Differentiable Co-Design of Ultrasound Probe Geometry and Acquisition for Plane-Wave Imaging

Somashekhar Dehury, Mahesh Raveendranatha Panicker

eess.SP ·submitted 2026-10-12 12:03 UTC · v1
code/data

Abstract

Ultrasound image quality depends on both the probe geometry and the transmit and receive processing, but engineers usually design the two separately. Physics-based simulators (Field II, k-Wave, MUST/SIMUS) compute the radio-frequency channel data but give no gradients, so design studies must use grid or black-box searches that become slow as the number of parameters increases. We present a fully differentiable end-to-end pipeline for plane-wave ultrasound imaging, with synthetic speckle phantoms, an analytical frequency-domain RF simulator, delay-and-sum beamforming and an extensible image-quality objective. The element pitch, the transmit and receive apodization and the global speed of sound are differentiable parameters of one computation graph, so one reverse-mode pass gives their exact joint gradients. Studies of each parameter rank the design parameters: the optimized pitch decreases the grating lobes, and the speed-of-sound autofocus gives clear contrast and resolution gains. Over 15 test phantoms, the optimized pitch increases the cyst gCNR from 0.55 to 0.66 (about 20%), and the speed-of-sound correction adds a further 0.05. The lateral point-spread function agrees with Field II within about 10%, and on 49 real CUBDL plane-wave acquisitions from five institutions, the framework finds the global speed of sound with a mean absolute difference of 16.6 m/s from the published corrections and phantom speed of sound.

Comment
11 pages, 4 figures, 2 tables. Code: https://github.com/sdehury34/jax-ultrasound-codesign . Interactive demo: https://pulseecho.in/demos/diffUS/

Closest papers

Nearest neighbours by meaning across the whole index.

Common Offset Gathers for Efficient Multistatic Beamforming in Frequency Domain

eess.SP physics.med-ph ·2026-10-29 ·Marko Jakovljevic, Xiaohong Wang, Ion Candel, Anthony E. Samir, Kai Thomenius
similarity 0.63
Abstract · 11 pages

Frequency-domain beamformers reconstruct ultrasound images with far fewer operations than delay-and-sum (DAS). However, for multistatic synthetic aperture data, they require Fourier transforms over the full transmit-receive aperture and large amounts of memory, limiting their use for volumetric imaging. Here, we adapt common offset gathers (COG) from seismic imaging to frequency-domain beamforming: transmit-receive pairs with the same offset are beamformed independently using a bistatic formulation, holding only one offset gather in memory at a time. We combine COG with the range-Doppler algorithm (RDA) and evaluate COG RDA against DAS and conventional multistatic RDA using simulated, phantom, and in vivo acquisitions. For linear arrays, COG RDA reduced memory requirements by two orders of magnitude relative to conventional RDA and beamformed up to 25x faster than DAS, with comparable image quality. For a 32x32-element matrix array, COG RDA requires only 2-D instead of 4-D Fourier transforms, and reconstructed a volume on a GPU 32x faster and with 67x less energy than DAS. COG RDA makes frequency-domain multistatic beamforming practical for matrix arrays and power-constrained devices.

0.32

Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter Imaging

eess.SP physics.med-ph ·2026-09-19 ·Esteban Avilés, Michael Oelze, Roberto Lavarello, Andres Coila
published similarity 0.61
Abstract · 4 pages, 3 figures, accepted to 22nd International Symposium on Medical Information Processing and Analysis (SIPAIM)

The depletion method (DM) enables the estimation of the acoustic nonlinearity parameter, B/A, in pulse-echo ultrasound using fundamental-band signals from low- and high-pressure acquisitions, but requires accurate knowledge of the acoustic-pressure scaling factor. This requirement limits experimental translation because of the slightly nonlinear scaling between excitation voltage set on the ultrasound system and pressure at the probe. We propose a complementary-aperture pulse sequence (CAPS) that recovers the linearly scaled low-pressure signal by coherently summing RF acquisitions from interleaved transmit subapertures whereas a full-aperture transmission provides the high-pressure signal maintaining the same excitation voltage configured in the ultrasound system. In silico phantoms of a homogeneous sample and a reference evaluated four aperture partitions of CAPS and compared them with the DM under relative scaling-factor errors from 0.01% to 10%. Using the DM, errors of 0.1% and 1% produced absolute B/A biases of up to 27.3% and 58.3%, respectively. CAPS yielded mean B/A values of 8.9 for a prescribed value of 9, with biases of less than 1.3% across all partitions, matching an ideally calibrated DM. These results demonstrate that CAPS preserves depletion-based B/A estimation without using the pressure scaling factor as a reconstruction input, supporting experimental and in vivo implementation without pressure-ratio calibration.

