CS Paper Radar

Wireless Power Transfer in Titanium Implants

R. W. Porto, L. Murliky, F. R. de Sousa, A. S. de Almeida, H. M. de Albuquerque, V. J. Brusamarello

eess.SY ·submitted 2026-08-20 20:31 UTC · v1
unusual

Abstract

This work addresses the problem of applications that require wireless power transfer to devices embedded within conductive blocks. The emergence of eddy currents tends to generate a magnetic field opposing that of the transmitter, drastically reducing the coupling coefficient and, consequently, the efficiency of the process. A case study involving a titanium implant is presented and analyzed with a load designed to ensure a constant current of 1 mA. An experimental setup is assembled and evaluated to compare differences in the measured electrical parameters when the receiver is embedded in the implant, considering several solutions that mitigate or avoid the effects of eddy currents. The results show that such a system may be technically unfeasible if careful attention is not paid to the geometry of the cavity in which the receiver is positioned.

Closest papers

Nearest neighbours by meaning across the whole index.

Focusing Efficiency for Large Intelligent Surfaces

eess.SP ·2026-10-26 ·Jiawang Li, Alireza Saberkari, Buon Kiong Lau, Mats Gustafsson
similarity 0.47
Abstract

This paper studies wireless power transmission (WPT) with large intelligent surfaces (LISs) in a single-user scenario, with emphasis on power-transfer efficiency. The optimal current distribution on the LIS is derived under a total-power constraint for an arbitrary receiving antenna, providing a theoretical characterization of the maximum achievable focusing efficiency. The analysis shows that the efficiency is strongly affected by the receiving radiation pattern, antenna orientation, and transmission distance. Clear performance differences are observed among practical antenna types for specific orientations, whereas directional averaging makes their average focusing efficiencies much closer. The theoretical findings are further validated by full-wave FEKO simulations, showing good agreement with the predictions of the proposed formulation. These results provide insight into the impact of receiver characteristics and propagation geometry on LIS-enabled WPT and offer a theoretical basis for efficient single-user power-transfer design.

0.26

Effective Range and Optimal Frequency of Through-the-Earth Magnetic Induction Communication

eess.SY ·2026-09-02 ·Honglei Ma, Erwu Liu, Wei Ni, Jun Zhu, Zhijun Fang +2 more
published similarity 0.46
Abstract · This work has been accepted by IEEE Transactions on Vehicular Technology for publication, DOI: 10.1109/TVT.2026.3721188

Magnetic induction communication (MIC) is a promising technology for through-the-earth (TTE) communication. Previous studies on the MIC range have often overlooked the impact of eddy losses caused by underground materials. For TTE MIC, significant eddy losses complicate the analysis of the effective MIC range, which is vital for optimizing performance but has never been addressed in the literature. Accounting for the conductivity and permittivity of the underground medium, this paper derives the effective MIC range in TTE MIC, along with a closed-from expression that predicts the optimal carrier frequency to maximize this range. Finite element simulations validate the analysis, demonstrating that the optimal carrier frequency can significantly enhance the MIC range. It is also revealed that optimizing the antenna radius is effective in extending the MIC range for TTE and vehicle MIC applications.

0.38

Bidirectional Wireless Communication for Weakly Coupled Implantable Brain-Computer Interfaces

eess.SY ·2026-09-05 ·Shreyas Sen, Baibhab Chatterjee, Gourab Barik, Anirudh Roy
benchmark similarity 0.45
Abstract · 33 pages, 4 figures, 2 Tables

