and vision foundation models. We propose AnatoProto, a lightweight sequence-level framework that adapts a frozen BiomedCLIP encoder to fetal blind sweeps through four components: (i) anatomy-weighted spatial pooling that uses nnU-Net abdominal-region probabilities as a spatial prior to reweight BiomedCLIP patch tokens, so frozen semantic features are aggregated onto anatomically meaningful regions; (ii) a within-case prototype loss that pulls each frame embedding toward the mean of positive frames of the same sweep, exploiting case-level structure unavailable at the frame level; (iii) a three-stage cascade refinement (frame->segment->case-level rejecter) that lifts the prediction unit from noisy frames to structurally-constrained segments; and (iv) a hybrid prediction head that jointly models per-frame stability and inter-frame boundary transitions to suppress boundary false positives. On the ACOUSLIC-AI benchmark, AnatoProto reaches a test F1 of 67.72, outperforming the strongest foundation-model baseline (FetalCLIP + PRS, F1 = 54.52) by +13.20 F1 and the strongest video temporal-action-detection baseline (TriDet + PRS) by +15.76 F1. A synergy study, backed by embedding geometry and paired-bootstrap confidence intervals, shows that the prototype loss and anatomy-weighted pooling are not additive: applied alone the prototype loss reduces recall by 12 points, but combined with anatomy-weighted pooling it increases recall by 6.5 points -- a sign-flip we trace to the accuracy of the within-case prototype.

0.31

Style-Driven Data Synthesis and Degradation-Aware Enhancement for Ultrasound Image Restoration

cs.CV ·2026-09-28 ·Yu-Kai Wang, Chun-Xin Tan, Manh-Hung Nguyen, Ching-Chun Huang
code/dataunusual similarity 0.56
Abstract

Low-cost handheld ultrasound devices can be widely deployed compared to professional hospital ultrasound machines. However, their images suffer from compound degradation that can mislead clinical judgment. Motivated by this observation, mapping handheld low-quality (LQ) to hospital high-quality (HQ) images has been considered a valuable research question. Conventionally, the mapping requires pixel-aligned LQ-HQ pairs. This requirement is unsatisfactory in practical scenarios because real scans at different times are never pixel-aligned. This paper addresses the challenge with a two-stage framework. The first stage generates pixel-aligned LQ-HQ datasets, and the second stage trains an enhancement model that improves LQ images. The first stage trains a cycle-consistent style-transfer model on unaligned real LQ-HQ pairs to learn a HQ-to-LQ model. Then, the model transforms real HQ images into pixel-aligned LQ images. Based on the dataset generated by the first stage, the second stage uses the Dual Degradation-Guided (DDG) Low-Rank Adaptation (LoRA) method to fine-tune an LQ-to-HQ model based on aligned pairs. In this stage, the model is based on the well known PiSA-SR framework but inserts a degradation-conditioned correction matrix. Experimental results on the USenhance2023 dataset show that the FID metric is improved by 16.7% over the strongest baseline while other metrics indicate that our enhanced outputs are well aligned with the real HQ distribution. The source code of our method is available at https://github.com/Jason0411202/DDG_LoRA.

0.38