Event-Based Estimation for Detrusor Pressure Using a Single Bladder Catheter Augmented with Abdominal Electromyography

Mohamed Abdelhady, Reilly Burhanna, Lee Brody, Margot S. Damaser, Steve J. A. Majerus

eess.SP ·submitted 2026-09-17 18:45 UTC · v1
unusual

Abstract

Objective: The aim of this study was to enhance patient comfort and diagnostic accuracy in urodynamic studies (UDS) by developing a novel framework that uses a single vesical pressure (Pves) channel combined with a surface abdominal electromyography (AEMG) electrode array to estimate detrusor pressure (Pdet) without the need for an abdominal catheter. The proposed method was designed to operate in real-time, and to be compatible with current UDS clinical hardware. Methods: 40 UDS studies were performed with AEMG measured simultaneously with both Pves and Pabd. A predefined, provocative abdominal maneuver was used before each study to ensure accurate coupling between the AEMG signal and associated abdominal pressures seen in Pves. Criteria were set to determine if coupling was sufficient for Pdet estimation from Pves and AEMG. Post-hoc signal processing was used to further refine and evaluate estimated Pdet using automated, non-causal event detection. An event-based scoring metric was used to determine how accurately this approach estimated Pdet by calculating the percentage of signal contained in the appropriate channel for that event (Pdet vs abdominal pressure). Results: The results showed that 30 of the 40 cases indicated a "Go" status after the provocative maneuver, demonstrating feasibility of this method in a majority of UDS studies. The scoring metric, along with correlation analysis, confirmed accuracy of estimated Pdet with scores ranging from 85% to 95% with median at 93%. Significance: This innovative approach has the potential to revolutionize urodynamic studies by providing a more comfortable and accurate diagnostic tool for assessing bladder function. This is particularly beneficial for vulnerable populations, such as the elderly and children, reducing the need for multiple catheters and improving the overall patient experience during UDS.

Closest papers

Nearest neighbours by meaning across the whole index.

OUCH: Objective, Uniform, Controlled Haptic Stimulation for Automated Neurological Assessment

cs.RO ·2026-11-03 ·Sherry Choi, Sitao Sunny Li, Joseph Romero, Lillian Chin, Ashwin G. Ramayya +2 more
benchmark similarity 0.47
Abstract · 8 pages, 7 figures. Supplementary video in ancillary files. Sherry Choi and Sitao Sunny Li contributed equally to this work

Motor responses evoked by noxious stimulation are used to assess neurological function in hospitalized patients with disorders of consciousness. In standard bedside assessments, these stimuli are applied manually, resulting in substantial variation in force, duration, frequency, and location. Here, we present OUCH (Objective, Uniform, Controlled Haptic stimulation), a wrist-worn device that uses pneumatic actuation to deliver noxious mechanical stimuli. OUCH combines stacked pneumatic pouch actuators, a rigid indenter, and a compliant spacer designed to buckle at a set load, so that the indenter contacts the skin only during actuation. We characterize the device using blocked force tests, clinician evaluation, finite element analysis, and a deformable pressure testbed, and benchmark it against a commercially available handheld pressure algometer. At 200 kPa pouch inflation pressure, our indenter concentrated its load into an estimated contact pressure of 18,500 kPa, over 25 times that of the algometer tip, while requiring less total force. On the deformable testbed, it localized deformation near the surface, while deformation at depth remained comparable to the algometer. As a result, in preliminary perceptual testing, our indenter elicited discomfort at approximately 89% lower applied force than the algometer, demonstrating the feasibility of evoking discomfort with a compact, wearable soft actuator. These results show that combining pneumatic actuation with designed contact geometry enables controlled, repeatable noxious stimulation at the wrist, providing a foundation for objective neurological assessment.