0.32

Physics-Grounded Objective-Conditioned Deep Beamforming for Adaptive Resolution-Contrast Control in Plane-Wave Ultrasound Imaging

eess.SP ·2026-10-05 ·Gopika Gopikrishnan, Mahesh Raveendranatha Panicker
similarity 0.61
Abstract

Plane-wave ultrasound enables ultrafast imaging but remains constrained by a receive-aperture-dependent resolution- contrast trade-off. Fine structures benefit from high spatial resolution, whereas diffuse tissue and lesions often require improved contrast and clutter suppression. These objectives favor different receive-aperture configurations, yet conventional beamformers operate at a fixed imaging point. We propose a physics-grounded, objective-conditioned deep beamformer that enables controllable image formation within a single network. From single-plane-wave channel data, the network predicts spatially varying receive-apodization weights, conditioned on the desired imaging objective through feature-wise linear modulation. The conditioning variable is explicitly tied to a physically defined receive-aperture configuration and supervised using the correspondingcoherent-plane-wave-compounding target. Differentiable, region-of-interest-aware, physics-guided quality losses further promote objective-specific resolution and contrast during end-to-end training. The model is trained exclusively on simulated channel data spanning diverse target morphologies and echogenicities, and evaluated zero-shot on the PICMUS and CUBDL datasets, including unseen probe geometries. The proposed framework generally improves upon single-plane-wave delay-and-sum reconstruction while providing objective-dependent control. On PICMUS, the proposed method improves gCNR by up to 16.8% and axial resolution by 22% compared with the reference beamforming method. Moreover, switching between the contrast- and resolution-oriented objectives enables controllable changes of up to 4.9% in lesion contrast and 11.4% in lateral resolution. Compared with the best-performing learning-based baselines, the proposed method reduces lateral full-width-at-half-maximum by 5.9% and improves generalized contrast-to-noise ratio by 6.6%.

0.29

Open-H-RF: A Large-Scale Ultrasound Raw Channel Dataset Initiative

eess.SP ·2026-10-09 ·T. S. W. Stevens, S. W. Penninga, W. L. van Nierop, O. Nolan, F. Duelmer +181 more
dataset similarity 0.59
Abstract · Preprint: 15 pages, 15 figures, 3 tables. Dataset (v1.0): https://open-h.github.io/Open-H-RF

Medical ultrasound is a widely accessible, real-time imaging modality, yet progress in data-driven ultrasound research is constrained by the scarcity of large, open collections of raw channel data. Existing public datasets are predominantly based on reconstructed images, while the smaller number of open raw channel datasets is generally specialized to individual reconstruction or imaging tasks, limiting the development of general-purpose methods and foundation models that operate directly on ultrasound signals. While interpretive tasks can be performed on reconstructed images, many of the key technical challenges in ultrasound imaging, such as aberration correction and reverberation reduction, require the raw channel data itself. We introduce Open-H-RF, a community-driven open dataset initiative comprising approximately 40 TB of medical ultrasound raw channel data in 19,332 acquisitions, contributed by 38 institutions across academia and industry, spanning in-vivo, phantom, and simulated data. Each dataset is annotated with a number of supported tasks, ranging from generalized reconstruction, speed-of-sound estimation, and aberration correction to blood flow imaging, motion estimation, quantitative ultrasound, and image interpretation. Open-H-RF provides a unified, self-contained data format that captures raw channel data together with the acquisition, probe, metadata, and reconstruction information required for reproducible processing across scanners and acquisition schemes. The corpus is publicly hosted under a permissive CC BY 4.0 license and can be streamed programmatically and selectively at scale. By bringing heterogeneous ultrasound data into a common representation, backed by modern data infrastructure, Open-H-RF lowers the barrier to working with raw channel data and aims to open medical ultrasound research to new communities. See https://open-h.github.io/Open-H-RF.

0.47

Laplace-Domain Beamforming for Ultrafast Plane-Wave Imaging

eess.IV physics.med-pheess.SP ·2026-10-06 ·Martin F. Schiffner
theoryunusual similarity 0.57
Abstract · 5 pages, 3 figures, 1 table; submitted to 2026 IEEE International Ultrasonics Symposium (IUS), Raleigh, NC, USA, Oct. 2026

Fourier-domain beamforming methods, such as filtered backpropagation (FBP), are beneficial in ultrafast plane-wave imaging due to their computational efficiency and high image quality. These methods use the Fourier diffraction theorem (FDT) for steered plane waves to exactly invert linear scattering models. However, these models rely on simplifying assumptions that degrade image quality. The use of linear transducer arrays, in particular, requires: (i) a two-dimensional space, and (ii) lossless tissues. Herein, Laplace-domain beamforming is proposed. This method retains the computational efficiency of the Fourier-domain methods while improving image quality for linear arrays. Specifically, the method uses a generalization of the FDT, here called Laplace diffraction theorem, to account for three spatial dimensions, finite array element heights, and lossy tissues. A phantom experiment showed that the proposed method improves image uniformity, spatial resolution, and contrast compared to FBP. Using three steering angles, the lateral and axial -6 dB-widths of wires reduced by up to 12% and 25%, respectively, while the contrast of anechoic regions improved by up to 9%.

