Unified Projection-Domain Material Decomposition for Energy-Integrating and Photon-Counting CT

Dale Black, Surya Naidu, Hamidreza Khodajou-Chokami, Sabee Molloi

eess.IV physics.med-ph ·submitted 2026-10-21 20:49 UTC · v1
theory

Abstract

Objective: To develop a unified and efficient method for projection-domain material decomposition in spectral computed tomography (CT) that incorporates an arbitrary number of spectral measurements, independent of the acquisition mechanism, without reducing the measurements to two effective energies. Approach: A nested optimization method was developed for two-material basis decomposition from K>=2 spectral measurements. It extends an existing two-measurement univariate method by replacing its polynomial approximation with profile minimization of the Poisson negative log-likelihood, reducing the inversion to two one-dimensional searches that use all measurements without spectral rebinning. The method was evaluated in simulation for energy-integrating (GE Revolution Apex Elite, Siemens SOMATOM Flash) and photon-counting (Siemens NAEOTOM Alpha) CT. Main results: The method recovered line integrals with noiseless iodine errors below 3e-6 g/cm^2 and noise within 3% of the Cramer-Rao lower bound (CRLB). Virtual monoenergetic images (VMIs) agreed, at 70 keV and above, with reference CT numbers and, on all three systems, with vendor VMIs of physical scans (mean absolute error 6-22 HU). For photon counting, four measurements instead of two reduced the iodine CRLB by 19% and image noise by 8-16% at every VMI energy, and lowered the regression error at 70 keV and above. Iteration and tolerance settings changed CT numbers by less than 0.04 HU, and decomposition took 0.34-0.67 s per million rays on an A100 GPU. Significance: The method provides fast, accurate decomposition for any spectral CT acquisition mechanism and number of measurements, so one method can serve every scanner in a virtual imaging trial.

Comment
27 pages, 10 figures, 6 tables

Closest papers

Nearest neighbours by meaning across the whole index.

SpectralCTGaussians: Projection-Domain Reconstruction and Basis Material Decomposition for Spectral CT using 3D Gaussian Splatting

cs.CV ·2026-08-29 ·Reinout Vos, Saptarshi Neil Sinha, Michael Weinmann
similarity 0.63
Abstract

Spectral computed tomography (CT) extends conventional CT by measuring attenuation across multiple energy channels, allowing improved modeling of physical X-ray interactions and energy-dependent material behavior and leading to richer scene understanding. We present a novel method for spectral CT reconstruction and basis material decomposition using 3D Gaussian Splatting by adding per-Gaussian basis material fractions to the set of learnable parameters, which together with a set of energy-dependent basis functions define the attenuation across the full spectral range. The basis functions represent various physical attenuation models such as photoelectric absorption and Compton scattering, and are jointly optimized across all energy channels through a differentiable polychromatic forward model, with material decomposition performed via mean-shift clustering of the resulting coefficients. We evaluate our method on a baseline real-world dataset as well as a synthetic dataset that we introduce, comparing against traditional reconstruction algorithms and state-of-the-art learning-based CT reconstruction methods. Our approach outperforms all traditional baselines in novel view synthesis and achieves the best PSNR among all compared methods for spectral CT volume reconstruction, while describing all energy channels with a single shared representation that requires a number of Gaussians comparable to single-channel Gaussian splatting-based CT reconstruction approaches. For basis material decomposition, no traditional or learning-based baseline offers one-step decomposition with direct RGB material segmentation, and our method additionally recovers the photoelectric basis with higher PSNR than traditional pipelines.

0.25

Gaussian Material Fields for Volumetric Multi-Energy CT Decomposition

cs.CV ·2026-10-07 ·Jian Lin, Jiancheng Fang, Hongming Shan, Shaoyu Wang, Yang Chen +1 more
similarity 0.62
Abstract · 13 pages, 12 figures

Volumetric material decomposition in multi-energy computed tomography requires a representation that organizes multiple three-dimensional material fields in a common spatial domain while retaining differences in composition and local structure. We observe that spatial primitives can be shared across materials without tying their coefficients, but their local capacity must respond to material-specific reconstruction needs. We introduce Gaussian material fields, which represent multiple material distributions with shared anisotropic 3D Gaussian primitives and independent nonnegative material coefficients. The shared geometry defines a continuous spatial basis, while the coefficients determine each primitive's contribution to the individual material fields. To reconstruct this representation from multi-energy projections, a differentiable spectral forward model combines Gaussian material path integrals with a calibrated basis matrix, enabling joint optimization of spatial geometry and material composition. Material-aware adaptive density control retains material-specific refinement evidence before aggregation and adjusts local representation capacity to accommodate both spatially extensive components and sparse details. Experiments use synthesized multi-energy projections generated from pseudo-reference material maps constructed by conventional methods from publicly available CT data. Across 15 cases, our approach improves average PSNR by 4.03 dB and SSIM by 4.96% over the strongest baseline, while reducing NRMSE by 33.45%. Material-wise comparisons and component ablations support improved recovery of localized structures, while runtime and memory measurements show favorable computational scaling. These results establish Gaussian material fields as an explicit, adaptive representation for volumetric multi-material reconstruction.

