Ultrasound and Magnetic Resonance Elastography of the Neck: Biomechanical Basis, Clinical Applications and Methodological Challenges

Rashedul Hasan, Anna Yang, Kishani Kandasamy, Hossein Bahreinizad, Md Ashikuzzaman, Suman K. Chowdhury, Shahed Mohammed

eess.IV ·submitted 2026-10-08 18:06 UTC · v1
unusual

Abstract

Many neck disorders, from chronic neck pain and headache to thyroid and nodal malignancy and radiation-induced fibrosis, both arise from and alter the mechanical properties of neck tissues, yet clinical assessment still relies on morphology, palpation and symptoms. Ultrasound and magnetic resonance elastography (MRE) now allow these properties to be measured in vivo, but the neck literature differs widely in methodology and remains dispersed across modalities and tissues. This review brings together the biomechanical basis of neck tissue behavior, the clinical conditions that alter it, and the principles and applications of ultrasound elastography and MRE in the neck. We show that ultrasound elastography has demonstrated diagnostic value for thyroid nodules and cervical lymph nodes, while neck MRE has been shown to be feasible but remains at an early stage. Across both modalities, reported values depend strongly on fiber orientation, posture, muscle activation and acquisition conditions. We therefore propose reporting recommendations and discuss how measurements beyond stiffness, including anisotropy, viscosity and nonlinearity, could improve disease management and inform computational models of the neck.

Closest papers

Nearest neighbours by meaning across the whole index.

Ultrasound Elastography: Excitation Methods and Low-Cost Alternative

cs.CV eess.IV ·2026-11-11 ·Corin F. Otesteanu
published similarity 0.54
Abstract

Ultrasound elastography (or elasticity imaging) is a method that characterizes the tissue mechanical properties with the purpose of aiding the diagnosis of certain pathologies. Many novel clinical applications use the assumption that these tissue properties are changed in the presence of pathological processes. Characterization of tissue mechanical properties is achieved by imaging the response of the tissue to an induced stress or excitation. A plethora of approaches have been used in the scientific community to induce excitation in tissue, which usually require additional tools or devices than are costly. Furthermore these are usually integrated in the imaging pipeline using additional software or hardware for control. An alternative that is simpler, faster to integrate (and low-cost) is proposed in this work, by using readily available smartphones as a source of mechanical excitation. The induced displacements are used to solve the elasticity inverse problem (IP) using a finite element method (FEM) approach. Preliminary elasticity reconstructions of a phantom that mimics tissue are similar to when using a dedicated actuator as means of inducing harmonic displacements, illustrating the feasibility of our low-cost ultrasound elasography approach.

0.36

WAVES: Wearable Active Vibroacoustic Elasticity Sensing for Low-Cost and Non-Invasive Muscle Stiffness Monitoring

cs.ET ·2026-11-04 ·Shubham Rohal, Dong Yoon Lee, Shijia Pan
similarity 0.49
Abstract

Muscle stiffness is a physiologically meaningful biomarker associated with a wide range of musculoskeletal and neurological conditions. However, existing muscle stiffness assessment methods, such as shear wave elastography (SWE), manual palpation, and handheld mechanical probes, remain impractical for continuous, everyday monitoring due to bulky hardware, operator dependence, and limited self-use. Meanwhile, wearable muscle-sensing modalities such as Electromyography (EMG) and Mechanomyography (MMG) primarily capture muscle activation rather than passive mechanical properties of tissue at rest, leaving a critical gap in portable monitoring of resting muscle stiffness. In this work, we present WAVES, a wearable system for continuous monitoring of muscle stiffness via active vibroacoustic sensing. WAVES injects controlled vibroacoustic excitation into soft tissue and estimates muscle stiffness by measuring shear-wave propagation velocity through Time-Difference-of-Arrival (TDoA) analysis across spatially distributed wearable sensors. To achieve accurate TDoA estimation with low-cost wearable hardware, we develop a physiology- and quality-guided vibroacoustic sensing framework that effectively screens and refines TDoA for accurate stiffness estimation. We evaluate WAVES through characterization experiments using ballistic jelly and end-to-end on-body evaluation with human participants. Results demonstrate that WAVES can robustly estimate muscle stiffness across multiple users and different superficial muscles while enabling lightweight, wearable deployment for everyday use. WAVES is able to achieve a muscle stiffness estimation error of 13.1% or less across multiple users and multiple muscle groups.

