Certificate-Based Verification of Pulsed Field Ablation Models

Konstantinos Dovelos

cs.CE math.NAphysics.med-ph ·submitted 2026-11-13 13:22 UTC · v1
theoryunusual

Abstract

Computational models of pulsed field ablation predict the lesion as the tissue where the electric field exceeds a lethal threshold. Their numerical error is rarely bounded, and automated pipelines increasingly run them without an expert checking each result. To verify each run, we separate the question "was the stated model solved correctly?" from the solver that answers it. An untrusted solver returns with its answer a certificate, consisting of a potential with the correct electrode and ground values and an exactly charge-conserving current. A small independent checker verifies the certificate against a written specification of the physics. Using the Prager-Synge hypercircle theorem evaluated in ball arithmetic, it turns the certificate into guaranteed two-sided bounds on the impedance and on the root-mean-square field of every tile of a lesion map, also over a range of blood/tissue conductivity ratios. A vectorized version of the checker takes about a tenth of the solve time. For a focal electrode in contact, the tile-resolved depth at 1000 V and 700 V/cm is proven to lie in 3.50-3.75 mm for every ratio from 1.7 to 3.75. For a bipolar catheter, certified depths overlap published in-vivo depths. Where they differ from a published reference model, the certificates prove that the difference is not numerical error in our computation. We compare the certificate with a grid convergence study on the same solver. On graded grids the study is sharper for local fields but remains an estimate. On uniform grids it fails to converge at the electrode edge, where the certificate still gives a valid bracket. Solvers that deliberately solve equations other than the specified ones pass the grid convergence study, but the certificates refute their claims or leave them unsupported.

Comment
18 pages, 5 figures, 7 tables

Closest papers

Nearest neighbours by meaning across the whole index.

Characterising cardiac tissue properties with graph neural networks

cs.AI q-bio.QMeess.SP ·2026-08-16 ·Ching-En Chiu, Yoo Ri Kim, Magdi Saba, Danilo Mandic, Marta Varela
similarity 0.43
Abstract · Accepted at The Statistical Atlases and Computational Modeling of the Heart (STACOM) workshop 2026

Characterising electrophysiological properties of cardiac tissue efficiently and accurately from spatially sparse intracardiac measurements is clinically important for localising ablation targets and improving arrhythmia treatment. We developed a graph neural network-based framework trained on synthetic electrogram signals on 2D flat surfaces to identify areas of interest in the context of cardiac ablation for premature ventricular complexes (PVCs). Our method achieved an average precision of 0.96, 0.97, and 0.95 for the detection of single-patch fibrosis, rapid depolarisation and high excitability, respectively. The trained model can then be applied to 2D curved surfaces with few-shot fine-tuning, demonstrating its generalisation capability. Future work will develop this framework further for clinical use in PVC ablation.

0.13

Certified by Abstention: Distribution-Free Guarantees for Chain-of-Thought Verifiers at Small Calibration Budgets

cs.CL stat.MLcs.LG ·2026-10-07 ·Arjun Balaji
publishedbenchmark similarity 0.42
Abstract · 22 pages, 7 figures, 12 tables. Under submission at AISTATS 2027

Signals that predict whether a chain-of-thought (CoT) trace is correct are compared by AUC, but deploying one requires a threshold with a guarantee. We ask what distribution-free selective guarantees deliver for CoT verifiers at realistic calibration budgets of tens to a few hundred labelled problems, using seven open models, five verifier signals and 37,000 graded traces. The central observation is validity by abstention: an $(α,δ)$-valid procedure that issues a certificate with probability $P_{\rm fire}$ bounds the failure probability of an issued certificate only by $δ/P_{\rm fire}$, so a certificate that rarely fires can be valid and wrong every time it is used. In a simulation with known risk the standard certificate fails in at most 0.3% of calibration draws but in up to 69% of those in which it fires. A certification floor and a lattice condition for Benjamini-Hochberg conformal selection explain why certificates abstain at these budgets, and the data bear them out: the standard certificate returns nothing or a large accepted set, and an unreadable residual-stream probe buys two to three times the coverage of the readable signals, an edge a cross-fitted reconstruction cannot recover linearly from the readable features. We then give a floor-started fixed-sequence certificate, valid without monotonicity assumptions, that covers more than the Bonferroni certificate on every model-signal pair and raises coverage at the non-vacuous target $0.75π_0$ from 0.05 to 0.16, although the floor keeps absolute coverage small. Finally, a certificate cannot see what matters after deployment: under benchmark shift the error among accepted traces tracks the new task's base error, and under best-of-$n$ selection against the verifier it rises past the target while the empirical failure frequency stays below $δ$, because abstention absorbs the failures.