0.27

TRACE: Two-Stage Detector-Response Estimation With Angular Cosine Expansion for Ring Artifact Correction in Photon-Counting CT

cs.CV ·2026-09-14 ·Jigang Duan, Heran Wang, Ligen Shi, Zheng Sun, Ping Yang +1 more
similarity 0.57
Abstract

Detector response nonuniformity introduces systematic projection errors and ring artifacts in photon-counting detector computed tomography (PCD-CT). In measured PCD-CT data, residual stripe amplitudes vary slowly with projection angle, which fixed-bias models cannot adequately capture. We propose TRACE, a two-stage unsupervised sinogram decomposition method for estimating and correcting these response-related errors. TRACE represents stripes as a fixed bias plus low-order discrete cosine transform (DCT) components, using a small number of coefficients to describe angular variations at each detector element. A learnable analysis--synthesis architecture represents the ideal projections, while two-stage optimization separates them from fixed and then dynamic stripes. An angular-gradient soft orthogonality constraint suppresses correlated variations within the shared DCT gradient subspace, reducing the leakage of object structures into the artifact estimate. All parameters are optimized directly on the measured sinogram without paired training data. Experiments on measured QRM mouse phantom and porcine trotter data show that TRACE suppresses ring artifacts and improves image uniformity while preserving edge sharpness, soft-tissue texture, and trabecular detail.

0.34

EPC-RawDC: Equivariant Physics Constrained 3D Diffusion with Raw Data Consistency for CBCT to CT Image Synthesis

cs.CV cs.AIphysics.med-ph ·2026-10-29 ·Alzahra Altalib, Chunhui Li, Alessandro Perelli
code/datahot direction similarity 0.56
Abstract · 10 pages, 4 figures, 8 tables

Cone beam computed tomography (CBCT) is widely used for image guided and adaptive radiotherapy, but scatter, noise, reconstruction artifacts, and inaccurate Hounsfield units limit its quantitative use. Recent diffusion approaches for CBCT to CT synthesis remain largely image driven, with limited use of the underlying acquisition physics. We propose EPC-RawDC, a conditional 3D latent diffusion framework that combines volumetric generative modeling with two complementary constraints derived from the same scanner specific CBCT forward model. First, geometry equivariant projection constraint exploits the rotational equivariance of the CBCT forward operator by supervising the synthesized CT through correspondingly transformed reference projections. Second, raw data consistency uses noisy projection measurements of the individual acquisition as scan specific supervision during training. The framework was evaluated on two independent clinical cohorts with distinct scanner geometries and projection sources. For JUST, the physics constraints used measured clinical raw CBCT projections acquired with an Elekta system, whereas for SynthRAD2025 the projection data were simulated from the reference CT using a Varian geometry with Poisson photon statistics. EPC-RawDC achieved an SSIM of 0.97 in both cohorts and improved PSNR over DDPM by 2.66 dB on SynthRAD2025 and 2.10 dB on JUST, while reducing mean MAE by 20.3% and 19.0%, respectively. On SynthRAD2025, projection consistency and equivariant errors were reduced by 55.0% and 41.4% relative to DDPM. These results demonstrate that combining a volumetric diffusion prior with scanner geometry and scan specific projection measurements improves quantitative CBCT to CT synthesis across both simulated and measured projection domains. Code available after publication at https://github.com/ALZAHRAALTALIB/EPC-RawDC.

0.37

Feed-Forward Hierarchical Gaussian Diffusion for Extreme CT Reconstruction

cs.CV ·2026-08-15 ·Yuezhe Yang, Li Cheng
benchmarkcode/data similarity 0.55
Abstract

Reconstructing three-dimensional computed tomography (CT) from severely constrained projections is highly ill-posed. Sparse angular sampling, restricted angular coverage, and low photon counts can occur individually or jointly, obscuring global anatomy and local tissue detail. Many learned CT reconstruction methods are tailored to a single dominant degradation. Existing diffusion and Gaussian approaches commonly recover global structure and local detail within a shared representation. We propose HiGDiff, a feed-forward hierarchical Gaussian diffusion framework that decomposes reconstruction both spatially and from structure to detail. Physics-conditioned anatomical anchors and a foreground capacity field allocate learnable Gaussian primitives to informative regions. A structure diffusion stage first recovers global attenuation geometry, and its learned representation conditions a detail diffusion stage for residual boundaries and tissue transitions. The resulting Gaussian banks are rendered as attenuation fields and further refined by a gradient-isolated residual module. Experiments on three distinct CT benchmark datasets demonstrate state-of-the-art reconstruction performance across isolated, paired, and joint degradation settings, including improvements of 5.81 dB in macro-average peak signal-to-noise ratio (PSNR) and 0.113 in structural similarity index measure (SSIM) on the Low Dose CT Image and Projection Data (LDCT-PD) collection. Code and experimental configurations are openly available at https://github.com/Bean-Young/HiGDiff.

