Perturbation-Based Reconstruction of Depth-Resolved OCT Attenuation without Requiring a Fully Decayed Signal Tail

Jinglun Yu, Yaning Wang, Shuwen Wei, Jin U. Kang

eess.IV physics.optics ·submitted 2026-11-04 05:19 UTC · v1
unusual

Abstract

Depth-resolved attenuation in optical coherence tomography (OCT) can provide quantitative optical contrast beyond structural imaging. However, conventional reconstruction methods can be biased when the OCT signal does not fully decay within the imaging range or when the distal signal tail is dominated by noise or residual reflections from distal sample or support interfaces. Here, we present a perturbation-based Beer-Lambert framework that combines a zeroth-order exponential estimate with a first-order correction, reducing reliance on the unreliable distal signal tail below the current depth. Numerical A-line simulations, Monte Carlo modeling, multilayer silicone-dye-TiO2 phantoms, and ex vivo bovine retina experiments were used to evaluate the method under finite-depth and heterogeneous imaging conditions. The numerical A-line tests showed reduced depth-dependent error against prescribed attenuation profiles; the Monte Carlo comparisons showed less distal drift relative to model-defined benchmarks, while the phantom profiles were compared with separately acquired single-layer OCT calibration values. In a Monte Carlo component test that withheld vessel-mask and vessel-boundary information from reconstruction and used that information only for post hoc evaluation, the first-order correction recovered 61% of the segmentation-informed target vessel contrast and reduced vessel ROI mean absolute error by 65% relative to the fixed zeroth-order baseline. These results suggest that perturbation-based reconstruction may provide a practical approach for depth-resolved reconstruction of effective OCT attenuation when the fully decayed A-line condition is not satisfied.

Comment
26 pages, 10 figures, including 1 supplementary figure. Submitted to Biomedical Optics Express

Closest papers

Nearest neighbours by meaning across the whole index.

Differentiable Inverse Rendering Recovers Intrinsic Properties from Intravascular Optical Coherence Tomography

cs.CV eess.IVphysics.med-phphysics.optics ·2026-11-09 ·Ajay Manicka, Karim Kadry, Sebastian Ruiz-Lopera, Maryam Alsharqi, Sanjana Arji +5 more
similarity 0.67
Abstract · 27 Pages, 6 Figures, Under Review

Medical images encode intrinsic tissue properties that indicate disease progression, yet these properties remain entangled with the system-dependent physics of image formation. Differentiable inverse rendering asks whether they are recoverable by inverting the imaging system's forward physics rather than through post-hoc correction. We address this in optical coherence tomography (OCT), focusing on the optical attenuation coefficient, a per-pixel marker of pathology and cellular composition. Conventional recovery relies on closed-form inversions limited to simplified light-tissue models, or learned methods coupled to specific imaging configurations. We instead recover per-pixel attenuation from a single OCT image by gradient descent through a differentiable physics-based forward model. We verified the method against analytical solutions, validated it on five in vitro phantoms, and applied it without modification to 29{,}269 clinical intravascular OCT images from 128 patients. The inversion needs no training data or segmentations and is portable across platforms and retrospective archives. Replacing the forward model or recovering another optical parameter leaves the optimization procedure unchanged. We demonstrate per-pixel recovery is tractable because gradient information scales to this high-dimensional problem, where gradient-free methods become intractable. Our work supports quantitative tissue characterization from OCT images and provides a framework for extension across imaging modalities and organ systems.

0.26

High-resolution retinal matrix imaging via computational motion and aberration correction

eess.IV physics.opticsphysics.app-phphysics.med-ph ·2026-10-13 ·Aymeric Melt, Léo Puyo, Victor Barolle, Claude Boccara, Gereon Hüttmann +1 more
unusual similarity 0.58
Abstract · 46 pages, 9 figures

High-resolution optical imaging of the human retina is essential for diagnosing ophthalmic and neurodegenerative diseases. While adaptive optics combined with coherence gating can resolve retinal structures at the cellular scale, it requires complex, costly, and cumbersome hardware to physically correct for ocular aberrations and motion. Matrix imaging offers a powerful digital alternative by leveraging wave linearity to undo aberrations and multiple scattering; however, its in vivo deployment has been hindered by continuous, involuntary eye movements that destroy phase coherence during acquisition. Here, we overcome this fundamental barrier and demonstrate the first in vivo reflection matrix imaging of the human retina using an ultra-simplified optical setup without physical correctors. We introduce a fully computational framework that tracks and compensates for multi-axial ocular motion directly from raw interferometric data, thereby restoring the phase coherence of a multi-spectral distortion matrix. This enables local, high-order aberration correction to achieve diffraction-limited, volumetric imaging of retinal tissues. By shifting the hardware burden to physics-guided numerical computation, this paradigm opens the way for accessible, non-invasive, and high-resolution digital ophthalmology.