Implantable brain-computer interfaces (BCIs) promise transformative societal impact, from restoring lost motor, sensory, and speech function in patients with paralysis, stroke, sclerosis, and sensory deficits to serving as a high-bandwidth conduit between human cognition and machine intelligence. Realizing these visions requires moving from laboratory prototypes to chronic clinical systems that support thousands to millions of electrodes, several millimeters to centimeters deep beneath the brain surface, under strict heating and size limits. Almost every clinically relevant implant therefore operates in a weakly coupled regime, with coupling coefficients of $10^{-3}$ to $10^{-1}$ across centimetres of lossy tissue. In this regime, the wireless channel sets the limits of power-transfer efficiency, communication bandwidth, and energy per bit. We review the recent progress of bidirectional wireless links across inductive, mid-field, RF, ultrasonic, magnetoelectric, optical, UWB, and electro-quasistatic modalities, and benchmark them against the clinical axes of depth, size, and data rate. Within an approximately 10 mW communication budget set by the approximately $1^\circ$C tissue-heating ceiling, narrowband high-Q links are well suited for power transfer and low-speed data, but at 1-10 nJ/b cannot reach the greater than 10 Mbps to tens of Gbps uplinks that thousand- to million-channel interfaces demand, even with aggressive on-implant compression. This calls for sub-10 pJ/b and ultimately sub-1 pJ/b wireless links, where wideband techniques, such as ultra-wideband (UWB) and brain-channel communication (BCC), are suitable. We close with a quantitative framework for analyzing such wireless links and a co-design roadmap across electromagnetics, circuits, packaging, security, and regulation, toward secure, networked, million-channel brain interfaces.

0.34

HDR-DTCP: High-Dynamic-Range Differential Tissue-Coupled Wireless Powering for Tunable Biphasic Stimulation in Injectable Implants

eess.SP eess.SY ·2026-11-14 ·Asif Iftekhar Omi, Anyu Jiang, Han Wu, Sultan Mahmud, Mali Halac +4 more
similarity 0.45
Abstract

Next-generation implantable neurostimulators require efficient wireless power delivery in smaller, less invasive form factors, yet variations in tissue coupling, implantation conditions, and stimulation loads can alter received power (or requirements) and degrade energy-harvesting performance. This work presents high-dynamic-range (HDR) differential tissue-coupled powering (DTCP), termed HDR-DTCP, a body-coupled architecture in which noninvasive differential electrodes establish an AC electric field in tissue and a thread-like implant extracts power from the longitudinal potential gradient for on-chip harvesting and stimulation. Distributed coupling along the implant length, rather than receiver cross-section, enables minimally invasive injectable geometries, with harvested power scaling with implant length. A 65-nm complementary metal-oxide-semiconductor (CMOS) prototype integrates a reconfigurable four-stage energy harvester with automatic stage selection and a tunable constant-current biphasic stimulator, achieving 63.5\% peak power-conversion efficiency (PCE) at $-16.3$ dBm and $>50\%$ PCE over a 9-dB input-power range. Preliminary in-vivo validation of DTCP was done with a second application-specific integrated circuit (ASIC) on an injectable flexible substrate, capable of being delivered through a 14-gauge needle adjacent to the sciatic nerve of Sprague-Dawley rats (n=4). External copper-pad DTCP powering evoked repeatable hind-limb twitches and electromyography (EMG) response; ultrasound confirmed sub-millimeter nerve proximity, video-based paw-tip tracking captured the response, and 8-day post-implant dissection verified placement along the sciatic-nerve trajectory. Overall, this work provides an integrated wireless powering, energy-harvesting, and stimulation platform toward fully battery-free injectable neuromodulation implants.

0.21

Gradient-Field Frequency Encoding for 1D Localization of Magneto-Mechanical Resonators

eess.SP eess.SY ·2026-10-05 ·Fabian Mohn, Martin Möddel, Tobias Knopp
similarity 0.40
Abstract · 4 pages, 3 figures

Magneto-mechanical resonators (MMRs) have emerged as promising passive sensors for wireless localization and remote inductive sensing in industrial and medical environments. However, fundamental limits on achievable localization accuracy using the spatial sensitivity profiles of receive coils remain a key challenge. In this work, we demonstrate that an MMR - comprising two permanent magnets in an anti-parallel arrangement - exhibits a measurable shift in its mechanical resonance frequency in response to an externally applied magnetic field, a property we exploit for spatial encoding. We derive and experimentally validate a method that maps frequency shifts of 1D positions within a known magnetic gradient field, demonstrating accurate localization along a linear trajectory. This gradient-field frequency encoding approach constitutes a conceptually distinct alternative to conventional sensitivity encoding with receiver coil arrays, with the potential to simplify hardware requirements while improving localization accuracy.