0.26

Pulse-Echo Ultrasound Methods for Speed-of-Sound Estimation: A Review

eess.IV physics.med-ph ·2026-08-23 ·Can Deniz Bezek, Orcun Goksel
survey similarity 0.57
Abstract

Speed of sound (SoS) is fundamental to ultrasound beamforming and a promising quantitative biomarker for tissue characterization. Although transmission- and reflector-based methods have demonstrated the diagnostic value of SoS, their need for dedicated hardware or constrained acquisition geometries limits clinical usability. Pulse-echo SoS imaging addresses these limitations by estimating SoS with conventional ultrasound probes and data acquisition paradigms. This review provides a structured overview of pulse-echo SoS estimation methods, categorized by the source of information used, the estimated SoS representation, and the reconstruction strategy. We review methods by their locality, e.g., global, layer-wise, and spatially-resolved predictions, as well as in terms of prediction strategies including gradient-free search, optimization-based solutions, analytical closed-form approximations, and deep learning-based methods. Finally, we summarize emerging diagnostic applications in the breast, liver, and muscle, while discussing challenges and possibilities with SoS imaging.

0.28

UltraDiff: Differentiable Ray Tracing in Ultrasound for Shape Optimization

cs.GR cs.CV ·2026-10-06 ·Felix Duelmer, Magdalena Wysocki, Nassir Navab, Mohammad Farid Azampour
top venuehot direction similarity 0.57
Abstract · 4 pages, 4 figures, 1 table. Accepted at SIGGRAPH Asia 2026 Technical Communications

Physically-based differentiable rendering enables gradient-based optimization of scene parameters by matching rendered images to measurements, but has so far mainly focused on light transport. We extend this paradigm to medical ultrasound, where image formation resembles transient rendering: echoes are binned by time-of-flight rather than projected onto an image plane. We present UltraDiff, a modular framework for differentiable ultrasound ray tracing. UltraDiff formulates ultrasound image formation as a path-space integral, gated by travel time between the transducer and tissue interfaces, and derives a Monte Carlo estimator of both the forward model and its gradients with respect to scene parameters. We demonstrate this on an inverse geometry estimation: starting from a sphere, an SDF is optimized until simulated echoes match measured ones, recovering vertebral surfaces from simulated B-mode sweeps and from a real robotic acquisition of a spine phantom. Unlike state-of-the-art ultrasound shape reconstruction methods, which rely on pre-segmented images, our approach operates unsupervised on B-mode images through analysis-by-synthesis, while achieving competitive geometric accuracy. Implemented on top of Mitsuba 3, UltraDiff brings differentiable path tracing to a new sensing modality and provides a foundation for inverse problems in acoustic imaging.

0.47

Shadow Reduction in Ultrasound Imaging Using Differentiable Simulation and Radiance Field Decomposition

cs.CV ·2026-09-24 ·Valentin Bacher, Pak Hei Yeung, Bernhard Kainz, Madeleine K. Wyburd, Nicola K. Dinsdale +2 more
similarity 0.56
Abstract · 39 pages, 19 figures, submitted to Medical Image Analysis

Acoustic shadows from bone and other highly attenuating tissues obscure clinically important structures in ultrasound. In fetal brain imaging, skull-induced artefacts disproportionately degrade the hemisphere closer to the transducer (proximal), limiting symmetric assessment of the two hemispheres. Existing correction methods require raw scanner data, impose restrictive assumptions on tissue properties, or rely on generative models that may hallucinate anatomy. We present RFlash, a physics-informed post-processing method that decomposes beamformed ultrasound images into explicit attenuation and scatter-intensity maps using a differentiable radiance-field formulation of image formation. Attenuation-adaptive re-rendering then removes the dependence of the signal at each depth on the intervening tissue, equivalent to virtually advancing the transducer into the tissue. Across 1,261 3D fetal brain volumes, 143 real 2D curvilinear abdominal scans, and 1,200 simulated 2D linear-probe liver scans, RFlash reduces shadow-related intensity differences more effectively than classical Hughes-Duck attenuation correction. For a gestational-age model trained on the distal hemisphere (further from the transducer) and applied to the proximal hemisphere, prediction error decreases by 5.1 days (40%) relative to the original images. The estimated attenuation maps also yield shadow-confidence maps that improve random-forest bone-shadow segmentation over the image alone and receive greater SHAP importance than an existing neural confidence-map baseline, suggesting greater physical consistency. RFlash requires neither hardware modification nor access to raw scanner data and supports 2D and 3D acquisitions with linear and curvilinear probes, making it widely applicable allowing clinicians to use our method on their already acquired scanners and images.

0.30