0.25

Uncertainty Quantification in Cardiac Model Personalisation from Ultrafast Ultrasound

cs.LG q-bio.QMphysics.med-ph ·2026-09-28 ·Camilla Ferrario, Maelys Venet, Olivier Villemain, Maxime Sermesant
top venueunusual similarity 0.42
Abstract

Cardiac model personalisation requires inferring mechanical parameters that are not directly measurable in vivo. Ultrafast ultrasound shear wave elastography (SWE) enables non-invasive tracking of myocardial stiffness dynamics over the cardiac cycle, providing a target for personalisation. However, mapping these observations to subject specific model parameters remains ill-posed, as multiple parameter sets can reproduce the same stiffness dynamics. We formulate SWE-informed personalisation as a statistical inference problem using simulation-based inference (SBI). Using a subject-adapted 0D cardiovascular model and neural posterior estimation, we estimate model-conditional posterior distributions over active stiffness scale k0, contraction rate kATP, and relaxation rate kSR, conditioned on SWE-derived curve features and subject specific context. Among six healthy volunteers, four passed objective prior-support diagnostics and were retained for quantitative posterior analysis. Curve-level RMSE against the observed SWE target decreased from 12.61 $\pm$ 5.55 kPa for the prior predictive median to 1.14 $\pm$ 0.38 kPa for the posterior predictive median, an 89.7 $\pm$ 4.2% reduction. Posterior analysis revealed parameter-specific uncertainty, k0-kATP compensation, weaker constraint of kSR, and the importance of prior-predictive diagnostics for assessing whether each subject is represented within the modelled SWE feature space. These results support SBI for uncertainty aware SWE-based personalisation, while identifying prior support and forward-model adequacy as key diagnostics.

0.43

Finite Element Elastographic Analysis of Kidney MRI for Computational Assessment of Chronic Kidney Disease

eess.IV ·2026-11-04 ·A. Bhattacharjee, M. M. Rahman, A. Srivastava, E. H. Bhuiyan
unusual similarity 0.42
Abstract · Original research article

Background and Objective: Bridging kidney MRI with patient-specific computational biomechanics may enable accurate, noninvasive assessment of chronic kidney disease. To develop a finite element method (FEM)-based computational elastography framework for estimating renal biomechanical properties from conventional kidney MRI and to evaluate its ability to differentiate chronic kidney disease (CKD) from healthy controls. Methods: Kidney segmentation masks from healthy control and CKD subjects were converted into triangulated surface meshes, and watertight surfaces were tetrahedralized to create volumetric FEM meshes. Kidney tissue was modeled as a nearly incompressible Kelvin-Voigt viscoelastic material. Harmonic wave propagation simulations were performed using FEM, followed by Helmholtz-based elastographic inversion to estimate complex shear modulus distributions. Derived biomechanical parameters were used for CKD classification using machine learning models. Results: FEM-derived stiffness estimates showed consistency with stiffness ranges previously reported in human renal MRE studies and also showed measurable differences between healthy and CKD groups. Machine learning models trained on FEM-derived features achieved classification performance up to 86.26% cross-validation accuracy with an AUC of 0.92. Conclusion: FEM-derived elastographic parameters showed potential for distinguishing CKD from healthy tissue and may provide an alternative for renal biomechanical assessment without dedicated MRE acquisition.

0.26

Pulse-Echo Ultrasound Methods for Speed-of-Sound Estimation: A Review

eess.IV physics.med-ph ·2026-08-23 ·Can Deniz Bezek, Orcun Goksel
survey similarity 0.42
Abstract

Speed of sound (SoS) is fundamental to ultrasound beamforming and a promising quantitative biomarker for tissue characterization. Although transmission- and reflector-based methods have demonstrated the diagnostic value of SoS, their need for dedicated hardware or constrained acquisition geometries limits clinical usability. Pulse-echo SoS imaging addresses these limitations by estimating SoS with conventional ultrasound probes and data acquisition paradigms. This review provides a structured overview of pulse-echo SoS estimation methods, categorized by the source of information used, the estimated SoS representation, and the reconstruction strategy. We review methods by their locality, e.g., global, layer-wise, and spatially-resolved predictions, as well as in terms of prediction strategies including gradient-free search, optimization-based solutions, analytical closed-form approximations, and deep learning-based methods. Finally, we summarize emerging diagnostic applications in the breast, liver, and muscle, while discussing challenges and possibilities with SoS imaging.

0.28

Mechanical behaviour of the intact brain-skull interface under tension, compression and shear

cs.CE ·2026-11-12 ·Sajjad Arzemanzadeh, Karol Miller, Ron Kikinis, Adam Wittek
similarity 0.41
Abstract