0.25

In Defense of OCTA: The Reconstruction-Utility Gap in OCT-to-OCTA Synthesis

cs.LG ·2026-08-16 ·Michael Chertok, Alon Tiosano, Orly Gal-Or, Lior Kramarski, Einav Baharav Shlezinger +2 more
published similarity 0.54
Abstract · 10 pages, 2 figures, 3 tables. Accepted at OMIA 2026 (13th Ophthalmic Medical Image Analysis Workshop, MICCAI 2026). This is the pre-peer-re…

Optical coherence tomography angiography (OCTA) images retinal blood flow, giving capillary-perfusion and foveal-avascular-zone biomarkers that grade diabetic-retinopathy ischemia. Because OCTA hardware is less common than structural OCT, recent work synthesizes it from OCT, reporting strong reconstruction (3D PSNR > 31 dB, SSIM > 0.9). We ask not whether the synthetic image looks similar, but whether it supports the measurements OCTA is acquired for. A frozen real-OCTA segmenter, applied as a probe to two synthesizers (XOCT, TransPro), shows downstream Dice falling with structural fineness: large vessels survive (0.862 -> 0.831) while the fine capillary network collapses (0.798 -> 0.635, five times the large-vessel loss; paired Wilcoxon p < 1e-3), TransPro worse throughout. A matched-blur control shows this detail is fabricated, not blurred. Retrained on a private Spectralis dataset, neither synthesizer reproduces the neovascular lesion (qualitative, n=3). Reconstruction fidelity is not clinical utility; we establish downstream-task fidelity as the evaluation OCT-to-OCTA synthesis needs.

0.24

An Ultra-Widefield Swept-Source OCTA Dataset and a Polar-Gated Mamba Network for Retinal Vessel Segmentation

cs.CV ·2026-09-11 ·Yang Liu, Yibing Shen, Keming Zhao, Cenk Jiang, Zhenghang Qian +10 more
code/data similarity 0.51
Abstract · Submitted to IEEE Journal of Biomedical and Health Informatics. Under review

Ultra-widefield (UWF) swept-source optical coherence tomography angiography (SS-OCTA) enables large-area retinal vascular imaging, yet vessel segmentation at this scale lacks dedicated public benchmarks and comprehensive evaluation for quantitative vascular analysis. We introduce WOIVES, to our knowledge the first publicly available UWF SS-OCTA vessel-segmentation dataset, comprising 206 eyes from 152 participants with a 24x20mm^2 field of view. WOIVES spans emmetropia to high myopia and provides soft probability vessel annotations. We further propose PG-Mamba, a visual state space model that enhances conventional directional scans with two complementary polar-coordinate scan orders. An auxiliary Dynamic FOV Gating module performs spatial modulation at the bottleneck. PG-Mamba outperformed seven competitive approaches on broad segmentation metrics under cross-validation. It achieved the lowest median absolute errors for vessel density, fractal dimension, and vessel length density. WOIVES is publicly available on Zenodo (DOI: 10.5281/zenodo.21904672), and the PG-Mamba code is available at https://github.com/syb1234567/PG-Mamba.

0.36

ARCOS: Zero-shot Boundary Localization for Corneal Layer Segmentation Across Optical Coherence Tomography Devices

cs.CV ·2026-09-03 ·Nuno Vivas Brás, Benjamin Memmi, Maëlle Bouhassane, Cristina Georgeon, Vincent Borderie +2 more
similarity 0.51
Abstract · 18 pages, 9 figures, 7 tables

Accurate segmentation of corneal layers in optical coherence tomography (OCT) is essential for quantitative assessment of corneal morphology, including layer thickness and structural changes associated with disease or surgery. However, automatic segmentation remains challenging because corneal interfaces are thin, affected by speckle noise, and variable across acquisition devices. In this work, we propose ARCOS, a patch-based zero-shot boundary localization framework for corneal layer segmentation in clinical anterior-segment OCT images. Rather than performing conventional region classification, the method predicts boundary heatmaps for the main corneal interfaces from overlapping native-resolution patches. Patch-level predictions are stitched across the full B-scan and converted into boundary locations to obtain continuous, anatomically ordered layer segmentations. The network combines multi-scale feature fusion with a self-conditioned refinement module that uses intermediate boundary information to improve local heatmap predictions while preserving spatial detail. The method was evaluated on clinical OCT images acquired from multiple devices and compared with representative segmentation baselines using boundary localization and derived thickness metrics. The proposed method achieved an off-by-one boundary localization accuracy of 95.1% and a mean absolute boundary error of 0.514 pixels on the matched-device test set. In zero-shot cross-device evaluation, it maintained an average off-by-one accuracy of 84.3% and a mean absolute boundary error of 0.855 pixels across unseen acquisition devices, outperforming the baseline models. Thickness estimates derived from the predicted boundaries showed low error across corneal regions, supporting the method's use for quantitative corneal OCT analysis.