0.27

Electromagnetic Micro-Guidewire Control in Large Workspaces

eess.SY ·2026-09-14 ·Jasan Zughaibi, Elia Jaggy, Valentin Gantenbein, Denis von Arx, Cristiano Sartini +7 more
similarity 0.39
Abstract

Electromagnetic navigation requires sufficient actuation at clinically relevant distances due to limited magnetic volumes and coil currents. We combine real-time pose feedback with constrained convex optimization, dynamic feedback, and repetitive control to achieve energy-efficient micro-guidewire steering inside realistic anatomical models. Experiments with a clinically oriented, three-coil electromagnetic navigation system and a 0.6 mm-diameter tip magnet demonstrate angular tracking with root-mean-square errors below 0.25 degrees at distances up to 55 cm from the coil cover. Nullspace current redistribution maintains accurate tracking under active 45 A coil-current constraints. Compared with conventional field alignment, we demonstrate that pose-dependent torque-based allocation substantially reduces current demand, with the efficiency benefit retained at a pose-feedback rate of 15 Hz. These results demonstrate how real-time state information and optimization can extend electromagnetic guidewire control toward clinically relevant working distances.

0.23

A Hierarchically Compressed Immittance Operator Formulation for Fast Statics-to-Full-Wave Electromagnetic Simulation with the Boundary Element Method

cs.CE physics.comp-ph ·2026-11-16 ·Shashwat Sharma
similarity 0.38
Abstract · 15 pages, 12 figures. Submitted to the IEEE Transactions on Antennas and Propagation

The boundary element method allows the electromagnetic simulation of three-dimensional problems with a two-dimensional discretization, using a surface integral representation of Maxwell's equations. For applications involving broadband simulations or multiscale geometries, the well-known low-frequency breakdown of conventional integral equation formulations must be addressed, while ensuring that dielectric, semiconducting, and lossy metallic regions are modeled rigorously, including the skin effect. In this work, a new boundary element formulation based on an immittance operator is proposed, where the underlying integral equations are inherently valid from statics to the full-wave regime. To enable efficient matrix assembly and solution, a hierarchical matrix compression strategy is devised, leveraging fast direct factorization as an intermediate step, and fast matrix-vector products for the iterative solution of the final system of equations. The proposed formulation is capable of electrostatic, magnetostatic, quasistatic, and full-wave simulations of dielectrics, semiconductors, and lossy metals. Representative numerical examples drawn from a variety of applications demonstrate the accuracy and efficiency of the proposed method.

0.40

DC-TORQ: Fast Assessment of Data Center-Induced Generator Torsional Oscillation Risk Quantification

eess.SY ·2026-10-15 ·Yingyi Tang, Bin Wang, Junbo Zhao, Qiang Zhang, Pradip Vijayan +2 more
benchmark similarity 0.38
Abstract

The periodic power demand of large data centers can introduce forced oscillations into the power system. When these oscillations contain frequency components near a turbine-generator torsional mode, they may excite sustained shaft torque oscillations with amplitudes exceeding the allowable cyclic torque limits. Characterizing the impact of data center power oscillations on generator shaft torque requires accounting for frequency-dependent network coupling among the data center, the transmission network, and the generators. Although electromagnetic transient (EMT) simulations can capture these interactions in detail, repeated simulations across excitation frequencies and load locations can be computationally demanding. This paper proposes an analytical framework for data center induced torsional oscillations that retains the coupled electrical and shaft dynamics. A detailed generator-network model is established, and closed-form transfer functions that map data center power perturbations to the torque in each shaft section are derived. Admissible shaft torque amplitudes are then translated into frequency-dependent limits on data center power oscillations. The proposed framework enables fast assessment of the shaft torque response while accounting for network dynamics and provides a quantitative screening criterion for candidate data center connection buses. The framework is validated on the IEEE benchmark model and applied to the IEEE 39-bus system.

0.19