0.33

Automated Optimization of Patient-Specific Electro-Anatomical Models Fitted to Cochlear Implant SCINSEV

eess.SP ·2026-10-29 ·Minh Q. Vu, Erin L. Bratu, Jack H. Noble
similarity 0.40
Abstract

Purpose: Patient-specific electro-anatomical models (EAMs) of the implanted cochlea are crucial for auditory nerve fiber simulation in cochlear implant (CI) programming and neural health estimation. Fitting an EAM to the stimulation current induced non-stimulating electrode voltage (SCINSEV) currently requires manual expert adjustment or exhaustive search over precomputed forward solutions, neither of which scales to routine clinical use. We present an automated optimization pipeline that fits patient specific EAMs in minutes. Methods: Four compartment tissue label maps (perilymph, soft tissue, modiolus, and bone) derived from postoperative CT were used to construct a finite difference forward model, fitted to the measured SCINSEV by Gaussian process Bayesian optimization. Three reductions cut fitting cost: a coarsened grid with trilinear injection and readout at sub-voxel contact positions; solving half the stimulating contacts and completing three quarters of the matrix by reciprocity; and a Jacobian-based identifiability analysis reducing four resistivities to two. Each reduction was validated separately against the measurement's reciprocity residual and the reproducibility of the fit. Results: Across 43 implanted ears, fitting took a median of 2.6 minutes per ear. Median SCINSEV RMSE on the reduced-resolution objective was 0.069 mV, against 0.084 mV for grid search at a matched evaluation budget. Against an exhaustive full-resolution grid search requiring a median of 9.1 hours per ear, the pipeline reached statistically indistinguishable fitting error roughly 210x faster. Conclusion: Patient-specific cochlear EAM fitting can be performed at clinically practical timescales without expert supervision, removing a major computational barrier to patient-specific cochlear implant programming and cochlear simulation.

0.35

MedPhysBench: A Benchmark for Evaluating Language Models on Medical Physics Calculations and Error Detection

cs.AI physics.med-phcs.CL ·2026-11-03 ·Omar Nusrat
benchmark similarity 0.39
Abstract

Medical physicists turn a prescription into a delivered dose and check one another's calculations, because one error in a beam calibration or a monitor-unit (MU) calculation can affect every patient treated with it. MedPhysBench has ten seeded generators of calculation items in radiation therapy, diagnostic imaging, nuclear medicine and radiation protection, up to clinical tasks such as a TG-51 calibration worksheet. Reference answers come from code implementing the governing protocols, and a grader converts units and applies tolerance bands. From the same inputs, separate checking items test error detection, showing a colleague's worksheet or only the colleague's result, half correct and half with one planted error. On a held-out test split of 1,200 items (half for two models), 12 models scored 33.3 to 100.0% on calculation. Of the nine answers graded incorrect for the four best (99.3-100.0%), five contained model errors and four followed NCRP Report 151 on points our items misstate or leave open. Below them, most models did worst on level-3 clinical tasks, and Claude Sonnet 5.5 omitted the inverse-square correction in 19 of 20 isocentric MU items. Five models detected at least 96.7% of mostly large planted errors with at most 4% false alarms; for four of the other seven, checking only the result raised false alarms significantly. In an exploratory targeted test with 20 public cases of an omitted inverse-square correction, the three best models tested caught it in every worksheet, and three weaker models accepted 16 to 18 of 20 planted worksheets, one of them while applying the correction in most of its own calculations. Claude Opus 5.5 accepted 6 of 20 wrong results, although it applied the correction in all 20 of its own. We report the effect of two problems in our shielding items, found after the runs, and release the code, per-item outcomes and targeted-test responses.

0.23

An Evidence-Guided Framework for Tissue-Barrier Impedance Review: Deletion and Resistance Sensitivity

eess.SP ·2026-10-23 ·Leo Yang, Yingtong Zhou
similarity 0.39
Abstract · 5 pages, 3 figures, 3 tables

Reviewing tissue-barrier impedance requires connecting spectral anomalies with resistance estimates and experimental context. We present an evidence-guided framework that coordinates numerical checks and traceable actions for human review. In-house simulations assess the effect of removing flagged frequencies, public datasets test model adaptation, and local measurements demonstrate the review workflow. Removing flagged frequencies reduced resistance error under quadratic objectives, whereas robust fitting showed little average benefit from deletion and a mean error increase for endpoint-pair perturbations. High errors clustered around a few fitted circuit parameter sets, highlighting the value of examining parameter identifiability. Under the tested noise and perturbation conditions, these findings support robust full-spectrum estimation, flag-guided inspection, and matched sensitivity checks.