0.28

Low-rank MERA Decomposition of 3D OCT Image Volume for Compression, Denoising, Feature Extraction and Diagnosis of Retinal Diseases

cs.CV ·2026-10-26 ·Ivica Kopriva, Fei Shi, Marija Štanfel, Qin Peng, Jianhao Bao +2 more
similarity 0.51
Abstract · 10 pages, 4 figures, 5 tables

Optical coherence tomography (OCT) is a non-invasive imaging modality used for visualization of retinal tissue. However, OCT volumes are affected by speckle noise, which degrades image contrast and obscures fine structural details. Furthermore, modern clinical practice requires large field-of-view, high-resolution OCT volumes, whose storage and transmission are demanding for archiving and teleophthalmology applications. To address aforementioned challenges, we propose a low-rank multi-scale entanglement renormalization ansatz (MERA) tensor network approximation of 3D OCT image volume, where the image volume is represented as contractions of a top core factor encoding global features and multiple interacting low-rank orthogonal or semi-orthogonal factors modeling inter- and intra-B scan correlations. This representation jointly enables compression, denoising, and feature extraction. Unlike deep learning approaches, the proposed approach is training-free and processes each OCT volume independently. We evaluate the proposed framework on OCT image volumes acquired by two different scanners and demonstrate improved visual quality and reduced retinal layer segmentation errors in comparison with several competing methods. We also extract MERA features from two clinical OCT volume datasets. They are subsequently classified using a Grassmann-manifold nearest-subspace classifier. With the subspace dimension as its only hyperparameter, the classifier is theoretically shown to be robust in small-sample clinical scenario. We demonstrate competitive accuracy of classification of retinal diseases in these small-sample settings, where deep learning methods are often limited by the scarcity of labeled data. The proposed structure-aware model-driven representation learning for high-dimensional data under limited supervision provides good generalization performance while maintaining high interpretability.

0.35

Multi-Aperture PPG with MAPIS: Spatial Optical and Temporal Coherence Fields

cs.CV physics.optics ·2026-09-27 ·Shuguang Wang, Yuanjing Wang
unusual similarity 0.50
Abstract · 10 pages, 3 figures

Conventional reflectance photoplethysmography (PPG) typically reduces tissue optical responses to one or a few detector channels. Matrix Pinhole Image Sensing (MAPIS) instead simultaneously samples multiple aperture-dependent optical signals. We analyzed 44 recordings of 30 seconds each from a single MAPIS device to characterize the spatial relationships among DC intensity, pulsatile AC, perfusion index (PI), and temporal coherence across Red and infrared channels. AC increased sublinearly with DC at both wavelengths, resulting in higher PI toward lower-DC apertures, consistent with a first-order relative path-sensitivity interpretation. A second counterintuitive spatial trend was observed in temporal coherence: lower-DC and lower-AC apertures exhibited higher autocorrelation and greater cardiac-periodic spectral concentration. IR signals also showed consistently higher temporal coherence than corresponding Red signals. These observations demonstrate that multi-aperture PPG resolves reproducible spatial relationships among optical intensity, fractional pulsatile sensitivity, and waveform coherence that are largely averaged together in conventional single-channel PPG. The present study establishes within-device relationships and motivates controlled studies of the underlying photon-path mechanisms.

0.30

Label-Free Deep-Tissue Peripheral Nerve Detection with a Handheld Multimodal OCT Probe and NerveDetNet

cs.CV physics.opticsphysics.med-ph ·2026-08-13 ·Yihan Wang, Ruilin You, Shaobai Li, Jiabin Chen, Bofan Song +2 more
similarity 0.50
Abstract

Peripheral nerves buried beneath intact tissue are difficult to visualize during surgery and remain inaccessible to white light wide-field imaging and other surface optical imaging methods. Existing OCT nerve studies have largely relied on exposed nerves or polarization contrast with limited depth penetration, restricting their value for subsurface intraoperative guidance. Here, we introduce, to our knowledge, the first label-free framework for detecting peripheral nerves beneath unopened tissue and resolving their depth using intensity-based OCT structural signatures alone. The framework combines a handheld multimodal probe, integrating swept-source OCT with co-registered white light and autofluorescence imaging, with a ``confirm-then-capture'' workflow designed for practical surgical use. To enable efficient analysis of sparsely sampled OCT volumes, we develop NerveDetNet, a lightweight 2.5D segmentation network that recovers weak and spatially displaced nerve signals by incorporating spatial context, frame-order information, and shift-tolerant correlations across frames through a dedicated nerve feature correlation module. In ex vivo tissue experiments, NerveDetNet consistently outperformed six representative 2D baselines across all frame spacings, achieving a Dice score of 0.725 under the sparsest sampling condition while using approximately half the model parameters. End-to-end validation demonstrated localization of nerves invisible at the surface and depth-resolved detection up to 1.3--1.4~mm below the tissue surface, with OCT derived depth maps overlaid directly onto the surgical view. Together, these results establish a practical label-free approach for subsurface nerve visualization that supports intraoperative compatibility, enables efficient sparse-volume analysis, and provides depth-resolved guidance without tissue opening, contrast agents, or nerve exposure.

0.04