The brain-skull interface governs the boundary/contact conditions of the brain in computational head models, yet its mechanical behaviour remains poorly characterised experimentally. We characterised the brain-skull interface by testing brain-skull complex samples (skull bone, meninges and brain tissue in their native arrangement) excised from 19 sheep cadaver heads under uniaxial tension and compression at strain rates of 0.002, 0.02 and 0.2 s$^{-1}$ and simple shear at 0.02 s$^{-1}$. Histology confirmed that the meningeal layers remained intact and attached to the brain and skull after excision. We acquired magnetic resonance images of every sample, reconstructed their geometry and developed a sample-specific finite element model of each experiment. We represented the brain tissue using a second-order Ogden hyperelastic model and the brain-skull interface using a cohesive zone model. Using inverse finite element analysis, we first calibrated the brain tissue parameters for each subject. Keeping these parameters fixed, we then identified the interface parameters from the corresponding brain-skull complex specimen. The brain tissue shear modulus increased with strain rate from 0.94 $\pm$ 0.16 kPa at 0.002 s$^{-1}$ to 1.60 $\pm$ 0.30 kPa at 0.2 s$^{-1}$. In contrast, the interface properties (maximum normal traction 2.35 $\pm$ 0.81 kPa, maximum tangential traction 4.14 $\pm$ 0.97 kPa, and fracture energy 0.35 $\pm$ 0.14 N m$^{-1}$) showed no significant dependence on strain rate or loading mode. In all samples, failure occurred within the subarachnoid space. Neither a rigid brain-skull connection nor frictionless sliding contact with separation reproduced the measured mechanical response or failure mode of the brain-skull interface. Our experimentally derived cohesive zone model provides a realistic representation of the brain-skull interface for computational head models.

0.32

Anthropomimetic Soft Robotic Forearm with Independently Articulated Carpal Bones Enabling Human-Like Adaptive Stiffness Modulability

cs.RO ·2026-09-24 ·Yoshinobu Obata, Yinlai Jiang, Hiroshi Yokoi, Shunta Togo
code/data similarity 0.41
Abstract · 30 pages, 17 figures, 5 tables. Supplementary material included. Data, CAD and code: https://github.com/TogoLab/anthropomimetic-forearm-carp…

The human wrist exhibits adaptive stiffness modulability: joint stiffness anisotropy can be actively regulated through muscle co-contraction. This functionality is essential for stable manipulation, yet the underlying morphological factors remain unclear. To identify these factors, we developed an anatomically accurate anthropomimetic soft robotic forearm comprising eight independently movable carpal bones interconnected by ligaments, 22 actuated muscles, and compliant fingertips. We measured wrist joint stiffness under four muscle activation patterns across three skeletal configurations: anatomically normal carpal bones, a fused proximal carpal row, and a geometric ellipsoidal skeleton. The stiffness ellipse exhibited low stiffness along the dart-throwing motion (DTM) direction when finger muscles were activated, but high stiffness along the same direction when wrist and finger muscles were activated simultaneously. These results agree with previously reported human measurements, demonstrating that precise anatomical replication reproduces human-like stiffness modulability. Fusing the proximal carpal row eliminated the low DTM-direction stiffness under finger muscle activation, while the geometric ellipsoidal skeleton showed poor stiffness ellipse reorientation across all conditions. Carpal bone motion analysis revealed significantly opposing coupling patterns between wrist and finger muscles at the proximal carpal row, accompanied by a consistent but non-significant trend at the midcarpal joint, providing a mechanical explanation for this modulation. These findings demonstrate that carpal bone morphology plays a dominant role in human wrist stiffness modulation and provide design principles for humanoid robot wrists.

0.37

A hybrid CNN-adjoint optimization framework for reconstruction of viscoelastic tissue properties in magnetic resonance elastography

cs.LG math.OCmath.NA ·2026-10-02 ·Anwesa Dey, Johann Rudi, Elena Cherkaev
similarity 0.41
Abstract

Magnetic resonance elastography (MRE) is a noninvasive imaging modality for quantifying the viscoelastic properties of soft tissues from shear wave propagation. Recovering the complex-valued shear modulus from measured displacement fields leads to a severely ill-posed inverse problem, particularly in the presence of noise and limited boundary excitations. We investigate adjoint-based optimization, convolutional neural network (CNN) reconstruction, and a hybrid framework combining both approaches. The forward model is based on a scalar form of the modified stationary Stokes system with a complex shear modulus. We establish well-posedness of the forward problem, existence of minimizers, and first-order optimality conditions for the adjoint-based formulation, and implement a nonlinear conjugate-gradient method with Armijo line search. While PDE-constrained optimization can accurately refine coefficient reconstructions, its performance depends strongly on initialization. We therefore construct a two-dimensional CNN that maps complex-valued displacement measurements to spatially varying complex shear modulus fields and provides rapid, informative initial reconstructions. The proposed hybrid method uses the CNN reconstruction to initialize the adjoint-based optimization, yielding faster convergence and improved accuracy. The CNN is trained on coefficient fields containing individual perturbations and tested on both individual and previously unseen combined configurations. Numerical experiments demonstrate that the CNN generalizes to these more challenging configurations, while subsequent PDE-constrained optimization further refines the reconstructed coefficient. These results demonstrate the potential of combining data-driven initialization with physics-based optimization for efficient and accurate MRE reconstruction.

0.12