0.24

Physics-Driven Independent Pair Generation for Iterative Self-Supervised Low-Dose CT Denoising

cs.CV ·2026-09-01 ·Xianlei Han, Shaoyu Wang, Jiancheng Fang, Weiwen Wu, Qiegen Liu
similarity 0.54
Abstract · 10 pages, 11 figures, 3 tables. Includes a 1-page appendix. Submitted to IEEE Transactions on Circuits and Systems for Video Technology

Low-dose computed tomography (LDCT) measurements contain mixed Poisson-Gaussian noise. However, most self-supervised methods rely on generic image statistics and do not explicitly model this noise, which may limit their ability to effectively suppress realistic LDCT noise. To address this issue, we propose a physics-driven framework with cross-domain iteration for self-supervised LDCT denoising. The proposed framework proceeds in three main steps. First, a learned sinogram prior and the LDCT noise model guide posterior inference of photon counts, enabling separation of the Poisson and Gaussian components. Second, the separated Poisson and Gaussian components are respectively processed by binomial thinning and Gaussian data thinning to construct two branches, and residual scaling matches each branch's noise level to that of the observation, yielding a training pair with approximately independent noise realizations from one low-dose measurement. Finally, the pair is used to train an image-domain network whose forward-projected outputs update the prior. Through cross-domain iteration, the prior and the training pair are progressively refined while maintaining consistency with CT acquisition physics. Experiments on simulated data from AAPM, LIDC-IDRI, and LoDoPaB-CT and on real LDCT data show consistent gains over the evaluated self-supervised baselines across dose levels, with performance comparable to the evaluated supervised baseline.

0.25

Bayesian Joint Reconstruction and Uncertainty Quantification for Limited-Data Multi-Contrast X-Ray CT

eess.IV physics.optics ·2026-10-15 ·Hassan Yazdanian, Henrik Mäkinen
similarity 0.53
Abstract

Multi-contrast X-ray computed tomography (CT) combines attenuation contrast with phase and dark-field contrast, which are sensitive to refraction and ultra-small-angle scattering, respectively. Common implementations record several intensity measurements per projection angle by phase or mask stepping, which lengthens acquisition and, at a fixed exposure per measurement, raises the dose. Limited-data acquisition via sparse angular sampling or reduced stepping lowers these costs but makes the data less informative, calling for prior information and uncertainty quantification. We study joint Bayesian reconstruction of attenuation, phase, and dark-field images from separately retrieved sinograms. A joint total variation term in the prior couples edge locations across the images but not their values. Posterior sampling with the No-U-Turn Sampler complements maximum a posteriori estimates with pixel-wise uncertainty. We compare this joint prior with channel-wise total variation alone on synthetic and experimental Talbot-Lau interferometry data under full, sparse-angle, sparse-step, and combined acquisition (up to 24 times fewer exposures). Both maximum a posteriori reconstructions suppressed undersampling artifacts, with synthetic relative errors 1.6-11 times lower than those of filtered backprojection. The joint prior gave modest synthetic accuracy gains, a cleaner experimental dark-field background, 23-30 % lower mean dark-field posterior standard deviation, and 4-38% higher median effective sample size in experimental limited-data runs. In the synthetic study, posterior standard deviations matched or exceeded the errors in flat regions but underestimated them at boundaries, especially in phase. Experimental phase chains mixed slowly under angular undersampling. The formulation also applies to edge illumination with adapted forward and noise models.

0.20

Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel Detectors

eess.IV ·2026-11-02 ·Seemeen Karimi, Joseph Bendahan, K. Aditya Mohan
publishedunusual similarity 0.52
Abstract · This has been presented at a conference, proceedings are pending

Accurate spatial-resolution characterization of flat-panel detectors at megavoltage energies is difficult due to long-tailed cross-talk. Conventional measurements at high energy using line-pair gauges or knife-edge phantoms require expensive, carefully fabricated objects and precise alignment, and despite these, scatter from the phantom, detector, and imaging room can further confound the measurement. We present a method for estimating the detector point spread function (PSF) from the mean noise power spectrum (NPS) calculated from repeated radiographs in air. The approach is mathematically equivalent to an autocovariance-based method but is more practical for long-tailed PSFs. We validate the method using simulated air data and Monte Carlo N-particle (MCNP6) calculations of energy deposition in a flat-panel detector. We show that deconvolving the experimental PSF increases the accuracy of transmission radiographs of step wedges. The method requires no phantom or alignments and is straightforward to reproduce.

0.34