0.26

Single-visit screening for intracochlear tissue growth validated against multi-visit electrical measurement changes

eess.SP ·2026-11-02 ·Minh Q. Vu, Erin L. Bratu, Jared A. Rybarczyk, Jack H. Noble
similarity 0.39
Abstract

Purpose: Tissue growth (TG) around cochlear implant electrodes raises impedance but is not reliably visible on CT scans. Routinely recorded stimulus current-induced non-stimulating electrode voltages (SCINSEVs) carry this information, but a single recording has no patient-specific reference. We present an algorithm for detecting signs of significant TG from a single SCINSEV measurement. Methods: TG is assumed to concentrate basally near the electrode insertion site, leaving the apical region mostly unaffected. A patient-specific forward model built from CT is fitted to the apical measurements and predicts the basal ones. The misfit score, T, measures each basal contact's signed relative misfit against the apical misfit, accounting for apical fit quality and measurement noise. An ear is flagged when two adjacent basal contacts cross a predetermined threshold. We compare T with T-delta, a model-free reference, tracking basal relative to apical SCINSEV change since the intra-operative recording, persisting over the last two visits. Results: T had a false-positive rate of 3% on growth-free ears. It agreed with T-delta in 81% of 68 ears (sensitivity 71%; specificity 86%; p = 6e-6). Over 200 patient-level two-fold splits, median held-out agreement was 78% at a false-positive rate of 5.1%. Conclusion: An apically fitted patient-specific model predicts basal SCINSEVs well enough to flag signs of significant TG at a single visit, without an intra-operative baseline. Detections largely agree with persistent multi-visit basal-versus-apical change, a surrogate ground truth for significant TG.

0.43

Risk Control Certifies Rates, Not Magnitudes: An Audit of Selective Fetal Biometry

cs.AI cs.CV ·2026-11-04 ·Zhengshuai Yang, Linna Jiang, Changyun Li, Guihu Zhao
theory similarity 0.38
Abstract · 10 pages, 2 figures, 3 tables

Prenatal ultrasound biometry must return a head circumference (HC) and refuse when the image cannot support one. Selective prediction trades coverage against error, but its finite-sample guarantees bound the rate of gross errors, not their magnitude. We audit a tool-grounded biometry pipeline on two separately pre-registered four-way splits of the 999 labelled HC18 images, retraining the segmenter three times on each. Calibrating the contour-completeness threshold by exact Clopper-Pearson risk control returns a certificate indistinguishable from a hand-set floor, yet the pipeline still emits a 271.3 mm error: a small, geometrically complete ring fitted to an image whose true HC is 324.6 mm. The accepted candidate scores 0.94 completeness, so the score cannot see the failure, and one outlier among $\approx 150$ answers leaves the certified risk bound ($α= 0.10$) intact. A second, untouched split reproduces the mechanism on a different image (233.3 mm). Constraining the admissibility region---the anatomical range in which the measurand is defined---removes the family instead: raising the lower bound from 50 to 55 mm costs 0.4 coverage points and cuts the worst-case error on both splits to 11.4 and 11.1 mm, whereas suppressing it through the completeness threshold costs 19.0 and 12.1 points. At an anatomically defined 70 mm the frozen tests give 1.91 and 1.81 mm MAE.

0.27

Data Archive for 2D and 3D Electrical Impedance Tomography with moderate conductivity contrast phantoms

eess.IV cs.LG ·2026-11-12 ·S. J. Hamilton, A. Hauptmann, A. Hänninen, V. Kolehmainen, J. Niskanen +2 more
code/data similarity 0.38
Abstract · 11 pages, 6 figures. Data available at https://doi.org/10.5281/zenodo.24120439

This document describes open Electrical Impedance Tomography (EIT) measurement data that were collected from biomedical imaging inspired phantoms with moderate conductivity contrasts. The data were collected with the KIT4 and KIT5 EIT measurement devices at the University of Eastern Finland, and are intended for EIT image reconstruction algorithm development and testing. The data from thoracic imaging inspired cylindrical and chest-shaped phantoms are suitable for 2D algorithms, and the data from an intracerebral hemorrhage imaging inspired head-shaped phantom with resistive skull are suitable for 3D algorithms. In addition to the measurement data, computational meshes and example reconstruction codes are provided and sample reconstructions included.

0.34