0.25

Sensor-Free Shockable Rhythm Classification During CPR Using the Existing ECG Channel of Public-Access Automated External Defibrillators

eess.SP ·2026-11-19 ·Jaechan Lim, David Hicks, Matt Valentine, Ki H. Chon
similarity 0.47
Abstract

Automated external defibrillators (AEDs) are public-access devices intended for use by lay rescuers, yet they must pause chest compressions to analyze the rhythm because compressions corrupt the electrocardiogram (ECG). Removing this pause requires a method that uses the hardware already present in the device, fits within its processor budget, and meets the rhythm-recognition performance required for defibrillator approval. Existing approaches rely either on an auxiliary sensor, which entails a hardware redesign, or on an additional network that reconstructs the ECG before classification. This work classifies the compression-contaminated ECG directly, using only the existing ECG channel of the AED. A cascade of three compact convolutional networks trained with different class ratios verifies each decision with a second network biased toward the proposed outcome; disagreement yields an indeterminate outcome. Evaluated on an independent manufacturer-provided cohort of 464 subjects over 15 train-validation partitions, the cascade satisfies both adjudicable American Heart Association criteria under all six noise conditions at an eight-second window and a signal-to-artifact ratio of 0 dB, with sinus-rhythm specificity of 99.20-99.43% and other non-shockable specificity of 95.48-96.20%, leaving only 1.07-1.48% of segments indeterminate. The criteria are not met at -3 dB. Single networks matched in parameter count and inference cost do not reproduce this performance. Each decision evaluates two of the three networks and requires 6.68 ms on a single thread of a general-purpose processor, and eight-bit quantization reduces the model to 1.45 MiB without changing any compliance verdict. Thus, rhythm analysis during ongoing CPR can be implemented with the ECG hardware already present in public-access AEDs, without auxiliary sensing hardware, subject to validation on the target device.

0.17

Spectral Features Dominate BCG Respiratory-Event Detection: A Large-Scale Patient-Independent Comparison of Feature Groups in Sleep Apnea Patients

cs.LG ·2026-08-28 ·Israel Campero Jurado, Zoe Bousraou, Lara Benning, Sara Padilla Neira, Alexander Breuss +3 more
similarity 0.46
Abstract · 36 pages, 7 figures

Unobtrusive ballistocardiographic (BCG) sensing is a promising modality for long-term sleep-apnea monitoring, yet it remains unclear which signal features are most discriminative for respiratory-event detection. We present a literature-guided, patient-independent comparison of ten BCG feature groups using a 512-sensor capacitive pressure mat recorded simultaneously with respiratory polygraphy in 155 patients (52 female, 103 male) undergoing in-hospital evaluation for obstructive sleep apnea. Features were extracted from six spatially distinct signal channels, yielding a 191-dimensional feature vector spanning general statistical, time-domain, frequency-domain, wavelet, frame-energy, and nonlinear complexity descriptors. Under strict leave-one-patient-out cross-validation for binary classification of respiratory-event windows versus event-free reference windows, Random Forest and Histogram Gradient Boosting achieved AUC-ROC of 0.967 and 0.969 and AUC-PR of 0.977 and 0.979, respectively. Feature-importance analysis revealed that frequency-domain features dominate discrimination: breathing-band power in the 0.1-0.4 Hz range accounted for 30.3% of total discriminative information across all spatial channels, and Fast Fourier Transform spectral-shape descriptors of the adaptively preprocessed channel contributed a further 15.1%. AUC and curve-length features provided the main complementary time-domain evidence (21.5%), whereas wavelet-derived and nonlinear features contributed smaller secondary effects (10.4% combined across 59 features). Frequency-domain and time-domain features together accounted for 67% of total discriminative information, demonstrating that a compact, interpretable subset of the full feature library achieves clinically relevant performance under patient-independent validation and providing an empirical basis for feature selection in future BCG systems.

0.26

Disentangling Monitor Hardware Artifacts and Smoothing Lag for Shift-Robust Intraoperative EEG and Hemodynamic Risk Calibration

cs.LG ·2026-11-19 ·Zeyu Jia
unusual similarity 0.46
Abstract · 16 pages, 4 figures, 5 tables

Commercial intraoperative electroencephalography (EEG) and blood pressure monitors introduce cyclic impedance pulses, smoothing delays, direct-current (DC) drift, and cuff holds that confound risk forecasting and uncertainty calibration. We developed a causal signal-conditioning and shift-robust conformal calibration framework across 5,439 public anesthesia procedures. Streaming filters blank 10.13-min 64-Hz impedance checks, apply unclipped post-notch conditioning, separate 63-s suppression-ratio lag, and distinguish three-channel cuff inflations from dropouts across propofol-remifentanil total intravenous anesthesia (TIVA; N = 3,214), volatile anesthesia (N = 2,046), European endoscopy sedation (N = 171), and DC-coupled surgery (N = 8). De-impedanced cortical tree ensembles were compared against deep sequence models, paired with patient-cluster Subgroup-Covariate-Weighted Conformal Risk Control (CRC) and lag-buffered online Conformal Proportional-Integral-Derivative (PID) tracking. Pre-filter blanking removed 97.90% and 97.60% of false flatline conflicts while preserving 99.23% and 98.96% of deep suppression. On unseen TIVA surgeries (N = 525), cortical features achieved area under the precision-recall curve (AUPRC) 0.3135 (1.94% prevalence) and 0.1627 (0.82% lag-free prevalence), with no detectable difference from deep encoders (p >= 0.36) and no cortical gain for hypotension (0.3432 vs. 0.3437, p = 0.784). Zero-retraining volatile transfer (N = 1,048) showed no detectable difference from an end-tidal gas oracle on lag-free suppression (0.0820 vs. 0.0736, p = 0.472). Subgroup-Covariate-Weighted CRC bounded all 70 TIVA shift cells at <= 4.38% (alpha = 5.0%) in 66.32 ms per 30-s epoch. Causal artifact disentangling and shift-aware conformal calibration support reliable real-time intraoperative risk bounds within explicit physiological observability limits.

0.31

Context-Augmented, Session-Calibrated Blood Pressure Tracking from Photoplethysmography

cs.AI ·2026-10-21 ·Farouk Ganiyu Adewumi, Timothy Oladunni
similarity 0.46
Abstract

Objective.To distinguish the contributions of acquisition context and photoplethysmography (PPG) to session-calibrated blood pressure (BP) tracking.Approach. The original comparison used three repeats of five-fold participant-level cross-validation in 653 AuroraBP participants, excluding calibration-labelled prediction targets. Matched analyses separated context and PPG and removed elapsed-time terms. Later exploratory follow-ups retained PPG, accelerometer signals and three initial cuff readings. Separate studies evaluated pulse-rate fusion and an intraoperative VitalDB model. Main results.The original wearable-state model achieved participant-level mean absolute errors (MAEs) of 7.697/6.139 mmHg for systolic/diastolic BP (SBP/DBP). Added context reduced MAE by 0.436/0.126 mmHg (97.5% intervals 0.305-0.573/0.053-0.200). In the matched factorial without the external PPG prior, PPG improved on calibration, time and protocol-provided posture/activity by 0.214/0.135 mmHg; removing time left the DBP interval including zero. Standard PPG descriptors contributed, but additional nonlinear/attractor benefit was not established. In a separate follow-up series, a sensor-history ensemble achieved 7.299/5.445 mmHg versus its static-rich reference's 7.363/5.470, small gains of 0.064/0.025 mmHg. Larger DBP errors did not improve uniformly. Quality-weighted fusion reduced pulse-rate MAE from 3.985 to 3.690 beats per minute. VitalDB MAEs remained 15.375/8.833 mmHg despite small PPG gains. Significance. The contribution is a calibrated evaluation framework with explicit controls, reproducible feature-cache-to-results analyses and retained negative findings. The sensor-history gain is exploratory after repeated dataset reuse. Independently estimated wearable states, complete raw-signal reproduction of the supplied richer features, external transfer and clinical adequacy remain unestablished.

0.24

ECGJev: Question-Conditioned Option Scoring for 12-Lead ECG Interpretation

cs.LG eess.SP ·2026-11-01 ·Zhuoyuan Li, Xidong Wu, Hanyu Cui, Jiansong Gao, Rongrong Fu +2 more
code/datahot direction similarity 0.46
Abstract · 5 pages (4 pages plus ethics statement and references), 2 figures. Submitted to IEEE ISBI 2027. Zhuoyuan Li and Xidong Wu contributed equall…

Clinical interpretation of a 12-lead electrocardiogram (ECG) often requires answering targeted diagnostic questions with clear, quantifiable decisions. We introduce ECGJev, the first ECG framework built on a decision decoder, a pretrained language model trained to return a probability for every listed option. A question-conditioned bridge compresses frozen waveform and spectrogram features into 16 ECG tokens, and a two-stage procedure adapts the bridge and low-rank adapters on the decoder's native option readout using diagnostic labels alone, scoring all options for a question in one forward pass. Evaluated across four ECG benchmarks, ECGJev achieves the highest macro-F1 on CPSC2018 (80.46%), PTB-XL (71.68%) and Chapman (96.86%), and generalizes competitively to the small-sample PTB cohort. Concept probes and perturbation analyses further show that its predictions rely directly on the ECG input, remain stable under option-order and wording variations, and transfer zero-shot to unseen waveform concepts. Our code is available at https://github.com/zhuoyuanli-lab/ECGJev.

0.34

A Co-Located MMG-FES Wearable Interface for Action-Updated State-Gated Stimulation

cs.HC ·2026-10-11 ·Cong Liu, Xiao Jin, Zhenhua Yu
similarity 0.46
Abstract · 9 pages, 8 figures. Submitted to IEEE Robotics and Automation Letters

Functional electrical stimulation (FES) assists movement, but fixed schedules cannot track how a repeatedly activated muscle changes, and surface electromyography (EMG) near the stimulation electrodes is often obscured by artifacts. Prior work recorded mechanomyography (MMG) passively during closed-loop FES cycling and estimated FES-induced fatigue or predicted fatigue for FES control, without gating stimulation with stimulated-site MMG. In this study, we present VG-FES, a wearable module co-locating microphone-based MMG, programmable charge-balanced biphasic stimulation, and embedded state-gated control at one on-skin thigh site. After each action, we compare rectus femoris (RF) mean power frequency (MPF) with an early-session baseline of the same direction, giving a relative state. A threshold on its window means enables fixed-duration bursts separated by rest periods. During onboard stimulation, MMG remained recordable, whereas near-site EMG was artifact-dominated. Under prolonged external stimulation, cycle-wise MMG tracked an independent local-pressure reference (Spearman rho = 0.968). In leave-one-session-out evaluation on 2284 squat actions (11 sessions, eight participants), MMG-based action segmentation reached an event-detection F1 of 0.923 and direction classification a balanced accuracy of 0.925. Online, over 15.69 min of self-paced squats, one participant received four bursts while MMG acquisition and action monitoring continued. Mechanical feedback from the stimulated site offers wearable FES and assistive robotics a per-action basis for state-gated stimulation.

0.40

RAG-TRACE: Tracing Measurement Evidence and Diagnostic Completeness in AI-Generated ECG Reports

cs.CL ·2026-11-01 ·Elyas Irankhah, Shahed Salehzehi, Parisa Hosseini
code/data similarity 0.45
Abstract · Code and study materials available at https://github.com/Elyasirankhah/RAG-TRACE

Purpose: Retrieval-augmented generation (RAG) can introduce electrocardiogram (ECG) criteria that lack patient-specific measurement support. This study evaluates the reliability of retrieved diagnostic criteria in AI-generated ECG reports. Methods: We analyzed 1,083 human-validated PTB-XL patient records using a library of 47 criteria cards, including 33 without numeric verification rules. Language models generated structured reports with and without retrieval. Criterion applicability was evaluated using Marquette 12SL measurements, with independent assessment using Uni-G and comparison against expert-validated diagnostic labels. Results: Keyword retrieval introduced untestable or measurement-disconfirmed criteria in 46.8% of GPT-4.1-mini reports. A deterministic verification rule withdrew 647 criterion applications, including 639 without numeric checks, while preserving 268 of 269 applications independently supported by Uni-G. Restricting retrieval to 14 testable criteria reduced the initial unsupported-criterion rate to 13.8%. Under matched instructions and an identical output schema, retrieval increased unsupported-criterion reports from 17.4% to 31.1%, while hypertrophy recall remained unchanged at 62.6%. Conclusion: These findings identify a fundamental limitation of retrieval-based clinical report generation: retrieved criteria do not necessarily provide valid patient-specific evidence. Measurement-based verification can reduce unsupported criterion applications, but diagnostic validity requires independent assessment.

0.41